Development of a Haptic Elbow Spasticity Simulator (HESS) for Improving Accuracy and Reliability of Clinical Assessment of Spasticity

Identifieur interne : 001718 ( Pmc/Corpus ); précédent : 001717; suivant : 001719

Development of a Haptic Elbow Spasticity Simulator (HESS) for Improving Accuracy and Reliability of Clinical Assessment of Spasticity

Auteurs : Hyung-Soon Park ; Jonghyun Kim ; Diane L. Damiano

Source :

RBID : PMC:3579668

Abstract

This paper presents the framework for developing a robotic system to improve accuracy and reliability of clinical assessment. Clinical assessment of spasticity tends to have poor reliability because of the nature of the in-person assessment. To improve accuracy and reliability of spasticity assessment, a haptic device, named the HESS (Haptic Elbow Spasticity Simulator) has been designed and constructed to recreate the clinical “feel” of elbow spasticity based on quantitative measurements. A mathematical model representing the spastic elbow joint was proposed based on clinical assessment using the Modified Ashworth Scale (MAS) and quantitative data (position, velocity, and torque) collected on subjects with elbow spasticity. Four haptic models (HMs) were created to represent the haptic feel of MAS 1, 1+, 2, and 3. The four HMs were assessed by experienced clinicians; three clinicians performed both in-person and haptic assessments, and had 100% agreement in MAS scores; and eight clinicians who were experienced with MAS assessed the four HMs without receiving any training prior to the test. Inter-rater reliability among the eight clinicians had substantial agreement (κ = 0.626). The eight clinicians also rated the level of realism (7.63 ± 0.92 out of 10) as compared to their experience with real patients.


Url:
DOI: 10.1109/TNSRE.2012.2195330
PubMed: 22562769
PubMed Central: 3579668

Links to Exploration step

PMC:3579668***** Acces problem to record *****\

Le document en format XML


Pour manipuler ce document sous Unix (Dilib)

EXPLOR_STEP=$WICRI_ROOT/Ticri/CIDE/explor/HapticV1/Data/Pmc/Corpus
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 001718 | SxmlIndent | more

Ou

HfdSelect -h $EXPLOR_AREA/Data/Pmc/Corpus/biblio.hfd -nk 001718 | SxmlIndent | more

Pour mettre un lien sur cette page dans le réseau Wicri

{{Explor lien
   |wiki=    Ticri/CIDE
   |area=    HapticV1
   |flux=    Pmc
   |étape=   Corpus
   |type=    RBID
   |clé=     PMC:3579668
   |texte=   Development of a Haptic Elbow Spasticity Simulator (HESS) for Improving Accuracy and Reliability of Clinical Assessment of Spasticity
}}

Pour générer des pages wiki

HfdIndexSelect -h $EXPLOR_AREA/Data/Pmc/Corpus/RBID.i   -Sk "pubmed:22562769" \
       | HfdSelect -Kh $EXPLOR_AREA/Data/Pmc/Corpus/biblio.hfd   \
       | NlmPubMed2Wicri -a HapticV1 

Wicri

This area was generated with Dilib version V0.6.23.
Data generation: Mon Jun 13 01:09:46 2016. Site generation: Wed Mar 6 09:54:07 2024