Serveur d'exploration sur les dispositifs haptiques

Attention, ce site est en cours de développement !
Attention, site généré par des moyens informatiques à partir de corpus bruts.
Les informations ne sont donc pas validées.

Early rotational stability of the longer Staar toric intraocular lens: Fifty consecutive cases

Identifieur interne : 001146 ( PascalFrancis/Corpus ); précédent : 001145; suivant : 001147

Early rotational stability of the longer Staar toric intraocular lens: Fifty consecutive cases

Auteurs : David F. Chang

Source :

RBID : Pascal:03-0396582

Descripteurs français

English descriptors

Abstract

Purpose: To determine whether rotational stability is improved with the longer Staar toric intraocular lens (IOL). Setting: A private practice, Los Altos, California, USA. Methods: Staar Surgical manufactures toric plate-haptic lOLs under +24.0 diopters (D) in 2 lengths. Fifty consecutive eyes requiring toric IOL spherical powers less than +24.0 D received the longer toric IOL (TL; 11.2 mm) using a specific surgical protocol. The IOL axis was recorded at 1 day, 1 week, and a final follow-up visit. Five additional eyes requiring powers greater than +24.0 D received the shorter toric IOL (TF; 10.8 mm), which is the only available length in these powers. Results were compared with those in an initial consecutive series of 6 eyes having implantation of the TF IOL with a power less than +24.0 D before the TL model became available. Results: Rotational stability of all 50 TL lOLs and the 5 TF lOLs of higher dioptric powers was excellent. One IOL rotated as much as 20 degrees off axis. The repositioning rate was zero in both groups. In the initial 6 eyes receiving the shorter toric IOL in powers below +24.0 D, the repositioning rate was 50%. Conclusions: The longer toric IOL had excellent early rotational stability. Rotational and repositioning rates with the TL IOL were superior to those in a previous series of the TF IOL. Adequate overall length is a critical factor in the rotational stability of plate-haptic toric lOLs, and the longer IOL should be used when available.

Notice en format standard (ISO 2709)

Pour connaître la documentation sur le format Inist Standard.

pA  
A01 01  1    @0 0886-3350
A02 01      @0 JCSUEV
A03   1    @0 J. cataract refractive surg.
A05       @2 29
A06       @2 5
A08 01  1  ENG  @1 Early rotational stability of the longer Staar toric intraocular lens: Fifty consecutive cases
A11 01  1    @1 CHANG (David F.)
A14 01      @2 762 Altos Oaks Drive, Suite 1, Los Altos, California 94024 @3 USA @Z 1 aut.
A20       @1 935-940
A21       @1 2003
A23 01      @0 ENG
A43 01      @1 INIST @2 20937 @5 354000118296380130
A44       @0 0000 @1 © 2003 INIST-CNRS. All rights reserved.
A45       @0 14 ref.
A47 01  1    @0 03-0396582
A60       @1 P
A61       @0 A
A64 01  1    @0 Journal of cataract and refractive surgery
A66 01      @0 USA
C01 01    ENG  @0 Purpose: To determine whether rotational stability is improved with the longer Staar toric intraocular lens (IOL). Setting: A private practice, Los Altos, California, USA. Methods: Staar Surgical manufactures toric plate-haptic lOLs under +24.0 diopters (D) in 2 lengths. Fifty consecutive eyes requiring toric IOL spherical powers less than +24.0 D received the longer toric IOL (TL; 11.2 mm) using a specific surgical protocol. The IOL axis was recorded at 1 day, 1 week, and a final follow-up visit. Five additional eyes requiring powers greater than +24.0 D received the shorter toric IOL (TF; 10.8 mm), which is the only available length in these powers. Results were compared with those in an initial consecutive series of 6 eyes having implantation of the TF IOL with a power less than +24.0 D before the TL model became available. Results: Rotational stability of all 50 TL lOLs and the 5 TF lOLs of higher dioptric powers was excellent. One IOL rotated as much as 20 degrees off axis. The repositioning rate was zero in both groups. In the initial 6 eyes receiving the shorter toric IOL in powers below +24.0 D, the repositioning rate was 50%. Conclusions: The longer toric IOL had excellent early rotational stability. Rotational and repositioning rates with the TL IOL were superior to those in a previous series of the TF IOL. Adequate overall length is a critical factor in the rotational stability of plate-haptic toric lOLs, and the longer IOL should be used when available.
C02 01  X    @0 002B25B
C03 01  X  FRE  @0 Postopératoire @5 01
C03 01  X  ENG  @0 Postoperative @5 01
C03 01  X  SPA  @0 Postoperatorio @5 01
C03 02  X  FRE  @0 Homme @5 02
C03 02  X  ENG  @0 Human @5 02
C03 02  X  SPA  @0 Hombre @5 02
C03 03  X  FRE  @0 Oeil @5 03
C03 03  X  ENG  @0 Eye @5 03
C03 03  X  SPA  @0 Ojo @5 03
C03 04  X  FRE  @0 Intraoculaire @5 04
C03 04  X  ENG  @0 Intraocular @5 04
C03 04  X  SPA  @0 Intraocular @5 04
C03 05  X  FRE  @0 Précoce @5 08
C03 05  X  ENG  @0 Early @5 08
C03 05  X  SPA  @0 Precoz @5 08
C03 06  X  FRE  @0 Stabilité @5 09
C03 06  X  ENG  @0 Stability @5 09
C03 06  X  SPA  @0 Estabilidad @5 09
C03 07  X  FRE  @0 Lentille torique @5 11
C03 07  X  ENG  @0 Toric lens @5 11
C03 07  X  SPA  @0 Lente tórico @5 11
C03 08  X  FRE  @0 Lentille intraoculaire @5 12
C03 08  X  ENG  @0 Intraocular lens @5 12
C03 08  X  SPA  @0 Lente intraocular @5 12
N21       @1 279
N82       @1 PSI

Format Inist (serveur)

NO : PASCAL 03-0396582 INIST
ET : Early rotational stability of the longer Staar toric intraocular lens: Fifty consecutive cases
AU : CHANG (David F.)
AF : 762 Altos Oaks Drive, Suite 1, Los Altos, California 94024/Etats-Unis (1 aut.)
DT : Publication en série; Niveau analytique
SO : Journal of cataract and refractive surgery; ISSN 0886-3350; Coden JCSUEV; Etats-Unis; Da. 2003; Vol. 29; No. 5; Pp. 935-940; Bibl. 14 ref.
LA : Anglais
EA : Purpose: To determine whether rotational stability is improved with the longer Staar toric intraocular lens (IOL). Setting: A private practice, Los Altos, California, USA. Methods: Staar Surgical manufactures toric plate-haptic lOLs under +24.0 diopters (D) in 2 lengths. Fifty consecutive eyes requiring toric IOL spherical powers less than +24.0 D received the longer toric IOL (TL; 11.2 mm) using a specific surgical protocol. The IOL axis was recorded at 1 day, 1 week, and a final follow-up visit. Five additional eyes requiring powers greater than +24.0 D received the shorter toric IOL (TF; 10.8 mm), which is the only available length in these powers. Results were compared with those in an initial consecutive series of 6 eyes having implantation of the TF IOL with a power less than +24.0 D before the TL model became available. Results: Rotational stability of all 50 TL lOLs and the 5 TF lOLs of higher dioptric powers was excellent. One IOL rotated as much as 20 degrees off axis. The repositioning rate was zero in both groups. In the initial 6 eyes receiving the shorter toric IOL in powers below +24.0 D, the repositioning rate was 50%. Conclusions: The longer toric IOL had excellent early rotational stability. Rotational and repositioning rates with the TL IOL were superior to those in a previous series of the TF IOL. Adequate overall length is a critical factor in the rotational stability of plate-haptic toric lOLs, and the longer IOL should be used when available.
CC : 002B25B
FD : Postopératoire; Homme; Oeil; Intraoculaire; Précoce; Stabilité; Lentille torique; Lentille intraoculaire
ED : Postoperative; Human; Eye; Intraocular; Early; Stability; Toric lens; Intraocular lens
SD : Postoperatorio; Hombre; Ojo; Intraocular; Precoz; Estabilidad; Lente tórico; Lente intraocular
LO : INIST-20937.354000118296380130
ID : 03-0396582

Links to Exploration step

Pascal:03-0396582

Le document en format XML

<record>
<TEI>
<teiHeader>
<fileDesc>
<titleStmt>
<title xml:lang="en" level="a">Early rotational stability of the longer Staar toric intraocular lens: Fifty consecutive cases</title>
<author>
<name sortKey="Chang, David F" sort="Chang, David F" uniqKey="Chang D" first="David F." last="Chang">David F. Chang</name>
<affiliation>
<inist:fA14 i1="01">
<s2>762 Altos Oaks Drive, Suite 1, Los Altos, California 94024</s2>
<s3>USA</s3>
<sZ>1 aut.</sZ>
</inist:fA14>
</affiliation>
</author>
</titleStmt>
<publicationStmt>
<idno type="wicri:source">INIST</idno>
<idno type="inist">03-0396582</idno>
<date when="2003">2003</date>
<idno type="stanalyst">PASCAL 03-0396582 INIST</idno>
<idno type="RBID">Pascal:03-0396582</idno>
<idno type="wicri:Area/PascalFrancis/Corpus">001146</idno>
</publicationStmt>
<sourceDesc>
<biblStruct>
<analytic>
<title xml:lang="en" level="a">Early rotational stability of the longer Staar toric intraocular lens: Fifty consecutive cases</title>
<author>
<name sortKey="Chang, David F" sort="Chang, David F" uniqKey="Chang D" first="David F." last="Chang">David F. Chang</name>
<affiliation>
<inist:fA14 i1="01">
<s2>762 Altos Oaks Drive, Suite 1, Los Altos, California 94024</s2>
<s3>USA</s3>
<sZ>1 aut.</sZ>
</inist:fA14>
</affiliation>
</author>
</analytic>
<series>
<title level="j" type="main">Journal of cataract and refractive surgery</title>
<title level="j" type="abbreviated">J. cataract refractive surg.</title>
<idno type="ISSN">0886-3350</idno>
<imprint>
<date when="2003">2003</date>
</imprint>
</series>
</biblStruct>
</sourceDesc>
<seriesStmt>
<title level="j" type="main">Journal of cataract and refractive surgery</title>
<title level="j" type="abbreviated">J. cataract refractive surg.</title>
<idno type="ISSN">0886-3350</idno>
</seriesStmt>
</fileDesc>
<profileDesc>
<textClass>
<keywords scheme="KwdEn" xml:lang="en">
<term>Early</term>
<term>Eye</term>
<term>Human</term>
<term>Intraocular</term>
<term>Intraocular lens</term>
<term>Postoperative</term>
<term>Stability</term>
<term>Toric lens</term>
</keywords>
<keywords scheme="Pascal" xml:lang="fr">
<term>Postopératoire</term>
<term>Homme</term>
<term>Oeil</term>
<term>Intraoculaire</term>
<term>Précoce</term>
<term>Stabilité</term>
<term>Lentille torique</term>
<term>Lentille intraoculaire</term>
</keywords>
</textClass>
</profileDesc>
</teiHeader>
<front>
<div type="abstract" xml:lang="en">Purpose: To determine whether rotational stability is improved with the longer Staar toric intraocular lens (IOL). Setting: A private practice, Los Altos, California, USA. Methods: Staar Surgical manufactures toric plate-haptic lOLs under +24.0 diopters (D) in 2 lengths. Fifty consecutive eyes requiring toric IOL spherical powers less than +24.0 D received the longer toric IOL (TL; 11.2 mm) using a specific surgical protocol. The IOL axis was recorded at 1 day, 1 week, and a final follow-up visit. Five additional eyes requiring powers greater than +24.0 D received the shorter toric IOL (TF; 10.8 mm), which is the only available length in these powers. Results were compared with those in an initial consecutive series of 6 eyes having implantation of the TF IOL with a power less than +24.0 D before the TL model became available. Results: Rotational stability of all 50 TL lOLs and the 5 TF lOLs of higher dioptric powers was excellent. One IOL rotated as much as 20 degrees off axis. The repositioning rate was zero in both groups. In the initial 6 eyes receiving the shorter toric IOL in powers below +24.0 D, the repositioning rate was 50%. Conclusions: The longer toric IOL had excellent early rotational stability. Rotational and repositioning rates with the TL IOL were superior to those in a previous series of the TF IOL. Adequate overall length is a critical factor in the rotational stability of plate-haptic toric lOLs, and the longer IOL should be used when available.</div>
</front>
</TEI>
<inist>
<standard h6="B">
<pA>
<fA01 i1="01" i2="1">
<s0>0886-3350</s0>
</fA01>
<fA02 i1="01">
<s0>JCSUEV</s0>
</fA02>
<fA03 i2="1">
<s0>J. cataract refractive surg.</s0>
</fA03>
<fA05>
<s2>29</s2>
</fA05>
<fA06>
<s2>5</s2>
</fA06>
<fA08 i1="01" i2="1" l="ENG">
<s1>Early rotational stability of the longer Staar toric intraocular lens: Fifty consecutive cases</s1>
</fA08>
<fA11 i1="01" i2="1">
<s1>CHANG (David F.)</s1>
</fA11>
<fA14 i1="01">
<s2>762 Altos Oaks Drive, Suite 1, Los Altos, California 94024</s2>
<s3>USA</s3>
<sZ>1 aut.</sZ>
</fA14>
<fA20>
<s1>935-940</s1>
</fA20>
<fA21>
<s1>2003</s1>
</fA21>
<fA23 i1="01">
<s0>ENG</s0>
</fA23>
<fA43 i1="01">
<s1>INIST</s1>
<s2>20937</s2>
<s5>354000118296380130</s5>
</fA43>
<fA44>
<s0>0000</s0>
<s1>© 2003 INIST-CNRS. All rights reserved.</s1>
</fA44>
<fA45>
<s0>14 ref.</s0>
</fA45>
<fA47 i1="01" i2="1">
<s0>03-0396582</s0>
</fA47>
<fA60>
<s1>P</s1>
</fA60>
<fA61>
<s0>A</s0>
</fA61>
<fA64 i1="01" i2="1">
<s0>Journal of cataract and refractive surgery</s0>
</fA64>
<fA66 i1="01">
<s0>USA</s0>
</fA66>
<fC01 i1="01" l="ENG">
<s0>Purpose: To determine whether rotational stability is improved with the longer Staar toric intraocular lens (IOL). Setting: A private practice, Los Altos, California, USA. Methods: Staar Surgical manufactures toric plate-haptic lOLs under +24.0 diopters (D) in 2 lengths. Fifty consecutive eyes requiring toric IOL spherical powers less than +24.0 D received the longer toric IOL (TL; 11.2 mm) using a specific surgical protocol. The IOL axis was recorded at 1 day, 1 week, and a final follow-up visit. Five additional eyes requiring powers greater than +24.0 D received the shorter toric IOL (TF; 10.8 mm), which is the only available length in these powers. Results were compared with those in an initial consecutive series of 6 eyes having implantation of the TF IOL with a power less than +24.0 D before the TL model became available. Results: Rotational stability of all 50 TL lOLs and the 5 TF lOLs of higher dioptric powers was excellent. One IOL rotated as much as 20 degrees off axis. The repositioning rate was zero in both groups. In the initial 6 eyes receiving the shorter toric IOL in powers below +24.0 D, the repositioning rate was 50%. Conclusions: The longer toric IOL had excellent early rotational stability. Rotational and repositioning rates with the TL IOL were superior to those in a previous series of the TF IOL. Adequate overall length is a critical factor in the rotational stability of plate-haptic toric lOLs, and the longer IOL should be used when available.</s0>
</fC01>
<fC02 i1="01" i2="X">
<s0>002B25B</s0>
</fC02>
<fC03 i1="01" i2="X" l="FRE">
<s0>Postopératoire</s0>
<s5>01</s5>
</fC03>
<fC03 i1="01" i2="X" l="ENG">
<s0>Postoperative</s0>
<s5>01</s5>
</fC03>
<fC03 i1="01" i2="X" l="SPA">
<s0>Postoperatorio</s0>
<s5>01</s5>
</fC03>
<fC03 i1="02" i2="X" l="FRE">
<s0>Homme</s0>
<s5>02</s5>
</fC03>
<fC03 i1="02" i2="X" l="ENG">
<s0>Human</s0>
<s5>02</s5>
</fC03>
<fC03 i1="02" i2="X" l="SPA">
<s0>Hombre</s0>
<s5>02</s5>
</fC03>
<fC03 i1="03" i2="X" l="FRE">
<s0>Oeil</s0>
<s5>03</s5>
</fC03>
<fC03 i1="03" i2="X" l="ENG">
<s0>Eye</s0>
<s5>03</s5>
</fC03>
<fC03 i1="03" i2="X" l="SPA">
<s0>Ojo</s0>
<s5>03</s5>
</fC03>
<fC03 i1="04" i2="X" l="FRE">
<s0>Intraoculaire</s0>
<s5>04</s5>
</fC03>
<fC03 i1="04" i2="X" l="ENG">
<s0>Intraocular</s0>
<s5>04</s5>
</fC03>
<fC03 i1="04" i2="X" l="SPA">
<s0>Intraocular</s0>
<s5>04</s5>
</fC03>
<fC03 i1="05" i2="X" l="FRE">
<s0>Précoce</s0>
<s5>08</s5>
</fC03>
<fC03 i1="05" i2="X" l="ENG">
<s0>Early</s0>
<s5>08</s5>
</fC03>
<fC03 i1="05" i2="X" l="SPA">
<s0>Precoz</s0>
<s5>08</s5>
</fC03>
<fC03 i1="06" i2="X" l="FRE">
<s0>Stabilité</s0>
<s5>09</s5>
</fC03>
<fC03 i1="06" i2="X" l="ENG">
<s0>Stability</s0>
<s5>09</s5>
</fC03>
<fC03 i1="06" i2="X" l="SPA">
<s0>Estabilidad</s0>
<s5>09</s5>
</fC03>
<fC03 i1="07" i2="X" l="FRE">
<s0>Lentille torique</s0>
<s5>11</s5>
</fC03>
<fC03 i1="07" i2="X" l="ENG">
<s0>Toric lens</s0>
<s5>11</s5>
</fC03>
<fC03 i1="07" i2="X" l="SPA">
<s0>Lente tórico</s0>
<s5>11</s5>
</fC03>
<fC03 i1="08" i2="X" l="FRE">
<s0>Lentille intraoculaire</s0>
<s5>12</s5>
</fC03>
<fC03 i1="08" i2="X" l="ENG">
<s0>Intraocular lens</s0>
<s5>12</s5>
</fC03>
<fC03 i1="08" i2="X" l="SPA">
<s0>Lente intraocular</s0>
<s5>12</s5>
</fC03>
<fN21>
<s1>279</s1>
</fN21>
<fN82>
<s1>PSI</s1>
</fN82>
</pA>
</standard>
<server>
<NO>PASCAL 03-0396582 INIST</NO>
<ET>Early rotational stability of the longer Staar toric intraocular lens: Fifty consecutive cases</ET>
<AU>CHANG (David F.)</AU>
<AF>762 Altos Oaks Drive, Suite 1, Los Altos, California 94024/Etats-Unis (1 aut.)</AF>
<DT>Publication en série; Niveau analytique</DT>
<SO>Journal of cataract and refractive surgery; ISSN 0886-3350; Coden JCSUEV; Etats-Unis; Da. 2003; Vol. 29; No. 5; Pp. 935-940; Bibl. 14 ref.</SO>
<LA>Anglais</LA>
<EA>Purpose: To determine whether rotational stability is improved with the longer Staar toric intraocular lens (IOL). Setting: A private practice, Los Altos, California, USA. Methods: Staar Surgical manufactures toric plate-haptic lOLs under +24.0 diopters (D) in 2 lengths. Fifty consecutive eyes requiring toric IOL spherical powers less than +24.0 D received the longer toric IOL (TL; 11.2 mm) using a specific surgical protocol. The IOL axis was recorded at 1 day, 1 week, and a final follow-up visit. Five additional eyes requiring powers greater than +24.0 D received the shorter toric IOL (TF; 10.8 mm), which is the only available length in these powers. Results were compared with those in an initial consecutive series of 6 eyes having implantation of the TF IOL with a power less than +24.0 D before the TL model became available. Results: Rotational stability of all 50 TL lOLs and the 5 TF lOLs of higher dioptric powers was excellent. One IOL rotated as much as 20 degrees off axis. The repositioning rate was zero in both groups. In the initial 6 eyes receiving the shorter toric IOL in powers below +24.0 D, the repositioning rate was 50%. Conclusions: The longer toric IOL had excellent early rotational stability. Rotational and repositioning rates with the TL IOL were superior to those in a previous series of the TF IOL. Adequate overall length is a critical factor in the rotational stability of plate-haptic toric lOLs, and the longer IOL should be used when available.</EA>
<CC>002B25B</CC>
<FD>Postopératoire; Homme; Oeil; Intraoculaire; Précoce; Stabilité; Lentille torique; Lentille intraoculaire</FD>
<ED>Postoperative; Human; Eye; Intraocular; Early; Stability; Toric lens; Intraocular lens</ED>
<SD>Postoperatorio; Hombre; Ojo; Intraocular; Precoz; Estabilidad; Lente tórico; Lente intraocular</SD>
<LO>INIST-20937.354000118296380130</LO>
<ID>03-0396582</ID>
</server>
</inist>
</record>

Pour manipuler ce document sous Unix (Dilib)

EXPLOR_STEP=$WICRI_ROOT/Ticri/CIDE/explor/HapticV1/Data/PascalFrancis/Corpus
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 001146 | SxmlIndent | more

Ou

HfdSelect -h $EXPLOR_AREA/Data/PascalFrancis/Corpus/biblio.hfd -nk 001146 | SxmlIndent | more

Pour mettre un lien sur cette page dans le réseau Wicri

{{Explor lien
   |wiki=    Ticri/CIDE
   |area=    HapticV1
   |flux=    PascalFrancis
   |étape=   Corpus
   |type=    RBID
   |clé=     Pascal:03-0396582
   |texte=   Early rotational stability of the longer Staar toric intraocular lens: Fifty consecutive cases
}}

Wicri

This area was generated with Dilib version V0.6.23.
Data generation: Mon Jun 13 01:09:46 2016. Site generation: Wed Mar 6 09:54:07 2024