Serveur d'exploration sur les dispositifs haptiques

Attention, ce site est en cours de développement !
Attention, site généré par des moyens informatiques à partir de corpus bruts.
Les informations ne sont donc pas validées.

Forces Applied during Transvenous Implantable Cardioverter Defibrillator Lead Removal

Identifieur interne : 001C08 ( Pmc/Corpus ); précédent : 001C07; suivant : 001C09

Forces Applied during Transvenous Implantable Cardioverter Defibrillator Lead Removal

Auteurs : Carsten Lennerz ; Herribert Pavaci ; Christian Grebmer ; Gesa Von Olshausen ; Verena Semmler ; Alessandra Buiatti ; Tilko Reents ; Sonia Ammar ; Isabel Deisenhofer ; Christof Kolb

Source :

RBID : PMC:4055293

Abstract

Methods. 17 physicians, experienced in transvenous lead removal, performed a lead extraction manoeuvre of an ICD lead on a torso phantom. They were advised to stop traction only when further traction would be considered as harmful to the patient or when—based on their experience—a change in the extraction strategy was indicated. Traction forces were recorded with a digital precision gauge. Results. Median traction forces on the endocardium were 10.9 N (range from 3.0 N to 24.7 N and interquartile range from 7.9 to 15.3). Forces applied to the proximal end were estimated to be 10% higher than those measured at the tip of the lead due to a friction loss. Conclusion. A traction force of around 11 N is typically exerted during standard transvenous extraction of ICD leads. A traction threshold for a safe procedure derived from a pool of experienced extractionists may be helpful for the development of required adequate simulator trainings.


Url:
DOI: 10.1155/2014/183483
PubMed: 24967337
PubMed Central: 4055293

Links to Exploration step

PMC:4055293***** Acces problem to record *****\

Le document en format XML


Pour manipuler ce document sous Unix (Dilib)

EXPLOR_STEP=$WICRI_ROOT/Ticri/CIDE/explor/HapticV1/Data/Pmc/Corpus
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 001C08 | SxmlIndent | more

Ou

HfdSelect -h $EXPLOR_AREA/Data/Pmc/Corpus/biblio.hfd -nk 001C08 | SxmlIndent | more

Pour mettre un lien sur cette page dans le réseau Wicri

{{Explor lien
   |wiki=    Ticri/CIDE
   |area=    HapticV1
   |flux=    Pmc
   |étape=   Corpus
   |type=    RBID
   |clé=     PMC:4055293
   |texte=   Forces Applied during Transvenous Implantable Cardioverter Defibrillator Lead Removal
}}

Pour générer des pages wiki

HfdIndexSelect -h $EXPLOR_AREA/Data/Pmc/Corpus/RBID.i   -Sk "pubmed:24967337" \
       | HfdSelect -Kh $EXPLOR_AREA/Data/Pmc/Corpus/biblio.hfd   \
       | NlmPubMed2Wicri -a HapticV1 

Wicri

This area was generated with Dilib version V0.6.23.
Data generation: Mon Jun 13 01:09:46 2016. Site generation: Wed Mar 6 09:54:07 2024