Intraocular pressure changes after phacoemulsification and foldable silicone lens implantation using Healon GV®
Identifieur interne : 001588 ( PascalFrancis/Corpus ); précédent : 001587; suivant : 001589Intraocular pressure changes after phacoemulsification and foldable silicone lens implantation using Healon GV®
Auteurs : L. Mastropasqua ; P. Carpineto ; M. Ciancaglini ; G. FalconioSource :
- Ophthalmologica : (Basel) [ 0030-3755 ] ; 1998.
Descripteurs français
- Pascal (Inist)
English descriptors
- KwdEn :
Abstract
The authors studied the intraocular pressure (IOP) changes after phacoemulsification with two different kinds of foldable silicone lens implantation using Healon GV® as viscoelastic substance. One hundred patients undergoing cataract surgery were enrolled in this study. Inclusion criteria were: absence of ocular hypertension or glaucoma. Patients were randomly assigned to plate haptic (Silens PH®, Domilens® Chiron Vision, Lyon, France) or three-piece (CeeOn 920, Pharmacia spa, Milano, Italia) silicone intraocular lens (IOL) implantation. Following phacoemulsification and cortical aspiration, the capsular bag was reinflated with Healon GV After foldable silicone IOL insertion residual Healon GV was removed from the anterior chamber. IOP was measured preoperatively and at 6, 24 h and 1 week postoperatively. Six hours postoperatively IOP was higher in the Silens PH group than in the CeeOn group (20.85±5.42 vs. 18.88±2.95 mm Hg, p=0.026). The difference was confirmed after 24h (21.02±5.18 vs. 17.34 ±3.18 mm Hg, p<0.01). Despite the medical treatment (acetazolamide 250mg orally every 6 h), at the 24-hour control 2 eyes with plate haptic silicone lens showed IOP values higher than 30 mm Hg. Slitlamp examination showed in both eyes a shallowing of the anterior chamber together with the evidence of capsular bag hyperdistension and capsular block resulting from the occlusion of the circular anterior capsule opening by the plate haptic IOL. Residual Healon GV removal from the anterior chamber was performed. At 1-week control both eyes showed normalization of IOP. The use of a plate haptic silicone lens may be associated with a more consistent retention of Healon GV in the eye with trapping in the capsular bag. Retained viscoelastic may cause either trabecular meshwork blockage by viscoelastic substance or postoperative capsular bag hyperdistension, anteroplacement of the IOL optic and capsular block from occlusion of the circular anterior opening by the IOL optic.
Notice en format standard (ISO 2709)
Pour connaître la documentation sur le format Inist Standard.
pA |
|
---|
Format Inist (serveur)
NO : | PASCAL 98-0406307 INIST |
---|---|
ET : | Intraocular pressure changes after phacoemulsification and foldable silicone lens implantation using Healon GV® |
AU : | MASTROPASQUA (L.); CARPINETO (P.); CIANCAGLINI (M.); FALCONIO (G.) |
AF : | Institute of Ophthalmology, University 'G. D'Annunzio'/Chieti/Italie (1 aut., 2 aut., 3 aut., 4 aut.) |
DT : | Publication en série; Niveau analytique |
SO : | Ophthalmologica : (Basel); ISSN 0030-3755; Suisse; Da. 1998; Vol. 212; No. 5; Pp. 318-321; Bibl. 24 ref. |
LA : | Anglais |
EA : | The authors studied the intraocular pressure (IOP) changes after phacoemulsification with two different kinds of foldable silicone lens implantation using Healon GV® as viscoelastic substance. One hundred patients undergoing cataract surgery were enrolled in this study. Inclusion criteria were: absence of ocular hypertension or glaucoma. Patients were randomly assigned to plate haptic (Silens PH®, Domilens® Chiron Vision, Lyon, France) or three-piece (CeeOn 920, Pharmacia spa, Milano, Italia) silicone intraocular lens (IOL) implantation. Following phacoemulsification and cortical aspiration, the capsular bag was reinflated with Healon GV After foldable silicone IOL insertion residual Healon GV was removed from the anterior chamber. IOP was measured preoperatively and at 6, 24 h and 1 week postoperatively. Six hours postoperatively IOP was higher in the Silens PH group than in the CeeOn group (20.85±5.42 vs. 18.88±2.95 mm Hg, p=0.026). The difference was confirmed after 24h (21.02±5.18 vs. 17.34 ±3.18 mm Hg, p<0.01). Despite the medical treatment (acetazolamide 250mg orally every 6 h), at the 24-hour control 2 eyes with plate haptic silicone lens showed IOP values higher than 30 mm Hg. Slitlamp examination showed in both eyes a shallowing of the anterior chamber together with the evidence of capsular bag hyperdistension and capsular block resulting from the occlusion of the circular anterior capsule opening by the plate haptic IOL. Residual Healon GV removal from the anterior chamber was performed. At 1-week control both eyes showed normalization of IOP. The use of a plate haptic silicone lens may be associated with a more consistent retention of Healon GV in the eye with trapping in the capsular bag. Retained viscoelastic may cause either trabecular meshwork blockage by viscoelastic substance or postoperative capsular bag hyperdistension, anteroplacement of the IOL optic and capsular block from occlusion of the circular anterior opening by the IOL optic. |
CC : | 002B25B |
FD : | Phacoémulsification; Cristallin; Lentille intraoculaire; Chambre postérieure; Tonus oculaire; Cataracte; Traitement; Matériau viscoélastique; Postopératoire; Hypertonie oculaire; Complication; Iatrogène; Homme |
FG : | Chirurgie; Oeil pathologie; Cristallin pathologie; Segment antérieur pathologie |
ED : | Phacoemulsification; Lens; Intraocular lens; Posterior chamber; Intraocular pressure; Cataract; Treatment; Viscoelastic material; Postoperative; Ocular hypertension; Complication; Iatrogenic; Human |
EG : | Surgery; Eye disease; Lens disease; Anterior segment disease |
GD : | Aufbereiten |
SD : | Facoemulsificación; Cristalino; Lente intraocular; Cámara posterior; Tono ocular; Catarata; Tratamiento; Material viscoelástico; Postoperatorio; Hipertensión ocular; Complicación; Iatrógeno; Hombre |
LO : | INIST-4874.354000072801760050 |
ID : | 98-0406307 |
Links to Exploration step
Pascal:98-0406307Le document en format XML
<record><TEI><teiHeader><fileDesc><titleStmt><title xml:lang="en" level="a">Intraocular pressure changes after phacoemulsification and foldable silicone lens implantation using Healon GV®</title>
<author><name sortKey="Mastropasqua, L" sort="Mastropasqua, L" uniqKey="Mastropasqua L" first="L." last="Mastropasqua">L. Mastropasqua</name>
<affiliation><inist:fA14 i1="01"><s1>Institute of Ophthalmology, University 'G. D'Annunzio'</s1>
<s2>Chieti</s2>
<s3>ITA</s3>
<sZ>1 aut.</sZ>
<sZ>2 aut.</sZ>
<sZ>3 aut.</sZ>
<sZ>4 aut.</sZ>
</inist:fA14>
</affiliation>
</author>
<author><name sortKey="Carpineto, P" sort="Carpineto, P" uniqKey="Carpineto P" first="P." last="Carpineto">P. Carpineto</name>
<affiliation><inist:fA14 i1="01"><s1>Institute of Ophthalmology, University 'G. D'Annunzio'</s1>
<s2>Chieti</s2>
<s3>ITA</s3>
<sZ>1 aut.</sZ>
<sZ>2 aut.</sZ>
<sZ>3 aut.</sZ>
<sZ>4 aut.</sZ>
</inist:fA14>
</affiliation>
</author>
<author><name sortKey="Ciancaglini, M" sort="Ciancaglini, M" uniqKey="Ciancaglini M" first="M." last="Ciancaglini">M. Ciancaglini</name>
<affiliation><inist:fA14 i1="01"><s1>Institute of Ophthalmology, University 'G. D'Annunzio'</s1>
<s2>Chieti</s2>
<s3>ITA</s3>
<sZ>1 aut.</sZ>
<sZ>2 aut.</sZ>
<sZ>3 aut.</sZ>
<sZ>4 aut.</sZ>
</inist:fA14>
</affiliation>
</author>
<author><name sortKey="Falconio, G" sort="Falconio, G" uniqKey="Falconio G" first="G." last="Falconio">G. Falconio</name>
<affiliation><inist:fA14 i1="01"><s1>Institute of Ophthalmology, University 'G. D'Annunzio'</s1>
<s2>Chieti</s2>
<s3>ITA</s3>
<sZ>1 aut.</sZ>
<sZ>2 aut.</sZ>
<sZ>3 aut.</sZ>
<sZ>4 aut.</sZ>
</inist:fA14>
</affiliation>
</author>
</titleStmt>
<publicationStmt><idno type="wicri:source">INIST</idno>
<idno type="inist">98-0406307</idno>
<date when="1998">1998</date>
<idno type="stanalyst">PASCAL 98-0406307 INIST</idno>
<idno type="RBID">Pascal:98-0406307</idno>
<idno type="wicri:Area/PascalFrancis/Corpus">001588</idno>
</publicationStmt>
<sourceDesc><biblStruct><analytic><title xml:lang="en" level="a">Intraocular pressure changes after phacoemulsification and foldable silicone lens implantation using Healon GV®</title>
<author><name sortKey="Mastropasqua, L" sort="Mastropasqua, L" uniqKey="Mastropasqua L" first="L." last="Mastropasqua">L. Mastropasqua</name>
<affiliation><inist:fA14 i1="01"><s1>Institute of Ophthalmology, University 'G. D'Annunzio'</s1>
<s2>Chieti</s2>
<s3>ITA</s3>
<sZ>1 aut.</sZ>
<sZ>2 aut.</sZ>
<sZ>3 aut.</sZ>
<sZ>4 aut.</sZ>
</inist:fA14>
</affiliation>
</author>
<author><name sortKey="Carpineto, P" sort="Carpineto, P" uniqKey="Carpineto P" first="P." last="Carpineto">P. Carpineto</name>
<affiliation><inist:fA14 i1="01"><s1>Institute of Ophthalmology, University 'G. D'Annunzio'</s1>
<s2>Chieti</s2>
<s3>ITA</s3>
<sZ>1 aut.</sZ>
<sZ>2 aut.</sZ>
<sZ>3 aut.</sZ>
<sZ>4 aut.</sZ>
</inist:fA14>
</affiliation>
</author>
<author><name sortKey="Ciancaglini, M" sort="Ciancaglini, M" uniqKey="Ciancaglini M" first="M." last="Ciancaglini">M. Ciancaglini</name>
<affiliation><inist:fA14 i1="01"><s1>Institute of Ophthalmology, University 'G. D'Annunzio'</s1>
<s2>Chieti</s2>
<s3>ITA</s3>
<sZ>1 aut.</sZ>
<sZ>2 aut.</sZ>
<sZ>3 aut.</sZ>
<sZ>4 aut.</sZ>
</inist:fA14>
</affiliation>
</author>
<author><name sortKey="Falconio, G" sort="Falconio, G" uniqKey="Falconio G" first="G." last="Falconio">G. Falconio</name>
<affiliation><inist:fA14 i1="01"><s1>Institute of Ophthalmology, University 'G. D'Annunzio'</s1>
<s2>Chieti</s2>
<s3>ITA</s3>
<sZ>1 aut.</sZ>
<sZ>2 aut.</sZ>
<sZ>3 aut.</sZ>
<sZ>4 aut.</sZ>
</inist:fA14>
</affiliation>
</author>
</analytic>
<series><title level="j" type="main">Ophthalmologica : (Basel)</title>
<title level="j" type="abbreviated">Ophthalmologica : (Basel)</title>
<idno type="ISSN">0030-3755</idno>
<imprint><date when="1998">1998</date>
</imprint>
</series>
</biblStruct>
</sourceDesc>
<seriesStmt><title level="j" type="main">Ophthalmologica : (Basel)</title>
<title level="j" type="abbreviated">Ophthalmologica : (Basel)</title>
<idno type="ISSN">0030-3755</idno>
</seriesStmt>
</fileDesc>
<profileDesc><textClass><keywords scheme="KwdEn" xml:lang="en"><term>Cataract</term>
<term>Complication</term>
<term>Human</term>
<term>Iatrogenic</term>
<term>Intraocular lens</term>
<term>Intraocular pressure</term>
<term>Lens</term>
<term>Ocular hypertension</term>
<term>Phacoemulsification</term>
<term>Posterior chamber</term>
<term>Postoperative</term>
<term>Treatment</term>
<term>Viscoelastic material</term>
</keywords>
<keywords scheme="Pascal" xml:lang="fr"><term>Phacoémulsification</term>
<term>Cristallin</term>
<term>Lentille intraoculaire</term>
<term>Chambre postérieure</term>
<term>Tonus oculaire</term>
<term>Cataracte</term>
<term>Traitement</term>
<term>Matériau viscoélastique</term>
<term>Postopératoire</term>
<term>Hypertonie oculaire</term>
<term>Complication</term>
<term>Iatrogène</term>
<term>Homme</term>
</keywords>
</textClass>
</profileDesc>
</teiHeader>
<front><div type="abstract" xml:lang="en">The authors studied the intraocular pressure (IOP) changes after phacoemulsification with two different kinds of foldable silicone lens implantation using Healon GV® as viscoelastic substance. One hundred patients undergoing cataract surgery were enrolled in this study. Inclusion criteria were: absence of ocular hypertension or glaucoma. Patients were randomly assigned to plate haptic (Silens PH®, Domilens® Chiron Vision, Lyon, France) or three-piece (CeeOn 920, Pharmacia spa, Milano, Italia) silicone intraocular lens (IOL) implantation. Following phacoemulsification and cortical aspiration, the capsular bag was reinflated with Healon GV After foldable silicone IOL insertion residual Healon GV was removed from the anterior chamber. IOP was measured preoperatively and at 6, 24 h and 1 week postoperatively. Six hours postoperatively IOP was higher in the Silens PH group than in the CeeOn group (20.85±5.42 vs. 18.88±2.95 mm Hg, p=0.026). The difference was confirmed after 24h (21.02±5.18 vs. 17.34 ±3.18 mm Hg, p<0.01). Despite the medical treatment (acetazolamide 250mg orally every 6 h), at the 24-hour control 2 eyes with plate haptic silicone lens showed IOP values higher than 30 mm Hg. Slitlamp examination showed in both eyes a shallowing of the anterior chamber together with the evidence of capsular bag hyperdistension and capsular block resulting from the occlusion of the circular anterior capsule opening by the plate haptic IOL. Residual Healon GV removal from the anterior chamber was performed. At 1-week control both eyes showed normalization of IOP. The use of a plate haptic silicone lens may be associated with a more consistent retention of Healon GV in the eye with trapping in the capsular bag. Retained viscoelastic may cause either trabecular meshwork blockage by viscoelastic substance or postoperative capsular bag hyperdistension, anteroplacement of the IOL optic and capsular block from occlusion of the circular anterior opening by the IOL optic.</div>
</front>
</TEI>
<inist><standard h6="B"><pA><fA01 i1="01" i2="1"><s0>0030-3755</s0>
</fA01>
<fA03 i2="1"><s0>Ophthalmologica : (Basel)</s0>
</fA03>
<fA05><s2>212</s2>
</fA05>
<fA06><s2>5</s2>
</fA06>
<fA08 i1="01" i2="1" l="ENG"><s1>Intraocular pressure changes after phacoemulsification and foldable silicone lens implantation using Healon GV®</s1>
</fA08>
<fA11 i1="01" i2="1"><s1>MASTROPASQUA (L.)</s1>
</fA11>
<fA11 i1="02" i2="1"><s1>CARPINETO (P.)</s1>
</fA11>
<fA11 i1="03" i2="1"><s1>CIANCAGLINI (M.)</s1>
</fA11>
<fA11 i1="04" i2="1"><s1>FALCONIO (G.)</s1>
</fA11>
<fA14 i1="01"><s1>Institute of Ophthalmology, University 'G. D'Annunzio'</s1>
<s2>Chieti</s2>
<s3>ITA</s3>
<sZ>1 aut.</sZ>
<sZ>2 aut.</sZ>
<sZ>3 aut.</sZ>
<sZ>4 aut.</sZ>
</fA14>
<fA20><s1>318-321</s1>
</fA20>
<fA21><s1>1998</s1>
</fA21>
<fA23 i1="01"><s0>ENG</s0>
</fA23>
<fA43 i1="01"><s1>INIST</s1>
<s2>4874</s2>
<s5>354000072801760050</s5>
</fA43>
<fA44><s0>0000</s0>
<s1>© 1998 INIST-CNRS. All rights reserved.</s1>
</fA44>
<fA45><s0>24 ref.</s0>
</fA45>
<fA47 i1="01" i2="1"><s0>98-0406307</s0>
</fA47>
<fA60><s1>P</s1>
</fA60>
<fA61><s0>A</s0>
</fA61>
<fA64 i2="1"><s0>Ophthalmologica : (Basel)</s0>
</fA64>
<fA66 i1="01"><s0>CHE</s0>
</fA66>
<fC01 i1="01" l="ENG"><s0>The authors studied the intraocular pressure (IOP) changes after phacoemulsification with two different kinds of foldable silicone lens implantation using Healon GV® as viscoelastic substance. One hundred patients undergoing cataract surgery were enrolled in this study. Inclusion criteria were: absence of ocular hypertension or glaucoma. Patients were randomly assigned to plate haptic (Silens PH®, Domilens® Chiron Vision, Lyon, France) or three-piece (CeeOn 920, Pharmacia spa, Milano, Italia) silicone intraocular lens (IOL) implantation. Following phacoemulsification and cortical aspiration, the capsular bag was reinflated with Healon GV After foldable silicone IOL insertion residual Healon GV was removed from the anterior chamber. IOP was measured preoperatively and at 6, 24 h and 1 week postoperatively. Six hours postoperatively IOP was higher in the Silens PH group than in the CeeOn group (20.85±5.42 vs. 18.88±2.95 mm Hg, p=0.026). The difference was confirmed after 24h (21.02±5.18 vs. 17.34 ±3.18 mm Hg, p<0.01). Despite the medical treatment (acetazolamide 250mg orally every 6 h), at the 24-hour control 2 eyes with plate haptic silicone lens showed IOP values higher than 30 mm Hg. Slitlamp examination showed in both eyes a shallowing of the anterior chamber together with the evidence of capsular bag hyperdistension and capsular block resulting from the occlusion of the circular anterior capsule opening by the plate haptic IOL. Residual Healon GV removal from the anterior chamber was performed. At 1-week control both eyes showed normalization of IOP. The use of a plate haptic silicone lens may be associated with a more consistent retention of Healon GV in the eye with trapping in the capsular bag. Retained viscoelastic may cause either trabecular meshwork blockage by viscoelastic substance or postoperative capsular bag hyperdistension, anteroplacement of the IOL optic and capsular block from occlusion of the circular anterior opening by the IOL optic.</s0>
</fC01>
<fC02 i1="01" i2="X"><s0>002B25B</s0>
</fC02>
<fC03 i1="01" i2="X" l="FRE"><s0>Phacoémulsification</s0>
<s5>01</s5>
</fC03>
<fC03 i1="01" i2="X" l="ENG"><s0>Phacoemulsification</s0>
<s5>01</s5>
</fC03>
<fC03 i1="01" i2="X" l="SPA"><s0>Facoemulsificación</s0>
<s5>01</s5>
</fC03>
<fC03 i1="02" i2="X" l="FRE"><s0>Cristallin</s0>
<s5>02</s5>
</fC03>
<fC03 i1="02" i2="X" l="ENG"><s0>Lens</s0>
<s5>02</s5>
</fC03>
<fC03 i1="02" i2="X" l="SPA"><s0>Cristalino</s0>
<s5>02</s5>
</fC03>
<fC03 i1="03" i2="X" l="FRE"><s0>Lentille intraoculaire</s0>
<s5>04</s5>
</fC03>
<fC03 i1="03" i2="X" l="ENG"><s0>Intraocular lens</s0>
<s5>04</s5>
</fC03>
<fC03 i1="03" i2="X" l="SPA"><s0>Lente intraocular</s0>
<s5>04</s5>
</fC03>
<fC03 i1="04" i2="X" l="FRE"><s0>Chambre postérieure</s0>
<s5>05</s5>
</fC03>
<fC03 i1="04" i2="X" l="ENG"><s0>Posterior chamber</s0>
<s5>05</s5>
</fC03>
<fC03 i1="04" i2="X" l="SPA"><s0>Cámara posterior</s0>
<s5>05</s5>
</fC03>
<fC03 i1="05" i2="X" l="FRE"><s0>Tonus oculaire</s0>
<s5>06</s5>
</fC03>
<fC03 i1="05" i2="X" l="ENG"><s0>Intraocular pressure</s0>
<s5>06</s5>
</fC03>
<fC03 i1="05" i2="X" l="SPA"><s0>Tono ocular</s0>
<s5>06</s5>
</fC03>
<fC03 i1="06" i2="X" l="FRE"><s0>Cataracte</s0>
<s5>07</s5>
</fC03>
<fC03 i1="06" i2="X" l="ENG"><s0>Cataract</s0>
<s5>07</s5>
</fC03>
<fC03 i1="06" i2="X" l="SPA"><s0>Catarata</s0>
<s5>07</s5>
</fC03>
<fC03 i1="07" i2="X" l="FRE"><s0>Traitement</s0>
<s5>08</s5>
</fC03>
<fC03 i1="07" i2="X" l="ENG"><s0>Treatment</s0>
<s5>08</s5>
</fC03>
<fC03 i1="07" i2="X" l="GER"><s0>Aufbereiten</s0>
<s5>08</s5>
</fC03>
<fC03 i1="07" i2="X" l="SPA"><s0>Tratamiento</s0>
<s5>08</s5>
</fC03>
<fC03 i1="08" i2="X" l="FRE"><s0>Matériau viscoélastique</s0>
<s5>09</s5>
</fC03>
<fC03 i1="08" i2="X" l="ENG"><s0>Viscoelastic material</s0>
<s5>09</s5>
</fC03>
<fC03 i1="08" i2="X" l="SPA"><s0>Material viscoelástico</s0>
<s5>09</s5>
</fC03>
<fC03 i1="09" i2="X" l="FRE"><s0>Postopératoire</s0>
<s5>10</s5>
</fC03>
<fC03 i1="09" i2="X" l="ENG"><s0>Postoperative</s0>
<s5>10</s5>
</fC03>
<fC03 i1="09" i2="X" l="SPA"><s0>Postoperatorio</s0>
<s5>10</s5>
</fC03>
<fC03 i1="10" i2="X" l="FRE"><s0>Hypertonie oculaire</s0>
<s5>13</s5>
</fC03>
<fC03 i1="10" i2="X" l="ENG"><s0>Ocular hypertension</s0>
<s5>13</s5>
</fC03>
<fC03 i1="10" i2="X" l="SPA"><s0>Hipertensión ocular</s0>
<s5>13</s5>
</fC03>
<fC03 i1="11" i2="X" l="FRE"><s0>Complication</s0>
<s5>14</s5>
</fC03>
<fC03 i1="11" i2="X" l="ENG"><s0>Complication</s0>
<s5>14</s5>
</fC03>
<fC03 i1="11" i2="X" l="SPA"><s0>Complicación</s0>
<s5>14</s5>
</fC03>
<fC03 i1="12" i2="X" l="FRE"><s0>Iatrogène</s0>
<s5>15</s5>
</fC03>
<fC03 i1="12" i2="X" l="ENG"><s0>Iatrogenic</s0>
<s5>15</s5>
</fC03>
<fC03 i1="12" i2="X" l="SPA"><s0>Iatrógeno</s0>
<s5>15</s5>
</fC03>
<fC03 i1="13" i2="X" l="FRE"><s0>Homme</s0>
<s5>20</s5>
</fC03>
<fC03 i1="13" i2="X" l="ENG"><s0>Human</s0>
<s5>20</s5>
</fC03>
<fC03 i1="13" i2="X" l="SPA"><s0>Hombre</s0>
<s5>20</s5>
</fC03>
<fC07 i1="01" i2="X" l="FRE"><s0>Chirurgie</s0>
<s5>37</s5>
</fC07>
<fC07 i1="01" i2="X" l="ENG"><s0>Surgery</s0>
<s5>37</s5>
</fC07>
<fC07 i1="01" i2="X" l="SPA"><s0>Cirugía</s0>
<s5>37</s5>
</fC07>
<fC07 i1="02" i2="X" l="FRE"><s0>Oeil pathologie</s0>
<s5>53</s5>
</fC07>
<fC07 i1="02" i2="X" l="ENG"><s0>Eye disease</s0>
<s5>53</s5>
</fC07>
<fC07 i1="02" i2="X" l="SPA"><s0>Ojo patología</s0>
<s5>53</s5>
</fC07>
<fC07 i1="03" i2="X" l="FRE"><s0>Cristallin pathologie</s0>
<s5>54</s5>
</fC07>
<fC07 i1="03" i2="X" l="ENG"><s0>Lens disease</s0>
<s5>54</s5>
</fC07>
<fC07 i1="03" i2="X" l="SPA"><s0>Cristalino patología</s0>
<s5>54</s5>
</fC07>
<fC07 i1="04" i2="X" l="FRE"><s0>Segment antérieur pathologie</s0>
<s2>NM</s2>
<s5>55</s5>
</fC07>
<fC07 i1="04" i2="X" l="ENG"><s0>Anterior segment disease</s0>
<s2>NM</s2>
<s5>55</s5>
</fC07>
<fC07 i1="04" i2="X" l="SPA"><s0>Segmento anterior patología</s0>
<s2>NM</s2>
<s5>55</s5>
</fC07>
<fN21><s1>271</s1>
</fN21>
</pA>
</standard>
<server><NO>PASCAL 98-0406307 INIST</NO>
<ET>Intraocular pressure changes after phacoemulsification and foldable silicone lens implantation using Healon GV®</ET>
<AU>MASTROPASQUA (L.); CARPINETO (P.); CIANCAGLINI (M.); FALCONIO (G.)</AU>
<AF>Institute of Ophthalmology, University 'G. D'Annunzio'/Chieti/Italie (1 aut., 2 aut., 3 aut., 4 aut.)</AF>
<DT>Publication en série; Niveau analytique</DT>
<SO>Ophthalmologica : (Basel); ISSN 0030-3755; Suisse; Da. 1998; Vol. 212; No. 5; Pp. 318-321; Bibl. 24 ref.</SO>
<LA>Anglais</LA>
<EA>The authors studied the intraocular pressure (IOP) changes after phacoemulsification with two different kinds of foldable silicone lens implantation using Healon GV® as viscoelastic substance. One hundred patients undergoing cataract surgery were enrolled in this study. Inclusion criteria were: absence of ocular hypertension or glaucoma. Patients were randomly assigned to plate haptic (Silens PH®, Domilens® Chiron Vision, Lyon, France) or three-piece (CeeOn 920, Pharmacia spa, Milano, Italia) silicone intraocular lens (IOL) implantation. Following phacoemulsification and cortical aspiration, the capsular bag was reinflated with Healon GV After foldable silicone IOL insertion residual Healon GV was removed from the anterior chamber. IOP was measured preoperatively and at 6, 24 h and 1 week postoperatively. Six hours postoperatively IOP was higher in the Silens PH group than in the CeeOn group (20.85±5.42 vs. 18.88±2.95 mm Hg, p=0.026). The difference was confirmed after 24h (21.02±5.18 vs. 17.34 ±3.18 mm Hg, p<0.01). Despite the medical treatment (acetazolamide 250mg orally every 6 h), at the 24-hour control 2 eyes with plate haptic silicone lens showed IOP values higher than 30 mm Hg. Slitlamp examination showed in both eyes a shallowing of the anterior chamber together with the evidence of capsular bag hyperdistension and capsular block resulting from the occlusion of the circular anterior capsule opening by the plate haptic IOL. Residual Healon GV removal from the anterior chamber was performed. At 1-week control both eyes showed normalization of IOP. The use of a plate haptic silicone lens may be associated with a more consistent retention of Healon GV in the eye with trapping in the capsular bag. Retained viscoelastic may cause either trabecular meshwork blockage by viscoelastic substance or postoperative capsular bag hyperdistension, anteroplacement of the IOL optic and capsular block from occlusion of the circular anterior opening by the IOL optic.</EA>
<CC>002B25B</CC>
<FD>Phacoémulsification; Cristallin; Lentille intraoculaire; Chambre postérieure; Tonus oculaire; Cataracte; Traitement; Matériau viscoélastique; Postopératoire; Hypertonie oculaire; Complication; Iatrogène; Homme</FD>
<FG>Chirurgie; Oeil pathologie; Cristallin pathologie; Segment antérieur pathologie</FG>
<ED>Phacoemulsification; Lens; Intraocular lens; Posterior chamber; Intraocular pressure; Cataract; Treatment; Viscoelastic material; Postoperative; Ocular hypertension; Complication; Iatrogenic; Human</ED>
<EG>Surgery; Eye disease; Lens disease; Anterior segment disease</EG>
<GD>Aufbereiten</GD>
<SD>Facoemulsificación; Cristalino; Lente intraocular; Cámara posterior; Tono ocular; Catarata; Tratamiento; Material viscoelástico; Postoperatorio; Hipertensión ocular; Complicación; Iatrógeno; Hombre</SD>
<LO>INIST-4874.354000072801760050</LO>
<ID>98-0406307</ID>
</server>
</inist>
</record>
Pour manipuler ce document sous Unix (Dilib)
EXPLOR_STEP=$WICRI_ROOT/Ticri/CIDE/explor/HapticV1/Data/PascalFrancis/Corpus
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 001588 | SxmlIndent | more
Ou
HfdSelect -h $EXPLOR_AREA/Data/PascalFrancis/Corpus/biblio.hfd -nk 001588 | SxmlIndent | more
Pour mettre un lien sur cette page dans le réseau Wicri
{{Explor lien |wiki= Ticri/CIDE |area= HapticV1 |flux= PascalFrancis |étape= Corpus |type= RBID |clé= Pascal:98-0406307 |texte= Intraocular pressure changes after phacoemulsification and foldable silicone lens implantation using Healon GV® }}
This area was generated with Dilib version V0.6.23. |