Capsular bag shrinkage after implantation of an open-loop silicone lens and a poly(methyl methacrylate) capsule tension ring

Identifieur interne : 000229 ( Istex/Corpus ); précédent : 000228; suivant : 000230

Capsular bag shrinkage after implantation of an open-loop silicone lens and a poly(methyl methacrylate) capsule tension ring

Auteurs : Karin Strenn ; Rupert Menapace ; Clemens Vass

Source :

RBID : ISTEX:41C2290A664512053197B87509B758A5667696A8

Abstract

Purpose: To quantify the dynamics of capsular bag shrinkage after cataract surgery.Setting: University Eye Hospital, Vienna, Austria.Methods: Nineteen eyes in 13 patients had clear corneal cataract surgery with implantation of a foldable, open-loop, silicone posterior chamber intraocular lens (IOL) (AMO SI-30). At the same time, an open, poly(methyl methacrylate) capsule tension ring (Morcher type 14 or 14A) was inserted to allow measurement of capsular bag circumference and diameter. After surgery, capsular bag shrinkage was quantified by measuring anterior chamber depth (ACD), iris-lens distance (ILD), and distance between the eyelets of the capsule tension ring using optical methods. Capsular bag circumference (CBC) was deduced from these measurements, which were taken 1 day (baseline), 1 week, and 1 and 3 months postoperatively.Results: Between 1 day and 3 months, all three parameters decreased significantly. During the first postoperative week, ACID and ILD did not change, while both parameters decreased between 1 week and 1 month. After 1 month, ACID did not decrease further; ILD continued to decrease. The CBC decreased during the entire postoperative period.Conclusion: Besides ACID and ILD, CBC significantly decreased during the first 3 months after cataract surgery with in-the-bag posterior chamber IOL implantation. The CBC can be calculated from the distance between the ends of the capsule tension ring, which can be measured gonioscopically. From this, the diameter of the capsular bag can be calculated in the living eye. The influence of the capsule tension ring itself on capsular bag shrinkage remains to be established.

Url:
DOI: 10.1016/S0886-3350(97)80027-2

Links to Exploration step

ISTEX:41C2290A664512053197B87509B758A5667696A8

Le document en format XML

<record>
<TEI wicri:istexFullTextTei="biblStruct">
<teiHeader>
<fileDesc>
<titleStmt>
<title xml:lang="en">Capsular bag shrinkage after implantation of an open-loop silicone lens and a poly(methyl methacrylate) capsule tension ring</title>
<author>
<name sortKey="Strenn, Karin" sort="Strenn, Karin" uniqKey="Strenn K" first="Karin" last="Strenn">Karin Strenn</name>
<affiliation>
<mods:affiliation>From University Eye Hospital, Vienna, Austria.University Eye HospitalViennaAustria</mods:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Menapace, Rupert" sort="Menapace, Rupert" uniqKey="Menapace R" first="Rupert" last="Menapace">Rupert Menapace</name>
<affiliation>
<mods:affiliation>From University Eye Hospital, Vienna, Austria.University Eye HospitalViennaAustria</mods:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Vass, Clemens" sort="Vass, Clemens" uniqKey="Vass C" first="Clemens" last="Vass">Clemens Vass</name>
<affiliation>
<mods:affiliation>From University Eye Hospital, Vienna, Austria.University Eye HospitalViennaAustria</mods:affiliation>
</affiliation>
</author>
</titleStmt>
<publicationStmt>
<idno type="wicri:source">ISTEX</idno>
<idno type="RBID">ISTEX:41C2290A664512053197B87509B758A5667696A8</idno>
<date when="1997" year="1997">1997</date>
<idno type="doi">10.1016/S0886-3350(97)80027-2</idno>
<idno type="url">https://api.istex.fr/document/41C2290A664512053197B87509B758A5667696A8/fulltext/pdf</idno>
<idno type="wicri:Area/Istex/Corpus">000229</idno>
</publicationStmt>
<sourceDesc>
<biblStruct>
<analytic>
<title level="a" type="main" xml:lang="en">Capsular bag shrinkage after implantation of an open-loop silicone lens and a poly(methyl methacrylate) capsule tension ring</title>
<author>
<name sortKey="Strenn, Karin" sort="Strenn, Karin" uniqKey="Strenn K" first="Karin" last="Strenn">Karin Strenn</name>
<affiliation>
<mods:affiliation>From University Eye Hospital, Vienna, Austria.University Eye HospitalViennaAustria</mods:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Menapace, Rupert" sort="Menapace, Rupert" uniqKey="Menapace R" first="Rupert" last="Menapace">Rupert Menapace</name>
<affiliation>
<mods:affiliation>From University Eye Hospital, Vienna, Austria.University Eye HospitalViennaAustria</mods:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Vass, Clemens" sort="Vass, Clemens" uniqKey="Vass C" first="Clemens" last="Vass">Clemens Vass</name>
<affiliation>
<mods:affiliation>From University Eye Hospital, Vienna, Austria.University Eye HospitalViennaAustria</mods:affiliation>
</affiliation>
</author>
</analytic>
<monogr></monogr>
<series>
<title level="j">Journal of Cataract & Refractive Surgery</title>
<title level="j" type="abbrev">JCRS</title>
<idno type="ISSN">0886-3350</idno>
<imprint>
<publisher>ELSEVIER</publisher>
<date type="published" when="1997">1997</date>
<biblScope unit="volume">23</biblScope>
<biblScope unit="issue">10</biblScope>
<biblScope unit="page" from="1543">1543</biblScope>
<biblScope unit="page" to="1547">1547</biblScope>
</imprint>
<idno type="ISSN">0886-3350</idno>
</series>
<idno type="istex">41C2290A664512053197B87509B758A5667696A8</idno>
<idno type="DOI">10.1016/S0886-3350(97)80027-2</idno>
<idno type="PII">S0886-3350(97)80027-2</idno>
<idno type="ArticleID">80027</idno>
</biblStruct>
</sourceDesc>
<seriesStmt>
<idno type="ISSN">0886-3350</idno>
</seriesStmt>
</fileDesc>
<profileDesc>
<textClass></textClass>
<langUsage>
<language ident="en">en</language>
</langUsage>
</profileDesc>
</teiHeader>
<front>
<div type="abstract" xml:lang="en">Purpose: To quantify the dynamics of capsular bag shrinkage after cataract surgery.Setting: University Eye Hospital, Vienna, Austria.Methods: Nineteen eyes in 13 patients had clear corneal cataract surgery with implantation of a foldable, open-loop, silicone posterior chamber intraocular lens (IOL) (AMO SI-30). At the same time, an open, poly(methyl methacrylate) capsule tension ring (Morcher type 14 or 14A) was inserted to allow measurement of capsular bag circumference and diameter. After surgery, capsular bag shrinkage was quantified by measuring anterior chamber depth (ACD), iris-lens distance (ILD), and distance between the eyelets of the capsule tension ring using optical methods. Capsular bag circumference (CBC) was deduced from these measurements, which were taken 1 day (baseline), 1 week, and 1 and 3 months postoperatively.Results: Between 1 day and 3 months, all three parameters decreased significantly. During the first postoperative week, ACID and ILD did not change, while both parameters decreased between 1 week and 1 month. After 1 month, ACID did not decrease further; ILD continued to decrease. The CBC decreased during the entire postoperative period.Conclusion: Besides ACID and ILD, CBC significantly decreased during the first 3 months after cataract surgery with in-the-bag posterior chamber IOL implantation. The CBC can be calculated from the distance between the ends of the capsule tension ring, which can be measured gonioscopically. From this, the diameter of the capsular bag can be calculated in the living eye. The influence of the capsule tension ring itself on capsular bag shrinkage remains to be established.</div>
</front>
</TEI>
<istex>
<corpusName>elsevier</corpusName>
<author>
<json:item>
<name>Karin Strenn MD</name>
<affiliations>
<json:string>From University Eye Hospital, Vienna, Austria.University Eye HospitalViennaAustria</json:string>
</affiliations>
</json:item>
<json:item>
<name>Rupert Menapace MD</name>
<affiliations>
<json:string>From University Eye Hospital, Vienna, Austria.University Eye HospitalViennaAustria</json:string>
</affiliations>
</json:item>
<json:item>
<name>Clemens Vass MD</name>
<affiliations>
<json:string>From University Eye Hospital, Vienna, Austria.University Eye HospitalViennaAustria</json:string>
</affiliations>
</json:item>
</author>
<articleId>
<json:string>80027</json:string>
</articleId>
<language>
<json:string>eng</json:string>
</language>
<abstract>Purpose: To quantify the dynamics of capsular bag shrinkage after cataract surgery.Setting: University Eye Hospital, Vienna, Austria.Methods: Nineteen eyes in 13 patients had clear corneal cataract surgery with implantation of a foldable, open-loop, silicone posterior chamber intraocular lens (IOL) (AMO SI-30). At the same time, an open, poly(methyl methacrylate) capsule tension ring (Morcher type 14 or 14A) was inserted to allow measurement of capsular bag circumference and diameter. After surgery, capsular bag shrinkage was quantified by measuring anterior chamber depth (ACD), iris-lens distance (ILD), and distance between the eyelets of the capsule tension ring using optical methods. Capsular bag circumference (CBC) was deduced from these measurements, which were taken 1 day (baseline), 1 week, and 1 and 3 months postoperatively.Results: Between 1 day and 3 months, all three parameters decreased significantly. During the first postoperative week, ACID and ILD did not change, while both parameters decreased between 1 week and 1 month. After 1 month, ACID did not decrease further; ILD continued to decrease. The CBC decreased during the entire postoperative period.Conclusion: Besides ACID and ILD, CBC significantly decreased during the first 3 months after cataract surgery with in-the-bag posterior chamber IOL implantation. The CBC can be calculated from the distance between the ends of the capsule tension ring, which can be measured gonioscopically. From this, the diameter of the capsular bag can be calculated in the living eye. The influence of the capsule tension ring itself on capsular bag shrinkage remains to be established.</abstract>
<qualityIndicators>
<score>7.068</score>
<pdfVersion>1.7</pdfVersion>
<pdfPageSize>576 x 828 pts</pdfPageSize>
<refBibsNative>true</refBibsNative>
<keywordCount>0</keywordCount>
<abstractCharCount>1657</abstractCharCount>
<pdfWordCount>2128</pdfWordCount>
<pdfCharCount>12923</pdfCharCount>
<pdfPageCount>5</pdfPageCount>
<abstractWordCount>245</abstractWordCount>
</qualityIndicators>
<title>Capsular bag shrinkage after implantation of an open-loop silicone lens and a poly(methyl methacrylate) capsule tension ring</title>
<pii>
<json:string>S0886-3350(97)80027-2</json:string>
</pii>
<genre>
<json:string>research-article</json:string>
</genre>
<host>
<volume>23</volume>
<pii>
<json:string>S0886-3350(97)X8001-0</json:string>
</pii>
<pages>
<last>1547</last>
<first>1543</first>
</pages>
<issn>
<json:string>0886-3350</json:string>
</issn>
<issue>10</issue>
<genre>
<json:string>Journal</json:string>
</genre>
<language>
<json:string>unknown</json:string>
</language>
<title>Journal of Cataract & Refractive Surgery</title>
<publicationDate>1997</publicationDate>
</host>
<categories>
<wos>
<json:string>OPHTHALMOLOGY</json:string>
<json:string>SURGERY</json:string>
</wos>
</categories>
<publicationDate>1997</publicationDate>
<copyrightDate>1997</copyrightDate>
<doi>
<json:string>10.1016/S0886-3350(97)80027-2</json:string>
</doi>
<id>41C2290A664512053197B87509B758A5667696A8</id>
<score>1</score>
<fulltext>
<json:item>
<original>true</original>
<mimetype>application/pdf</mimetype>
<extension>pdf</extension>
<uri>https://api.istex.fr/document/41C2290A664512053197B87509B758A5667696A8/fulltext/pdf</uri>
</json:item>
<json:item>
<original>true</original>
<mimetype>text/plain</mimetype>
<extension>txt</extension>
<uri>https://api.istex.fr/document/41C2290A664512053197B87509B758A5667696A8/fulltext/txt</uri>
</json:item>
<json:item>
<original>false</original>
<mimetype>application/zip</mimetype>
<extension>zip</extension>
<uri>https://api.istex.fr/document/41C2290A664512053197B87509B758A5667696A8/fulltext/zip</uri>
</json:item>
<istex:fulltextTEI uri="https://api.istex.fr/document/41C2290A664512053197B87509B758A5667696A8/fulltext/tei">
<teiHeader>
<fileDesc>
<titleStmt>
<title level="a" type="main" xml:lang="en">Capsular bag shrinkage after implantation of an open-loop silicone lens and a poly(methyl methacrylate) capsule tension ring</title>
</titleStmt>
<publicationStmt>
<authority>ISTEX</authority>
<publisher>ELSEVIER</publisher>
<availability>
<p>ELSEVIER</p>
</availability>
<date>1997</date>
</publicationStmt>
<sourceDesc>
<biblStruct type="inbook">
<analytic>
<title level="a" type="main" xml:lang="en">Capsular bag shrinkage after implantation of an open-loop silicone lens and a poly(methyl methacrylate) capsule tension ring</title>
<author>
<persName>
<forename type="first">Karin</forename>
<surname>Strenn</surname>
</persName>
<roleName type="degree">MD</roleName>
<note type="biography">Reprint requests to Karin Strenn, MD, Universitätsklinik für Augenheilkunde, Allgemeines Krankenhaus, Währinger Gürtel 18-20, A-1090 Wien, Austria.</note>
<affiliation>Reprint requests to Karin Strenn, MD, Universitätsklinik für Augenheilkunde, Allgemeines Krankenhaus, Währinger Gürtel 18-20, A-1090 Wien, Austria.</affiliation>
<affiliation>From University Eye Hospital, Vienna, Austria.University Eye HospitalViennaAustria</affiliation>
</author>
<author>
<persName>
<forename type="first">Rupert</forename>
<surname>Menapace</surname>
</persName>
<roleName type="degree">MD</roleName>
<affiliation>From University Eye Hospital, Vienna, Austria.University Eye HospitalViennaAustria</affiliation>
</author>
<author>
<persName>
<forename type="first">Clemens</forename>
<surname>Vass</surname>
</persName>
<roleName type="degree">MD</roleName>
<affiliation>From University Eye Hospital, Vienna, Austria.University Eye HospitalViennaAustria</affiliation>
</author>
</analytic>
<monogr>
<title level="j">Journal of Cataract & Refractive Surgery</title>
<title level="j" type="abbrev">JCRS</title>
<idno type="pISSN">0886-3350</idno>
<idno type="PII">S0886-3350(97)X8001-0</idno>
<imprint>
<publisher>ELSEVIER</publisher>
<date type="published" when="1997"></date>
<biblScope unit="volume">23</biblScope>
<biblScope unit="issue">10</biblScope>
<biblScope unit="page" from="1543">1543</biblScope>
<biblScope unit="page" to="1547">1547</biblScope>
</imprint>
</monogr>
<idno type="istex">41C2290A664512053197B87509B758A5667696A8</idno>
<idno type="DOI">10.1016/S0886-3350(97)80027-2</idno>
<idno type="PII">S0886-3350(97)80027-2</idno>
<idno type="ArticleID">80027</idno>
</biblStruct>
</sourceDesc>
</fileDesc>
<profileDesc>
<creation>
<date>1997</date>
</creation>
<langUsage>
<language ident="en">en</language>
</langUsage>
<abstract xml:lang="en">
<p>Purpose: To quantify the dynamics of capsular bag shrinkage after cataract surgery.Setting: University Eye Hospital, Vienna, Austria.Methods: Nineteen eyes in 13 patients had clear corneal cataract surgery with implantation of a foldable, open-loop, silicone posterior chamber intraocular lens (IOL) (AMO SI-30). At the same time, an open, poly(methyl methacrylate) capsule tension ring (Morcher type 14 or 14A) was inserted to allow measurement of capsular bag circumference and diameter. After surgery, capsular bag shrinkage was quantified by measuring anterior chamber depth (ACD), iris-lens distance (ILD), and distance between the eyelets of the capsule tension ring using optical methods. Capsular bag circumference (CBC) was deduced from these measurements, which were taken 1 day (baseline), 1 week, and 1 and 3 months postoperatively.Results: Between 1 day and 3 months, all three parameters decreased significantly. During the first postoperative week, ACID and ILD did not change, while both parameters decreased between 1 week and 1 month. After 1 month, ACID did not decrease further; ILD continued to decrease. The CBC decreased during the entire postoperative period.Conclusion: Besides ACID and ILD, CBC significantly decreased during the first 3 months after cataract surgery with in-the-bag posterior chamber IOL implantation. The CBC can be calculated from the distance between the ends of the capsule tension ring, which can be measured gonioscopically. From this, the diameter of the capsular bag can be calculated in the living eye. The influence of the capsule tension ring itself on capsular bag shrinkage remains to be established.</p>
</abstract>
</profileDesc>
<revisionDesc>
<change when="1997">Published</change>
</revisionDesc>
</teiHeader>
</istex:fulltextTEI>
</fulltext>
<metadata>
<istex:metadataXml wicri:clean="Elsevier doc found" wicri:toSee="Elsevier, no converted or simple article">
<istex:xmlDeclaration>version="1.0" encoding="UTF-8" </istex:xmlDeclaration>
<istex:docType PUBLIC="-//ES//DTD journal article DTD version 5.2.0//EN//XML" URI="art520.dtd" name="istex:docType"></istex:docType>
<istex:document>
<article version="5.2" xml:lang="en" docsubtype="fla">
<item-info>
<jid>JCRS</jid>
<aid>80027</aid>
<ce:pii>S0886-3350(97)80027-2</ce:pii>
<ce:doi>10.1016/S0886-3350(97)80027-2</ce:doi>
<ce:copyright type="society" year="1997">American Society of Cataract and Refractive Surgery and European Society of Cataract and Refractive Surgeons. All rights reserved</ce:copyright>
<ce:copyright-line>Copyright 1997 American Society of Cataract and Refractive Surgery and European Society of Cataract and Refractive Surgeons. All rights reserved</ce:copyright-line>
</item-info>
<head>
<ce:title id="ttl005">Capsular bag shrinkage after implantation of an open-loop silicone lens and a poly(methyl methacrylate) capsule tension ring</ce:title>
<ce:author-group id="AUG005">
<ce:author id="AUT005">
<ce:given-name>Karin</ce:given-name>
<ce:surname>Strenn</ce:surname>
<ce:degrees>MD</ce:degrees>
<ce:cross-ref id="crr005" refid="cor1">
<ce:sup>*</ce:sup>
</ce:cross-ref>
</ce:author>
<ce:author id="AUT010">
<ce:given-name>Rupert</ce:given-name>
<ce:surname>Menapace</ce:surname>
<ce:degrees>MD</ce:degrees>
</ce:author>
<ce:author id="AUT015">
<ce:given-name>Clemens</ce:given-name>
<ce:surname>Vass</ce:surname>
<ce:degrees>MD</ce:degrees>
</ce:author>
<ce:affiliation id="aff1">
<ce:textfn>From University Eye Hospital, Vienna, Austria.</ce:textfn>
<sa:affiliation>
<sa:organization>University Eye Hospital</sa:organization>
<sa:city>Vienna</sa:city>
<sa:country>Austria</sa:country>
</sa:affiliation>
</ce:affiliation>
<ce:correspondence id="cor1">
<ce:label>*</ce:label>
<ce:text>Reprint requests to Karin Strenn, MD, Universitätsklinik für Augenheilkunde, Allgemeines Krankenhaus, Währinger Gürtel 18-20, A-1090 Wien, Austria.</ce:text>
</ce:correspondence>
</ce:author-group>
<ce:abstract id="ab1">
<ce:section-title id="SET005">Abstract</ce:section-title>
<ce:abstract-sec id="abs005">
<ce:simple-para id="SP0005">
<ce:bold>Purpose:</ce:bold>
To quantify the dynamics of capsular bag shrinkage after cataract surgery.</ce:simple-para>
<ce:simple-para id="SP0010">
<ce:bold>Setting:</ce:bold>
University Eye Hospital, Vienna, Austria.</ce:simple-para>
<ce:simple-para id="SP0015">
<ce:bold>Methods:</ce:bold>
Nineteen eyes in 13 patients had clear corneal cataract surgery with implantation of a foldable, open-loop, silicone posterior chamber intraocular lens (IOL) (AMO SI-30). At the same time, an open, poly(methyl methacrylate) capsule tension ring (Morcher type 14 or 14A) was inserted to allow measurement of capsular bag circumference and diameter. After surgery, capsular bag shrinkage was quantified by measuring anterior chamber depth (ACD), iris-lens distance (ILD), and distance between the eyelets of the capsule tension ring using optical methods. Capsular bag circumference (CBC) was deduced from these measurements, which were taken 1 day (baseline), 1 week, and 1 and 3 months postoperatively.</ce:simple-para>
<ce:simple-para id="SP0020">
<ce:bold>Results:</ce:bold>
Between 1 day and 3 months, all three parameters decreased significantly. During the first postoperative week, ACID and ILD did not change, while both parameters decreased between 1 week and 1 month. After 1 month, ACID did not decrease further; ILD continued to decrease. The CBC decreased during the entire postoperative period.</ce:simple-para>
<ce:simple-para id="SP0025">
<ce:bold>Conclusion:</ce:bold>
Besides ACID and ILD, CBC significantly decreased during the first 3 months after cataract surgery with in-the-bag posterior chamber IOL implantation. The CBC can be calculated from the distance between the ends of the capsule tension ring, which can be measured gonioscopically. From this, the diameter of the capsular bag can be calculated in the living eye. The influence of the capsule tension ring itself on capsular bag shrinkage remains to be established.</ce:simple-para>
</ce:abstract-sec>
</ce:abstract>
</head>
<tail>
<ce:bibliography id="R0005">
<ce:section-title id="SET010">References</ce:section-title>
<ce:bibliography-sec id="RS0005">
<ce:bib-reference id="bib1">
<ce:label>1.</ce:label>
<sb:reference id="SB0005">
<sb:contribution>
<sb:authors>
<sb:author>
<ce:given-name>R</ce:given-name>
<ce:surname>Guthoff</ce:surname>
</sb:author>
<sb:author>
<ce:given-name>F</ce:given-name>
<ce:surname>Abramo</ce:surname>
</sb:author>
<sb:author>
<ce:given-name>J</ce:given-name>
<ce:surname>Draeger</ce:surname>
</sb:author>
</sb:authors>
<sb:title>
<sb:maintitle>Forces on intraocular lens haptics induced by capsular fibrosis; an experimental study</sb:maintitle>
</sb:title>
</sb:contribution>
<sb:host>
<sb:issue>
<sb:series>
<sb:title>
<sb:maintitle>Graefes Arch Clin Exp Ophthalmol</sb:maintitle>
</sb:title>
<sb:volume-nr>228</sb:volume-nr>
</sb:series>
<sb:date>1990</sb:date>
</sb:issue>
<sb:pages>
<sb:first-page>363</sb:first-page>
<sb:last-page>368</sb:last-page>
</sb:pages>
</sb:host>
</sb:reference>
</ce:bib-reference>
<ce:bib-reference id="bib2">
<ce:label>2.</ce:label>
<sb:reference id="SB0010">
<sb:contribution>
<sb:authors>
<sb:author>
<ce:given-name>SO</ce:given-name>
<ce:surname>Hansen</ce:surname>
</sb:author>
<sb:author>
<ce:given-name>KD</ce:given-name>
<ce:surname>Solomon</ce:surname>
</sb:author>
<sb:author>
<ce:given-name>GT</ce:given-name>
<ce:surname>McKnight</ce:surname>
</sb:author>
</sb:authors>
<sb:title>
<sb:maintitle>Posterior capsular opacification and intraocular lens decentration. Part 1: comparison of various posterior chamber lens designs implanted in the rabbit model</sb:maintitle>
</sb:title>
</sb:contribution>
<sb:host>
<sb:issue>
<sb:series>
<sb:title>
<sb:maintitle>J Cataract Refract Surg</sb:maintitle>
</sb:title>
<sb:volume-nr>14</sb:volume-nr>
</sb:series>
<sb:date>1988</sb:date>
</sb:issue>
<sb:pages>
<sb:first-page>605</sb:first-page>
<sb:last-page>613</sb:last-page>
</sb:pages>
</sb:host>
</sb:reference>
</ce:bib-reference>
<ce:bib-reference id="bib3">
<ce:label>3.</ce:label>
<sb:reference id="SB0015">
<sb:contribution>
<sb:authors>
<sb:author>
<ce:given-name>FA</ce:given-name>
<ce:surname>Richburg</ce:surname>
</sb:author>
<sb:author>
<ce:given-name>HS</ce:given-name>
<ce:surname>Sun</ce:surname>
</sb:author>
</sb:authors>
<sb:title>
<sb:maintitle>Size of the crushed cataractous capsule bag</sb:maintitle>
</sb:title>
</sb:contribution>
<sb:host>
<sb:issue>
<sb:series>
<sb:title>
<sb:maintitle>Am Intra-Ocular Implant Soc J</sb:maintitle>
</sb:title>
<sb:volume-nr>9</sb:volume-nr>
</sb:series>
<sb:date>1983</sb:date>
</sb:issue>
<sb:pages>
<sb:first-page>333</sb:first-page>
<sb:last-page>335</sb:last-page>
</sb:pages>
</sb:host>
</sb:reference>
</ce:bib-reference>
<ce:bib-reference id="bib4">
<ce:label>4.</ce:label>
<sb:reference id="SB0020">
<sb:contribution>
<sb:authors>
<sb:author>
<ce:given-name>A</ce:given-name>
<ce:surname>Galand</ce:surname>
</sb:author>
<sb:author>
<ce:given-name>L</ce:given-name>
<ce:surname>Bonhomme</ce:surname>
</sb:author>
<sb:author>
<ce:given-name>M</ce:given-name>
<ce:surname>Collee</ce:surname>
</sb:author>
</sb:authors>
<sb:title>
<sb:maintitle>Direct measurement of the capsular bag</sb:maintitle>
</sb:title>
</sb:contribution>
<sb:host>
<sb:issue>
<sb:series>
<sb:title>
<sb:maintitle>Am Intra-Ocular Implant Soc J</sb:maintitle>
</sb:title>
<sb:volume-nr>10</sb:volume-nr>
</sb:series>
<sb:date>1984</sb:date>
</sb:issue>
<sb:pages>
<sb:first-page>475</sb:first-page>
<sb:last-page>476</sb:last-page>
</sb:pages>
</sb:host>
</sb:reference>
</ce:bib-reference>
<ce:bib-reference id="bib5">
<ce:label>5.</ce:label>
<sb:reference id="SB0025">
<sb:contribution>
<sb:authors>
<sb:author>
<ce:given-name>H</ce:given-name>
<ce:surname>Nagamoto T Bissen-Miyajima</ce:surname>
</sb:author>
</sb:authors>
<sb:title>
<sb:maintitle>A ring to support the capsular bag after continuous curvilinear capsulorhexis</sb:maintitle>
</sb:title>
</sb:contribution>
<sb:host>
<sb:issue>
<sb:series>
<sb:title>
<sb:maintitle>J Cataract Refract Surg</sb:maintitle>
</sb:title>
<sb:volume-nr>20</sb:volume-nr>
</sb:series>
<sb:date>1994</sb:date>
</sb:issue>
<sb:pages>
<sb:first-page>417</sb:first-page>
<sb:last-page>420</sb:last-page>
</sb:pages>
</sb:host>
</sb:reference>
</ce:bib-reference>
<ce:bib-reference id="bib6">
<ce:label>6.</ce:label>
<sb:reference id="SB0030">
<sb:contribution>
<sb:authors>
<sb:author>
<ce:given-name>T</ce:given-name>
<ce:surname>Hara T Hara</ce:surname>
</sb:author>
<sb:author>
<ce:given-name>Y</ce:given-name>
<ce:surname>Yamada</ce:surname>
</sb:author>
</sb:authors>
<sb:title>
<sb:maintitle>“Equator ring” for maintenance of the completely circular contour of the capsular bag equator after cataract removal</sb:maintitle>
</sb:title>
</sb:contribution>
<sb:host>
<sb:issue>
<sb:series>
<sb:title>
<sb:maintitle>Ophthalmic Surg</sb:maintitle>
</sb:title>
<sb:volume-nr>22</sb:volume-nr>
</sb:series>
<sb:date>1991</sb:date>
</sb:issue>
<sb:pages>
<sb:first-page>358</sb:first-page>
<sb:last-page>359</sb:last-page>
</sb:pages>
</sb:host>
</sb:reference>
</ce:bib-reference>
<ce:bib-reference id="bib7">
<ce:label>7.</ce:label>
<sb:reference id="SB0035">
<sb:contribution>
<sb:authors>
<sb:author>
<ce:given-name>T</ce:given-name>
<ce:surname>Hara</ce:surname>
</sb:author>
<sb:author>
<ce:given-name>K</ce:given-name>
<ce:surname>Hara T Sakanishi</ce:surname>
</sb:author>
<sb:author>
<ce:given-name>Y</ce:given-name>
<ce:surname>Yamada</ce:surname>
</sb:author>
</sb:authors>
<sb:title>
<sb:maintitle>Efficacy of equator rings in an experimental rabbit study</sb:maintitle>
</sb:title>
</sb:contribution>
<sb:host>
<sb:issue>
<sb:series>
<sb:title>
<sb:maintitle>Arch Ophthalmol</sb:maintitle>
</sb:title>
<sb:volume-nr>113</sb:volume-nr>
</sb:series>
<sb:date>1995</sb:date>
</sb:issue>
<sb:pages>
<sb:first-page>1060</sb:first-page>
<sb:last-page>1065</sb:last-page>
</sb:pages>
</sb:host>
</sb:reference>
</ce:bib-reference>
<ce:bib-reference id="bib8">
<ce:label>8.</ce:label>
<sb:reference id="SB0040">
<sb:contribution>
<sb:authors>
<sb:author>
<ce:given-name>A</ce:given-name>
<ce:surname>Galand</ce:surname>
</sb:author>
<sb:author>
<ce:given-name>M</ce:given-name>
<ce:surname>Delmelle</ce:surname>
</sb:author>
</sb:authors>
<sb:title>
<sb:maintitle>Preliminary report on the rigid disc lens</sb:maintitle>
</sb:title>
</sb:contribution>
<sb:host>
<sb:issue>
<sb:series>
<sb:title>
<sb:maintitle>J Cataract Refract Surg</sb:maintitle>
</sb:title>
<sb:volume-nr>12</sb:volume-nr>
</sb:series>
<sb:date>1986</sb:date>
</sb:issue>
<sb:pages>
<sb:first-page>394</sb:first-page>
<sb:last-page>397</sb:last-page>
</sb:pages>
</sb:host>
</sb:reference>
</ce:bib-reference>
<ce:bib-reference id="bib9">
<ce:label>9.</ce:label>
<sb:reference id="SB0045">
<sb:contribution>
<sb:authors>
<sb:author>
<ce:given-name>O</ce:given-name>
<ce:surname>Nishi</ce:surname>
</sb:author>
</sb:authors>
<sb:title>
<sb:maintitle>New posterior chamber lens with open circular haptic for a small capsular opening: preliminary report</sb:maintitle>
</sb:title>
</sb:contribution>
<sb:host>
<sb:issue>
<sb:series>
<sb:title>
<sb:maintitle>J Cataract Refract Surg</sb:maintitle>
</sb:title>
<sb:volume-nr>16</sb:volume-nr>
</sb:series>
<sb:date>1990</sb:date>
</sb:issue>
<sb:pages>
<sb:first-page>640</sb:first-page>
<sb:last-page>643</sb:last-page>
</sb:pages>
</sb:host>
</sb:reference>
</ce:bib-reference>
<ce:bib-reference id="bib10">
<ce:label>10.</ce:label>
<sb:reference id="SB0050">
<sb:contribution>
<sb:authors>
<sb:author>
<ce:given-name>MR</ce:given-name>
<ce:surname>Tetz</ce:surname>
</sb:author>
<sb:author>
<ce:given-name>DJC</ce:given-name>
<ce:surname>O'Morchoe</ce:surname>
</sb:author>
<sb:author>
<ce:given-name>TD</ce:given-name>
<ce:surname>Gwin</ce:surname>
</sb:author>
<sb:et-al></sb:et-al>
</sb:authors>
<sb:title>
<sb:maintitle>Posterior capsular opacification and intraocular lens decentration. Part II: experimental findings on a prototype circular intraocular lens design</sb:maintitle>
</sb:title>
</sb:contribution>
<sb:host>
<sb:issue>
<sb:series>
<sb:title>
<sb:maintitle>J Cataract Refract Surg</sb:maintitle>
</sb:title>
<sb:volume-nr>14</sb:volume-nr>
</sb:series>
<sb:date>1988</sb:date>
</sb:issue>
<sb:pages>
<sb:first-page>614</sb:first-page>
<sb:last-page>623</sb:last-page>
</sb:pages>
</sb:host>
</sb:reference>
</ce:bib-reference>
</ce:bibliography-sec>
</ce:bibliography>
</tail>
</article>
</istex:document>
</istex:metadataXml>
<mods version="3.6">
<titleInfo lang="en">
<title>Capsular bag shrinkage after implantation of an open-loop silicone lens and a poly(methyl methacrylate) capsule tension ring</title>
</titleInfo>
<titleInfo type="alternative" lang="en" contentType="CDATA">
<title>Capsular bag shrinkage after implantation of an open-loop silicone lens and a poly(methyl methacrylate) capsule tension ring</title>
</titleInfo>
<name type="personal">
<namePart type="given">Karin</namePart>
<namePart type="family">Strenn</namePart>
<namePart type="termsOfAddress">MD</namePart>
<affiliation>From University Eye Hospital, Vienna, Austria.University Eye HospitalViennaAustria</affiliation>
<description>Reprint requests to Karin Strenn, MD, Universitätsklinik für Augenheilkunde, Allgemeines Krankenhaus, Währinger Gürtel 18-20, A-1090 Wien, Austria.</description>
<role>
<roleTerm type="text">author</roleTerm>
</role>
</name>
<name type="personal">
<namePart type="given">Rupert</namePart>
<namePart type="family">Menapace</namePart>
<namePart type="termsOfAddress">MD</namePart>
<affiliation>From University Eye Hospital, Vienna, Austria.University Eye HospitalViennaAustria</affiliation>
<role>
<roleTerm type="text">author</roleTerm>
</role>
</name>
<name type="personal">
<namePart type="given">Clemens</namePart>
<namePart type="family">Vass</namePart>
<namePart type="termsOfAddress">MD</namePart>
<affiliation>From University Eye Hospital, Vienna, Austria.University Eye HospitalViennaAustria</affiliation>
<role>
<roleTerm type="text">author</roleTerm>
</role>
</name>
<typeOfResource>text</typeOfResource>
<genre type="research-article" displayLabel="Full-length article"></genre>
<originInfo>
<publisher>ELSEVIER</publisher>
<dateIssued encoding="w3cdtf">1997</dateIssued>
<copyrightDate encoding="w3cdtf">1997</copyrightDate>
</originInfo>
<language>
<languageTerm type="code" authority="iso639-2b">eng</languageTerm>
<languageTerm type="code" authority="rfc3066">en</languageTerm>
</language>
<physicalDescription>
<internetMediaType>text/html</internetMediaType>
</physicalDescription>
<abstract lang="en">Purpose: To quantify the dynamics of capsular bag shrinkage after cataract surgery.Setting: University Eye Hospital, Vienna, Austria.Methods: Nineteen eyes in 13 patients had clear corneal cataract surgery with implantation of a foldable, open-loop, silicone posterior chamber intraocular lens (IOL) (AMO SI-30). At the same time, an open, poly(methyl methacrylate) capsule tension ring (Morcher type 14 or 14A) was inserted to allow measurement of capsular bag circumference and diameter. After surgery, capsular bag shrinkage was quantified by measuring anterior chamber depth (ACD), iris-lens distance (ILD), and distance between the eyelets of the capsule tension ring using optical methods. Capsular bag circumference (CBC) was deduced from these measurements, which were taken 1 day (baseline), 1 week, and 1 and 3 months postoperatively.Results: Between 1 day and 3 months, all three parameters decreased significantly. During the first postoperative week, ACID and ILD did not change, while both parameters decreased between 1 week and 1 month. After 1 month, ACID did not decrease further; ILD continued to decrease. The CBC decreased during the entire postoperative period.Conclusion: Besides ACID and ILD, CBC significantly decreased during the first 3 months after cataract surgery with in-the-bag posterior chamber IOL implantation. The CBC can be calculated from the distance between the ends of the capsule tension ring, which can be measured gonioscopically. From this, the diameter of the capsular bag can be calculated in the living eye. The influence of the capsule tension ring itself on capsular bag shrinkage remains to be established.</abstract>
<relatedItem type="host">
<titleInfo>
<title>Journal of Cataract & Refractive Surgery</title>
</titleInfo>
<titleInfo type="abbreviated">
<title>JCRS</title>
</titleInfo>
<genre type="Journal">journal</genre>
<originInfo>
<dateIssued encoding="w3cdtf">199712</dateIssued>
</originInfo>
<identifier type="ISSN">0886-3350</identifier>
<identifier type="PII">S0886-3350(97)X8001-0</identifier>
<part>
<detail type="volume">
<number>23</number>
<caption>vol.</caption>
</detail>
<detail type="issue">
<number>10</number>
<caption>no.</caption>
</detail>
<extent unit="issue pages">
<start>1429</start>
<end>1622</end>
</extent>
<extent unit="pages">
<start>1543</start>
<end>1547</end>
</extent>
</part>
</relatedItem>
<identifier type="istex">41C2290A664512053197B87509B758A5667696A8</identifier>
<identifier type="DOI">10.1016/S0886-3350(97)80027-2</identifier>
<identifier type="PII">S0886-3350(97)80027-2</identifier>
<identifier type="ArticleID">80027</identifier>
<accessCondition type="use and reproduction" contentType="">© 1997American Society of Cataract and Refractive Surgery and European Society of Cataract and Refractive Surgeons. All rights reserved</accessCondition>
<recordInfo>
<recordContentSource>ELSEVIER</recordContentSource>
<recordOrigin>American Society of Cataract and Refractive Surgery and European Society of Cataract and Refractive Surgeons. All rights reserved, ©1997</recordOrigin>
</recordInfo>
</mods>
</metadata>
<enrichments>
<istex:catWosTEI uri="https://api.istex.fr/document/41C2290A664512053197B87509B758A5667696A8/enrichments/catWos">
<teiHeader>
<profileDesc>
<textClass>
<classCode scheme="WOS">OPHTHALMOLOGY</classCode>
<classCode scheme="WOS">SURGERY</classCode>
</textClass>
</profileDesc>
</teiHeader>
</istex:catWosTEI>
</enrichments>
<serie></serie>
</istex>
</record>

Pour manipuler ce document sous Unix (Dilib)

EXPLOR_STEP=$WICRI_ROOT/Ticri/CIDE/explor/HapticV1/Data/Istex/Corpus
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 000229 | SxmlIndent | more

Ou

HfdSelect -h $EXPLOR_AREA/Data/Istex/Corpus/biblio.hfd -nk 000229 | SxmlIndent | more

Pour mettre un lien sur cette page dans le réseau Wicri

{{Explor lien
   |wiki=    Ticri/CIDE
   |area=    HapticV1
   |flux=    Istex
   |étape=   Corpus
   |type=    RBID
   |clé=     ISTEX:41C2290A664512053197B87509B758A5667696A8
   |texte=   Capsular bag shrinkage after implantation of an open-loop silicone lens and a poly(methyl methacrylate) capsule tension ring
}}

Wicri

This area was generated with Dilib version V0.6.23.
Data generation: Mon Jun 13 01:09:46 2016. Site generation: Wed Mar 6 09:54:07 2024