The Reliability of a Simplified Water Displacement instrument: A method for measuring arm volume

Identifieur interne : 000579 ( PascalFrancis/Corpus ); précédent : 000578; suivant : 000580

The Reliability of a Simplified Water Displacement instrument: A method for measuring arm volume

Auteurs : Ase Sagen ; Rolf Karesen ; May Arna Risberg

Source :

RBID : Pascal:05-0235133

Descripteurs français

English descriptors

Abstract

Objectives: To present a new water displacement measurement, the Simplified Water Displacement Instrument (SWDI), and to evaluate its intra- and intertester reliability. Design: Reliability design. Setting: Hospital setting. Participants: Fifty-six healthy people were studied. Intratester reliability was evaluated once a week for 4 weeks in 20 women and 10 men. Intertester reliability was assessed by 2 physical therapists in 26 people. Interventions: Not applicable. Main Outcome Measures: Coefficients of variation (CVs) and intraclass correlation coefficients (ICCs). Results: The intratester reliability showed a CV range of 2.2% to 2.6% and an ICC range of .98 to .99. The intertester reliability showed a CV of 1.3% and an ICC of .99. There was a significant increase in arm volume in men compared with women. There were no significant differences in changes in volume over the 4 weeks. There was a significant greater right arm volume (3.3%) among the right-handed subjects (P<.001). Conclusions: Both intra- and intertester reliability were satisfactory for the SWDI.

Notice en format standard (ISO 2709)

Pour connaître la documentation sur le format Inist Standard.

pA  
A01 01  1    @0 0003-9993
A02 01      @0 APMHAI
A03   1    @0 Arch. phys. med. rehabil.
A05       @2 86
A06       @2 1
A08 01  1  ENG  @1 The Reliability of a Simplified Water Displacement instrument: A method for measuring arm volume
A11 01  1    @1 SAGEN (Ase)
A11 02  1    @1 KARESEN (Rolf)
A11 03  1    @1 RISBERG (May Arna)
A14 01      @1 Surgical Division, Ullevaal University Hospital @2 Oslo @3 NOR @Z 1 aut. @Z 2 aut.
A14 02      @1 Center for Clinical Research, Ullevaal University Hospital @2 Oslo @3 NOR @Z 3 aut.
A20       @1 86-89
A21       @1 2005
A23 01      @0 ENG
A43 01      @1 INIST @2 8199 @5 354000126605780170
A44       @0 0000 @1 © 2005 INIST-CNRS. All rights reserved.
A45       @0 33 ref.
A47 01  1    @0 05-0235133
A60       @1 P
A61       @0 A
A64 01  1    @0 Archives of physical medicine and rehabilitation
A66 01      @0 USA
C01 01    ENG  @0 Objectives: To present a new water displacement measurement, the Simplified Water Displacement Instrument (SWDI), and to evaluate its intra- and intertester reliability. Design: Reliability design. Setting: Hospital setting. Participants: Fifty-six healthy people were studied. Intratester reliability was evaluated once a week for 4 weeks in 20 women and 10 men. Intertester reliability was assessed by 2 physical therapists in 26 people. Interventions: Not applicable. Main Outcome Measures: Coefficients of variation (CVs) and intraclass correlation coefficients (ICCs). Results: The intratester reliability showed a CV range of 2.2% to 2.6% and an ICC range of .98 to .99. The intertester reliability showed a CV of 1.3% and an ICC of .99. There was a significant increase in arm volume in men compared with women. There were no significant differences in changes in volume over the 4 weeks. There was a significant greater right arm volume (3.3%) among the right-handed subjects (P<.001). Conclusions: Both intra- and intertester reliability were satisfactory for the SWDI.
C02 01  X    @0 002B15
C02 02  X    @0 002B12B04
C02 03  X    @0 002B26O
C03 01  X  FRE  @0 Tumeur maligne @5 01
C03 01  X  ENG  @0 Malignant tumor @5 01
C03 01  X  SPA  @0 Tumor maligno @5 01
C03 02  X  FRE  @0 Glande mammaire pathologie @2 NM @5 02
C03 02  X  ENG  @0 Mammary gland diseases @2 NM @5 02
C03 02  X  SPA  @0 Glándula mamaria patología @2 NM @5 02
C03 03  X  FRE  @0 Fiabilité @5 07
C03 03  X  ENG  @0 Reliability @5 07
C03 03  X  SPA  @0 Fiabilidad @5 07
C03 04  X  FRE  @0 Eau @5 08
C03 04  X  ENG  @0 Water @5 08
C03 04  X  SPA  @0 Agua @5 08
C03 05  X  FRE  @0 Déplacement @5 09
C03 05  X  ENG  @0 Displacement @5 09
C03 05  X  SPA  @0 Desplazamiento @5 09
C03 06  X  FRE  @0 Appareil mesure @5 13
C03 06  X  ENG  @0 Measuring instrument @5 13
C03 06  X  SPA  @0 Aparato medición @5 13
C03 07  X  FRE  @0 Méthode @5 14
C03 07  X  ENG  @0 Method @5 14
C03 07  X  SPA  @0 Método @5 14
C03 08  X  FRE  @0 Bras @5 15
C03 08  X  ENG  @0 Arm @5 15
C03 08  X  SPA  @0 Brazo @5 15
C03 09  X  FRE  @0 Femme @5 16
C03 09  X  ENG  @0 Woman @5 16
C03 09  X  SPA  @0 Mujer @5 16
C03 10  X  FRE  @0 Volume @5 17
C03 10  X  ENG  @0 Volume @5 17
C03 10  X  SPA  @0 Volumen @5 17
C03 11  X  FRE  @0 Instrumentation @5 18
C03 11  X  ENG  @0 Instruments @5 18
C03 11  X  SPA  @0 Instrumentación @5 18
C03 12  X  FRE  @0 Hôpital @5 19
C03 12  X  ENG  @0 Hospital @5 19
C03 12  X  SPA  @0 Hospital @5 19
C03 13  X  FRE  @0 Sexe @5 20
C03 13  X  ENG  @0 Sex @5 20
C03 13  X  SPA  @0 Sexo @5 20
C03 14  X  FRE  @0 Pronostic @5 21
C03 14  X  ENG  @0 Prognosis @5 21
C03 14  X  SPA  @0 Pronóstico @5 21
C03 15  X  FRE  @0 Coefficient corrélation @5 22
C03 15  X  ENG  @0 Correlation coefficient @5 22
C03 15  X  SPA  @0 Coeficiente correlación @5 22
C03 16  X  FRE  @0 Augmentation @5 23
C03 16  X  ENG  @0 Increase @5 23
C03 16  X  SPA  @0 Aumentación @5 23
C03 17  X  FRE  @0 Femelle @5 25
C03 17  X  ENG  @0 Female @5 25
C03 17  X  SPA  @0 Hembra @5 25
C03 18  X  FRE  @0 Etude comparative @5 26
C03 18  X  ENG  @0 Comparative study @5 26
C03 18  X  SPA  @0 Estudio comparativo @5 26
C03 19  X  FRE  @0 Main @5 27
C03 19  X  ENG  @0 Hand @5 27
C03 19  X  SPA  @0 Mano @5 27
C03 20  X  FRE  @0 Glande mammaire @5 28
C03 20  X  ENG  @0 Mammary gland @5 28
C03 20  X  SPA  @0 Glándula mamaria @5 28
C03 21  X  FRE  @0 Lymphoedème @5 29
C03 21  X  ENG  @0 Lymphedema @5 29
C03 21  X  SPA  @0 Linfedema @5 29
C03 22  X  FRE  @0 Mâle @5 35
C03 22  X  ENG  @0 Male @5 35
C03 22  X  SPA  @0 Macho @5 35
C03 23  X  FRE  @0 Cancer sein @4 CD @5 96
C03 23  X  ENG  @0 Breast cancer @4 CD @5 96
C03 23  X  SPA  @0 Cáncer del pecho @4 CD @5 96
C07 01  X  FRE  @0 Homme
C07 01  X  ENG  @0 Human
C07 01  X  SPA  @0 Hombre
C07 02  X  FRE  @0 Appareil circulatoire pathologie @5 37
C07 02  X  ENG  @0 Cardiovascular disease @5 37
C07 02  X  SPA  @0 Aparato circulatorio patología @5 37
C07 03  X  FRE  @0 Lymphatique pathologie @5 38
C07 03  X  ENG  @0 Lymphatic vessel disease @5 38
C07 03  X  SPA  @0 Linfático patología @5 38
N21       @1 164
N44 01      @1 OTO
N82       @1 OTO

Format Inist (serveur)

NO : PASCAL 05-0235133 INIST
ET : The Reliability of a Simplified Water Displacement instrument: A method for measuring arm volume
AU : SAGEN (Ase); KARESEN (Rolf); RISBERG (May Arna)
AF : Surgical Division, Ullevaal University Hospital/Oslo/Norvège (1 aut., 2 aut.); Center for Clinical Research, Ullevaal University Hospital/Oslo/Norvège (3 aut.)
DT : Publication en série; Niveau analytique
SO : Archives of physical medicine and rehabilitation; ISSN 0003-9993; Coden APMHAI; Etats-Unis; Da. 2005; Vol. 86; No. 1; Pp. 86-89; Bibl. 33 ref.
LA : Anglais
EA : Objectives: To present a new water displacement measurement, the Simplified Water Displacement Instrument (SWDI), and to evaluate its intra- and intertester reliability. Design: Reliability design. Setting: Hospital setting. Participants: Fifty-six healthy people were studied. Intratester reliability was evaluated once a week for 4 weeks in 20 women and 10 men. Intertester reliability was assessed by 2 physical therapists in 26 people. Interventions: Not applicable. Main Outcome Measures: Coefficients of variation (CVs) and intraclass correlation coefficients (ICCs). Results: The intratester reliability showed a CV range of 2.2% to 2.6% and an ICC range of .98 to .99. The intertester reliability showed a CV of 1.3% and an ICC of .99. There was a significant increase in arm volume in men compared with women. There were no significant differences in changes in volume over the 4 weeks. There was a significant greater right arm volume (3.3%) among the right-handed subjects (P<.001). Conclusions: Both intra- and intertester reliability were satisfactory for the SWDI.
CC : 002B15; 002B12B04; 002B26O
FD : Tumeur maligne; Glande mammaire pathologie; Fiabilité; Eau; Déplacement; Appareil mesure; Méthode; Bras; Femme; Volume; Instrumentation; Hôpital; Sexe; Pronostic; Coefficient corrélation; Augmentation; Femelle; Etude comparative; Main; Glande mammaire; Lymphoedème; Mâle; Cancer sein
FG : Homme; Appareil circulatoire pathologie; Lymphatique pathologie
ED : Malignant tumor; Mammary gland diseases; Reliability; Water; Displacement; Measuring instrument; Method; Arm; Woman; Volume; Instruments; Hospital; Sex; Prognosis; Correlation coefficient; Increase; Female; Comparative study; Hand; Mammary gland; Lymphedema; Male; Breast cancer
EG : Human; Cardiovascular disease; Lymphatic vessel disease
SD : Tumor maligno; Glándula mamaria patología; Fiabilidad; Agua; Desplazamiento; Aparato medición; Método; Brazo; Mujer; Volumen; Instrumentación; Hospital; Sexo; Pronóstico; Coeficiente correlación; Aumentación; Hembra; Estudio comparativo; Mano; Glándula mamaria; Linfedema; Macho; Cáncer del pecho
LO : INIST-8199.354000126605780170
ID : 05-0235133

Links to Exploration step

Pascal:05-0235133

Le document en format XML

<record>
<TEI>
<teiHeader>
<fileDesc>
<titleStmt>
<title xml:lang="en" level="a">The Reliability of a Simplified Water Displacement instrument: A method for measuring arm volume</title>
<author>
<name sortKey="Sagen, Ase" sort="Sagen, Ase" uniqKey="Sagen A" first="Ase" last="Sagen">Ase Sagen</name>
<affiliation>
<inist:fA14 i1="01">
<s1>Surgical Division, Ullevaal University Hospital</s1>
<s2>Oslo</s2>
<s3>NOR</s3>
<sZ>1 aut.</sZ>
<sZ>2 aut.</sZ>
</inist:fA14>
</affiliation>
</author>
<author>
<name sortKey="Karesen, Rolf" sort="Karesen, Rolf" uniqKey="Karesen R" first="Rolf" last="Karesen">Rolf Karesen</name>
<affiliation>
<inist:fA14 i1="01">
<s1>Surgical Division, Ullevaal University Hospital</s1>
<s2>Oslo</s2>
<s3>NOR</s3>
<sZ>1 aut.</sZ>
<sZ>2 aut.</sZ>
</inist:fA14>
</affiliation>
</author>
<author>
<name sortKey="Risberg, May Arna" sort="Risberg, May Arna" uniqKey="Risberg M" first="May Arna" last="Risberg">May Arna Risberg</name>
<affiliation>
<inist:fA14 i1="02">
<s1>Center for Clinical Research, Ullevaal University Hospital</s1>
<s2>Oslo</s2>
<s3>NOR</s3>
<sZ>3 aut.</sZ>
</inist:fA14>
</affiliation>
</author>
</titleStmt>
<publicationStmt>
<idno type="wicri:source">INIST</idno>
<idno type="inist">05-0235133</idno>
<date when="2005">2005</date>
<idno type="stanalyst">PASCAL 05-0235133 INIST</idno>
<idno type="RBID">Pascal:05-0235133</idno>
<idno type="wicri:Area/PascalFrancis/Corpus">000579</idno>
</publicationStmt>
<sourceDesc>
<biblStruct>
<analytic>
<title xml:lang="en" level="a">The Reliability of a Simplified Water Displacement instrument: A method for measuring arm volume</title>
<author>
<name sortKey="Sagen, Ase" sort="Sagen, Ase" uniqKey="Sagen A" first="Ase" last="Sagen">Ase Sagen</name>
<affiliation>
<inist:fA14 i1="01">
<s1>Surgical Division, Ullevaal University Hospital</s1>
<s2>Oslo</s2>
<s3>NOR</s3>
<sZ>1 aut.</sZ>
<sZ>2 aut.</sZ>
</inist:fA14>
</affiliation>
</author>
<author>
<name sortKey="Karesen, Rolf" sort="Karesen, Rolf" uniqKey="Karesen R" first="Rolf" last="Karesen">Rolf Karesen</name>
<affiliation>
<inist:fA14 i1="01">
<s1>Surgical Division, Ullevaal University Hospital</s1>
<s2>Oslo</s2>
<s3>NOR</s3>
<sZ>1 aut.</sZ>
<sZ>2 aut.</sZ>
</inist:fA14>
</affiliation>
</author>
<author>
<name sortKey="Risberg, May Arna" sort="Risberg, May Arna" uniqKey="Risberg M" first="May Arna" last="Risberg">May Arna Risberg</name>
<affiliation>
<inist:fA14 i1="02">
<s1>Center for Clinical Research, Ullevaal University Hospital</s1>
<s2>Oslo</s2>
<s3>NOR</s3>
<sZ>3 aut.</sZ>
</inist:fA14>
</affiliation>
</author>
</analytic>
<series>
<title level="j" type="main">Archives of physical medicine and rehabilitation</title>
<title level="j" type="abbreviated">Arch. phys. med. rehabil.</title>
<idno type="ISSN">0003-9993</idno>
<imprint>
<date when="2005">2005</date>
</imprint>
</series>
</biblStruct>
</sourceDesc>
<seriesStmt>
<title level="j" type="main">Archives of physical medicine and rehabilitation</title>
<title level="j" type="abbreviated">Arch. phys. med. rehabil.</title>
<idno type="ISSN">0003-9993</idno>
</seriesStmt>
</fileDesc>
<profileDesc>
<textClass>
<keywords scheme="KwdEn" xml:lang="en">
<term>Arm</term>
<term>Breast cancer</term>
<term>Comparative study</term>
<term>Correlation coefficient</term>
<term>Displacement</term>
<term>Female</term>
<term>Hand</term>
<term>Hospital</term>
<term>Increase</term>
<term>Instruments</term>
<term>Lymphedema</term>
<term>Male</term>
<term>Malignant tumor</term>
<term>Mammary gland</term>
<term>Mammary gland diseases</term>
<term>Measuring instrument</term>
<term>Method</term>
<term>Prognosis</term>
<term>Reliability</term>
<term>Sex</term>
<term>Volume</term>
<term>Water</term>
<term>Woman</term>
</keywords>
<keywords scheme="Pascal" xml:lang="fr">
<term>Tumeur maligne</term>
<term>Glande mammaire pathologie</term>
<term>Fiabilité</term>
<term>Eau</term>
<term>Déplacement</term>
<term>Appareil mesure</term>
<term>Méthode</term>
<term>Bras</term>
<term>Femme</term>
<term>Volume</term>
<term>Instrumentation</term>
<term>Hôpital</term>
<term>Sexe</term>
<term>Pronostic</term>
<term>Coefficient corrélation</term>
<term>Augmentation</term>
<term>Femelle</term>
<term>Etude comparative</term>
<term>Main</term>
<term>Glande mammaire</term>
<term>Lymphoedème</term>
<term>Mâle</term>
<term>Cancer sein</term>
</keywords>
</textClass>
</profileDesc>
</teiHeader>
<front>
<div type="abstract" xml:lang="en">Objectives: To present a new water displacement measurement, the Simplified Water Displacement Instrument (SWDI), and to evaluate its intra- and intertester reliability. Design: Reliability design. Setting: Hospital setting. Participants: Fifty-six healthy people were studied. Intratester reliability was evaluated once a week for 4 weeks in 20 women and 10 men. Intertester reliability was assessed by 2 physical therapists in 26 people. Interventions: Not applicable. Main Outcome Measures: Coefficients of variation (CVs) and intraclass correlation coefficients (ICCs). Results: The intratester reliability showed a CV range of 2.2% to 2.6% and an ICC range of .98 to .99. The intertester reliability showed a CV of 1.3% and an ICC of .99. There was a significant increase in arm volume in men compared with women. There were no significant differences in changes in volume over the 4 weeks. There was a significant greater right arm volume (3.3%) among the right-handed subjects (P<.001). Conclusions: Both intra- and intertester reliability were satisfactory for the SWDI.</div>
</front>
</TEI>
<inist>
<standard h6="B">
<pA>
<fA01 i1="01" i2="1">
<s0>0003-9993</s0>
</fA01>
<fA02 i1="01">
<s0>APMHAI</s0>
</fA02>
<fA03 i2="1">
<s0>Arch. phys. med. rehabil.</s0>
</fA03>
<fA05>
<s2>86</s2>
</fA05>
<fA06>
<s2>1</s2>
</fA06>
<fA08 i1="01" i2="1" l="ENG">
<s1>The Reliability of a Simplified Water Displacement instrument: A method for measuring arm volume</s1>
</fA08>
<fA11 i1="01" i2="1">
<s1>SAGEN (Ase)</s1>
</fA11>
<fA11 i1="02" i2="1">
<s1>KARESEN (Rolf)</s1>
</fA11>
<fA11 i1="03" i2="1">
<s1>RISBERG (May Arna)</s1>
</fA11>
<fA14 i1="01">
<s1>Surgical Division, Ullevaal University Hospital</s1>
<s2>Oslo</s2>
<s3>NOR</s3>
<sZ>1 aut.</sZ>
<sZ>2 aut.</sZ>
</fA14>
<fA14 i1="02">
<s1>Center for Clinical Research, Ullevaal University Hospital</s1>
<s2>Oslo</s2>
<s3>NOR</s3>
<sZ>3 aut.</sZ>
</fA14>
<fA20>
<s1>86-89</s1>
</fA20>
<fA21>
<s1>2005</s1>
</fA21>
<fA23 i1="01">
<s0>ENG</s0>
</fA23>
<fA43 i1="01">
<s1>INIST</s1>
<s2>8199</s2>
<s5>354000126605780170</s5>
</fA43>
<fA44>
<s0>0000</s0>
<s1>© 2005 INIST-CNRS. All rights reserved.</s1>
</fA44>
<fA45>
<s0>33 ref.</s0>
</fA45>
<fA47 i1="01" i2="1">
<s0>05-0235133</s0>
</fA47>
<fA60>
<s1>P</s1>
</fA60>
<fA61>
<s0>A</s0>
</fA61>
<fA64 i1="01" i2="1">
<s0>Archives of physical medicine and rehabilitation</s0>
</fA64>
<fA66 i1="01">
<s0>USA</s0>
</fA66>
<fC01 i1="01" l="ENG">
<s0>Objectives: To present a new water displacement measurement, the Simplified Water Displacement Instrument (SWDI), and to evaluate its intra- and intertester reliability. Design: Reliability design. Setting: Hospital setting. Participants: Fifty-six healthy people were studied. Intratester reliability was evaluated once a week for 4 weeks in 20 women and 10 men. Intertester reliability was assessed by 2 physical therapists in 26 people. Interventions: Not applicable. Main Outcome Measures: Coefficients of variation (CVs) and intraclass correlation coefficients (ICCs). Results: The intratester reliability showed a CV range of 2.2% to 2.6% and an ICC range of .98 to .99. The intertester reliability showed a CV of 1.3% and an ICC of .99. There was a significant increase in arm volume in men compared with women. There were no significant differences in changes in volume over the 4 weeks. There was a significant greater right arm volume (3.3%) among the right-handed subjects (P<.001). Conclusions: Both intra- and intertester reliability were satisfactory for the SWDI.</s0>
</fC01>
<fC02 i1="01" i2="X">
<s0>002B15</s0>
</fC02>
<fC02 i1="02" i2="X">
<s0>002B12B04</s0>
</fC02>
<fC02 i1="03" i2="X">
<s0>002B26O</s0>
</fC02>
<fC03 i1="01" i2="X" l="FRE">
<s0>Tumeur maligne</s0>
<s5>01</s5>
</fC03>
<fC03 i1="01" i2="X" l="ENG">
<s0>Malignant tumor</s0>
<s5>01</s5>
</fC03>
<fC03 i1="01" i2="X" l="SPA">
<s0>Tumor maligno</s0>
<s5>01</s5>
</fC03>
<fC03 i1="02" i2="X" l="FRE">
<s0>Glande mammaire pathologie</s0>
<s2>NM</s2>
<s5>02</s5>
</fC03>
<fC03 i1="02" i2="X" l="ENG">
<s0>Mammary gland diseases</s0>
<s2>NM</s2>
<s5>02</s5>
</fC03>
<fC03 i1="02" i2="X" l="SPA">
<s0>Glándula mamaria patología</s0>
<s2>NM</s2>
<s5>02</s5>
</fC03>
<fC03 i1="03" i2="X" l="FRE">
<s0>Fiabilité</s0>
<s5>07</s5>
</fC03>
<fC03 i1="03" i2="X" l="ENG">
<s0>Reliability</s0>
<s5>07</s5>
</fC03>
<fC03 i1="03" i2="X" l="SPA">
<s0>Fiabilidad</s0>
<s5>07</s5>
</fC03>
<fC03 i1="04" i2="X" l="FRE">
<s0>Eau</s0>
<s5>08</s5>
</fC03>
<fC03 i1="04" i2="X" l="ENG">
<s0>Water</s0>
<s5>08</s5>
</fC03>
<fC03 i1="04" i2="X" l="SPA">
<s0>Agua</s0>
<s5>08</s5>
</fC03>
<fC03 i1="05" i2="X" l="FRE">
<s0>Déplacement</s0>
<s5>09</s5>
</fC03>
<fC03 i1="05" i2="X" l="ENG">
<s0>Displacement</s0>
<s5>09</s5>
</fC03>
<fC03 i1="05" i2="X" l="SPA">
<s0>Desplazamiento</s0>
<s5>09</s5>
</fC03>
<fC03 i1="06" i2="X" l="FRE">
<s0>Appareil mesure</s0>
<s5>13</s5>
</fC03>
<fC03 i1="06" i2="X" l="ENG">
<s0>Measuring instrument</s0>
<s5>13</s5>
</fC03>
<fC03 i1="06" i2="X" l="SPA">
<s0>Aparato medición</s0>
<s5>13</s5>
</fC03>
<fC03 i1="07" i2="X" l="FRE">
<s0>Méthode</s0>
<s5>14</s5>
</fC03>
<fC03 i1="07" i2="X" l="ENG">
<s0>Method</s0>
<s5>14</s5>
</fC03>
<fC03 i1="07" i2="X" l="SPA">
<s0>Método</s0>
<s5>14</s5>
</fC03>
<fC03 i1="08" i2="X" l="FRE">
<s0>Bras</s0>
<s5>15</s5>
</fC03>
<fC03 i1="08" i2="X" l="ENG">
<s0>Arm</s0>
<s5>15</s5>
</fC03>
<fC03 i1="08" i2="X" l="SPA">
<s0>Brazo</s0>
<s5>15</s5>
</fC03>
<fC03 i1="09" i2="X" l="FRE">
<s0>Femme</s0>
<s5>16</s5>
</fC03>
<fC03 i1="09" i2="X" l="ENG">
<s0>Woman</s0>
<s5>16</s5>
</fC03>
<fC03 i1="09" i2="X" l="SPA">
<s0>Mujer</s0>
<s5>16</s5>
</fC03>
<fC03 i1="10" i2="X" l="FRE">
<s0>Volume</s0>
<s5>17</s5>
</fC03>
<fC03 i1="10" i2="X" l="ENG">
<s0>Volume</s0>
<s5>17</s5>
</fC03>
<fC03 i1="10" i2="X" l="SPA">
<s0>Volumen</s0>
<s5>17</s5>
</fC03>
<fC03 i1="11" i2="X" l="FRE">
<s0>Instrumentation</s0>
<s5>18</s5>
</fC03>
<fC03 i1="11" i2="X" l="ENG">
<s0>Instruments</s0>
<s5>18</s5>
</fC03>
<fC03 i1="11" i2="X" l="SPA">
<s0>Instrumentación</s0>
<s5>18</s5>
</fC03>
<fC03 i1="12" i2="X" l="FRE">
<s0>Hôpital</s0>
<s5>19</s5>
</fC03>
<fC03 i1="12" i2="X" l="ENG">
<s0>Hospital</s0>
<s5>19</s5>
</fC03>
<fC03 i1="12" i2="X" l="SPA">
<s0>Hospital</s0>
<s5>19</s5>
</fC03>
<fC03 i1="13" i2="X" l="FRE">
<s0>Sexe</s0>
<s5>20</s5>
</fC03>
<fC03 i1="13" i2="X" l="ENG">
<s0>Sex</s0>
<s5>20</s5>
</fC03>
<fC03 i1="13" i2="X" l="SPA">
<s0>Sexo</s0>
<s5>20</s5>
</fC03>
<fC03 i1="14" i2="X" l="FRE">
<s0>Pronostic</s0>
<s5>21</s5>
</fC03>
<fC03 i1="14" i2="X" l="ENG">
<s0>Prognosis</s0>
<s5>21</s5>
</fC03>
<fC03 i1="14" i2="X" l="SPA">
<s0>Pronóstico</s0>
<s5>21</s5>
</fC03>
<fC03 i1="15" i2="X" l="FRE">
<s0>Coefficient corrélation</s0>
<s5>22</s5>
</fC03>
<fC03 i1="15" i2="X" l="ENG">
<s0>Correlation coefficient</s0>
<s5>22</s5>
</fC03>
<fC03 i1="15" i2="X" l="SPA">
<s0>Coeficiente correlación</s0>
<s5>22</s5>
</fC03>
<fC03 i1="16" i2="X" l="FRE">
<s0>Augmentation</s0>
<s5>23</s5>
</fC03>
<fC03 i1="16" i2="X" l="ENG">
<s0>Increase</s0>
<s5>23</s5>
</fC03>
<fC03 i1="16" i2="X" l="SPA">
<s0>Aumentación</s0>
<s5>23</s5>
</fC03>
<fC03 i1="17" i2="X" l="FRE">
<s0>Femelle</s0>
<s5>25</s5>
</fC03>
<fC03 i1="17" i2="X" l="ENG">
<s0>Female</s0>
<s5>25</s5>
</fC03>
<fC03 i1="17" i2="X" l="SPA">
<s0>Hembra</s0>
<s5>25</s5>
</fC03>
<fC03 i1="18" i2="X" l="FRE">
<s0>Etude comparative</s0>
<s5>26</s5>
</fC03>
<fC03 i1="18" i2="X" l="ENG">
<s0>Comparative study</s0>
<s5>26</s5>
</fC03>
<fC03 i1="18" i2="X" l="SPA">
<s0>Estudio comparativo</s0>
<s5>26</s5>
</fC03>
<fC03 i1="19" i2="X" l="FRE">
<s0>Main</s0>
<s5>27</s5>
</fC03>
<fC03 i1="19" i2="X" l="ENG">
<s0>Hand</s0>
<s5>27</s5>
</fC03>
<fC03 i1="19" i2="X" l="SPA">
<s0>Mano</s0>
<s5>27</s5>
</fC03>
<fC03 i1="20" i2="X" l="FRE">
<s0>Glande mammaire</s0>
<s5>28</s5>
</fC03>
<fC03 i1="20" i2="X" l="ENG">
<s0>Mammary gland</s0>
<s5>28</s5>
</fC03>
<fC03 i1="20" i2="X" l="SPA">
<s0>Glándula mamaria</s0>
<s5>28</s5>
</fC03>
<fC03 i1="21" i2="X" l="FRE">
<s0>Lymphoedème</s0>
<s5>29</s5>
</fC03>
<fC03 i1="21" i2="X" l="ENG">
<s0>Lymphedema</s0>
<s5>29</s5>
</fC03>
<fC03 i1="21" i2="X" l="SPA">
<s0>Linfedema</s0>
<s5>29</s5>
</fC03>
<fC03 i1="22" i2="X" l="FRE">
<s0>Mâle</s0>
<s5>35</s5>
</fC03>
<fC03 i1="22" i2="X" l="ENG">
<s0>Male</s0>
<s5>35</s5>
</fC03>
<fC03 i1="22" i2="X" l="SPA">
<s0>Macho</s0>
<s5>35</s5>
</fC03>
<fC03 i1="23" i2="X" l="FRE">
<s0>Cancer sein</s0>
<s4>CD</s4>
<s5>96</s5>
</fC03>
<fC03 i1="23" i2="X" l="ENG">
<s0>Breast cancer</s0>
<s4>CD</s4>
<s5>96</s5>
</fC03>
<fC03 i1="23" i2="X" l="SPA">
<s0>Cáncer del pecho</s0>
<s4>CD</s4>
<s5>96</s5>
</fC03>
<fC07 i1="01" i2="X" l="FRE">
<s0>Homme</s0>
</fC07>
<fC07 i1="01" i2="X" l="ENG">
<s0>Human</s0>
</fC07>
<fC07 i1="01" i2="X" l="SPA">
<s0>Hombre</s0>
</fC07>
<fC07 i1="02" i2="X" l="FRE">
<s0>Appareil circulatoire pathologie</s0>
<s5>37</s5>
</fC07>
<fC07 i1="02" i2="X" l="ENG">
<s0>Cardiovascular disease</s0>
<s5>37</s5>
</fC07>
<fC07 i1="02" i2="X" l="SPA">
<s0>Aparato circulatorio patología</s0>
<s5>37</s5>
</fC07>
<fC07 i1="03" i2="X" l="FRE">
<s0>Lymphatique pathologie</s0>
<s5>38</s5>
</fC07>
<fC07 i1="03" i2="X" l="ENG">
<s0>Lymphatic vessel disease</s0>
<s5>38</s5>
</fC07>
<fC07 i1="03" i2="X" l="SPA">
<s0>Linfático patología</s0>
<s5>38</s5>
</fC07>
<fN21>
<s1>164</s1>
</fN21>
<fN44 i1="01">
<s1>OTO</s1>
</fN44>
<fN82>
<s1>OTO</s1>
</fN82>
</pA>
</standard>
<server>
<NO>PASCAL 05-0235133 INIST</NO>
<ET>The Reliability of a Simplified Water Displacement instrument: A method for measuring arm volume</ET>
<AU>SAGEN (Ase); KARESEN (Rolf); RISBERG (May Arna)</AU>
<AF>Surgical Division, Ullevaal University Hospital/Oslo/Norvège (1 aut., 2 aut.); Center for Clinical Research, Ullevaal University Hospital/Oslo/Norvège (3 aut.)</AF>
<DT>Publication en série; Niveau analytique</DT>
<SO>Archives of physical medicine and rehabilitation; ISSN 0003-9993; Coden APMHAI; Etats-Unis; Da. 2005; Vol. 86; No. 1; Pp. 86-89; Bibl. 33 ref.</SO>
<LA>Anglais</LA>
<EA>Objectives: To present a new water displacement measurement, the Simplified Water Displacement Instrument (SWDI), and to evaluate its intra- and intertester reliability. Design: Reliability design. Setting: Hospital setting. Participants: Fifty-six healthy people were studied. Intratester reliability was evaluated once a week for 4 weeks in 20 women and 10 men. Intertester reliability was assessed by 2 physical therapists in 26 people. Interventions: Not applicable. Main Outcome Measures: Coefficients of variation (CVs) and intraclass correlation coefficients (ICCs). Results: The intratester reliability showed a CV range of 2.2% to 2.6% and an ICC range of .98 to .99. The intertester reliability showed a CV of 1.3% and an ICC of .99. There was a significant increase in arm volume in men compared with women. There were no significant differences in changes in volume over the 4 weeks. There was a significant greater right arm volume (3.3%) among the right-handed subjects (P<.001). Conclusions: Both intra- and intertester reliability were satisfactory for the SWDI.</EA>
<CC>002B15; 002B12B04; 002B26O</CC>
<FD>Tumeur maligne; Glande mammaire pathologie; Fiabilité; Eau; Déplacement; Appareil mesure; Méthode; Bras; Femme; Volume; Instrumentation; Hôpital; Sexe; Pronostic; Coefficient corrélation; Augmentation; Femelle; Etude comparative; Main; Glande mammaire; Lymphoedème; Mâle; Cancer sein</FD>
<FG>Homme; Appareil circulatoire pathologie; Lymphatique pathologie</FG>
<ED>Malignant tumor; Mammary gland diseases; Reliability; Water; Displacement; Measuring instrument; Method; Arm; Woman; Volume; Instruments; Hospital; Sex; Prognosis; Correlation coefficient; Increase; Female; Comparative study; Hand; Mammary gland; Lymphedema; Male; Breast cancer</ED>
<EG>Human; Cardiovascular disease; Lymphatic vessel disease</EG>
<SD>Tumor maligno; Glándula mamaria patología; Fiabilidad; Agua; Desplazamiento; Aparato medición; Método; Brazo; Mujer; Volumen; Instrumentación; Hospital; Sexo; Pronóstico; Coeficiente correlación; Aumentación; Hembra; Estudio comparativo; Mano; Glándula mamaria; Linfedema; Macho; Cáncer del pecho</SD>
<LO>INIST-8199.354000126605780170</LO>
<ID>05-0235133</ID>
</server>
</inist>
</record>

Pour manipuler ce document sous Unix (Dilib)

EXPLOR_STEP=$WICRI_ROOT/Wicri/Sante/explor/LymphedemaV1/Data/PascalFrancis/Corpus
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 000579 | SxmlIndent | more

Ou

HfdSelect -h $EXPLOR_AREA/Data/PascalFrancis/Corpus/biblio.hfd -nk 000579 | SxmlIndent | more

Pour mettre un lien sur cette page dans le réseau Wicri

{{Explor lien
   |wiki=    Wicri/Sante
   |area=    LymphedemaV1
   |flux=    PascalFrancis
   |étape=   Corpus
   |type=    RBID
   |clé=     Pascal:05-0235133
   |texte=   The Reliability of a Simplified Water Displacement instrument: A method for measuring arm volume
}}

Wicri

This area was generated with Dilib version V0.6.31.
Data generation: Sat Nov 4 17:40:35 2017. Site generation: Tue Feb 13 16:42:16 2024