Optimization of beam quality for photon-counting spectral computed tomography in head imaging: simulation study.

Identifieur interne : 000015 ( PubMed/Corpus ); précédent : 000014; suivant : 000016

Optimization of beam quality for photon-counting spectral computed tomography in head imaging: simulation study.

Auteurs : Han Chen ; Cheng Xu ; Mats Persson ; Mats Danielsson

Source :

RBID : pubmed:26835495

Abstract

Head computed tomography (CT) plays an important role in the comprehensive evaluation of acute stroke. Photon-counting spectral detectors, as promising candidates for use in the next generation of x-ray CT systems, allow for assigning more weight to low-energy x-rays that generally contain more contrast information. Most importantly, the spectral information can be utilized to decompose the original set of energy-selective images into several basis function images that are inherently free of beam-hardening artifacts, a potential advantage for further improving the diagnosis accuracy. We are developing a photon-counting spectral detector for CT applications. The purpose of this work is to determine the optimal beam quality for material decomposition in two head imaging cases: nonenhanced imaging and K-edge imaging. A cylindrical brain tissue of 16-cm diameter, coated by a 6-mm-thick bone layer and 2-mm-thick skin layer, was used as a head phantom. The imaging target was a 5-mm-thick blood vessel centered in the head phantom. In K-edge imaging, two contrast agents, iodine and gadolinium, with the same concentration ([Formula: see text]) were studied. Three parameters that affect beam quality were evaluated: kVp settings (50 to 130 kVp), filter materials ([Formula: see text] to 83), and filter thicknesses [0 to 2 half-value layer (HVL)]. The image qualities resulting from the varying x-ray beams were compared in terms of two figures of merit (FOMs): squared signal-difference-to-noise ratio normalized by brain dose ([Formula: see text]) and that normalized by skin dose ([Formula: see text]). For nonenhanced imaging, the results show that the use of the 120-kVp spectrum filtered by 2 HVL copper ([Formula: see text]) provides the best performance in both FOMs. When iodine is used in K-edge imaging, the optimal filter is 2 HVL iodine ([Formula: see text]) and the optimal kVps are 60 kVp in terms of [Formula: see text] and 75 kVp in terms of [Formula: see text]. A tradeoff of 65 kVp was proposed to lower the potential risk of skin injuries if a relatively long exposure time is necessarily performed in the iodinated imaging. In the case of gadolinium imaging, both SD and BD can be minimized at 120 kVp filtered with 2 HVL thulium ([Formula: see text]). The results also indicate that with the same concentration and their respective optimal spectrum, the values of [Formula: see text] and [Formula: see text] in gadolinium imaging are, respectively, around 3 and 10 times larger than those in iodine imaging. However, since gadolinium is used in much lower concentrations than iodine in the clinic, iodine may be a preferable candidate for K-edge imaging.

DOI: 10.1117/1.JMI.2.4.043504
PubMed: 26835495

Links to Exploration step

pubmed:26835495

Le document en format XML

<record>
<TEI>
<teiHeader>
<fileDesc>
<titleStmt>
<title xml:lang="en">Optimization of beam quality for photon-counting spectral computed tomography in head imaging: simulation study.</title>
<author>
<name sortKey="Chen, Han" sort="Chen, Han" uniqKey="Chen H" first="Han" last="Chen">Han Chen</name>
<affiliation>
<nlm:affiliation>Royal Institute of Technology (KTH) , Department of Physics, Stockholm 106 91, Sweden.</nlm:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Xu, Cheng" sort="Xu, Cheng" uniqKey="Xu C" first="Cheng" last="Xu">Cheng Xu</name>
<affiliation>
<nlm:affiliation>Royal Institute of Technology (KTH) , Department of Physics, Stockholm 106 91, Sweden.</nlm:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Persson, Mats" sort="Persson, Mats" uniqKey="Persson M" first="Mats" last="Persson">Mats Persson</name>
<affiliation>
<nlm:affiliation>Royal Institute of Technology (KTH) , Department of Physics, Stockholm 106 91, Sweden.</nlm:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Danielsson, Mats" sort="Danielsson, Mats" uniqKey="Danielsson M" first="Mats" last="Danielsson">Mats Danielsson</name>
<affiliation>
<nlm:affiliation>Royal Institute of Technology (KTH) , Department of Physics, Stockholm 106 91, Sweden.</nlm:affiliation>
</affiliation>
</author>
</titleStmt>
<publicationStmt>
<idno type="wicri:source">PubMed</idno>
<date when="2015">2015</date>
<idno type="doi">10.1117/1.JMI.2.4.043504</idno>
<idno type="RBID">pubmed:26835495</idno>
<idno type="pmid">26835495</idno>
<idno type="wicri:Area/PubMed/Corpus">000015</idno>
<idno type="wicri:explorRef" wicri:stream="PubMed" wicri:step="Corpus" wicri:corpus="PubMed">000015</idno>
</publicationStmt>
<sourceDesc>
<biblStruct>
<analytic>
<title xml:lang="en">Optimization of beam quality for photon-counting spectral computed tomography in head imaging: simulation study.</title>
<author>
<name sortKey="Chen, Han" sort="Chen, Han" uniqKey="Chen H" first="Han" last="Chen">Han Chen</name>
<affiliation>
<nlm:affiliation>Royal Institute of Technology (KTH) , Department of Physics, Stockholm 106 91, Sweden.</nlm:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Xu, Cheng" sort="Xu, Cheng" uniqKey="Xu C" first="Cheng" last="Xu">Cheng Xu</name>
<affiliation>
<nlm:affiliation>Royal Institute of Technology (KTH) , Department of Physics, Stockholm 106 91, Sweden.</nlm:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Persson, Mats" sort="Persson, Mats" uniqKey="Persson M" first="Mats" last="Persson">Mats Persson</name>
<affiliation>
<nlm:affiliation>Royal Institute of Technology (KTH) , Department of Physics, Stockholm 106 91, Sweden.</nlm:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Danielsson, Mats" sort="Danielsson, Mats" uniqKey="Danielsson M" first="Mats" last="Danielsson">Mats Danielsson</name>
<affiliation>
<nlm:affiliation>Royal Institute of Technology (KTH) , Department of Physics, Stockholm 106 91, Sweden.</nlm:affiliation>
</affiliation>
</author>
</analytic>
<series>
<title level="j">Journal of medical imaging (Bellingham, Wash.)</title>
<idno type="ISSN">2329-4302</idno>
<imprint>
<date when="2015" type="published">2015</date>
</imprint>
</series>
</biblStruct>
</sourceDesc>
</fileDesc>
<profileDesc>
<textClass></textClass>
</profileDesc>
</teiHeader>
<front>
<div type="abstract" xml:lang="en">Head computed tomography (CT) plays an important role in the comprehensive evaluation of acute stroke. Photon-counting spectral detectors, as promising candidates for use in the next generation of x-ray CT systems, allow for assigning more weight to low-energy x-rays that generally contain more contrast information. Most importantly, the spectral information can be utilized to decompose the original set of energy-selective images into several basis function images that are inherently free of beam-hardening artifacts, a potential advantage for further improving the diagnosis accuracy. We are developing a photon-counting spectral detector for CT applications. The purpose of this work is to determine the optimal beam quality for material decomposition in two head imaging cases: nonenhanced imaging and K-edge imaging. A cylindrical brain tissue of 16-cm diameter, coated by a 6-mm-thick bone layer and 2-mm-thick skin layer, was used as a head phantom. The imaging target was a 5-mm-thick blood vessel centered in the head phantom. In K-edge imaging, two contrast agents, iodine and gadolinium, with the same concentration ([Formula: see text]) were studied. Three parameters that affect beam quality were evaluated: kVp settings (50 to 130 kVp), filter materials ([Formula: see text] to 83), and filter thicknesses [0 to 2 half-value layer (HVL)]. The image qualities resulting from the varying x-ray beams were compared in terms of two figures of merit (FOMs): squared signal-difference-to-noise ratio normalized by brain dose ([Formula: see text]) and that normalized by skin dose ([Formula: see text]). For nonenhanced imaging, the results show that the use of the 120-kVp spectrum filtered by 2 HVL copper ([Formula: see text]) provides the best performance in both FOMs. When iodine is used in K-edge imaging, the optimal filter is 2 HVL iodine ([Formula: see text]) and the optimal kVps are 60 kVp in terms of [Formula: see text] and 75 kVp in terms of [Formula: see text]. A tradeoff of 65 kVp was proposed to lower the potential risk of skin injuries if a relatively long exposure time is necessarily performed in the iodinated imaging. In the case of gadolinium imaging, both SD and BD can be minimized at 120 kVp filtered with 2 HVL thulium ([Formula: see text]). The results also indicate that with the same concentration and their respective optimal spectrum, the values of [Formula: see text] and [Formula: see text] in gadolinium imaging are, respectively, around 3 and 10 times larger than those in iodine imaging. However, since gadolinium is used in much lower concentrations than iodine in the clinic, iodine may be a preferable candidate for K-edge imaging.</div>
</front>
</TEI>
<pubmed>
<MedlineCitation Owner="NLM" Status="PubMed-not-MEDLINE">
<PMID Version="1">26835495</PMID>
<DateCreated>
<Year>2016</Year>
<Month>02</Month>
<Day>02</Day>
</DateCreated>
<DateCompleted>
<Year>2016</Year>
<Month>02</Month>
<Day>02</Day>
</DateCompleted>
<DateRevised>
<Year>2016</Year>
<Month>02</Month>
<Day>04</Day>
</DateRevised>
<Article PubModel="Print-Electronic">
<Journal>
<ISSN IssnType="Print">2329-4302</ISSN>
<JournalIssue CitedMedium="Print">
<Volume>2</Volume>
<Issue>4</Issue>
<PubDate>
<Year>2015</Year>
<Month>Oct</Month>
</PubDate>
</JournalIssue>
<Title>Journal of medical imaging (Bellingham, Wash.)</Title>
<ISOAbbreviation>J Med Imaging (Bellingham)</ISOAbbreviation>
</Journal>
<ArticleTitle>Optimization of beam quality for photon-counting spectral computed tomography in head imaging: simulation study.</ArticleTitle>
<Pagination>
<MedlinePgn>043504</MedlinePgn>
</Pagination>
<ELocationID EIdType="doi" ValidYN="Y">10.1117/1.JMI.2.4.043504</ELocationID>
<Abstract>
<AbstractText>Head computed tomography (CT) plays an important role in the comprehensive evaluation of acute stroke. Photon-counting spectral detectors, as promising candidates for use in the next generation of x-ray CT systems, allow for assigning more weight to low-energy x-rays that generally contain more contrast information. Most importantly, the spectral information can be utilized to decompose the original set of energy-selective images into several basis function images that are inherently free of beam-hardening artifacts, a potential advantage for further improving the diagnosis accuracy. We are developing a photon-counting spectral detector for CT applications. The purpose of this work is to determine the optimal beam quality for material decomposition in two head imaging cases: nonenhanced imaging and K-edge imaging. A cylindrical brain tissue of 16-cm diameter, coated by a 6-mm-thick bone layer and 2-mm-thick skin layer, was used as a head phantom. The imaging target was a 5-mm-thick blood vessel centered in the head phantom. In K-edge imaging, two contrast agents, iodine and gadolinium, with the same concentration ([Formula: see text]) were studied. Three parameters that affect beam quality were evaluated: kVp settings (50 to 130 kVp), filter materials ([Formula: see text] to 83), and filter thicknesses [0 to 2 half-value layer (HVL)]. The image qualities resulting from the varying x-ray beams were compared in terms of two figures of merit (FOMs): squared signal-difference-to-noise ratio normalized by brain dose ([Formula: see text]) and that normalized by skin dose ([Formula: see text]). For nonenhanced imaging, the results show that the use of the 120-kVp spectrum filtered by 2 HVL copper ([Formula: see text]) provides the best performance in both FOMs. When iodine is used in K-edge imaging, the optimal filter is 2 HVL iodine ([Formula: see text]) and the optimal kVps are 60 kVp in terms of [Formula: see text] and 75 kVp in terms of [Formula: see text]. A tradeoff of 65 kVp was proposed to lower the potential risk of skin injuries if a relatively long exposure time is necessarily performed in the iodinated imaging. In the case of gadolinium imaging, both SD and BD can be minimized at 120 kVp filtered with 2 HVL thulium ([Formula: see text]). The results also indicate that with the same concentration and their respective optimal spectrum, the values of [Formula: see text] and [Formula: see text] in gadolinium imaging are, respectively, around 3 and 10 times larger than those in iodine imaging. However, since gadolinium is used in much lower concentrations than iodine in the clinic, iodine may be a preferable candidate for K-edge imaging.</AbstractText>
</Abstract>
<AuthorList CompleteYN="Y">
<Author ValidYN="Y">
<LastName>Chen</LastName>
<ForeName>Han</ForeName>
<Initials>H</Initials>
<AffiliationInfo>
<Affiliation>Royal Institute of Technology (KTH) , Department of Physics, Stockholm 106 91, Sweden.</Affiliation>
</AffiliationInfo>
</Author>
<Author ValidYN="Y">
<LastName>Xu</LastName>
<ForeName>Cheng</ForeName>
<Initials>C</Initials>
<AffiliationInfo>
<Affiliation>Royal Institute of Technology (KTH) , Department of Physics, Stockholm 106 91, Sweden.</Affiliation>
</AffiliationInfo>
</Author>
<Author ValidYN="Y">
<LastName>Persson</LastName>
<ForeName>Mats</ForeName>
<Initials>M</Initials>
<AffiliationInfo>
<Affiliation>Royal Institute of Technology (KTH) , Department of Physics, Stockholm 106 91, Sweden.</Affiliation>
</AffiliationInfo>
</Author>
<Author ValidYN="Y">
<LastName>Danielsson</LastName>
<ForeName>Mats</ForeName>
<Initials>M</Initials>
<AffiliationInfo>
<Affiliation>Royal Institute of Technology (KTH) , Department of Physics, Stockholm 106 91, Sweden.</Affiliation>
</AffiliationInfo>
</Author>
</AuthorList>
<Language>eng</Language>
<PublicationTypeList>
<PublicationType UI="D016428">Journal Article</PublicationType>
</PublicationTypeList>
<ArticleDate DateType="Electronic">
<Year>2015</Year>
<Month>11</Month>
<Day>06</Day>
</ArticleDate>
</Article>
<MedlineJournalInfo>
<Country>United States</Country>
<MedlineTA>J Med Imaging (Bellingham)</MedlineTA>
<NlmUniqueID>101643461</NlmUniqueID>
<ISSNLinking>2329-4302</ISSNLinking>
</MedlineJournalInfo>
<OtherID Source="NLM">PMC4718445 [Available on 11/06/16]</OtherID>
<KeywordList Owner="NOTNLM">
<Keyword MajorTopicYN="N">K-edge imaging</Keyword>
<Keyword MajorTopicYN="N">material decomposition</Keyword>
<Keyword MajorTopicYN="N">photon counting</Keyword>
<Keyword MajorTopicYN="N">spectral computed tomography</Keyword>
</KeywordList>
</MedlineCitation>
<PubmedData>
<History>
<PubMedPubDate PubStatus="received">
<Year>2015</Year>
<Month>5</Month>
<Day>20</Day>
</PubMedPubDate>
<PubMedPubDate PubStatus="accepted">
<Year>2015</Year>
<Month>10</Month>
<Day>9</Day>
</PubMedPubDate>
<PubMedPubDate PubStatus="epublish">
<Year>2015</Year>
<Month>11</Month>
<Day>6</Day>
</PubMedPubDate>
<PubMedPubDate PubStatus="entrez">
<Year>2016</Year>
<Month>2</Month>
<Day>3</Day>
<Hour>6</Hour>
<Minute>0</Minute>
</PubMedPubDate>
<PubMedPubDate PubStatus="pubmed">
<Year>2016</Year>
<Month>2</Month>
<Day>3</Day>
<Hour>6</Hour>
<Minute>0</Minute>
</PubMedPubDate>
<PubMedPubDate PubStatus="medline">
<Year>2016</Year>
<Month>2</Month>
<Day>3</Day>
<Hour>6</Hour>
<Minute>1</Minute>
</PubMedPubDate>
<PubMedPubDate PubStatus="pmc-release">
<Year>2016</Year>
<Month>11</Month>
<Day>6</Day>
<Hour>0</Hour>
<Minute>0</Minute>
</PubMedPubDate>
</History>
<PublicationStatus>ppublish</PublicationStatus>
<ArticleIdList>
<ArticleId IdType="doi">10.1117/1.JMI.2.4.043504</ArticleId>
<ArticleId IdType="pii">15107RR</ArticleId>
<ArticleId IdType="pubmed">26835495</ArticleId>
<ArticleId IdType="pmc">PMC4718445</ArticleId>
</ArticleIdList>
</PubmedData>
</pubmed>
</record>

Pour manipuler ce document sous Unix (Dilib)

EXPLOR_STEP=$WICRI_ROOT/Wicri/Terre/explor/ThuliumV1/Data/PubMed/Corpus
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 000015 | SxmlIndent | more

Ou

HfdSelect -h $EXPLOR_AREA/Data/PubMed/Corpus/biblio.hfd -nk 000015 | SxmlIndent | more

Pour mettre un lien sur cette page dans le réseau Wicri

{{Explor lien
   |wiki=    Wicri/Terre
   |area=    ThuliumV1
   |flux=    PubMed
   |étape=   Corpus
   |type=    RBID
   |clé=     pubmed:26835495
   |texte=   Optimization of beam quality for photon-counting spectral computed tomography in head imaging: simulation study.
}}

Pour générer des pages wiki

HfdIndexSelect -h $EXPLOR_AREA/Data/PubMed/Corpus/RBID.i   -Sk "pubmed:26835495" \
       | HfdSelect -Kh $EXPLOR_AREA/Data/PubMed/Corpus/biblio.hfd   \
       | NlmPubMed2Wicri -a ThuliumV1 

Wicri

This area was generated with Dilib version V0.6.21.
Data generation: Thu May 12 08:27:09 2016. Site generation: Thu Mar 7 22:33:44 2024