Early rotational stability of the longer Staar toric intraocular lens: Fifty consecutive cases
Identifieur interne : 000365 ( PascalFrancis/Curation ); précédent : 000364; suivant : 000366Early rotational stability of the longer Staar toric intraocular lens: Fifty consecutive cases
Auteurs : David F. Chang [États-Unis]Source :
- Journal of cataract and refractive surgery [ 0886-3350 ] ; 2003.
Descripteurs français
- Pascal (Inist)
- Wicri :
- topic : Homme.
English descriptors
- KwdEn :
Abstract
Purpose: To determine whether rotational stability is improved with the longer Staar toric intraocular lens (IOL). Setting: A private practice, Los Altos, California, USA. Methods: Staar Surgical manufactures toric plate-haptic lOLs under +24.0 diopters (D) in 2 lengths. Fifty consecutive eyes requiring toric IOL spherical powers less than +24.0 D received the longer toric IOL (TL; 11.2 mm) using a specific surgical protocol. The IOL axis was recorded at 1 day, 1 week, and a final follow-up visit. Five additional eyes requiring powers greater than +24.0 D received the shorter toric IOL (TF; 10.8 mm), which is the only available length in these powers. Results were compared with those in an initial consecutive series of 6 eyes having implantation of the TF IOL with a power less than +24.0 D before the TL model became available. Results: Rotational stability of all 50 TL lOLs and the 5 TF lOLs of higher dioptric powers was excellent. One IOL rotated as much as 20 degrees off axis. The repositioning rate was zero in both groups. In the initial 6 eyes receiving the shorter toric IOL in powers below +24.0 D, the repositioning rate was 50%. Conclusions: The longer toric IOL had excellent early rotational stability. Rotational and repositioning rates with the TL IOL were superior to those in a previous series of the TF IOL. Adequate overall length is a critical factor in the rotational stability of plate-haptic toric lOLs, and the longer IOL should be used when available.
pA |
|
---|
Links toward previous steps (curation, corpus...)
- to stream PascalFrancis, to step Corpus: Pour aller vers cette notice dans l'étape Curation :001146
Links to Exploration step
Pascal:03-0396582Le document en format XML
<record><TEI><teiHeader><fileDesc><titleStmt><title xml:lang="en" level="a">Early rotational stability of the longer Staar toric intraocular lens: Fifty consecutive cases</title>
<author><name sortKey="Chang, David F" sort="Chang, David F" uniqKey="Chang D" first="David F." last="Chang">David F. Chang</name>
<affiliation wicri:level="1"><inist:fA14 i1="01"><s2>762 Altos Oaks Drive, Suite 1, Los Altos, California 94024</s2>
<s3>USA</s3>
<sZ>1 aut.</sZ>
</inist:fA14>
<country>États-Unis</country>
</affiliation>
</author>
</titleStmt>
<publicationStmt><idno type="wicri:source">INIST</idno>
<idno type="inist">03-0396582</idno>
<date when="2003">2003</date>
<idno type="stanalyst">PASCAL 03-0396582 INIST</idno>
<idno type="RBID">Pascal:03-0396582</idno>
<idno type="wicri:Area/PascalFrancis/Corpus">001146</idno>
<idno type="wicri:Area/PascalFrancis/Curation">000365</idno>
</publicationStmt>
<sourceDesc><biblStruct><analytic><title xml:lang="en" level="a">Early rotational stability of the longer Staar toric intraocular lens: Fifty consecutive cases</title>
<author><name sortKey="Chang, David F" sort="Chang, David F" uniqKey="Chang D" first="David F." last="Chang">David F. Chang</name>
<affiliation wicri:level="1"><inist:fA14 i1="01"><s2>762 Altos Oaks Drive, Suite 1, Los Altos, California 94024</s2>
<s3>USA</s3>
<sZ>1 aut.</sZ>
</inist:fA14>
<country>États-Unis</country>
</affiliation>
</author>
</analytic>
<series><title level="j" type="main">Journal of cataract and refractive surgery</title>
<title level="j" type="abbreviated">J. cataract refractive surg.</title>
<idno type="ISSN">0886-3350</idno>
<imprint><date when="2003">2003</date>
</imprint>
</series>
</biblStruct>
</sourceDesc>
<seriesStmt><title level="j" type="main">Journal of cataract and refractive surgery</title>
<title level="j" type="abbreviated">J. cataract refractive surg.</title>
<idno type="ISSN">0886-3350</idno>
</seriesStmt>
</fileDesc>
<profileDesc><textClass><keywords scheme="KwdEn" xml:lang="en"><term>Early</term>
<term>Eye</term>
<term>Human</term>
<term>Intraocular</term>
<term>Intraocular lens</term>
<term>Postoperative</term>
<term>Stability</term>
<term>Toric lens</term>
</keywords>
<keywords scheme="Pascal" xml:lang="fr"><term>Postopératoire</term>
<term>Homme</term>
<term>Oeil</term>
<term>Intraoculaire</term>
<term>Précoce</term>
<term>Stabilité</term>
<term>Lentille torique</term>
<term>Lentille intraoculaire</term>
</keywords>
<keywords scheme="Wicri" type="topic" xml:lang="fr"><term>Homme</term>
</keywords>
</textClass>
</profileDesc>
</teiHeader>
<front><div type="abstract" xml:lang="en">Purpose: To determine whether rotational stability is improved with the longer Staar toric intraocular lens (IOL). Setting: A private practice, Los Altos, California, USA. Methods: Staar Surgical manufactures toric plate-haptic lOLs under +24.0 diopters (D) in 2 lengths. Fifty consecutive eyes requiring toric IOL spherical powers less than +24.0 D received the longer toric IOL (TL; 11.2 mm) using a specific surgical protocol. The IOL axis was recorded at 1 day, 1 week, and a final follow-up visit. Five additional eyes requiring powers greater than +24.0 D received the shorter toric IOL (TF; 10.8 mm), which is the only available length in these powers. Results were compared with those in an initial consecutive series of 6 eyes having implantation of the TF IOL with a power less than +24.0 D before the TL model became available. Results: Rotational stability of all 50 TL lOLs and the 5 TF lOLs of higher dioptric powers was excellent. One IOL rotated as much as 20 degrees off axis. The repositioning rate was zero in both groups. In the initial 6 eyes receiving the shorter toric IOL in powers below +24.0 D, the repositioning rate was 50%. Conclusions: The longer toric IOL had excellent early rotational stability. Rotational and repositioning rates with the TL IOL were superior to those in a previous series of the TF IOL. Adequate overall length is a critical factor in the rotational stability of plate-haptic toric lOLs, and the longer IOL should be used when available.</div>
</front>
</TEI>
<inist><standard h6="B"><pA><fA01 i1="01" i2="1"><s0>0886-3350</s0>
</fA01>
<fA02 i1="01"><s0>JCSUEV</s0>
</fA02>
<fA03 i2="1"><s0>J. cataract refractive surg.</s0>
</fA03>
<fA05><s2>29</s2>
</fA05>
<fA06><s2>5</s2>
</fA06>
<fA08 i1="01" i2="1" l="ENG"><s1>Early rotational stability of the longer Staar toric intraocular lens: Fifty consecutive cases</s1>
</fA08>
<fA11 i1="01" i2="1"><s1>CHANG (David F.)</s1>
</fA11>
<fA14 i1="01"><s2>762 Altos Oaks Drive, Suite 1, Los Altos, California 94024</s2>
<s3>USA</s3>
<sZ>1 aut.</sZ>
</fA14>
<fA20><s1>935-940</s1>
</fA20>
<fA21><s1>2003</s1>
</fA21>
<fA23 i1="01"><s0>ENG</s0>
</fA23>
<fA43 i1="01"><s1>INIST</s1>
<s2>20937</s2>
<s5>354000118296380130</s5>
</fA43>
<fA44><s0>0000</s0>
<s1>© 2003 INIST-CNRS. All rights reserved.</s1>
</fA44>
<fA45><s0>14 ref.</s0>
</fA45>
<fA47 i1="01" i2="1"><s0>03-0396582</s0>
</fA47>
<fA60><s1>P</s1>
</fA60>
<fA61><s0>A</s0>
</fA61>
<fA64 i1="01" i2="1"><s0>Journal of cataract and refractive surgery</s0>
</fA64>
<fA66 i1="01"><s0>USA</s0>
</fA66>
<fC01 i1="01" l="ENG"><s0>Purpose: To determine whether rotational stability is improved with the longer Staar toric intraocular lens (IOL). Setting: A private practice, Los Altos, California, USA. Methods: Staar Surgical manufactures toric plate-haptic lOLs under +24.0 diopters (D) in 2 lengths. Fifty consecutive eyes requiring toric IOL spherical powers less than +24.0 D received the longer toric IOL (TL; 11.2 mm) using a specific surgical protocol. The IOL axis was recorded at 1 day, 1 week, and a final follow-up visit. Five additional eyes requiring powers greater than +24.0 D received the shorter toric IOL (TF; 10.8 mm), which is the only available length in these powers. Results were compared with those in an initial consecutive series of 6 eyes having implantation of the TF IOL with a power less than +24.0 D before the TL model became available. Results: Rotational stability of all 50 TL lOLs and the 5 TF lOLs of higher dioptric powers was excellent. One IOL rotated as much as 20 degrees off axis. The repositioning rate was zero in both groups. In the initial 6 eyes receiving the shorter toric IOL in powers below +24.0 D, the repositioning rate was 50%. Conclusions: The longer toric IOL had excellent early rotational stability. Rotational and repositioning rates with the TL IOL were superior to those in a previous series of the TF IOL. Adequate overall length is a critical factor in the rotational stability of plate-haptic toric lOLs, and the longer IOL should be used when available.</s0>
</fC01>
<fC02 i1="01" i2="X"><s0>002B25B</s0>
</fC02>
<fC03 i1="01" i2="X" l="FRE"><s0>Postopératoire</s0>
<s5>01</s5>
</fC03>
<fC03 i1="01" i2="X" l="ENG"><s0>Postoperative</s0>
<s5>01</s5>
</fC03>
<fC03 i1="01" i2="X" l="SPA"><s0>Postoperatorio</s0>
<s5>01</s5>
</fC03>
<fC03 i1="02" i2="X" l="FRE"><s0>Homme</s0>
<s5>02</s5>
</fC03>
<fC03 i1="02" i2="X" l="ENG"><s0>Human</s0>
<s5>02</s5>
</fC03>
<fC03 i1="02" i2="X" l="SPA"><s0>Hombre</s0>
<s5>02</s5>
</fC03>
<fC03 i1="03" i2="X" l="FRE"><s0>Oeil</s0>
<s5>03</s5>
</fC03>
<fC03 i1="03" i2="X" l="ENG"><s0>Eye</s0>
<s5>03</s5>
</fC03>
<fC03 i1="03" i2="X" l="SPA"><s0>Ojo</s0>
<s5>03</s5>
</fC03>
<fC03 i1="04" i2="X" l="FRE"><s0>Intraoculaire</s0>
<s5>04</s5>
</fC03>
<fC03 i1="04" i2="X" l="ENG"><s0>Intraocular</s0>
<s5>04</s5>
</fC03>
<fC03 i1="04" i2="X" l="SPA"><s0>Intraocular</s0>
<s5>04</s5>
</fC03>
<fC03 i1="05" i2="X" l="FRE"><s0>Précoce</s0>
<s5>08</s5>
</fC03>
<fC03 i1="05" i2="X" l="ENG"><s0>Early</s0>
<s5>08</s5>
</fC03>
<fC03 i1="05" i2="X" l="SPA"><s0>Precoz</s0>
<s5>08</s5>
</fC03>
<fC03 i1="06" i2="X" l="FRE"><s0>Stabilité</s0>
<s5>09</s5>
</fC03>
<fC03 i1="06" i2="X" l="ENG"><s0>Stability</s0>
<s5>09</s5>
</fC03>
<fC03 i1="06" i2="X" l="SPA"><s0>Estabilidad</s0>
<s5>09</s5>
</fC03>
<fC03 i1="07" i2="X" l="FRE"><s0>Lentille torique</s0>
<s5>11</s5>
</fC03>
<fC03 i1="07" i2="X" l="ENG"><s0>Toric lens</s0>
<s5>11</s5>
</fC03>
<fC03 i1="07" i2="X" l="SPA"><s0>Lente tórico</s0>
<s5>11</s5>
</fC03>
<fC03 i1="08" i2="X" l="FRE"><s0>Lentille intraoculaire</s0>
<s5>12</s5>
</fC03>
<fC03 i1="08" i2="X" l="ENG"><s0>Intraocular lens</s0>
<s5>12</s5>
</fC03>
<fC03 i1="08" i2="X" l="SPA"><s0>Lente intraocular</s0>
<s5>12</s5>
</fC03>
<fN21><s1>279</s1>
</fN21>
<fN82><s1>PSI</s1>
</fN82>
</pA>
</standard>
</inist>
</record>
Pour manipuler ce document sous Unix (Dilib)
EXPLOR_STEP=$WICRI_ROOT/Ticri/CIDE/explor/HapticV1/Data/PascalFrancis/Curation
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 000365 | SxmlIndent | more
Ou
HfdSelect -h $EXPLOR_AREA/Data/PascalFrancis/Curation/biblio.hfd -nk 000365 | SxmlIndent | more
Pour mettre un lien sur cette page dans le réseau Wicri
{{Explor lien |wiki= Ticri/CIDE |area= HapticV1 |flux= PascalFrancis |étape= Curation |type= RBID |clé= Pascal:03-0396582 |texte= Early rotational stability of the longer Staar toric intraocular lens: Fifty consecutive cases }}
This area was generated with Dilib version V0.6.23. |