A comparison of higher order aberrations following implantation of four foldable intraocular lens designs.

Identifieur interne : 002469 ( PubMed/Checkpoint ); précédent : 002468; suivant : 002470

A comparison of higher order aberrations following implantation of four foldable intraocular lens designs.

Auteurs : Robert G. Martin [États-Unis] ; Donald R. Sanders

Source :

RBID : pubmed:16329364

English descriptors

Abstract

To determine the higher order aberrations following implantation of four foldable intraocular lens (IOL) designs.

PubMed: 16329364


Affiliations:


Links toward previous steps (curation, corpus...)


Links to Exploration step

pubmed:16329364

Le document en format XML

<record>
<TEI>
<teiHeader>
<fileDesc>
<titleStmt>
<title xml:lang="en">A comparison of higher order aberrations following implantation of four foldable intraocular lens designs.</title>
<author>
<name sortKey="Martin, Robert G" sort="Martin, Robert G" uniqKey="Martin R" first="Robert G" last="Martin">Robert G. Martin</name>
<affiliation wicri:level="2">
<nlm:affiliation>Carolina Eye Associates, Southern Pines, NC, USA.</nlm:affiliation>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Carolina Eye Associates, Southern Pines, NC</wicri:regionArea>
<placeName>
<region type="state">Caroline du Nord</region>
</placeName>
</affiliation>
</author>
<author>
<name sortKey="Sanders, Donald R" sort="Sanders, Donald R" uniqKey="Sanders D" first="Donald R" last="Sanders">Donald R. Sanders</name>
</author>
</titleStmt>
<publicationStmt>
<idno type="wicri:source">PubMed</idno>
<date when="????">
<PubDate>
<MedlineDate>2005 Nov-Dec</MedlineDate>
</PubDate>
</date>
<idno type="RBID">pubmed:16329364</idno>
<idno type="pmid">16329364</idno>
<idno type="wicri:Area/PubMed/Corpus">001880</idno>
<idno type="wicri:Area/PubMed/Curation">001880</idno>
<idno type="wicri:Area/PubMed/Checkpoint">002469</idno>
</publicationStmt>
<sourceDesc>
<biblStruct>
<analytic>
<title xml:lang="en">A comparison of higher order aberrations following implantation of four foldable intraocular lens designs.</title>
<author>
<name sortKey="Martin, Robert G" sort="Martin, Robert G" uniqKey="Martin R" first="Robert G" last="Martin">Robert G. Martin</name>
<affiliation wicri:level="2">
<nlm:affiliation>Carolina Eye Associates, Southern Pines, NC, USA.</nlm:affiliation>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Carolina Eye Associates, Southern Pines, NC</wicri:regionArea>
<placeName>
<region type="state">Caroline du Nord</region>
</placeName>
</affiliation>
</author>
<author>
<name sortKey="Sanders, Donald R" sort="Sanders, Donald R" uniqKey="Sanders D" first="Donald R" last="Sanders">Donald R. Sanders</name>
</author>
</analytic>
<series>
<title level="j">Journal of refractive surgery (Thorofare, N.J. : 1995)</title>
<idno type="ISSN">1081-597X</idno>
</series>
</biblStruct>
</sourceDesc>
</fileDesc>
<profileDesc>
<textClass>
<keywords scheme="KwdEn" xml:lang="en">
<term>Aphakia, Postcataract (surgery)</term>
<term>Astigmatism (etiology)</term>
<term>Astigmatism (physiopathology)</term>
<term>Follow-Up Studies</term>
<term>Humans</term>
<term>Lens Implantation, Intraocular</term>
<term>Lenses, Intraocular</term>
<term>Postoperative Complications</term>
<term>Prosthesis Design</term>
<term>Refraction, Ocular</term>
<term>Severity of Illness Index</term>
</keywords>
<keywords scheme="MESH" qualifier="etiology" xml:lang="en">
<term>Astigmatism</term>
</keywords>
<keywords scheme="MESH" qualifier="physiopathology" xml:lang="en">
<term>Astigmatism</term>
</keywords>
<keywords scheme="MESH" qualifier="surgery" xml:lang="en">
<term>Aphakia, Postcataract</term>
</keywords>
<keywords scheme="MESH" xml:lang="en">
<term>Follow-Up Studies</term>
<term>Humans</term>
<term>Lens Implantation, Intraocular</term>
<term>Lenses, Intraocular</term>
<term>Postoperative Complications</term>
<term>Prosthesis Design</term>
<term>Refraction, Ocular</term>
<term>Severity of Illness Index</term>
</keywords>
</textClass>
</profileDesc>
</teiHeader>
<front>
<div type="abstract" xml:lang="en">To determine the higher order aberrations following implantation of four foldable intraocular lens (IOL) designs.</div>
</front>
</TEI>
<pubmed>
<MedlineCitation Owner="NLM" Status="MEDLINE">
<PMID Version="1">16329364</PMID>
<DateCreated>
<Year>2005</Year>
<Month>12</Month>
<Day>06</Day>
</DateCreated>
<DateCompleted>
<Year>2006</Year>
<Month>01</Month>
<Day>24</Day>
</DateCompleted>
<Article PubModel="Print">
<Journal>
<ISSN IssnType="Print">1081-597X</ISSN>
<JournalIssue CitedMedium="Print">
<Volume>21</Volume>
<Issue>6</Issue>
<PubDate>
<MedlineDate>2005 Nov-Dec</MedlineDate>
</PubDate>
</JournalIssue>
<Title>Journal of refractive surgery (Thorofare, N.J. : 1995)</Title>
<ISOAbbreviation>J Refract Surg</ISOAbbreviation>
</Journal>
<ArticleTitle>A comparison of higher order aberrations following implantation of four foldable intraocular lens designs.</ArticleTitle>
<Pagination>
<MedlinePgn>716-21</MedlinePgn>
</Pagination>
<Abstract>
<AbstractText Label="PURPOSE" NlmCategory="OBJECTIVE">To determine the higher order aberrations following implantation of four foldable intraocular lens (IOL) designs.</AbstractText>
<AbstractText Label="METHODS" NlmCategory="METHODS">One hundred twenty-two eyes were randomly assigned to undergo implantation with one of four foldable IOL designs--Collamer (STAAR Surgical, Monrovia, Calif), plate haptic silicone (STAAR Surgical), AcrySof (Alcon, Ft Worth, Tex), or the Sensar (Advanced Medical Optics, Santa Ana, Calif). All enrolled eyes underwent wavefront assessment using the Tracey Visual Function Analyzer (Tracey Technologies, Houston, Tex).</AbstractText>
<AbstractText Label="RESULTS" NlmCategory="RESULTS">Preoperatively, no significant differences were noted in higher order aberrations between the four lens groups. No significant differences were noted among the plate haptic silicone, Sensar, or AcrySof lenses at 1 week or 1 month postoperatively, or among the Collamer, plate haptic silicone, or Sensar lenses, with regard to total higher order aberrations, although differences were observed in individual aberration terms. At 1 week and 1 month postoperatively, the Collamer IOL had significantly less higher order aberrations than the AcrySof lens (P < .01). Significantly less third and fourth order aberrations, coma, trefoil, spherical aberration, and tetrafoil were observed postoperatively with the Collamer than the AcrySof IOL. There were no parameters where the AcrySof had significantly less optical aberrations than the Collamer.</AbstractText>
<AbstractText Label="CONCLUSIONS" NlmCategory="CONCLUSIONS">The Collamer IOL design is associated with fewer higher order aberrations than the AcrySof lens.</AbstractText>
</Abstract>
<AuthorList CompleteYN="Y">
<Author ValidYN="Y">
<LastName>Martin</LastName>
<ForeName>Robert G</ForeName>
<Initials>RG</Initials>
<AffiliationInfo>
<Affiliation>Carolina Eye Associates, Southern Pines, NC, USA.</Affiliation>
</AffiliationInfo>
</Author>
<Author ValidYN="Y">
<LastName>Sanders</LastName>
<ForeName>Donald R</ForeName>
<Initials>DR</Initials>
</Author>
</AuthorList>
<Language>eng</Language>
<PublicationTypeList>
<PublicationType UI="D016428">Journal Article</PublicationType>
<PublicationType UI="D016449">Randomized Controlled Trial</PublicationType>
</PublicationTypeList>
</Article>
<MedlineJournalInfo>
<Country>United States</Country>
<MedlineTA>J Refract Surg</MedlineTA>
<NlmUniqueID>9505927</NlmUniqueID>
<ISSNLinking>1081-597X</ISSNLinking>
</MedlineJournalInfo>
<CitationSubset>IM</CitationSubset>
<MeshHeadingList>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D001036">Aphakia, Postcataract</DescriptorName>
<QualifierName MajorTopicYN="Y" UI="Q000601">surgery</QualifierName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D001251">Astigmatism</DescriptorName>
<QualifierName MajorTopicYN="N" UI="Q000209">etiology</QualifierName>
<QualifierName MajorTopicYN="Y" UI="Q000503">physiopathology</QualifierName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D005500">Follow-Up Studies</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D006801">Humans</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="Y" UI="D019654">Lens Implantation, Intraocular</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="Y" UI="D007910">Lenses, Intraocular</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D011183">Postoperative Complications</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D011474">Prosthesis Design</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="Y" UI="D012029">Refraction, Ocular</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName MajorTopicYN="N" UI="D012720">Severity of Illness Index</DescriptorName>
</MeshHeading>
</MeshHeadingList>
</MedlineCitation>
<PubmedData>
<History>
<PubMedPubDate PubStatus="pubmed">
<Year>2005</Year>
<Month>12</Month>
<Day>7</Day>
<Hour>9</Hour>
<Minute>0</Minute>
</PubMedPubDate>
<PubMedPubDate PubStatus="medline">
<Year>2006</Year>
<Month>1</Month>
<Day>25</Day>
<Hour>9</Hour>
<Minute>0</Minute>
</PubMedPubDate>
<PubMedPubDate PubStatus="entrez">
<Year>2005</Year>
<Month>12</Month>
<Day>7</Day>
<Hour>9</Hour>
<Minute>0</Minute>
</PubMedPubDate>
</History>
<PublicationStatus>ppublish</PublicationStatus>
<ArticleIdList>
<ArticleId IdType="pubmed">16329364</ArticleId>
</ArticleIdList>
</PubmedData>
</pubmed>
<affiliations>
<list>
<country>
<li>États-Unis</li>
</country>
<region>
<li>Caroline du Nord</li>
</region>
</list>
<tree>
<noCountry>
<name sortKey="Sanders, Donald R" sort="Sanders, Donald R" uniqKey="Sanders D" first="Donald R" last="Sanders">Donald R. Sanders</name>
</noCountry>
<country name="États-Unis">
<region name="Caroline du Nord">
<name sortKey="Martin, Robert G" sort="Martin, Robert G" uniqKey="Martin R" first="Robert G" last="Martin">Robert G. Martin</name>
</region>
</country>
</tree>
</affiliations>
</record>

Pour manipuler ce document sous Unix (Dilib)

EXPLOR_STEP=$WICRI_ROOT/Ticri/CIDE/explor/HapticV1/Data/PubMed/Checkpoint
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 002469 | SxmlIndent | more

Ou

HfdSelect -h $EXPLOR_AREA/Data/PubMed/Checkpoint/biblio.hfd -nk 002469 | SxmlIndent | more

Pour mettre un lien sur cette page dans le réseau Wicri

{{Explor lien
   |wiki=    Ticri/CIDE
   |area=    HapticV1
   |flux=    PubMed
   |étape=   Checkpoint
   |type=    RBID
   |clé=     pubmed:16329364
   |texte=   A comparison of higher order aberrations following implantation of four foldable intraocular lens designs.
}}

Pour générer des pages wiki

HfdIndexSelect -h $EXPLOR_AREA/Data/PubMed/Checkpoint/RBID.i   -Sk "pubmed:16329364" \
       | HfdSelect -Kh $EXPLOR_AREA/Data/PubMed/Checkpoint/biblio.hfd   \
       | NlmPubMed2Wicri -a HapticV1 

Wicri

This area was generated with Dilib version V0.6.23.
Data generation: Mon Jun 13 01:09:46 2016. Site generation: Wed Mar 6 09:54:07 2024