Rotational stability and posterior capsule opacification of a plate-haptic and an open-loop-haptic intraocular lens.

Identifieur interne : 000F65 ( PubMed/Corpus ); précédent : 000F64; suivant : 000F66

Rotational stability and posterior capsule opacification of a plate-haptic and an open-loop-haptic intraocular lens.

Auteurs : Ana Prinz ; Thomas Neumayer ; Wolf Buehl ; Lorenz Vock ; Rupert Menapace ; Oliver Findl ; Michael Georgopoulos

Source :

RBID : pubmed:21241906

English descriptors

Abstract

To compare the rotational stability and posterior capsule opacification (PCO) rate in eyes with a 1-piece or 3-piece acrylic intraocular lens (IOL).

DOI: 10.1016/j.jcrs.2010.08.049
PubMed: 21241906

Links to Exploration step

pubmed:21241906

Le document en format XML

<record>
<TEI>
<teiHeader>
<fileDesc>
<titleStmt>
<title xml:lang="en">Rotational stability and posterior capsule opacification of a plate-haptic and an open-loop-haptic intraocular lens.</title>
<author>
<name sortKey="Prinz, Ana" sort="Prinz, Ana" uniqKey="Prinz A" first="Ana" last="Prinz">Ana Prinz</name>
<affiliation>
<nlm:affiliation>Department of Ophthalmology, Medical University of Vienna, Vienna, Austria.</nlm:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Neumayer, Thomas" sort="Neumayer, Thomas" uniqKey="Neumayer T" first="Thomas" last="Neumayer">Thomas Neumayer</name>
</author>
<author>
<name sortKey="Buehl, Wolf" sort="Buehl, Wolf" uniqKey="Buehl W" first="Wolf" last="Buehl">Wolf Buehl</name>
</author>
<author>
<name sortKey="Vock, Lorenz" sort="Vock, Lorenz" uniqKey="Vock L" first="Lorenz" last="Vock">Lorenz Vock</name>
</author>
<author>
<name sortKey="Menapace, Rupert" sort="Menapace, Rupert" uniqKey="Menapace R" first="Rupert" last="Menapace">Rupert Menapace</name>
</author>
<author>
<name sortKey="Findl, Oliver" sort="Findl, Oliver" uniqKey="Findl O" first="Oliver" last="Findl">Oliver Findl</name>
</author>
<author>
<name sortKey="Georgopoulos, Michael" sort="Georgopoulos, Michael" uniqKey="Georgopoulos M" first="Michael" last="Georgopoulos">Michael Georgopoulos</name>
</author>
</titleStmt>
<publicationStmt>
<idno type="wicri:source">PubMed</idno>
<date when="2011">2011</date>
<idno type="doi">10.1016/j.jcrs.2010.08.049</idno>
<idno type="RBID">pubmed:21241906</idno>
<idno type="pmid">21241906</idno>
<idno type="wicri:Area/PubMed/Corpus">000F65</idno>
</publicationStmt>
<sourceDesc>
<biblStruct>
<analytic>
<title xml:lang="en">Rotational stability and posterior capsule opacification of a plate-haptic and an open-loop-haptic intraocular lens.</title>
<author>
<name sortKey="Prinz, Ana" sort="Prinz, Ana" uniqKey="Prinz A" first="Ana" last="Prinz">Ana Prinz</name>
<affiliation>
<nlm:affiliation>Department of Ophthalmology, Medical University of Vienna, Vienna, Austria.</nlm:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Neumayer, Thomas" sort="Neumayer, Thomas" uniqKey="Neumayer T" first="Thomas" last="Neumayer">Thomas Neumayer</name>
</author>
<author>
<name sortKey="Buehl, Wolf" sort="Buehl, Wolf" uniqKey="Buehl W" first="Wolf" last="Buehl">Wolf Buehl</name>
</author>
<author>
<name sortKey="Vock, Lorenz" sort="Vock, Lorenz" uniqKey="Vock L" first="Lorenz" last="Vock">Lorenz Vock</name>
</author>
<author>
<name sortKey="Menapace, Rupert" sort="Menapace, Rupert" uniqKey="Menapace R" first="Rupert" last="Menapace">Rupert Menapace</name>
</author>
<author>
<name sortKey="Findl, Oliver" sort="Findl, Oliver" uniqKey="Findl O" first="Oliver" last="Findl">Oliver Findl</name>
</author>
<author>
<name sortKey="Georgopoulos, Michael" sort="Georgopoulos, Michael" uniqKey="Georgopoulos M" first="Michael" last="Georgopoulos">Michael Georgopoulos</name>
</author>
</analytic>
<series>
<title level="j">Journal of cataract and refractive surgery</title>
<idno type="eISSN">1873-4502</idno>
<imprint>
<date when="2011" type="published">2011</date>
</imprint>
</series>
</biblStruct>
</sourceDesc>
</fileDesc>
<profileDesc>
<textClass>
<keywords scheme="KwdEn" xml:lang="en">
<term>Acrylates</term>
<term>Aged</term>
<term>Capsule Opacification (etiology)</term>
<term>Cohort Studies</term>
<term>Double-Blind Method</term>
<term>Follow-Up Studies</term>
<term>Humans</term>
<term>Lenses, Intraocular (adverse effects)</term>
<term>Middle Aged</term>
<term>Postoperative Complications</term>
<term>Prospective Studies</term>
<term>Prosthesis Design</term>
<term>Reproducibility of Results</term>
<term>Rotation</term>
</keywords>
<keywords scheme="MESH" type="chemical" xml:lang="en">
<term>Acrylates</term>
</keywords>
<keywords scheme="MESH" qualifier="adverse effects" xml:lang="en">
<term>Lenses, Intraocular</term>
</keywords>
<keywords scheme="MESH" qualifier="etiology" xml:lang="en">
<term>Capsule Opacification</term>
</keywords>
<keywords scheme="MESH" xml:lang="en">
<term>Aged</term>
<term>Cohort Studies</term>
<term>Double-Blind Method</term>
<term>Follow-Up Studies</term>
<term>Humans</term>
<term>Middle Aged</term>
<term>Postoperative Complications</term>
<term>Prospective Studies</term>
<term>Prosthesis Design</term>
<term>Reproducibility of Results</term>
<term>Rotation</term>
</keywords>
</textClass>
</profileDesc>
</teiHeader>
<front>
<div type="abstract" xml:lang="en">To compare the rotational stability and posterior capsule opacification (PCO) rate in eyes with a 1-piece or 3-piece acrylic intraocular lens (IOL).</div>
</front>
</TEI>
<pubmed>
<MedlineCitation Owner="NLM" Status="MEDLINE">
<PMID Version="1">21241906</PMID>
<DateCreated>
<Year>2011</Year>
<Month>01</Month>
<Day>18</Day>
</DateCreated>
<DateCompleted>
<Year>2011</Year>
<Month>05</Month>
<Day>19</Day>
</DateCompleted>
<Article PubModel="Print">
<Journal>
<ISSN IssnType="Electronic">1873-4502</ISSN>
<JournalIssue CitedMedium="Internet">
<Volume>37</Volume>
<Issue>2</Issue>
<PubDate>
<Year>2011</Year>
<Month>Feb</Month>
</PubDate>
</JournalIssue>
<Title>Journal of cataract and refractive surgery</Title>
<ISOAbbreviation>J Cataract Refract Surg</ISOAbbreviation>
</Journal>
<ArticleTitle>Rotational stability and posterior capsule opacification of a plate-haptic and an open-loop-haptic intraocular lens.</ArticleTitle>
<Pagination>
<MedlinePgn>251-7</MedlinePgn>
</Pagination>
<ELocationID EIdType="doi" ValidYN="Y">10.1016/j.jcrs.2010.08.049</ELocationID>
<Abstract>
<AbstractText Label="PURPOSE" NlmCategory="OBJECTIVE">To compare the rotational stability and posterior capsule opacification (PCO) rate in eyes with a 1-piece or 3-piece acrylic intraocular lens (IOL).</AbstractText>
<AbstractText Label="SETTING" NlmCategory="METHODS">Department of Ophthalmology, Medical University of Vienna, Vienna, Austria.</AbstractText>
<AbstractText Label="DESIGN" NlmCategory="METHODS">Prospective randomized masked clinical trial.</AbstractText>
<AbstractText Label="METHODS" NlmCategory="METHODS">Patients with age-related cataract received a plate-haptic acrylic IOL (Acri.Smart 46S) in 1 eye and a 3-piece loop-haptic acrylic IOL (Acri.Lyc 53N) in the other eye. Retroillumination images were taken 1 hour, 1 week, and 1, 6, and 12 months postoperatively. Intraocular lens rotation was measured using standard software (Adobe Photoshop). The amount of PCO was assessed subjectively at the slitlamp and objectively using an automated image-analysis software (AQUA).</AbstractText>
<AbstractText Label="RESULTS" NlmCategory="RESULTS">The study enrolled 80 eyes of 40 patients. The IOL rotation measurements showed excellent reproducibility, with a deviation of less than 0.8 degrees. Both IOLs had comparable and good rotational stability; rotation was less than 4 degrees in 71% of eyes 1 year postoperatively. The mean absolute rotation was 2.6 degrees ± 1.9 (SD) in the plate-haptic IOL group and 3.1 ± 2.4 degrees in the loop-haptic IOL group. The mean AQUA PCO score (scale 0 to 10) was 0.4 in both IOL groups (P = .7).</AbstractText>
<AbstractText Label="CONCLUSION" NlmCategory="CONCLUSIONS">The 2 IOL models had comparable, excellent rotational stability and low PCO intensity 1 year postoperatively. Thus, the plate-haptic IOL may be a good platform for a toric model.</AbstractText>
<CopyrightInformation>Copyright © 2011 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.</CopyrightInformation>
</Abstract>
<AuthorList CompleteYN="Y">
<Author ValidYN="Y">
<LastName>Prinz</LastName>
<ForeName>Ana</ForeName>
<Initials>A</Initials>
<AffiliationInfo>
<Affiliation>Department of Ophthalmology, Medical University of Vienna, Vienna, Austria.</Affiliation>
</AffiliationInfo>
</Author>
<Author ValidYN="Y">
<LastName>Neumayer</LastName>
<ForeName>Thomas</ForeName>
<Initials>T</Initials>
</Author>
<Author ValidYN="Y">
<LastName>Buehl</LastName>
<ForeName>Wolf</ForeName>
<Initials>W</Initials>
</Author>
<Author ValidYN="Y">
<LastName>Vock</LastName>
<ForeName>Lorenz</ForeName>
<Initials>L</Initials>
</Author>
<Author ValidYN="Y">
<LastName>Menapace</LastName>
<ForeName>Rupert</ForeName>
<Initials>R</Initials>
</Author>
<Author ValidYN="Y">
<LastName>Findl</LastName>
<ForeName>Oliver</ForeName>
<Initials>O</Initials>
</Author>
<Author ValidYN="Y">
<LastName>Georgopoulos</LastName>
<ForeName>Michael</ForeName>
<Initials>M</Initials>
</Author>
</AuthorList>
<Language>eng</Language>
<PublicationTypeList>
<PublicationType UI="D003160">Comparative Study</PublicationType>
<PublicationType UI="D016428">Journal Article</PublicationType>
<PublicationType UI="D016449">Randomized Controlled Trial</PublicationType>
<PublicationType UI="D013485">Research Support, Non-U.S. Gov't</PublicationType>
</PublicationTypeList>
</Article>
<MedlineJournalInfo>
<Country>United States</Country>
<MedlineTA>J Cataract Refract Surg</MedlineTA>
<NlmUniqueID>8604171</NlmUniqueID>
<ISSNLinking>0886-3350</ISSNLinking>
</MedlineJournalInfo>
<ChemicalList>
<Chemical>
<RegistryNumber>0</RegistryNumber>
<NameOfSubstance UI="D000179">Acrylates</NameOfSubstance>
</Chemical>
</ChemicalList>
<CitationSubset>IM</CitationSubset>
<MeshHeadingList>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D000179">Acrylates</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D000368">Aged</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D058442">Capsule Opacification</DescriptorName>
<QualifierName MajorTopicYN="Y" UI="Q000209">etiology</QualifierName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D015331">Cohort Studies</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D004311">Double-Blind Method</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D005500">Follow-Up Studies</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D006801">Humans</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D007910">Lenses, Intraocular</DescriptorName>
<QualifierName MajorTopicYN="Y" UI="Q000009">adverse effects</QualifierName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D008875">Middle Aged</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D011183">Postoperative Complications</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D011446">Prospective Studies</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D011474">Prosthesis Design</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D015203">Reproducibility of Results</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D012399">Rotation</DescriptorName>
</MeshHeading>
</MeshHeadingList>
</MedlineCitation>
<PubmedData>
<History>
<PubMedPubDate PubStatus="received">
<Year>2010</Year>
<Month>7</Month>
<Day>1</Day>
</PubMedPubDate>
<PubMedPubDate PubStatus="revised">
<Year>2010</Year>
<Month>8</Month>
<Day>18</Day>
</PubMedPubDate>
<PubMedPubDate PubStatus="accepted">
<Year>2010</Year>
<Month>8</Month>
<Day>22</Day>
</PubMedPubDate>
<PubMedPubDate PubStatus="entrez">
<Year>2011</Year>
<Month>1</Month>
<Day>19</Day>
<Hour>6</Hour>
<Minute>0</Minute>
</PubMedPubDate>
<PubMedPubDate PubStatus="pubmed">
<Year>2011</Year>
<Month>1</Month>
<Day>19</Day>
<Hour>6</Hour>
<Minute>0</Minute>
</PubMedPubDate>
<PubMedPubDate PubStatus="medline">
<Year>2011</Year>
<Month>5</Month>
<Day>20</Day>
<Hour>6</Hour>
<Minute>0</Minute>
</PubMedPubDate>
</History>
<PublicationStatus>ppublish</PublicationStatus>
<ArticleIdList>
<ArticleId IdType="pii">S0886-3350(10)01723-2</ArticleId>
<ArticleId IdType="doi">10.1016/j.jcrs.2010.08.049</ArticleId>
<ArticleId IdType="pubmed">21241906</ArticleId>
</ArticleIdList>
</PubmedData>
</pubmed>
</record>

Pour manipuler ce document sous Unix (Dilib)

EXPLOR_STEP=$WICRI_ROOT/Ticri/CIDE/explor/HapticV1/Data/PubMed/Corpus
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 000F65 | SxmlIndent | more

Ou

HfdSelect -h $EXPLOR_AREA/Data/PubMed/Corpus/biblio.hfd -nk 000F65 | SxmlIndent | more

Pour mettre un lien sur cette page dans le réseau Wicri

{{Explor lien
   |wiki=    Ticri/CIDE
   |area=    HapticV1
   |flux=    PubMed
   |étape=   Corpus
   |type=    RBID
   |clé=     pubmed:21241906
   |texte=   Rotational stability and posterior capsule opacification of a plate-haptic and an open-loop-haptic intraocular lens.
}}

Pour générer des pages wiki

HfdIndexSelect -h $EXPLOR_AREA/Data/PubMed/Corpus/RBID.i   -Sk "pubmed:21241906" \
       | HfdSelect -Kh $EXPLOR_AREA/Data/PubMed/Corpus/biblio.hfd   \
       | NlmPubMed2Wicri -a HapticV1 

Wicri

This area was generated with Dilib version V0.6.23.
Data generation: Mon Jun 13 01:09:46 2016. Site generation: Wed Mar 6 09:54:07 2024