Abstract

Purpose

We tested the feasibility of using titanium to enhance adhesion of the Boston Keratoprosthesis (B-KPro), ultimately to decrease the risk of implant-associated complications.

Methods

Cylindrical rods were made of poly(methyl methacrylate) (PMMA), PMMA coated with titanium dioxide (TiO2) over a layer of polydopamine (PMMATiO2), smooth (Ti) and sandblasted (TiSB) titanium, and titanium treated with oxygen plasma (Tiox and TiSBox). Topography and surface chemistry were analyzed by scanning electron microscopy (SEM), atomic force microscopy (AFM), and X-ray photoelectron spectroscopy (XPS). Adhesion force between rods and porcine corneas was measured ex vivo. Titanium sleeves, smooth and sandblasted, were inserted around the stem of the B-KPro and implanted in rabbits. Tissue adhesion to the stem was assessed and compared to an unmodified B-Kpro after 1 month.

Results

X-ray photoelectron spectroscopy demonstrated successful deposition of TiO2 on polydopamine-coated PMMA. Oxygen plasma treatment did not change the XPS spectra of titanium rods (Ti and TiSB), although it increased their hydrophilicity. The materials did not show cell toxicity. After 14 days of incubation, PMMATiO2, smooth titanium treated with oxygen plasma (Tiox), and sandblasted titanium rods (TiSB, TiSBox) showed significantly higher adhesion forces than PMMA ex vivo. In vivo, the use of a TiSB sleeve around the stem of the B-KPro induced a significant increase in tissue adhesion compared to a Ti sleeve or bare PMMA.

Conclusions

Sandblasted titanium sleeves greatly enhanced adherence of the B-KPro to the rabbit cornea. This approach may improve adhesion with the donor cornea in humans as well.

Translational Relevance

This approach may improve adhesion with donor corneas in humans.


Url:
DOI: 10.1167/tvst.5.2.17
PubMed: 27152247
PubMed Central: 4855478

Links toward previous steps (curation, corpus...)


Links to Exploration step

PMC:4855478

Le document en format XML

<record>
<TEI>
<teiHeader>
<fileDesc>
<titleStmt>
<title xml:lang="en">Titanium Coating of the Boston Keratoprosthesis</title>
<author>
<name sortKey="Salvador Culla, Borja" sort="Salvador Culla, Borja" uniqKey="Salvador Culla B" first="Borja" last="Salvador-Culla">Borja Salvador-Culla</name>
<affiliation wicri:level="1">
<nlm:aff id="aff1">Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA</wicri:regionArea>
</affiliation>
<affiliation wicri:level="1">
<nlm:aff id="aff2">Laboratory for Biomaterials and Drug Delivery, Department of Anesthesiology, Division of Critical Care Medicine, Children's Hospital Boston, Harvard Medical School, Boston, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Laboratory for Biomaterials and Drug Delivery, Department of Anesthesiology, Division of Critical Care Medicine, Children's Hospital Boston, Harvard Medical School, Boston, MA</wicri:regionArea>
</affiliation>
<affiliation wicri:level="1">
<nlm:aff id="aff3">David H. Koch Institute for Integrative Cancer Research, Massachusetts Institute of Technology, Cambridge, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>David H. Koch Institute for Integrative Cancer Research, Massachusetts Institute of Technology, Cambridge, MA</wicri:regionArea>
</affiliation>
</author>
<author>
<name sortKey="Jeong, Kyung Jae" sort="Jeong, Kyung Jae" uniqKey="Jeong K" first="Kyung Jae" last="Jeong">Kyung Jae Jeong</name>
<affiliation wicri:level="1">
<nlm:aff id="aff4">Department of Chemical Engineering, University of New Hampshire, Durham, NH, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Department of Chemical Engineering, University of New Hampshire, Durham, NH</wicri:regionArea>
</affiliation>
</author>
<author>
<name sortKey="Kolovou, Paraskevi Evi" sort="Kolovou, Paraskevi Evi" uniqKey="Kolovou P" first="Paraskevi Evi" last="Kolovou">Paraskevi Evi Kolovou</name>
<affiliation wicri:level="1">
<nlm:aff id="aff1">Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA</wicri:regionArea>
</affiliation>
</author>
<author>
<name sortKey="Chiang, Homer H" sort="Chiang, Homer H" uniqKey="Chiang H" first="Homer H." last="Chiang">Homer H. Chiang</name>
<affiliation wicri:level="1">
<nlm:aff id="aff2">Laboratory for Biomaterials and Drug Delivery, Department of Anesthesiology, Division of Critical Care Medicine, Children's Hospital Boston, Harvard Medical School, Boston, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Laboratory for Biomaterials and Drug Delivery, Department of Anesthesiology, Division of Critical Care Medicine, Children's Hospital Boston, Harvard Medical School, Boston, MA</wicri:regionArea>
</affiliation>
<affiliation wicri:level="1">
<nlm:aff id="aff3">David H. Koch Institute for Integrative Cancer Research, Massachusetts Institute of Technology, Cambridge, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>David H. Koch Institute for Integrative Cancer Research, Massachusetts Institute of Technology, Cambridge, MA</wicri:regionArea>
</affiliation>
</author>
<author>
<name sortKey="Chodosh, James" sort="Chodosh, James" uniqKey="Chodosh J" first="James" last="Chodosh">James Chodosh</name>
<affiliation wicri:level="1">
<nlm:aff id="aff1">Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA</wicri:regionArea>
</affiliation>
</author>
<author>
<name sortKey="Dohlman, Claes H" sort="Dohlman, Claes H" uniqKey="Dohlman C" first="Claes H." last="Dohlman">Claes H. Dohlman</name>
<affiliation wicri:level="1">
<nlm:aff id="aff1">Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA</wicri:regionArea>
</affiliation>
</author>
<author>
<name sortKey="Kohane, Daniel S" sort="Kohane, Daniel S" uniqKey="Kohane D" first="Daniel S." last="Kohane">Daniel S. Kohane</name>
<affiliation wicri:level="1">
<nlm:aff id="aff2">Laboratory for Biomaterials and Drug Delivery, Department of Anesthesiology, Division of Critical Care Medicine, Children's Hospital Boston, Harvard Medical School, Boston, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Laboratory for Biomaterials and Drug Delivery, Department of Anesthesiology, Division of Critical Care Medicine, Children's Hospital Boston, Harvard Medical School, Boston, MA</wicri:regionArea>
</affiliation>
<affiliation>
<nlm:aff>NONE</nlm:aff>
</affiliation>
</author>
</titleStmt>
<publicationStmt>
<idno type="wicri:source">PMC</idno>
<idno type="pmid">27152247</idno>
<idno type="pmc">4855478</idno>
<idno type="url">http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4855478</idno>
<idno type="RBID">PMC:4855478</idno>
<idno type="doi">10.1167/tvst.5.2.17</idno>
<date when="2016">2016</date>
<idno type="wicri:Area/Pmc/Corpus">000580</idno>
<idno type="wicri:Area/Pmc/Curation">000580</idno>
</publicationStmt>
<sourceDesc>
<biblStruct>
<analytic>
<title xml:lang="en" level="a" type="main">Titanium Coating of the Boston Keratoprosthesis</title>
<author>
<name sortKey="Salvador Culla, Borja" sort="Salvador Culla, Borja" uniqKey="Salvador Culla B" first="Borja" last="Salvador-Culla">Borja Salvador-Culla</name>
<affiliation wicri:level="1">
<nlm:aff id="aff1">Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA</wicri:regionArea>
</affiliation>
<affiliation wicri:level="1">
<nlm:aff id="aff2">Laboratory for Biomaterials and Drug Delivery, Department of Anesthesiology, Division of Critical Care Medicine, Children's Hospital Boston, Harvard Medical School, Boston, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Laboratory for Biomaterials and Drug Delivery, Department of Anesthesiology, Division of Critical Care Medicine, Children's Hospital Boston, Harvard Medical School, Boston, MA</wicri:regionArea>
</affiliation>
<affiliation wicri:level="1">
<nlm:aff id="aff3">David H. Koch Institute for Integrative Cancer Research, Massachusetts Institute of Technology, Cambridge, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>David H. Koch Institute for Integrative Cancer Research, Massachusetts Institute of Technology, Cambridge, MA</wicri:regionArea>
</affiliation>
</author>
<author>
<name sortKey="Jeong, Kyung Jae" sort="Jeong, Kyung Jae" uniqKey="Jeong K" first="Kyung Jae" last="Jeong">Kyung Jae Jeong</name>
<affiliation wicri:level="1">
<nlm:aff id="aff4">Department of Chemical Engineering, University of New Hampshire, Durham, NH, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Department of Chemical Engineering, University of New Hampshire, Durham, NH</wicri:regionArea>
</affiliation>
</author>
<author>
<name sortKey="Kolovou, Paraskevi Evi" sort="Kolovou, Paraskevi Evi" uniqKey="Kolovou P" first="Paraskevi Evi" last="Kolovou">Paraskevi Evi Kolovou</name>
<affiliation wicri:level="1">
<nlm:aff id="aff1">Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA</wicri:regionArea>
</affiliation>
</author>
<author>
<name sortKey="Chiang, Homer H" sort="Chiang, Homer H" uniqKey="Chiang H" first="Homer H." last="Chiang">Homer H. Chiang</name>
<affiliation wicri:level="1">
<nlm:aff id="aff2">Laboratory for Biomaterials and Drug Delivery, Department of Anesthesiology, Division of Critical Care Medicine, Children's Hospital Boston, Harvard Medical School, Boston, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Laboratory for Biomaterials and Drug Delivery, Department of Anesthesiology, Division of Critical Care Medicine, Children's Hospital Boston, Harvard Medical School, Boston, MA</wicri:regionArea>
</affiliation>
<affiliation wicri:level="1">
<nlm:aff id="aff3">David H. Koch Institute for Integrative Cancer Research, Massachusetts Institute of Technology, Cambridge, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>David H. Koch Institute for Integrative Cancer Research, Massachusetts Institute of Technology, Cambridge, MA</wicri:regionArea>
</affiliation>
</author>
<author>
<name sortKey="Chodosh, James" sort="Chodosh, James" uniqKey="Chodosh J" first="James" last="Chodosh">James Chodosh</name>
<affiliation wicri:level="1">
<nlm:aff id="aff1">Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA</wicri:regionArea>
</affiliation>
</author>
<author>
<name sortKey="Dohlman, Claes H" sort="Dohlman, Claes H" uniqKey="Dohlman C" first="Claes H." last="Dohlman">Claes H. Dohlman</name>
<affiliation wicri:level="1">
<nlm:aff id="aff1">Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA</wicri:regionArea>
</affiliation>
</author>
<author>
<name sortKey="Kohane, Daniel S" sort="Kohane, Daniel S" uniqKey="Kohane D" first="Daniel S." last="Kohane">Daniel S. Kohane</name>
<affiliation wicri:level="1">
<nlm:aff id="aff2">Laboratory for Biomaterials and Drug Delivery, Department of Anesthesiology, Division of Critical Care Medicine, Children's Hospital Boston, Harvard Medical School, Boston, MA, USA</nlm:aff>
<country xml:lang="fr">États-Unis</country>
<wicri:regionArea>Laboratory for Biomaterials and Drug Delivery, Department of Anesthesiology, Division of Critical Care Medicine, Children's Hospital Boston, Harvard Medical School, Boston, MA</wicri:regionArea>
</affiliation>
<affiliation>
<nlm:aff>NONE</nlm:aff>
</affiliation>
</author>
</analytic>
<series>
<title level="j">Translational Vision Science & Technology</title>
<idno type="eISSN">2164-2591</idno>
<imprint>
<date when="2016">2016</date>
</imprint>
</series>
</biblStruct>
</sourceDesc>
</fileDesc>
<profileDesc>
<textClass></textClass>
</profileDesc>
</teiHeader>
<front>
<div type="abstract" xml:lang="en">
<sec id="st1">
<title>Purpose</title>
<p>We tested the feasibility of using titanium to enhance adhesion of the Boston Keratoprosthesis (B-KPro), ultimately to decrease the risk of implant-associated complications.</p>
</sec>
<sec id="st2">
<title>Methods</title>
<p>Cylindrical rods were made of poly(methyl methacrylate) (PMMA), PMMA coated with titanium dioxide (TiO
<sub>2</sub>
) over a layer of polydopamine (PMMA
<sub>TiO2</sub>
), smooth (Ti) and sandblasted (Ti
<sub>SB</sub>
) titanium, and titanium treated with oxygen plasma (Ti
<sub>ox</sub>
and Ti
<sub>SBox</sub>
). Topography and surface chemistry were analyzed by scanning electron microscopy (SEM), atomic force microscopy (AFM), and X-ray photoelectron spectroscopy (XPS). Adhesion force between rods and porcine corneas was measured ex vivo. Titanium sleeves, smooth and sandblasted, were inserted around the stem of the B-KPro and implanted in rabbits. Tissue adhesion to the stem was assessed and compared to an unmodified B-Kpro after 1 month.</p>
</sec>
<sec id="st3">
<title>Results</title>
<p>X-ray photoelectron spectroscopy demonstrated successful deposition of TiO
<sub>2</sub>
on polydopamine-coated PMMA. Oxygen plasma treatment did not change the XPS spectra of titanium rods (Ti and Ti
<sub>SB</sub>
), although it increased their hydrophilicity. The materials did not show cell toxicity. After 14 days of incubation, PMMA
<sub>TiO2</sub>
, smooth titanium treated with oxygen plasma (Ti
<sub>ox</sub>
), and sandblasted titanium rods (Ti
<sub>SB</sub>
, Ti
<sub>SBox</sub>
) showed significantly higher adhesion forces than PMMA ex vivo. In vivo, the use of a Ti
<sub>SB</sub>
sleeve around the stem of the B-KPro induced a significant increase in tissue adhesion compared to a Ti sleeve or bare PMMA.</p>
</sec>
<sec id="st4">
<title>Conclusions</title>
<p>Sandblasted titanium sleeves greatly enhanced adherence of the B-KPro to the rabbit cornea. This approach may improve adhesion with the donor cornea in humans as well.</p>
</sec>
<sec id="st5">
<title>Translational Relevance</title>
<p>This approach may improve adhesion with donor corneas in humans.</p>
</sec>
</div>
</front>
<back>
<div1 type="bibliography">
<listBibl>
<biblStruct>
<analytic>
<author>
<name sortKey="Dohlman, Ch" uniqKey="Dohlman C">CH, Dohlman</name>
</author>
<author>
<name sortKey="Schneider, Ha" uniqKey="Schneider H">HA, Schneider</name>
</author>
<author>
<name sortKey="Doane, Mg" uniqKey="Doane M">MG. Doane</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Todani, A" uniqKey="Todani A">A, Todani</name>
</author>
<author>
<name sortKey="Ciolino, Jb" uniqKey="Ciolino J">JB, Ciolino</name>
</author>
<author>
<name sortKey="Ament, Jd" uniqKey="Ament J">JD, Ament</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Doane, Mg" uniqKey="Doane M">MG, Doane</name>
</author>
<author>
<name sortKey="Dohlman, Ch" uniqKey="Dohlman C">CH, Dohlman</name>
</author>
<author>
<name sortKey="Bearse, G" uniqKey="Bearse G">G. Bearse</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Colby, Ka" uniqKey="Colby K">KA, Colby</name>
</author>
<author>
<name sortKey="Koo, Eb" uniqKey="Koo E">EB. Koo</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Shapiro, Bl" uniqKey="Shapiro B">BL, Shapiro</name>
</author>
<author>
<name sortKey="Cortes, De" uniqKey="Cortes D">DE, Cortés</name>
</author>
<author>
<name sortKey="Chin, Ek" uniqKey="Chin E">EK, Chin</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Behlau, I" uniqKey="Behlau I">I, Behlau</name>
</author>
<author>
<name sortKey="Martin, Kv" uniqKey="Martin K">KV, Martin</name>
</author>
<author>
<name sortKey="Martin, Jn" uniqKey="Martin J">JN, Martin</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Chan, Cc" uniqKey="Chan C">CC, Chan</name>
</author>
<author>
<name sortKey="Holland, Ej" uniqKey="Holland E">EJ. Holland</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Henry, Pj" uniqKey="Henry P">PJ. Henry</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Theis, Jc" uniqKey="Theis J">JC, Theis</name>
</author>
<author>
<name sortKey="Gambhir, S" uniqKey="Gambhir S">S, Gambhir</name>
</author>
<author>
<name sortKey="White, J" uniqKey="White J">J. White</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Hacking, Sa" uniqKey="Hacking S">SA, Hacking</name>
</author>
<author>
<name sortKey="Pauyo, T" uniqKey="Pauyo T">T, Pauyo</name>
</author>
<author>
<name sortKey="Lim, L" uniqKey="Lim L">L, Lim</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Branemark, Pi" uniqKey="Branemark P">PI, Branemark</name>
</author>
<author>
<name sortKey="Hansson, Bo" uniqKey="Hansson B">BO, Hansson</name>
</author>
<author>
<name sortKey="Adell, R" uniqKey="Adell R">R, Adell</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Baril, E" uniqKey="Baril E">E, Baril</name>
</author>
<author>
<name sortKey="Lefebvre, Lp" uniqKey="Lefebvre L">LP, Lefebvre</name>
</author>
<author>
<name sortKey="Hacking, Sa" uniqKey="Hacking S">SA. Hacking</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Tan, Xw" uniqKey="Tan X">XW, Tan</name>
</author>
<author>
<name sortKey="Perera, Ap" uniqKey="Perera A">AP, Perera</name>
</author>
<author>
<name sortKey="Tan, A" uniqKey="Tan A">A, Tan</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Tan, Xw" uniqKey="Tan X">XW, Tan</name>
</author>
<author>
<name sortKey="Beuerman, Rw" uniqKey="Beuerman R">RW, Beuerman</name>
</author>
<author>
<name sortKey="Shi, Zl" uniqKey="Shi Z">ZL, Shi</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Hacking, Sa" uniqKey="Hacking S">SA, Hacking</name>
</author>
<author>
<name sortKey="Harvey, E" uniqKey="Harvey E">E, Harvey</name>
</author>
<author>
<name sortKey="Roughley, P" uniqKey="Roughley P">P, Roughley</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Hacking, Sa" uniqKey="Hacking S">SA, Hacking</name>
</author>
<author>
<name sortKey="Boyraz, P" uniqKey="Boyraz P">P, Boyraz</name>
</author>
<author>
<name sortKey="Powers, Bm" uniqKey="Powers B">BM, Powers</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Paschalis, Ei" uniqKey="Paschalis E">EI, Paschalis</name>
</author>
<author>
<name sortKey="Chodosh, J" uniqKey="Chodosh J">J, Chodosh</name>
</author>
<author>
<name sortKey="Spurr Michaud, S" uniqKey="Spurr Michaud S">S, Spurr-Michaud</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Ament, Jd" uniqKey="Ament J">JD, Ament</name>
</author>
<author>
<name sortKey="Spurr Michaud, Sj" uniqKey="Spurr Michaud S">SJ, Spurr-Michaud</name>
</author>
<author>
<name sortKey="Dohlman, Ch" uniqKey="Dohlman C">CH, Dohlman</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Salvador Culla, B" uniqKey="Salvador Culla B">B, Salvador-Culla</name>
</author>
<author>
<name sortKey="Dohlman, Ch" uniqKey="Dohlman C">CH. Dohlman</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Linnola, Rj" uniqKey="Linnola R">RJ, Linnola</name>
</author>
<author>
<name sortKey="Happonen, Rp" uniqKey="Happonen R">RP, Happonen</name>
</author>
<author>
<name sortKey="Andersson, Oh" uniqKey="Andersson O">OH, Andersson</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Aronsson, Bo" uniqKey="Aronsson B">BO, Aronsson</name>
</author>
<author>
<name sortKey="Lausmaa, J" uniqKey="Lausmaa J">J, Lausmaa</name>
</author>
<author>
<name sortKey="Kasemo, B" uniqKey="Kasemo B">B. Kasemo</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Ou, J" uniqKey="Ou J">J, Ou</name>
</author>
<author>
<name sortKey="Wang, J" uniqKey="Wang J">J, Wang</name>
</author>
<author>
<name sortKey="Zhang, D" uniqKey="Zhang D">D, Zhang</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Wang, L" uniqKey="Wang L">L, Wang</name>
</author>
<author>
<name sortKey="Jeong, Kj" uniqKey="Jeong K">KJ, Jeong</name>
</author>
<author>
<name sortKey="Chiang, Hh" uniqKey="Chiang H">HH, Chiang</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Chen, Ch" uniqKey="Chen C">CH, Chen</name>
</author>
<author>
<name sortKey="Chen, Sc" uniqKey="Chen S">SC. Chen</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Chen, Ch" uniqKey="Chen C">CH, Chen</name>
</author>
<author>
<name sortKey="Rama, P" uniqKey="Rama P">P, Rama</name>
</author>
<author>
<name sortKey="Chen, Sc" uniqKey="Chen S">SC, Chen</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Wagner, D" uniqKey="Wagner D">D, Wagner</name>
</author>
<author>
<name sortKey="Naumkin, Av" uniqKey="Naumkin A">AV, Naumkin</name>
</author>
<author>
<name sortKey="Kraut Vass, A" uniqKey="Kraut Vass A">A, Kraut-Vass</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Martin, Jy" uniqKey="Martin J">JY, Martin</name>
</author>
<author>
<name sortKey="Schwartz, Z" uniqKey="Schwartz Z">Z, Schwartz</name>
</author>
<author>
<name sortKey="Hummert, Tw" uniqKey="Hummert T">TW, Hummert</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Kim, Mj" uniqKey="Kim M">MJ, Kim</name>
</author>
<author>
<name sortKey="Kim, Cw" uniqKey="Kim C">CW, Kim</name>
</author>
<author>
<name sortKey="Lim, Yj" uniqKey="Lim Y">YJ, Lim</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Jeong, Kj" uniqKey="Jeong K">KJ, Jeong</name>
</author>
<author>
<name sortKey="Kohane, Ds" uniqKey="Kohane D">DS. Kohane</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Crnej, A" uniqKey="Crnej A">A, Crnej</name>
</author>
<author>
<name sortKey="Paschalis, Ei" uniqKey="Paschalis E">EI, Paschalis</name>
</author>
<author>
<name sortKey="Salvador Culla, B" uniqKey="Salvador Culla B">B, Salvador-Culla</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Zerbe, Bl" uniqKey="Zerbe B">BL, Zerbe</name>
</author>
<author>
<name sortKey="Belin, Mw" uniqKey="Belin M">MW, Belin</name>
</author>
<author>
<name sortKey="Ciolino, Jb" uniqKey="Ciolino J">JB, Ciolino</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Aldave, Aj" uniqKey="Aldave A">AJ, Aldave</name>
</author>
<author>
<name sortKey="Sangwan, Vs" uniqKey="Sangwan V">VS, Sangwan</name>
</author>
<author>
<name sortKey="Basu, S" uniqKey="Basu S">S, Basu</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Sivaraman, Kr" uniqKey="Sivaraman K">KR, Sivaraman</name>
</author>
<author>
<name sortKey="Hou, Jh" uniqKey="Hou J">JH, Hou</name>
</author>
<author>
<name sortKey="Allemann, N" uniqKey="Allemann N">N, Allemann</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Srikumaran, D" uniqKey="Srikumaran D">D, Srikumaran</name>
</author>
<author>
<name sortKey="Munoz, B" uniqKey="Munoz B">B, Munoz</name>
</author>
<author>
<name sortKey="Aldave, Aj" uniqKey="Aldave A">AJ, Aldave</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Ciolino, Jb" uniqKey="Ciolino J">JB, Ciolino</name>
</author>
<author>
<name sortKey="Belin, Mw" uniqKey="Belin M">MW, Belin</name>
</author>
<author>
<name sortKey="Todani, A" uniqKey="Todani A">A, Todani</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Durand, Ml" uniqKey="Durand M">ML, Durand</name>
</author>
<author>
<name sortKey="Dohlman, Ch" uniqKey="Dohlman C">CH. Dohlman</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Jain, V" uniqKey="Jain V">V, Jain</name>
</author>
<author>
<name sortKey="Mhatre, K" uniqKey="Mhatre K">K, Mhatre</name>
</author>
<author>
<name sortKey="Shome, D" uniqKey="Shome D">D, Shome</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Andrei, C" uniqKey="Andrei C">C, Andrei</name>
</author>
<author>
<name sortKey="Lestini, E" uniqKey="Lestini E">E, Lestini</name>
</author>
<author>
<name sortKey="Crosbie, S" uniqKey="Crosbie S">S, Crosbie</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Terriza, A" uniqKey="Terriza A">A, Terriza</name>
</author>
<author>
<name sortKey="Diaz Cuenca, A" uniqKey="Diaz Cuenca A">A, Díaz-Cuenca</name>
</author>
<author>
<name sortKey="Yubero, F" uniqKey="Yubero F">F, Yubero</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Teixeira, Ai" uniqKey="Teixeira A">AI, Teixeira</name>
</author>
<author>
<name sortKey="Nealey, Pf" uniqKey="Nealey P">PF, Nealey</name>
</author>
<author>
<name sortKey="Murphy, Cj" uniqKey="Murphy C">CJ. Murphy</name>
</author>
</analytic>
</biblStruct>
</listBibl>
</div1>
</back>
</TEI>
<pmc article-type="research-article">
<pmc-dir>properties open_access</pmc-dir>
<front>
<journal-meta>
<journal-id journal-id-type="nlm-ta">Transl Vis Sci Technol</journal-id>
<journal-id journal-id-type="iso-abbrev">Transl Vis Sci Technol</journal-id>
<journal-id journal-id-type="hwp">tvst</journal-id>
<journal-id journal-id-type="pmc">tvst</journal-id>
<journal-id journal-id-type="publisher-id">tvst</journal-id>
<journal-title-group>
<journal-title>Translational Vision Science & Technology</journal-title>
</journal-title-group>
<issn pub-type="epub">2164-2591</issn>
<publisher>
<publisher-name>The Association for Research in Vision and Ophthalmology</publisher-name>
</publisher>
</journal-meta>
<article-meta>
<article-id pub-id-type="pmid">27152247</article-id>
<article-id pub-id-type="pmc">4855478</article-id>
<article-id pub-id-type="doi">10.1167/tvst.5.2.17</article-id>
<article-id pub-id-type="sici">tvst-05-02-18</article-id>
<article-id pub-id-type="publisher-id">MS#: TVST-15-0287</article-id>
<article-categories>
<subj-group subj-group-type="heading">
<subject>Articles</subject>
</subj-group>
</article-categories>
<title-group>
<article-title>Titanium Coating of the Boston Keratoprosthesis</article-title>
</title-group>
<contrib-group>
<contrib contrib-type="author">
<name>
<surname>Salvador-Culla</surname>
<given-names>Borja</given-names>
</name>
<xref ref-type="aff" rid="aff1">
<sup>1</sup>
</xref>
<xref ref-type="aff" rid="aff2">
<sup>2</sup>
</xref>
<xref ref-type="aff" rid="aff3">
<sup>3</sup>
</xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Jeong</surname>
<given-names>Kyung Jae</given-names>
</name>
<xref ref-type="aff" rid="aff4">
<sup>4</sup>
</xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Kolovou</surname>
<given-names>Paraskevi Evi</given-names>
</name>
<xref ref-type="aff" rid="aff1">
<sup>1</sup>
</xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Chiang</surname>
<given-names>Homer H.</given-names>
</name>
<xref ref-type="aff" rid="aff2">
<sup>2</sup>
</xref>
<xref ref-type="aff" rid="aff3">
<sup>3</sup>
</xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Chodosh</surname>
<given-names>James</given-names>
</name>
<xref ref-type="aff" rid="aff1">
<sup>1</sup>
</xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Dohlman</surname>
<given-names>Claes H.</given-names>
</name>
<xref ref-type="aff" rid="aff1">
<sup>1</sup>
</xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Kohane</surname>
<given-names>Daniel S.</given-names>
</name>
<xref ref-type="aff" rid="aff2">
<sup>2</sup>
</xref>
<xref ref-type="aff" rid="aff1"></xref>
<xref ref-type="author-notes" rid="n103"></xref>
</contrib>
<aff id="aff1">
<label>1</label>
Department of Ophthalmology Massachusetts Eye & Ear Infirmary, Harvard Medical School, Boston, MA, USA</aff>
<aff id="aff2">
<label>2</label>
Laboratory for Biomaterials and Drug Delivery, Department of Anesthesiology, Division of Critical Care Medicine, Children's Hospital Boston, Harvard Medical School, Boston, MA, USA</aff>
<aff id="aff3">
<label>3</label>
David H. Koch Institute for Integrative Cancer Research, Massachusetts Institute of Technology, Cambridge, MA, USA</aff>
<aff id="aff4">
<label>4</label>
Department of Chemical Engineering, University of New Hampshire, Durham, NH, USA</aff>
</contrib-group>
<author-notes>
<corresp id="cor1">
<bold>Correspondence:</bold>
Daniel S. Kohane, Boston Children's Hospital, 300 Longwood Avenue, Enders 361, Boston, MA 02115 USA. e-mail:
<email>Daniel.Kohane@childrens.harvard.edu</email>
</corresp>
<fn id="n103" fn-type="current-aff">
<p>
<email>Daniel.Kohane@childrens.harvard.edu</email>
</p>
</fn>
</author-notes>
<pub-date pub-type="epub">
<day>28</day>
<month>4</month>
<year>2016</year>
</pub-date>
<pub-date pub-type="collection">
<month>4</month>
<year>2016</year>
</pub-date>
<volume>5</volume>
<issue>2</issue>
<elocation-id>17</elocation-id>
<history>
<date date-type="received">
<day>28</day>
<month>9</month>
<year>2015</year>
</date>
<date date-type="accepted">
<day>29</day>
<month>2</month>
<year>2016</year>
</date>
</history>
<permissions>
<license license-type="cc-by-nc-nd" xlink:href="http://creativecommons.org/licenses/by-nc-nd/4.0/">
<license-p>This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.</license-p>
</license>
</permissions>
<self-uri xlink:title="pdf" xlink:href="i2164-2591-5-2-17.pdf"></self-uri>
<abstract>
<sec id="st1">
<title>Purpose</title>
<p>We tested the feasibility of using titanium to enhance adhesion of the Boston Keratoprosthesis (B-KPro), ultimately to decrease the risk of implant-associated complications.</p>
</sec>
<sec id="st2">
<title>Methods</title>
<p>Cylindrical rods were made of poly(methyl methacrylate) (PMMA), PMMA coated with titanium dioxide (TiO
<sub>2</sub>
) over a layer of polydopamine (PMMA
<sub>TiO2</sub>
), smooth (Ti) and sandblasted (Ti
<sub>SB</sub>
) titanium, and titanium treated with oxygen plasma (Ti
<sub>ox</sub>
and Ti
<sub>SBox</sub>
). Topography and surface chemistry were analyzed by scanning electron microscopy (SEM), atomic force microscopy (AFM), and X-ray photoelectron spectroscopy (XPS). Adhesion force between rods and porcine corneas was measured ex vivo. Titanium sleeves, smooth and sandblasted, were inserted around the stem of the B-KPro and implanted in rabbits. Tissue adhesion to the stem was assessed and compared to an unmodified B-Kpro after 1 month.</p>
</sec>
<sec id="st3">
<title>Results</title>
<p>X-ray photoelectron spectroscopy demonstrated successful deposition of TiO
<sub>2</sub>
on polydopamine-coated PMMA. Oxygen plasma treatment did not change the XPS spectra of titanium rods (Ti and Ti
<sub>SB</sub>
), although it increased their hydrophilicity. The materials did not show cell toxicity. After 14 days of incubation, PMMA
<sub>TiO2</sub>
, smooth titanium treated with oxygen plasma (Ti
<sub>ox</sub>
), and sandblasted titanium rods (Ti
<sub>SB</sub>
, Ti
<sub>SBox</sub>
) showed significantly higher adhesion forces than PMMA ex vivo. In vivo, the use of a Ti
<sub>SB</sub>
sleeve around the stem of the B-KPro induced a significant increase in tissue adhesion compared to a Ti sleeve or bare PMMA.</p>
</sec>
<sec id="st4">
<title>Conclusions</title>
<p>Sandblasted titanium sleeves greatly enhanced adherence of the B-KPro to the rabbit cornea. This approach may improve adhesion with the donor cornea in humans as well.</p>
</sec>
<sec id="st5">
<title>Translational Relevance</title>
<p>This approach may improve adhesion with donor corneas in humans.</p>
</sec>
</abstract>
<kwd-group>
<title>Keywords</title>
<kwd>keratoprosthesis</kwd>
<kwd>Boston Keratoprosthesis</kwd>
<kwd>artificial cornea</kwd>
<kwd>titanium coating</kwd>
<kwd>titanium oxide</kwd>
<kwd>titanium skirt</kwd>
<kwd>biointegration</kwd>
<kwd>corneal adhesion</kwd>
<kwd>titanium adhesion</kwd>
</kwd-group>
</article-meta>
</front>
<body>
<sec id="s1">
<title>Introduction</title>
<p>The Boston keratoprosthesis (B-Kpro; Massachusetts Eye and Ear Infirmary, Boston, MA) is currently (as of 2015) the most widely used artificial cornea, with more than 12,000 units implanted worldwide. Its current configuration consists of a carrier donor corneal graft sandwiched between a poly(methyl methacrylate) (PMMA) front plate and a PMMA or titanium back plate.
<sup>
<xref rid="i2164-2591-5-2-17-b01" ref-type="bibr">1</xref>
<xref rid="i2164-2591-5-2-17-b04" ref-type="bibr">4</xref>
</sup>
Since the stem of the B-Kpro crosses the cornea, inadequate attachment with corneal tissue could create a space
<sup>
<xref rid="i2164-2591-5-2-17-b05" ref-type="bibr">5</xref>
</sup>
through which microorganisms can penetrate into the eye and induce endophthalmitis.
<sup>
<xref rid="i2164-2591-5-2-17-b06" ref-type="bibr">6</xref>
,
<xref rid="i2164-2591-5-2-17-b07" ref-type="bibr">7</xref>
</sup>
Consequently, the recommended protocol after B-Kpro implantation includes the use of daily antibiotic drops, which have markedly decreased the risk of endophthalmitis. Nonetheless, severe infections still can occur if compliance with the antibiotic regimen is not observed.
<sup>
<xref rid="i2164-2591-5-2-17-b06" ref-type="bibr">6</xref>
</sup>
</p>
<p>We have hypothesized that modifying the surface of the B-Kpro stem could enhance adhesion with the donor corneal carrier, which could decrease the risk of endophthalmitis. We selected titanium to test this hypothesis, due to its long history of successful use in bone or dental prostheses,
<sup>
<xref rid="i2164-2591-5-2-17-b08" ref-type="bibr">8</xref>
<xref rid="i2164-2591-5-2-17-b12" ref-type="bibr">12</xref>
</sup>
and the fact that it is biologically very well tolerated, as it induces relatively little inflammation and foreign body reaction.
<sup>
<xref rid="i2164-2591-5-2-17-b13" ref-type="bibr">13</xref>
<xref rid="i2164-2591-5-2-17-b15" ref-type="bibr">15</xref>
</sup>
Furthermore, by modifying their topography, titanium surfaces can induce proliferation or differentiation of osteocytes, promoting better apposition and adhesion between the material and bone.
<sup>
<xref rid="i2164-2591-5-2-17-b12" ref-type="bibr">12</xref>
,
<xref rid="i2164-2591-5-2-17-b16" ref-type="bibr">16</xref>
</sup>
In vitro studies have shown very low toxicity to human corneal epithelial cell lines.
<sup>
<xref rid="i2164-2591-5-2-17-b17" ref-type="bibr">17</xref>
,
<xref rid="i2164-2591-5-2-17-b18" ref-type="bibr">18</xref>
</sup>
In addition, titanium has been used as a locking ring and back plate in the B-Kpro,
<sup>
<xref rid="i2164-2591-5-2-17-b02" ref-type="bibr">2</xref>
,
<xref rid="i2164-2591-5-2-17-b19" ref-type="bibr">19</xref>
</sup>
as an intrastromal haptic in the Fyodorov-Zuev keratoprosthesis,
<sup>
<xref rid="i2164-2591-5-2-17-b19" ref-type="bibr">19</xref>
</sup>
and in other experimental corneal devices in rabbits.
<sup>
<xref rid="i2164-2591-5-2-17-b19" ref-type="bibr">19</xref>
,
<xref rid="i2164-2591-5-2-17-b20" ref-type="bibr">20</xref>
</sup>
</p>
<p>We assessed the effectiveness of titanium-based materials in improving adherence to corneal tissue ex vivo and in vivo.</p>
</sec>
<sec sec-type="methods" id="s2">
<title>Materials and Methods</title>
<sec id="s2a">
<title>In Vitro Surface Characterization of the Candidate Materials</title>
<p>We studied several surface modifications of titanium as candidate materials for the B-Kpro in an attempt to enhance adhesion between the device and the cornea, and compared them to PMMA, the current manufacturing material of the device. All experimental samples were obtained from and processed by a local vendor (JG Machine Co., Inc., Wilmington, MA). For the in vitro surface characterization and cell viability studies, we used discs (0.2 mm thick × 10 mm diameter) made either of 100% medical-grade PMMA (Spartech Townsend, Pleasant Hill, IA) or titanium-grade 23 (Ti-6AL-4V-ELI or Ti 6-4 ELI), which contains 6% aluminum and 4% vanadium (ISO 9001:2008 certification; President Titanium Co., Inc., Hanson, MA). Before proceeding with any experiments, all the samples were cleaned ultrasonically with Alconox powdered detergent (Alconox, Inc., White Plains, NY) in two cycles of 1 minute each, after which they were rinsed profusely with deionized water to remove any remaining debris, and blow-dried with air.</p>
<p>Titanium initially was chosen based on its long-term compatibility and safety when used inside the eye.
<sup>
<xref rid="i2164-2591-5-2-17-b02" ref-type="bibr">2</xref>
,
<xref rid="i2164-2591-5-2-17-b14" ref-type="bibr">14</xref>
,
<xref rid="i2164-2591-5-2-17-b19" ref-type="bibr">19</xref>
,
<xref rid="i2164-2591-5-2-17-b20" ref-type="bibr">20</xref>
</sup>
In addition, we chose to manufacture the titanium samples with a rough surface achieved through sand blasting (Ti
<sub>SB</sub>
), and compared them to samples with a smooth surface (Ti). The sand blasting of the surface was intended to promote and enhance the adhesion between the titanium and corneal tissue, in the same way that previous studies have shown an enhanced tissue adhesion between implanted titanium and bone or tooth after increasing the surface roughness of the implant.
<sup>
<xref rid="i2164-2591-5-2-17-b16" ref-type="bibr">16</xref>
</sup>
Sand blasting currently is used on the back plate of the B-Kpro to eliminate its shiny-metallic appearance.
<sup>
<xref rid="i2164-2591-5-2-17-b17" ref-type="bibr">17</xref>
</sup>
Some samples underwent plasma oxidization (Ti
<sub>ox</sub>
, and Ti
<sub>SBox</sub>
) in an attempt to make the surfaces more hydrophilic, which also is believed to promote better cellular and even extracellular matrix adhesion.
<sup>
<xref rid="i2164-2591-5-2-17-b21" ref-type="bibr">21</xref>
</sup>
For this purpose, the samples were placed in a plasma oxidizer chamber at ≈ 250 millitorrs, for 5 minutes. Additionally, we included unmodified PMMA samples (PMMA) as well as plasma-treated PMMA (PMMA
<sub>ox</sub>
) as group controls.</p>
<p>Finally, we studied another surface modification by coating PMMA with titanium dioxide (PMMATiO
<sub>2</sub>
), which has been shown to promote keratocyte adhesion in cell culture.
<sup>
<xref rid="i2164-2591-5-2-17-b13" ref-type="bibr">13</xref>
</sup>
The rationale for using this material can be explained by the fact that TiO
<sub>2</sub>
implants possess a high corrosion resistance in vitro and in vivo with low inflammatory reaction.
<sup>
<xref rid="i2164-2591-5-2-17-b14" ref-type="bibr">14</xref>
</sup>
Furthermore, a TiO
<sub>2</sub>
coating can be easily applied directly on PMMA through liquid phase deposition,
<sup>
<xref rid="i2164-2591-5-2-17-b22" ref-type="bibr">22</xref>
</sup>
without adding additional components to the current design of the B-Kpro, which could potentially raise questions on safety. For this purpose, TiO
<sub>2</sub>
was deposited over a base-layer of polydopamine (PDA), which in turn allows the deposition of TiO
<sub>2</sub>
on virtually any surface, even those with complex topographies.
<sup>
<xref rid="i2164-2591-5-2-17-b22" ref-type="bibr">22</xref>
</sup>
Briefly, the PMMA rods were immersed in a dilute aqueous solution of dopamine (2 mg/mL dopamine in 10 mM Tris buffer, pH 8.5) and shaken with a vortex mixer overnight to create a layer of PDA. After a thorough wash with deionized water the next day, the PDA-coated rods were immersed in 0.1 M ammonium hexafluorotitanate ([NH
<sub>4</sub>
]
<sub>2</sub>
TiF
<sub>6</sub>
) and 0.3 M boric acid (H
<sub>3</sub>
BO
<sub>3</sub>
) solution at pH 3.9 and were left overnight on a shaker at room temperature. Finally the samples were rinsed carefully with deionized water and blow-dried with air.</p>
<p>For the initial characterization of the PMMA
<sub>TiO2</sub>
samples, we used a PHI VersaProbe II Scanning X-ray photoelectron spectroscope (XPS; Physical Electronics, Chanhassen, MN) equipped with a monochromatized Al Kα source. Furthermore, we used a drop shape analyzer (Drop Shape Analyzer DSA100B, Model: FM3200; Krüss GmbH, Hamburg, Germany) to measure the hydrophilicity of all the samples in our study by registering the water contact angle on their surface, while atomic force microscopy (AFM; MFP-3D AFM; Asylum Research, Santa Barbara, CA) was used to characterize their surface nano-topography. Finally, we analyzed the surface morphology with scanning electron microscopy (SEM; JEOL 5600LV; JEOL Inc., Peabody, MA), after coating the samples with gold-palladium.</p>
</sec>
<sec id="s2b">
<title>In Vitro Cell Viability Assay</title>
<p>To assess cell viability, we performed an alamarBlue assay (Invitrogen, Life Technologies, Grand Island, NY). This assay allows establishing relative cell toxicity by using resazurin and the reducing power of living cells to quantitatively measure proliferation. For this purpose, discs were cleaned, placed individually in 24-well plates, and seeded with primary human corneal fibroblasts at 5 × 10
<sup>3</sup>
cells/cm
<sup>2</sup>
(courtesy of James Zieske, PhD, from Schepens Eye Research Institute, Harvard Medical School, Boston, MA) in Minimum Essential Medium Eagle (Sigma-Aldrich Corp., St. Louis, MO) supplemented with 10% fetal bovine serum and 1% penicillin/streptomycin. At 7 and 14 days after seeding, we added alamarBlue reagent to each well (10% vol/vol), incubated for 2.5 hours, and measured the absorbance of the media at 570/600 nm. The cellular proliferation was measured by the degree of reduction of the reagent, which was calculated by the following equation provided in the manufacturer's manual:
<disp-formula id="i2164-2591-5-2-17-e01">
<graphic xlink:href="i2164-2591-5-2-17-e01.jpg" position="float" orientation="portrait"></graphic>
</disp-formula>
</p>
<p>
<italic>ε</italic>
<sub>OX</sub>
and
<italic>ε</italic>
<sub>RED</sub>
are the molar extinction coefficients of alamarBlue in oxidized and reduced form, and
<italic>A</italic>
and
<italic>A</italic>
′ are the absorbances of each well and of media only, respectively. The data were normalized to the cells cultured on tissue culture polystyrene (TCPS).</p>
</sec>
<sec id="s2c">
<title>Ex Vivo Pullout Force Technique</title>
<p>For the pullout ex vivo experiments we used rods (15 mm long × 3.35 mm diameter) made of the different materials, which previously were cleaned following the aforementioned procedure. To test the adhesion between the cornea and candidate materials, we used a mechanical “pullout” technique developed in our lab.
<sup>
<xref rid="i2164-2591-5-2-17-b23" ref-type="bibr">23</xref>
</sup>
Porcine eyes were obtained from the vendor, transported to the laboratory in a moisture chamber with ice, and processed within 24 hours of animal death. Before manipulation, the eyes were rinsed with 10% povidone iodine solution, and washed with phosphate-buffered saline (PBS 10X, pH 7.4; Invitrogen, Life Technologies). The porcine cornea then was trephined with a 10-mm diameter surgical trephine, and an additional 3-mm hole was punched through the center with an Acu·Punch biopsy punch (Acuderm, Inc., Fort Lauderdale, FL). The cornea then was slid over one of the 3.35-mm diameter rods, and placed individually in a 12-well plate filled with 2 mL of Chen cornea organ culture medium, which consists of Modified Medium 199 (Invitrogen), 1% penicillin-streptomycin, 7% dextran, 25 mM Hepes, and 10 nM D,L-β-hydroxybutyrate (Sigma-Aldrich Corp.).
<sup>
<xref rid="i2164-2591-5-2-17-b24" ref-type="bibr">24</xref>
,
<xref rid="i2164-2591-5-2-17-b25" ref-type="bibr">25</xref>
</sup>
Specimens were incubated at 37°C in 5% carbon dioxide for 2 weeks, and during that time half of the media was replaced every 12 hours.</p>
<p>After 14 days, the specimens were placed in a custom-designed holder,
<sup>
<xref rid="i2164-2591-5-2-17-b23" ref-type="bibr">23</xref>
</sup>
to fit a universal tensiometer (Instron 5542; Instron, Norwood, MA). The cornea was placed in a cavity between two plastic holders (acrylonitrile butadiene styrene; ABSPlus; Dimension, Inc., Eden Prairie, MN), with the rod protruding through the top of the holder. The space between the parts of the holder was kept fixed (500–1000 μm approximately) with the help of two screws, so the cornea would not be compressed. The Instron device gripped the holder from its base and the rod was pulled out from above at a constant velocity (10 mm/min), while the force required to sustain the pullout was measured. The test was considered finished when the cylinder was completely removed from the cornea. After completion of the mechanical testing, the rods were submerged in 4% formalin, and processed for SEM. Before imaging, the samples were dried in a critical point dryer, and 5 nm of gold/palladium were deposited on the surface to avoid unnecessary electric charging.</p>
</sec>
<sec id="s2d">
<title>In Vivo B-Kpro Implantation</title>
<p>All animal studies were approved by the Animal Care Committee of the Massachusetts Eye and Ear Infirmary and performed in accordance with the ARVO Statement for the Use of Animals in Ophthalmic and Vision Research.</p>
<p>A total of 7 New Zealand White albino rabbits were used (male, 2.5–3.5 Kg), each one having a type 1 B-Kpro surgically implanted into the right eye, using the same standard technique as used in humans. Briefly, in deeply anesthetized animals, the cornea was trephined (9 mm), excised and set aside, and the lens was removed through an open sky extracapsular extraction. The excised cornea then was further trephined centrally with a 3 mm-diameter punch. The device then was slid through the cornea carrier, and a titanium back plate was clicked-on. The corneal graft with the B-Kpro then was sutured back in the eye with twelve 10-0 interrupted nylon sutures. Two animals were implanted with a smooth Ti clad (Ti) around the stem of the B-Kpro, three animals received sand blasted titanium clads (Ti
<sub>SB</sub>
), and the remaining two animals received a standard B-Kpro. The titanium ring was inserted easily around the stem of the B-Kpro after placing the prosthesis in dry ice for 30 seconds. This was performed before sending the implants for sterilization. After the implantation, the animals were treated with topical antibiotics (polymyxin B/trimethoprim) and steroids (prednisolone acetate 1%), every 12 hours for the first week, and once daily for the following weeks. Prednisolone was discontinued at 2 weeks postoperatively. Six animals were monitored for 1 month while one of the Ti
<sub>SB</sub>
clad animals was kept for 10 months. At the end time point, the animals were euthanized and enucleated, and the B-Kpro pulled away manually from the cornea. The devices then were immersed in half-strength Karnovsky's fixative (2% paraformaldehyde; 2.5% glutaraldehyde) in 0.1 M phosphate buffer pH 7.4 overnight, dried in a critical point dryer, and coated with gold/palladium for SEM imaging.</p>
</sec>
<sec id="s2e">
<title>Statistics</title>
<p>In general, data are presented as means ± SD and were analyzed accordingly. For all statistical tests, 2-tailed values of
<italic>P</italic>
< 0.05 were considered statistically significant. All data analysis was performed using the statistical tool provided by IGOR Pro (version 6.3; WaveMetrics, Inc., Portland, OR) and GraphPad Prism for Mac OS X (version 5.0c, GraphPad Software, Inc., La Jolla, CA).</p>
</sec>
</sec>
<sec id="s3">
<title>Results</title>
<sec id="s3a">
<title>Production and Characterization Of Materials</title>
<p>The aim of this study was to assess the potential for titanium-based materials to enhance adhesion between the PMMA B-KPro stem and the adjacent donor cornea. For in vitro studies, titanium samples were manufactured as discs (0.2 mm thick × 10 mm diameter), and for ex vivo work as rods (15 mm long × 3.35 mm diameter). Sample surfaces were either smooth (Ti) or were roughened by sand blasting (Ti
<sub>SB</sub>
), since rougher surfaces are reported to promote cellular adhesion.
<sup>
<xref rid="i2164-2591-5-2-17-b16" ref-type="bibr">16</xref>
</sup>
Both types of experimental samples also were oxidized by oxygen plasma treatment (Ti
<sub>ox</sub>
, and Ti
<sub>SBox</sub>
) to make the surfaces more hydrophilic, which also is believed to promote better cellular attachment.
<sup>
<xref rid="i2164-2591-5-2-17-b21" ref-type="bibr">21</xref>
</sup>
Their performance was compared to that of unmodified PMMA (PMMA) as well as plasma-treated PMMA (PMMA
<sub>ox</sub>
), and PMMA coated with TiO
<sub>2</sub>
by deposition on top of PDA (PMMA
<sub>TiO2</sub>
). Polydopamine films can form on a wide range of substrates by the polymerization of dopamine in an alkaline solution. The quinone groups in PDA can chelate metal ions and have been used for the deposition of hydroxyapatite and titanium oxide.
<sup>
<xref rid="i2164-2591-5-2-17-b22" ref-type="bibr">22</xref>
,
<xref rid="i2164-2591-5-2-17-b23" ref-type="bibr">23</xref>
</sup>
</p>
<p>The surface chemistry of materials was characterized by their high-resolution XPS spectra to confirm the presence of titanium and its oxidation states (
<xref ref-type="fig" rid="i2164-2591-5-2-17-f01">Fig. 1</xref>
). The binding energies for titanium and its various oxidation states were identified: 453.9 eV = Ti, 455.5 eV = TiO, 457.3 eV = Ti
<sub>2</sub>
O
<sub>3</sub>
, and 458.7 eV = TiO
<sub>2</sub>
.
<sup>
<xref rid="i2164-2591-5-2-17-b26" ref-type="bibr">26</xref>
</sup>
High-resolution spectra of all titanium samples, regardless of surface roughness (smooth and sandblasted) and oxygen plasma treatment, could be assigned mainly to TiO
<sub>2</sub>
with minor variations of Ti, TiO, and Ti
<sub>2</sub>
O
<sub>3</sub>
peaks. Coating of PMMA with TiO
<sub>2</sub>
(PMMA
<sub>TiO2</sub>
) was demonstrated by the presence of a TiO
<sub>2</sub>
peak, along with negligible peaks for Ti and other oxidized states of titanium (TiO, Ti
<sub>2</sub>
O
<sub>3</sub>
). The PMMA samples did not show any traces of titanium.</p>
<fig id="i2164-2591-5-2-17-f01" orientation="portrait" position="float">
<label>Figure 1</label>
<caption>
<p>X-ray photoelectron spectroscopy of various materials. Peaks corresponding to Ti (453.9 eV), TiO (455.5 eV), Ti
<sub>2</sub>
O
<sub>3</sub>
(457.3 eV), and TiO
<sub>2</sub>
(458.7 eV) are indicated.</p>
</caption>
<graphic xlink:href="i2164-2591-5-2-17-f01"></graphic>
</fig>
<p>The hydrophilicity of the materials was characterized by measuring water contact angles. Samples treated with oxygen plasma showed a marked decrease in the contact angle measurements, indicating an increase in hydrophilicity due to oxidization of the outermost surface of the samples (
<xref ref-type="fig" rid="i2164-2591-5-2-17-f02">Fig. 2</xref>
). PMMA
<sub>TiO2</sub>
had a zero contact angle because it was coated with a highly hydrophilic TiO
<sub>2</sub>
layer (
<xref ref-type="fig" rid="i2164-2591-5-2-17-f01">Fig. 1</xref>
). All the groups in
<xref ref-type="fig" rid="i2164-2591-5-2-17-f02">Figure 2</xref>
were statistically significantly different from each other (
<italic>P</italic>
< 0.05), except where indicated.</p>
<fig id="i2164-2591-5-2-17-f02" orientation="portrait" position="float">
<label>Figure 2</label>
<caption>
<p>Water contact angle measurements. The
<italic>graph</italic>
shows the mean value of the contact angle measurements from every group. All samples were statistically significantly different from each other (
<italic>P</italic>
< 0.05 by ANOVA with Tukey's test), except where indicated (
<italic>ns</italic>
). Data are means ± SD (
<italic>N</italic>
= 4).</p>
</caption>
<graphic xlink:href="i2164-2591-5-2-17-f02"></graphic>
</fig>
<p>Atomic force microscopy imaging was used to quantify the nano-scale topography and the surface roughness of the materials, as shown in
<xref ref-type="fig" rid="i2164-2591-5-2-17-f03">Figure 3A</xref>
. The Ti
<sub>SB</sub>
samples were the roughest materials, while the Ti samples were the smoothest. The effect of the addition of TiO
<sub>2</sub>
on the roughness of the PMMA (PMMA
<sub>TiO2</sub>
samples) was not statistically significant (
<italic>P</italic>
> 0.05). Oxygen plasma treatment did not have a statistically significant effect on the surface roughness of PMMA, Ti, or Ti
<sub>SB</sub>
(
<italic>P</italic>
= 0.9593). Scanning electron microscopy imaging of the materials showed no obvious effect of oxidation on surface roughness, but confirmed that the sand blasted groups (Ti
<sub>SB</sub>
, Ti
<sub>SBox</sub>
) had greater surface roughness (
<xref ref-type="fig" rid="i2164-2591-5-2-17-f03">Fig. 3B</xref>
).</p>
<fig id="i2164-2591-5-2-17-f03" orientation="portrait" position="float">
<label>Figure 3</label>
<caption>
<p>Surface imaging. (A) Atomic force microscopy imaging provided quantitative data on the surface topography and roughness of the materials, summarized in (B) the graph. Data are means ± SD (
<italic>N</italic>
= 4,
<italic>t</italic>
-test). Oxidation did not show any statistical differences in any of the materials (
<italic>P</italic>
> 0.05). (C) Scanning electron microscopy of materials before (
<italic>top row</italic>
) and after plasma oxidization (
<italic>bottom row</italic>
).</p>
</caption>
<graphic xlink:href="i2164-2591-5-2-17-f03"></graphic>
</fig>
</sec>
<sec id="s3b">
<title>In Vitro Cell Viability Assay</title>
<p>Cell viability on the various surfaces was assessed by the alamarBlue assay. Human corneal fibroblasts proliferated on all the samples over 2 weeks (
<xref ref-type="fig" rid="i2164-2591-5-2-17-f04">Fig. 4</xref>
). In the Ti
<sub>SB</sub>
and Ti
<sub>SBox</sub>
groups, the number of cells that attached was lower than in other groups, possibly because of the high roughness of the materials.
<sup>
<xref rid="i2164-2591-5-2-17-b27" ref-type="bibr">27</xref>
,
<xref rid="i2164-2591-5-2-17-b28" ref-type="bibr">28</xref>
</sup>
Nonetheless, these results demonstrated that cell growth occurred on all the testing surfaces, indicating that the materials were not toxic to cells.</p>
<fig id="i2164-2591-5-2-17-f04" orientation="portrait" position="float">
<label>Figure 4</label>
<caption>
<p>Alamar blue assay for cell viability (human cornea fibroblasts). Data are means ± SD.</p>
</caption>
<graphic xlink:href="i2164-2591-5-2-17-f04"></graphic>
</fig>
</sec>
<sec id="s3c">
<title>Ex Vivo Measurements of Adhesion</title>
<p>We measured the strength of the adhesion between the cornea and the materials ex vivo. Rods of the various materials were implanted in holes punched into porcine corneas (see Methods). The assembly was maintained in corneal tissue culture for 14 days and then mounted in a device developed in our lab to measure the force required to pull the rods from the corneas (
<xref ref-type="fig" rid="i2164-2591-5-2-17-f05">Figs. 5A</xref>
,
<xref ref-type="fig" rid="i2164-2591-5-2-17-f05">5B</xref>
).
<sup>
<xref rid="i2164-2591-5-2-17-b23" ref-type="bibr">23</xref>
</sup>
After 14 days of incubation, the force required to pull Ti
<sub>ox</sub>
, PMMA
<sub>TiO2</sub>
, Ti
<sub>SB</sub>
, and Ti
<sub>SBox</sub>
rods out of the porcine corneas was statistically significantly higher than rods of plain PMMA or smooth titanium (
<italic>P</italic>
< 0.05;
<xref ref-type="fig" rid="i2164-2591-5-2-17-f05">Fig. 5C</xref>
). The increased adhesion after treating Ti surfaces with oxygen can be attributed to increased surface hydrophilicity, which improves adhesion by osteoblasts and keratocytes.
<sup>
<xref rid="i2164-2591-5-2-17-b13" ref-type="bibr">13</xref>
,
<xref rid="i2164-2591-5-2-17-b16" ref-type="bibr">16</xref>
,
<xref rid="i2164-2591-5-2-17-b21" ref-type="bibr">21</xref>
</sup>
In general, PMMA is a poor substrate for cell adhesion
<sup>
<xref rid="i2164-2591-5-2-17-b29" ref-type="bibr">29</xref>
</sup>
and no improvement was observed after treating it with oxygen plasma. In the case of sandblasted titanium, oxygen plasma did not further improve adhesion, most likely because the surface roughness dominated the enhanced adhesion. Scanning electron microscopy imaging of the surfaces of the rods following the pull out experiments (
<xref ref-type="fig" rid="i2164-2591-5-2-17-f06">Fig. 6</xref>
) showed more tissue adherent to the samples that required a higher force to be pulled out from the corneas (Ti
<sub>ox</sub>
, PMMA
<sub>TiO2</sub>
, Ti
<sub>SB</sub>
, and Ti
<sub>SBox</sub>
).</p>
<fig id="i2164-2591-5-2-17-f05" orientation="portrait" position="float">
<label>Figure 5</label>
<caption>
<p>Ex vivo pull out measurements. (A) Schematic of the customized holder containing a cornea-rod assembly. (B) Image of the customized holder within a universal (mechanical) testing machine. (C) Force required to pull rods of various materials from porcine corneas. * and # indicate
<italic>P</italic>
< 0.05 (by ANOVA with Tukey's test) in comparison to plain PMMA and smooth Ti, respectively. Data are means ± SD (
<italic>N</italic>
= 4).</p>
</caption>
<graphic xlink:href="i2164-2591-5-2-17-f05"></graphic>
</fig>
<fig id="i2164-2591-5-2-17-f06" orientation="portrait" position="float">
<label>Figure 6</label>
<caption>
<p>Representative scanning electron microscopy of implant surfaces after the ex vivo pullout experiments.</p>
</caption>
<graphic xlink:href="i2164-2591-5-2-17-f06"></graphic>
</fig>
<p>Our ex vivo results showed that when increasing the hydrophilicity of the surface of titanium, or when its surface roughness is increased, the corneal tissue attachment was enhanced. This is not the case for PMMA.</p>
</sec>
<sec id="s3d">
<title>In Vivo Results With a Ti-Clad Around the B-Kpro Stem</title>
<p>The four groups that showed the strongest adhesion to corneas ex vivo were Ti
<sub>ox</sub>
, PMMA
<sub>TiO2</sub>
, Ti
<sub>SB</sub>
, and Ti
<sub>SBox</sub>
. Although plasma oxidation greatly increased sample adhesion to the cornea ex vivo, we did not pursue that approach in vivo because the effect of plasma oxidation on titanium (Ti
<sub>ox</sub>
and Ti
<sub>SBox</sub>
) is only temporary in air, and would be lost during the manufacturing and storing stages before implantation in humans.
<sup>
<xref rid="i2164-2591-5-2-17-b21" ref-type="bibr">21</xref>
</sup>
We also did not proceed with the PMMA
<sub>TiO2</sub>
coating because we realized that it was easily scratched even with careful handling, which would present a problem during surgical implantation. Consequently, we decided to use solid titanium instead (Ti and Ti
<sub>SB</sub>
) for in vivo experimentation. These materials have the additional advantage of having well established and simple means of manufacture.</p>
<p>Boston keratoprostheses, unmodified or with a smooth Ti or Ti
<sub>SB</sub>
sleeve around the stem, were implanted in two rabbits each, using the same technique used in humans. All the devices were retained, without signs of infection or inflammation. One month after implantation, animals were euthanized, the eyes were enucleated and a corneal-scleral ring was excised from the globe. A full thickness tangential cut was performed on the cornea, and the tissue was removed manually from the stem. In the rabbits implanted with either a standard B-Kpro or with a smooth Ti sleeve, the cornea was removed easily from the stem. In the rabbits implanted with a Ti
<sub>SB</sub>
sleeve, the force required to remove the tissue was clearly (albeit subjectively) greater.</p>
<p>Scanning electron microscopy imaging of the explanted devices (
<xref ref-type="fig" rid="i2164-2591-5-2-17-f07">Fig. 7</xref>
) showed few tissue remnants on the standard B-Kpro (
<xref ref-type="fig" rid="i2164-2591-5-2-17-f07">Fig. 7A</xref>
) or the B-Kpro with a smooth Ti sleeve (
<xref ref-type="fig" rid="i2164-2591-5-2-17-f07">Fig. 7B</xref>
), with larger areas without any tissue residue. The B-Kpros with the Ti
<sub>SB</sub>
sleeves (
<xref ref-type="fig" rid="i2164-2591-5-2-17-f07">Fig. 7C</xref>
) showed much more tissue residue, covering the implants.</p>
<fig id="i2164-2591-5-2-17-f07" orientation="portrait" position="float">
<label>Figure 7</label>
<caption>
<p>Scanning electron microscopy of B-Kpros after implantation in rabbits. FP, front plate; BP, back plate.
<italic>Arrows</italic>
indicate the remains of corneal tissue attached to the stem.</p>
</caption>
<graphic xlink:href="i2164-2591-5-2-17-f07"></graphic>
</fig>
<p>One additional animal implanted with B-Kpro with a Ti
<sub>SB</sub>
sleeve around the stem was followed for 10 months. As at 1 month, this B-Kpro required greater force to remove. On SEM image (
<xref ref-type="fig" rid="i2164-2591-5-2-17-f07">Fig. 7D</xref>
), the sleeve was completely covered with a thick layer of fibrous material.</p>
</sec>
</sec>
<sec id="s4">
<title>Discussion</title>
<p>Since its introduction in 1974,
<sup>
<xref rid="i2164-2591-5-2-17-b01" ref-type="bibr">1</xref>
</sup>
the B-Kpro has become an important surgical option for cases where a standard corneal transplantation has an increased risk of graft failure. However, despite providing immediate restoration of visual acuity, thanks to its almost perfect optical qualities (it can allow vision of 20/15), B-Kpro implantation still carries important risks, including glaucoma progression,
<sup>
<xref rid="i2164-2591-5-2-17-b30" ref-type="bibr">30</xref>
</sup>
corneal necrosis or keratolysis,
<sup>
<xref rid="i2164-2591-5-2-17-b31" ref-type="bibr">31</xref>
<xref rid="i2164-2591-5-2-17-b35" ref-type="bibr">35</xref>
</sup>
and bacterial endophthalmitis.
<sup>
<xref rid="i2164-2591-5-2-17-b06" ref-type="bibr">6</xref>
,
<xref rid="i2164-2591-5-2-17-b07" ref-type="bibr">7</xref>
</sup>
Among these complications, tissue necrosis has a major role and is of high importance since it may lead to gaps in the interface between the B-Kpro and the carrier cornea,
<sup>
<xref rid="i2164-2591-5-2-17-b33" ref-type="bibr">33</xref>
</sup>
which in turn can affect retention and facilitate the entry of bacteria or other microorganisms into the eye, ultimately leading to endophthalmitis.</p>
<p>Although microbial endophthalmitis after B-Kpro has been drastically reduced in the last decade by use of daily low-dose prophylactic antibiotics,
<sup>
<xref rid="i2164-2591-5-2-17-b06" ref-type="bibr">6</xref>
,
<xref rid="i2164-2591-5-2-17-b36" ref-type="bibr">36</xref>
</sup>
infectious complications still occur, especially in the developing world.
<sup>
<xref rid="i2164-2591-5-2-17-b32" ref-type="bibr">32</xref>
,
<xref rid="i2164-2591-5-2-17-b37" ref-type="bibr">37</xref>
</sup>
It has long been suspected that the main contributing factor facilitating the increase in the risk of endophthalmitis could be inadequate integration between the B-KPro and the surrounding corneal tissue. Shapiro et al.
<sup>
<xref rid="i2164-2591-5-2-17-b05" ref-type="bibr">5</xref>
</sup>
have recently documented this phenomenon by imaging B-Kpro patients with anterior segment optical coherence tomography (AS-OCT), in which a significant gap could be observed between the stem of the device and the corneal graft tissue. The aim of this study has been to identify titanium-based materials that would improve the attachment (or biointegration)
<sup>
<xref rid="i2164-2591-5-2-17-b29" ref-type="bibr">29</xref>
</sup>
between the corneal donor tissue and the B-Kpro.</p>
<p>Biointegration or a tight attachment between implanted medical devices and body tissues directly influences clinical outcomes, including tissue breakdown and infection, ultimately having an impact on safety and efficacy.
<sup>
<xref rid="i2164-2591-5-2-17-b08" ref-type="bibr">8</xref>
<xref rid="i2164-2591-5-2-17-b10" ref-type="bibr">10</xref>
</sup>
The strength of attachment will depend on the chemical composition and the topography of the surface of the implanted material. It has been demonstrated that materials with a highly hydrophilic surface and high surface roughness induce better tissue attachment in vivo.
<sup>
<xref rid="i2164-2591-5-2-17-b16" ref-type="bibr">16</xref>
,
<xref rid="i2164-2591-5-2-17-b21" ref-type="bibr">21</xref>
</sup>
The ex vivo and in vivo data in our study also found that the materials with the most hydrophilic surface or the highest roughness demonstrated a stronger cellular attachment.</p>
<p>Titanium is very well tolerated in the eye.
<sup>
<xref rid="i2164-2591-5-2-17-b19" ref-type="bibr">19</xref>
</sup>
In 1993, the first successful use of titanium in a keratoprosthesis was reported. The device consisting of a PMMA optic with a tantalum-titanium alloy haptic was implanted in humans.
<sup>
<xref rid="i2164-2591-5-2-17-b19" ref-type="bibr">19</xref>
</sup>
The extrusion rate in this study was 3% to 13% depending upon the operative techniques used, and aseptic corneal necrosis presented in 11% of the cases. Three years later, an experimental keratoprosthesis was developed that included an intrastromal haptic made of titanium.
<sup>
<xref rid="i2164-2591-5-2-17-b20" ref-type="bibr">20</xref>
</sup>
The mild inflammatory reaction in rabbit corneas was consistent with that found after any medical device implantation. A more recent study has evaluated the use of titanium oxide in vitro. In this study, TiO
<sub>2</sub>
enhanced keratocyte cell integration and cell adhesion, while at the same time presenting antibacterial properties, suggesting it could be used as candidate material for an artificial corneal skirt.
<sup>
<xref rid="i2164-2591-5-2-17-b13" ref-type="bibr">13</xref>
</sup>
Titanium has recently become a major component of the B-Kpro; the most recent version includes a back plate made of medical grade titanium alloy (also containing aluminum and vanadium).
<sup>
<xref rid="i2164-2591-5-2-17-b02" ref-type="bibr">2</xref>
</sup>
This experience informed our decision to use titanium in this study. However, none of the previous studies addressed the potential for titanium to enhance mechanical adherence between cornea and keratoprosthesis in vivo.</p>
<p>Titanium can be found in several oxidative states, the most common one being TiO
<sub>2</sub>
. This was our first choice for the study, based on the fact that TiO
<sub>2</sub>
is highly hydrophilic, possesses a high corrosion resistance in vitro and in vivo, and causes little inflammatory response in vivo.
<sup>
<xref rid="i2164-2591-5-2-17-b14" ref-type="bibr">14</xref>
</sup>
We coated PMMA with PDA, which served as a base-layer for subsequently depositing TiO
<sub>2</sub>
through liquid phase deposition
<sup>
<xref rid="i2164-2591-5-2-17-b22" ref-type="bibr">22</xref>
</sup>
(PMMA
<sub>TiO2</sub>
). X-ray photoelectron spectroscopy analysis of the PMMA
<sub>TiO2</sub>
samples was consistent with previous publications,
<sup>
<xref rid="i2164-2591-5-2-17-b38" ref-type="bibr">38</xref>
</sup>
and confirmed the successful coating of TiO
<sub>2</sub>
. The high hydrophilicity of the coating, verified by the contact angle measurements, was predictive of the ex vivo findings that PMMA
<sub>TiO2</sub>
showed increased adhesion to corneal tissue. These results also correlated well with previous work on corneal keratocytes
<sup>
<xref rid="i2164-2591-5-2-17-b13" ref-type="bibr">13</xref>
</sup>
and osteoblasts,
<sup>
<xref rid="i2164-2591-5-2-17-b39" ref-type="bibr">39</xref>
</sup>
in which the increased hydrophilicity of TiO
<sub>2</sub>
facilitated cell attachment. Unfortunately, the PMMA
<sub>TiO2</sub>
coating was easily scratched with handling, making it less practical for use in vivo. Solid titanium is structurally robust and has the advantage of having well established and simple means of manufacture. Therefore, we decided to use solid titanium instead.</p>
<p>Here solid titanium was used in the form of a sleeve around the stem of the B-Kpro in in vivo studies. The sand blasted titanium sleeve was used because it had performed well in the ex vivo studies of adhesion to corneal tissue. Those results had been consistent with previous studies where osteoblast and corneal keratocyte proliferation and differentiation in vitro were promoted by an increase in surface roughness.
<sup>
<xref rid="i2164-2591-5-2-17-b15" ref-type="bibr">15</xref>
,
<xref rid="i2164-2591-5-2-17-b16" ref-type="bibr">16</xref>
,
<xref rid="i2164-2591-5-2-17-b27" ref-type="bibr">27</xref>
,
<xref rid="i2164-2591-5-2-17-b40" ref-type="bibr">40</xref>
</sup>
</p>
<p>Plasma oxidation increases the hydrophilicity of material surfaces and promotes cellular attachment.
<sup>
<xref rid="i2164-2591-5-2-17-b21" ref-type="bibr">21</xref>
</sup>
In our study, plasma-oxidation of titanium significantly increased adhesion to cornea tissue compared to nonoxidized titanium. However, the effects of plasma oxidation are short-lived and the equipment for it is not widely available, which would make it less practical to implement this approach in clinical situations. Our ex vivo study revealed that simple sand-blasting on titanium even without the plasma oxidation process enhanced adhesion as effectively as titanium oxide deposition or plasma oxidation on titanium. In vivo, the Ti
<sub>SB</sub>
sleeve around the B-Kpro stem had a tighter attachment to the cornea than did the unmodified PMMA stem or the smooth titanium sleeve, and had more tissue remnants adherent to the surface of the sleeve compared to the two other groups.</p>
<p>In conclusion, our findings supported the hypothesis that titanium-based materials would enhance adhesion between the B-Kpro and corneal tissue in vivo. Our ex vivo and in vivo results showed better tissue attachment when titanium was manufactured with a rough surface through sand blasting, compared to a smooth surface titanium sleeve or PMMA. Incorporating a sand blasted titanium sleeve around the stem of the B-Kpro has the potential to improve attachment to the carrier corneal graft in humans, and reduce the likelihood of postoperative complications such as keratolysis and endophthalmitis.</p>
</sec>
</body>
<back>
<ack>
<title>Acknowledgments</title>
<p>The authors thank Ann Tisdale, M.S., Staff Associate at Ilene K. Gipson laboratory, Schepens Eye Research Institute of Massachusetts Eye and Ear Infirmary, Harvard Medical School, Boston, MA, USA, for providing the SEM images of the B-Kpro devices; and John Granney, JG Machine, Precision Machining, the manufacturer of the B-Kpro, for providing all the necessary materials for this project.</p>
<p>Supported by the Keratoprosthesis Fund, Massachusetts Eye and Ear Infirmary, Boston, MA, USA, and R01 GM073626; DSK. The authors alone are responsible for the content and writing of this paper.</p>
<p>Borja Salvador-Culla and Kyung Jae Jeong contributed equally to this work.</p>
<p>B. Salvador-Culla and K.J. Jeong designed and conducted the study, B. Salvador-Culla, K.J. Jeong and H.H. Chiang collected the data; B. Salvador-Culla, K.J. Jeong, C.H. Dohlman, and D.S. Kohane were responsible for management; B. Salvador-Culla, J.J. Jeong, and P.E. Kolovou performed analysis; B. Salvador-Culla, K.J. Jeong, P.E. Kolovou, J. Chodosh, C.H. Dohlman, and D.S. Kohane interpreted the data; B. Salvador-Culla and P.E. Kolovou helped with manuscript preparation; all authors reviewed and approved the manuscript.</p>
<p>Disclosure:
<bold>B. Salvador-Culla</bold>
(E);
<bold>K.J. Jeong</bold>
, None;
<bold>P.E. Kolovou</bold>
(E);
<bold>H.H. Chiang</bold>
, None;
<bold>J. Chodosh</bold>
(E);
<bold>C.H. Dohlman</bold>
(E);
<bold>D.S. Kohane</bold>
, None</p>
</ack>
<ref-list>
<title>References</title>
<ref id="i2164-2591-5-2-17-b01">
<label>1.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Dohlman</surname>
<given-names>CH,</given-names>
</name>
<name>
<surname>Schneider</surname>
<given-names>HA,</given-names>
</name>
<name>
<surname>Doane</surname>
<given-names>MG.</given-names>
</name>
</person-group>
<article-title>Prosthokeratoplasty</article-title>
.
<source>
<italic>Am J Ophthalmol</italic>
</source>
.
<year>1974</year>
;
<volume>77</volume>
:
<fpage>694</fpage>
<lpage>670</lpage>
.
<pub-id pub-id-type="pmid">4596032</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b02">
<label>2.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Todani</surname>
<given-names>A,</given-names>
</name>
<name>
<surname>Ciolino</surname>
<given-names>JB,</given-names>
</name>
<name>
<surname>Ament</surname>
<given-names>JD,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Titanium back plate for a PMMA keratoprosthesis: clinical outcomes</article-title>
.
<source>
<italic>Graefes Arch Clin Exp Ophthalmol</italic>
</source>
.
<year>2011</year>
<comment>249;</comment>
<volume>10</volume>
:
<fpage>1515</fpage>
<lpage>1518</lpage>
.
<pub-id pub-id-type="pmid">21519940</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b03">
<label>3.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Doane</surname>
<given-names>MG,</given-names>
</name>
<name>
<surname>Dohlman</surname>
<given-names>CH,</given-names>
</name>
<name>
<surname>Bearse</surname>
<given-names>G.</given-names>
</name>
</person-group>
<article-title>Fabrication of a keratoprosthesis</article-title>
.
<source>
<italic>Cornea</italic>
</source>
.
<year>1996</year>
;
<volume>15</volume>
:
<fpage>179</fpage>
<lpage>184</lpage>
.
<pub-id pub-id-type="pmid">8925666</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b04">
<label>4.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Colby</surname>
<given-names>KA,</given-names>
</name>
<name>
<surname>Koo</surname>
<given-names>EB.</given-names>
</name>
</person-group>
<article-title>Expanding indications for the Boston keratoprosthesis</article-title>
.
<source>
<italic>Curr Opin Ophthalmol</italic>
</source>
.
<year>2011</year>
;
<volume>22</volume>
:
<fpage>267</fpage>
<lpage>273</lpage>
.
<pub-id pub-id-type="pmid">21537184</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b05">
<label>5.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Shapiro</surname>
<given-names>BL,</given-names>
</name>
<name>
<surname>Cortés</surname>
<given-names>DE,</given-names>
</name>
<name>
<surname>Chin</surname>
<given-names>EK,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>High-resolution spectral domain anterior segment optical coherence tomography in type 1 Boston keratoprosthesis</article-title>
.
<source>
<italic>Cornea</italic>
</source>
.
<year>2013</year>
;
<volume>32</volume>
:
<fpage>951</fpage>
<lpage>955</lpage>
.
<pub-id pub-id-type="pmid">23591146</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b06">
<label>6.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Behlau</surname>
<given-names>I,</given-names>
</name>
<name>
<surname>Martin</surname>
<given-names>KV,</given-names>
</name>
<name>
<surname>Martin</surname>
<given-names>JN,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Infectious endophthalmitis in Boston keratoprosthesis: incidence and prevention</article-title>
.
<source>
<italic>Acta Ophthalmol</italic>
</source>
.
<year>2014</year>
;
<volume>92</volume>
:
<fpage>546</fpage>
<lpage>555</lpage>
. </mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b07">
<label>7.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Chan</surname>
<given-names>CC,</given-names>
</name>
<name>
<surname>Holland</surname>
<given-names>EJ.</given-names>
</name>
</person-group>
<article-title>Infectious endophthalmitis after Boston type 1 keratoprosthesis implantation</article-title>
.
<source>
<italic>Cornea</italic>
</source>
.
<year>2012</year>
;
<volume>31</volume>
:
<fpage>346</fpage>
<lpage>349</lpage>
.
<pub-id pub-id-type="pmid">22314825</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b08">
<label>8.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Henry</surname>
<given-names>PJ.</given-names>
</name>
</person-group>
<article-title>Clinical experiences with dental implants</article-title>
.
<source>
<italic>Adv Dent Res</italic>
</source>
.
<year>1999</year>
;
<volume>13</volume>
:
<fpage>147</fpage>
<lpage>152</lpage>
.
<pub-id pub-id-type="pmid">11276737</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b09">
<label>9.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Theis</surname>
<given-names>JC,</given-names>
</name>
<name>
<surname>Gambhir</surname>
<given-names>S,</given-names>
</name>
<name>
<surname>White</surname>
<given-names>J.</given-names>
</name>
</person-group>
<article-title>Factors affecting implant retention in infected joint replacements</article-title>
.
<source>
<italic>ANZ J Surg</italic>
</source>
.
<year>2007</year>
;
<volume>77</volume>
:
<fpage>877</fpage>
<lpage>879</lpage>
.
<pub-id pub-id-type="pmid">17803553</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b10">
<label>10.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Hacking</surname>
<given-names>SA,</given-names>
</name>
<name>
<surname>Pauyo</surname>
<given-names>T,</given-names>
</name>
<name>
<surname>Lim</surname>
<given-names>L,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Tissue response to the components of a hydroxyapatite-coated composite femoral implant</article-title>
.
<source>
<italic>J Biomed Mater Res A</italic>
</source>
.
<year>2010</year>
;
<volume>94</volume>
:
<fpage>953</fpage>
<lpage>960</lpage>
.
<pub-id pub-id-type="pmid">20730932</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b11">
<label>11.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Branemark</surname>
<given-names>PI,</given-names>
</name>
<name>
<surname>Hansson</surname>
<given-names>BO,</given-names>
</name>
<name>
<surname>Adell</surname>
<given-names>R,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Osseointegrated implants in the treatment of the edentulous jaw. Experience from a 10-year period</article-title>
.
<source>
<italic>Scand J Plast Reconstr Surg Suppl</italic>
</source>
.
<year>1977</year>
;
<volume>16</volume>
:
<fpage>1</fpage>
<lpage>132</lpage>
.
<pub-id pub-id-type="pmid">356184</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b12">
<label>12.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Baril</surname>
<given-names>E,</given-names>
</name>
<name>
<surname>Lefebvre</surname>
<given-names>LP,</given-names>
</name>
<name>
<surname>Hacking</surname>
<given-names>SA.</given-names>
</name>
</person-group>
<article-title>Direct visualization and quantification of bone growth into porous titanium implants using micro computed tomography</article-title>
.
<source>
<italic>J Mater Sci Mater Med</italic>
</source>
.
<year>2011</year>
;
<volume>22</volume>
:
<fpage>1321</fpage>
<lpage>1332</lpage>
.
<pub-id pub-id-type="pmid">21512898</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b13">
<label>13.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Tan</surname>
<given-names>XW,</given-names>
</name>
<name>
<surname>Perera</surname>
<given-names>AP,</given-names>
</name>
<name>
<surname>Tan</surname>
<given-names>A,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Comparison of candidate materials for a synthetic osteo-odonto keratoprosthesis device</article-title>
.
<source>
<italic>Invest Ophthalmol Vis Sci</italic>
</source>
.
<year>2011</year>
;
<volume>52</volume>
:
<fpage>21</fpage>
<lpage>29</lpage>
.
<pub-id pub-id-type="pmid">20739467</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b14">
<label>14.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Tan</surname>
<given-names>XW,</given-names>
</name>
<name>
<surname>Beuerman</surname>
<given-names>RW,</given-names>
</name>
<name>
<surname>Shi</surname>
<given-names>ZL,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>In vivo evaluation of titanium oxide and hydroxyapatite as an artificial cornea skirt</article-title>
.
<source>
<italic>J Mater Sci Mater Med</italic>
</source>
.
<year>2012</year>
;
<volume>23</volume>
:
<fpage>1063</fpage>
<lpage>1072</lpage>
.
<pub-id pub-id-type="pmid">22426652</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b15">
<label>15.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Hacking</surname>
<given-names>SA,</given-names>
</name>
<name>
<surname>Harvey</surname>
<given-names>E,</given-names>
</name>
<name>
<surname>Roughley</surname>
<given-names>P,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>The response of mineralizing culture systems to microtextured and polished titanium surfaces</article-title>
.
<source>
<italic>J Orthop Res</italic>
</source>
.
<year>2008</year>
;
<volume>26</volume>
:
<fpage>1347</fpage>
<lpage>1354</lpage>
.
<pub-id pub-id-type="pmid">18404719</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b16">
<label>16.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Hacking</surname>
<given-names>SA,</given-names>
</name>
<name>
<surname>Boyraz</surname>
<given-names>P,</given-names>
</name>
<name>
<surname>Powers</surname>
<given-names>BM,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Surface roughness enhances the osseointegration of titanium headposts in non-human primates</article-title>
.
<source>
<italic>J Neurosci Methods</italic>
</source>
.
<year>2012</year>
;
<volume>211</volume>
:
<fpage>237</fpage>
<lpage>244</lpage>
.
<pub-id pub-id-type="pmid">22975472</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b17">
<label>17.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Paschalis</surname>
<given-names>EI,</given-names>
</name>
<name>
<surname>Chodosh</surname>
<given-names>J,</given-names>
</name>
<name>
<surname>Spurr-Michaud</surname>
<given-names>S,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>In vitro and in vivo assessment of titanium surface modification for coloring the backplate of the Boston keratoprosthesis</article-title>
.
<source>
<italic>Invest Ophthalmol Vis Sci</italic>
</source>
.
<year>2013</year>
;
<volume>54</volume>
:
<fpage>3863</fpage>
<lpage>3873</lpage>
.
<pub-id pub-id-type="pmid">23661373</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b18">
<label>18.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Ament</surname>
<given-names>JD,</given-names>
</name>
<name>
<surname>Spurr-Michaud</surname>
<given-names>SJ,</given-names>
</name>
<name>
<surname>Dohlman</surname>
<given-names>CH,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>The Boston Keratoprosthesis: comparing corneal epithelial cell compatibility with titanium and PMMA</article-title>
.
<source>
<italic>Cornea</italic>
</source>
.
<year>2009</year>
;
<volume>28</volume>
:
<fpage>808</fpage>
<lpage>811</lpage>
.
<pub-id pub-id-type="pmid">19574903</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b19">
<label>19.</label>
<mixed-citation publication-type="book">
<person-group person-group-type="author">
<name>
<surname>Salvador-Culla</surname>
<given-names>B,</given-names>
</name>
<name>
<surname>Dohlman</surname>
<given-names>CH.</given-names>
</name>
</person-group>
<article-title>Keratoprosthesis</article-title>
.
<person-group person-group-type="editor">
<name>
<surname>Lemp</surname>
<given-names>M,</given-names>
</name>
<name>
<surname>Benítezdel Castillo</surname>
<given-names>J,</given-names>
</name>
</person-group>
<source>
<italic>Ocular Surface Disorders</italic>
</source>
.
<publisher-loc>London</publisher-loc>
:
<publisher-name>JP Medical Ltd;</publisher-name>
<year>2013</year>
. </mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b20">
<label>20.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Linnola</surname>
<given-names>RJ,</given-names>
</name>
<name>
<surname>Happonen</surname>
<given-names>RP,</given-names>
</name>
<name>
<surname>Andersson</surname>
<given-names>OH,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Titanium and bioactive glass-ceramic coated titanium as materials for keratoprosthesis</article-title>
.
<source>
<italic>Exp Eye Res</italic>
</source>
.
<year>1996</year>
;
<volume>63</volume>
:
<fpage>471</fpage>
<lpage>478</lpage>
.
<pub-id pub-id-type="pmid">8944554</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b21">
<label>21.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Aronsson</surname>
<given-names>BO,</given-names>
</name>
<name>
<surname>Lausmaa</surname>
<given-names>J,</given-names>
</name>
<name>
<surname>Kasemo</surname>
<given-names>B.</given-names>
</name>
</person-group>
<article-title>Glow discharge plasma treatment for surface cleaning and modification of metallic biomaterials</article-title>
.
<source>
<italic>J Biomed Mater Res</italic>
</source>
.
<year>1997</year>
;
<volume>35</volume>
:
<fpage>49</fpage>
<lpage>73</lpage>
.
<pub-id pub-id-type="pmid">9104698</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b22">
<label>22.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Ou</surname>
<given-names>J,</given-names>
</name>
<name>
<surname>Wang</surname>
<given-names>J,</given-names>
</name>
<name>
<surname>Zhang</surname>
<given-names>D,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Fabrication and biocompatibility investigation of TiO(2) films on the polymer substrates obtained via a novel and versatile route</article-title>
.
<source>
<italic>Colloids Surf B Biointerfaces</italic>
</source>
.
<year>2010</year>
;
<volume>76</volume>
:
<fpage>123</fpage>
<lpage>127</lpage>
.
<pub-id pub-id-type="pmid">19914812</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b23">
<label>23.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Wang</surname>
<given-names>L,</given-names>
</name>
<name>
<surname>Jeong</surname>
<given-names>KJ,</given-names>
</name>
<name>
<surname>Chiang</surname>
<given-names>HH,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Hydroxyapatite for keratoprosthesis biointegration</article-title>
.
<source>
<italic>Invest Ophthalmol Vis Sci</italic>
</source>
.
<year>2011</year>
;
<volume>52</volume>
:
<fpage>7392</fpage>
<lpage>7399</lpage>
.
<pub-id pub-id-type="pmid">21849419</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b24">
<label>24.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Chen</surname>
<given-names>CH,</given-names>
</name>
<name>
<surname>Chen</surname>
<given-names>SC.</given-names>
</name>
</person-group>
<article-title>The efficacy of non-lactate-generating metabolites as substrates for maintaining donor tissues</article-title>
.
<source>
<italic>Transplantation</italic>
</source>
.
<year>1994</year>
;
<volume>57</volume>
:
<fpage>1778</fpage>
<lpage>1785</lpage>
.
<pub-id pub-id-type="pmid">8016884</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b25">
<label>25.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Chen</surname>
<given-names>CH,</given-names>
</name>
<name>
<surname>Rama</surname>
<given-names>P,</given-names>
</name>
<name>
<surname>Chen</surname>
<given-names>SC,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Efficacy of organ preservation media enriched with nonlactate-generating substrate for maintaining tissue viability: a transplantation study</article-title>
.
<source>
<italic>Transplantation</italic>
</source>
.
<year>1997</year>
;
<volume>63</volume>
:
<fpage>656</fpage>
<lpage>663</lpage>
.
<pub-id pub-id-type="pmid">9075834</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b26">
<label>26.</label>
<mixed-citation publication-type="other">
<person-group person-group-type="author">
<name>
<surname>Wagner</surname>
<given-names>D,</given-names>
</name>
<name>
<surname>Naumkin</surname>
<given-names>AV,</given-names>
</name>
<name>
<surname>Kraut-Vass</surname>
<given-names>A,</given-names>
</name>
<etal></etal>
</person-group>
<source>Rumble, NIST Standard Reference Database 20, version 3.4 (web version)</source>
,
<year>2003</year>
<comment>Available at: http:/srdata.nist.gov/xps/.</comment>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b27">
<label>27.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Martin</surname>
<given-names>JY,</given-names>
</name>
<name>
<surname>Schwartz</surname>
<given-names>Z,</given-names>
</name>
<name>
<surname>Hummert</surname>
<given-names>TW,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Effect of titanium surface roughness on proliferation, differentiation, and protein synthesis of human osteoblast-like cells (MG63)</article-title>
.
<source>
<italic>J Biomed Mater Res</italic>
</source>
.
<year>1995</year>
;
<volume>29</volume>
:
<fpage>389</fpage>
<lpage>401</lpage>
.
<pub-id pub-id-type="pmid">7542245</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b28">
<label>28.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Kim</surname>
<given-names>MJ,</given-names>
</name>
<name>
<surname>Kim</surname>
<given-names>CW,</given-names>
</name>
<name>
<surname>Lim</surname>
<given-names>YJ,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Microrough titanium surface affects biologic response in MG63 osteoblast-like cells</article-title>
.
<source>
<italic>J Biomed Mater Res A</italic>
</source>
.
<year>2006</year>
;
<volume>79</volume>
:
<fpage>1023</fpage>
<lpage>1032</lpage>
.
<pub-id pub-id-type="pmid">17034031</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b29">
<label>29.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Jeong</surname>
<given-names>KJ,</given-names>
</name>
<name>
<surname>Kohane</surname>
<given-names>DS.</given-names>
</name>
</person-group>
<article-title>Surface modification and drug delivery for biointegration</article-title>
.
<source>
<italic>Ther Deliv</italic>
</source>
.
<year>2011</year>
;
<volume>2</volume>
:
<fpage>737</fpage>
<lpage>752</lpage>
.
<pub-id pub-id-type="pmid">22822506</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b30">
<label>30.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Crnej</surname>
<given-names>A,</given-names>
</name>
<name>
<surname>Paschalis</surname>
<given-names>EI,</given-names>
</name>
<name>
<surname>Salvador-Culla</surname>
<given-names>B,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Glaucoma progression and role of glaucoma surgery in patients with Boston keratoprosthesis</article-title>
.
<source>
<italic>Cornea</italic>
</source>
.
<year>2014</year>
;
<volume>33</volume>
:
<fpage>349</fpage>
<lpage>354</lpage>
.
<pub-id pub-id-type="pmid">24531120</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b31">
<label>31.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Zerbe</surname>
<given-names>BL,</given-names>
</name>
<name>
<surname>Belin</surname>
<given-names>MW,</given-names>
</name>
<name>
<surname>Ciolino</surname>
<given-names>JB,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Results from the multicenter Boston Type 1 Keratoprosthesis Study</article-title>
.
<source>
<italic>Ophthalmology</italic>
</source>
.
<year>2006</year>
;
<volume>113</volume>
:
<fpage>1779.e1</fpage>
<lpage>e7</lpage>
.
<pub-id pub-id-type="pmid">16872678</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b32">
<label>32.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Aldave</surname>
<given-names>AJ,</given-names>
</name>
<name>
<surname>Sangwan</surname>
<given-names>VS,</given-names>
</name>
<name>
<surname>Basu</surname>
<given-names>S,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>International results with the Boston type I keratoprosthesis</article-title>
.
<source>
<italic>Ophthalmology</italic>
</source>
.
<year>2012</year>
;
<volume>119</volume>
:
<fpage>1530</fpage>
<lpage>1538</lpage>
.
<pub-id pub-id-type="pmid">22512986</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b33">
<label>33.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Sivaraman</surname>
<given-names>KR,</given-names>
</name>
<name>
<surname>Hou</surname>
<given-names>JH,</given-names>
</name>
<name>
<surname>Allemann</surname>
<given-names>N,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Retroprosthetic membrane and risk of sterile keratolysis in patients with type I Boston Keratoprosthesis</article-title>
.
<source>
<italic>Am J Ophthalmol</italic>
</source>
.
<year>2013</year>
;
<volume>155</volume>
:
<fpage>814</fpage>
<lpage>822</lpage>
.
<pub-id pub-id-type="pmid">23352344</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b34">
<label>34.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Srikumaran</surname>
<given-names>D,</given-names>
</name>
<name>
<surname>Munoz</surname>
<given-names>B,</given-names>
</name>
<name>
<surname>Aldave</surname>
<given-names>AJ,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Long-term outcomes of Boston Type 1 keratoprosthesis implantation: a retrospective multicenter cohort</article-title>
.
<source>
<italic>Ophthalmology</italic>
</source>
.
<year>2014</year>
;
<volume>121</volume>
:
<fpage>2159</fpage>
<lpage>2164</lpage>
.
<pub-id pub-id-type="pmid">25017414</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b35">
<label>35.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Ciolino</surname>
<given-names>JB,</given-names>
</name>
<name>
<surname>Belin</surname>
<given-names>MW,</given-names>
</name>
<name>
<surname>Todani</surname>
<given-names>A,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Retention of the Boston keratoprosthesis type 1: multicenter study results</article-title>
.
<source>
<italic>Ophthalmology</italic>
</source>
.
<year>2013</year>
;
<volume>120</volume>
:
<fpage>1195</fpage>
<lpage>1200</lpage>
.
<pub-id pub-id-type="pmid">23499061</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b36">
<label>36.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Durand</surname>
<given-names>ML,</given-names>
</name>
<name>
<surname>Dohlman</surname>
<given-names>CH.</given-names>
</name>
</person-group>
<article-title>Successful prevention of bacterial endophthalmitis in eyes with the Boston keratoprosthesis</article-title>
.
<source>
<italic>Cornea</italic>
</source>
.
<year>2009</year>
;
<volume>28</volume>
:
<fpage>896</fpage>
<lpage>901</lpage>
.
<pub-id pub-id-type="pmid">19654525</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b37">
<label>37.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Jain</surname>
<given-names>V,</given-names>
</name>
<name>
<surname>Mhatre</surname>
<given-names>K,</given-names>
</name>
<name>
<surname>Shome</surname>
<given-names>D,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Fungal keratitis with the type 1 Boston keratoprosthesis: early Indian experience</article-title>
.
<source>
<italic>Cornea</italic>
</source>
.
<year>2012</year>
;
<volume>31</volume>
:
<fpage>841</fpage>
<lpage>843</lpage>
.
<pub-id pub-id-type="pmid">22531432</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b38">
<label>38.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Andrei</surname>
<given-names>C,</given-names>
</name>
<name>
<surname>Lestini</surname>
<given-names>E,</given-names>
</name>
<name>
<surname>Crosbie</surname>
<given-names>S,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Plasmonic enhancement of dye sensitized solar cells via a tailored size-distribution of chemically functionalized gold nanoparticles</article-title>
.
<source>
<italic>PLoS One</italic>
</source>
.
<year>2014</year>
;
<volume>9</volume>
:
<fpage>e109836</fpage>
.
<pub-id pub-id-type="pmid">25354362</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b39">
<label>39.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Terriza</surname>
<given-names>A,</given-names>
</name>
<name>
<surname>Díaz-Cuenca</surname>
<given-names>A,</given-names>
</name>
<name>
<surname>Yubero</surname>
<given-names>F,</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Light induced hydrophilicity and osteoblast adhesion promotion on amorphous TiO2</article-title>
.
<source>
<italic>J Biomed Mater Res A</italic>
</source>
.
<year>2013</year>
;
<volume>101</volume>
:
<fpage>1026</fpage>
<lpage>1035</lpage>
.
<pub-id pub-id-type="pmid">22965473</pub-id>
</mixed-citation>
</ref>
<ref id="i2164-2591-5-2-17-b40">
<label>40.</label>
<mixed-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Teixeira</surname>
<given-names>AI,</given-names>
</name>
<name>
<surname>Nealey</surname>
<given-names>PF,</given-names>
</name>
<name>
<surname>Murphy</surname>
<given-names>CJ.</given-names>
</name>
</person-group>
<article-title>Responses of human keratocytes to micro- and nanostructured substrates</article-title>
.
<source>
<italic>J Biomed Mater Res A</italic>
</source>
.
<year>2004</year>
;
<volume>71</volume>
:
<fpage>369</fpage>
<lpage>376</lpage>
.
<pub-id pub-id-type="pmid">15470741</pub-id>
</mixed-citation>
</ref>
</ref-list>
</back>
</pmc>
</record>

Pour manipuler ce document sous Unix (Dilib)

EXPLOR_STEP=$WICRI_ROOT/Ticri/CIDE/explor/HapticV1/Data/Pmc/Curation
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 000580 | SxmlIndent | more

Ou

HfdSelect -h $EXPLOR_AREA/Data/Pmc/Curation/biblio.hfd -nk 000580 | SxmlIndent | more

Pour mettre un lien sur cette page dans le réseau Wicri

{{Explor lien
   |wiki=    Ticri/CIDE
   |area=    HapticV1
   |flux=    Pmc
   |étape=   Curation
   |type=    RBID
   |clé=     PMC:4855478
   |texte=   Titanium Coating of the Boston Keratoprosthesis
}}

Pour générer des pages wiki

HfdIndexSelect -h $EXPLOR_AREA/Data/Pmc/Curation/RBID.i   -Sk "pubmed:27152247" \
       | HfdSelect -Kh $EXPLOR_AREA/Data/Pmc/Curation/biblio.hfd   \
       | NlmPubMed2Wicri -a HapticV1 

Wicri

This area was generated with Dilib version V0.6.23.
Data generation: Mon Jun 13 01:09:46 2016. Site generation: Wed Mar 6 09:54:07 2024