The influence of thread geometry on implant osseointegration under immediate loading: a literature review

Identifieur interne : 002F07 ( Pmc/Corpus ); précédent : 002F06; suivant : 002F08

The influence of thread geometry on implant osseointegration under immediate loading: a literature review

Auteurs : Hyo-Sook Ryu ; Cheol Namgung ; Jong-Ho Lee ; Young-Jun Lim

Source :

RBID : PMC:4279055

Abstract

Implant success is achieved by the synergistic combination of numerous biomechanical factors. This report examines the mechanical aspect of implants. In particular, it is focused on macrodesign such as thread shape, pitch, width and depth, and crestal module of implants. This study reviews the literature regarding the effect of implant thread geometry on primary stability and osseointegration under immediate loading. The search strategy included both in vitro and in vivo studies published in the MEDLINE database from January 2000 to June 2014. Various geometrical parameters are analyzed to evaluate their significance for optimal stress distribution, implant surface area, and bone remodeling responses during the process of osseointegration.


Url:
DOI: 10.4047/jap.2014.6.6.547
PubMed: 25551016
PubMed Central: 4279055

Links to Exploration step

PMC:4279055

Le document en format XML

<record>
<TEI>
<teiHeader>
<fileDesc>
<titleStmt>
<title xml:lang="en">The influence of thread geometry on implant osseointegration under immediate loading: a literature review</title>
<author>
<name sortKey="Ryu, Hyo Sook" sort="Ryu, Hyo Sook" uniqKey="Ryu H" first="Hyo-Sook" last="Ryu">Hyo-Sook Ryu</name>
<affiliation>
<nlm:aff id="A1">Department of Prosthodontics and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Republic of Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Namgung, Cheol" sort="Namgung, Cheol" uniqKey="Namgung C" first="Cheol" last="Namgung">Cheol Namgung</name>
<affiliation>
<nlm:aff id="A1">Department of Prosthodontics and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Republic of Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Lee, Jong Ho" sort="Lee, Jong Ho" uniqKey="Lee J" first="Jong-Ho" last="Lee">Jong-Ho Lee</name>
<affiliation>
<nlm:aff id="A2">Department of Oral Maxillofacial Surgery and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Republic of Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Lim, Young Jun" sort="Lim, Young Jun" uniqKey="Lim Y" first="Young-Jun" last="Lim">Young-Jun Lim</name>
<affiliation>
<nlm:aff id="A1">Department of Prosthodontics and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Republic of Korea.</nlm:aff>
</affiliation>
</author>
</titleStmt>
<publicationStmt>
<idno type="wicri:source">PMC</idno>
<idno type="pmid">25551016</idno>
<idno type="pmc">4279055</idno>
<idno type="url">http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4279055</idno>
<idno type="RBID">PMC:4279055</idno>
<idno type="doi">10.4047/jap.2014.6.6.547</idno>
<date when="2014">2014</date>
<idno type="wicri:Area/Pmc/Corpus">002F07</idno>
<idno type="wicri:explorRef" wicri:stream="Pmc" wicri:step="Corpus" wicri:corpus="PMC">002F07</idno>
</publicationStmt>
<sourceDesc>
<biblStruct>
<analytic>
<title xml:lang="en" level="a" type="main">The influence of thread geometry on implant osseointegration under immediate loading: a literature review</title>
<author>
<name sortKey="Ryu, Hyo Sook" sort="Ryu, Hyo Sook" uniqKey="Ryu H" first="Hyo-Sook" last="Ryu">Hyo-Sook Ryu</name>
<affiliation>
<nlm:aff id="A1">Department of Prosthodontics and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Republic of Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Namgung, Cheol" sort="Namgung, Cheol" uniqKey="Namgung C" first="Cheol" last="Namgung">Cheol Namgung</name>
<affiliation>
<nlm:aff id="A1">Department of Prosthodontics and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Republic of Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Lee, Jong Ho" sort="Lee, Jong Ho" uniqKey="Lee J" first="Jong-Ho" last="Lee">Jong-Ho Lee</name>
<affiliation>
<nlm:aff id="A2">Department of Oral Maxillofacial Surgery and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Republic of Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Lim, Young Jun" sort="Lim, Young Jun" uniqKey="Lim Y" first="Young-Jun" last="Lim">Young-Jun Lim</name>
<affiliation>
<nlm:aff id="A1">Department of Prosthodontics and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Republic of Korea.</nlm:aff>
</affiliation>
</author>
</analytic>
<series>
<title level="j">The Journal of Advanced Prosthodontics</title>
<idno type="ISSN">2005-7806</idno>
<idno type="eISSN">2005-7814</idno>
<imprint>
<date when="2014">2014</date>
</imprint>
</series>
</biblStruct>
</sourceDesc>
</fileDesc>
<profileDesc>
<textClass></textClass>
</profileDesc>
</teiHeader>
<front>
<div type="abstract" xml:lang="en">
<p>Implant success is achieved by the synergistic combination of numerous biomechanical factors. This report examines the mechanical aspect of implants. In particular, it is focused on macrodesign such as thread shape, pitch, width and depth, and crestal module of implants. This study reviews the literature regarding the effect of implant thread geometry on primary stability and osseointegration under immediate loading. The search strategy included both
<italic>in vitro</italic>
and
<italic>in vivo</italic>
studies published in the MEDLINE database from January 2000 to June 2014. Various geometrical parameters are analyzed to evaluate their significance for optimal stress distribution, implant surface area, and bone remodeling responses during the process of osseointegration.</p>
</div>
</front>
<back>
<div1 type="bibliography">
<listBibl>
<biblStruct>
<analytic>
<author>
<name sortKey="Haas, R" uniqKey="Haas R">R Haas</name>
</author>
<author>
<name sortKey="Mensdorff Pouilly, N" uniqKey="Mensdorff Pouilly N">N Mensdorff-Pouilly</name>
</author>
<author>
<name sortKey="Mailath, G" uniqKey="Mailath G">G Mailath</name>
</author>
<author>
<name sortKey="Watzek, G" uniqKey="Watzek G">G Watzek</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Goodacre, Cj" uniqKey="Goodacre C">CJ Goodacre</name>
</author>
<author>
<name sortKey="Bernal, G" uniqKey="Bernal G">G Bernal</name>
</author>
<author>
<name sortKey="Rungcharassaeng, K" uniqKey="Rungcharassaeng K">K Rungcharassaeng</name>
</author>
<author>
<name sortKey="Kan, Jy" uniqKey="Kan J">JY Kan</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Fugazzotto, Pa" uniqKey="Fugazzotto P">PA Fugazzotto</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Atieh, Ma" uniqKey="Atieh M">MA Atieh</name>
</author>
<author>
<name sortKey="Atieh, Ah" uniqKey="Atieh A">AH Atieh</name>
</author>
<author>
<name sortKey="Payne, Ag" uniqKey="Payne A">AG Payne</name>
</author>
<author>
<name sortKey="Duncan, Wj" uniqKey="Duncan W">WJ Duncan</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Cochran, Dl" uniqKey="Cochran D">DL Cochran</name>
</author>
<author>
<name sortKey="Morton, D" uniqKey="Morton D">D Morton</name>
</author>
<author>
<name sortKey="Weber, Hp" uniqKey="Weber H">HP Weber</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Glauser, R" uniqKey="Glauser R">R Glauser</name>
</author>
<author>
<name sortKey="Zembic, A" uniqKey="Zembic A">A Zembic</name>
</author>
<author>
<name sortKey="H Mmerle, Ch" uniqKey="H Mmerle C">CH Hämmerle</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Romanos, Ge" uniqKey="Romanos G">GE Romanos</name>
</author>
<author>
<name sortKey="Toh, Cg" uniqKey="Toh C">CG Toh</name>
</author>
<author>
<name sortKey="Siar, Ch" uniqKey="Siar C">CH Siar</name>
</author>
<author>
<name sortKey="Swaminathan, D" uniqKey="Swaminathan D">D Swaminathan</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Iba Ez, Jc" uniqKey="Iba Ez J">JC Ibañez</name>
</author>
<author>
<name sortKey="Jalbout, Zn" uniqKey="Jalbout Z">ZN Jalbout</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Rocci, A" uniqKey="Rocci A">A Rocci</name>
</author>
<author>
<name sortKey="Rocci, M" uniqKey="Rocci M">M Rocci</name>
</author>
<author>
<name sortKey="Rocci, C" uniqKey="Rocci C">C Rocci</name>
</author>
<author>
<name sortKey="Scoccia, A" uniqKey="Scoccia A">A Scoccia</name>
</author>
<author>
<name sortKey="Gargari, M" uniqKey="Gargari M">M Gargari</name>
</author>
<author>
<name sortKey="Martignoni, M" uniqKey="Martignoni M">M Martignoni</name>
</author>
<author>
<name sortKey="Gottlow, J" uniqKey="Gottlow J">J Gottlow</name>
</author>
<author>
<name sortKey="Sennerby, L" uniqKey="Sennerby L">L Sennerby</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Ostman, Po" uniqKey="Ostman P">PO Ostman</name>
</author>
<author>
<name sortKey="Hellman, M" uniqKey="Hellman M">M Hellman</name>
</author>
<author>
<name sortKey="Sennerby, L" uniqKey="Sennerby L">L Sennerby</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Sennerby, L" uniqKey="Sennerby L">L Sennerby</name>
</author>
<author>
<name sortKey="Meredith, N" uniqKey="Meredith N">N Meredith</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Abuhussein, H" uniqKey="Abuhussein H">H Abuhussein</name>
</author>
<author>
<name sortKey="Pagni, G" uniqKey="Pagni G">G Pagni</name>
</author>
<author>
<name sortKey="Rebaudi, A" uniqKey="Rebaudi A">A Rebaudi</name>
</author>
<author>
<name sortKey="Wang, Hl" uniqKey="Wang H">HL Wang</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Boggan, Rs" uniqKey="Boggan R">RS Boggan</name>
</author>
<author>
<name sortKey="Strong, Jt" uniqKey="Strong J">JT Strong</name>
</author>
<author>
<name sortKey="Misch, Ce" uniqKey="Misch C">CE Misch</name>
</author>
<author>
<name sortKey="Bidez, Mw" uniqKey="Bidez M">MW Bidez</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Hansson, S" uniqKey="Hansson S">S Hansson</name>
</author>
<author>
<name sortKey="Werke, M" uniqKey="Werke M">M Werke</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Prendergast, Pj" uniqKey="Prendergast P">PJ Prendergast</name>
</author>
<author>
<name sortKey="Huiskes, R" uniqKey="Huiskes R">R Huiskes</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Brunski, Jb" uniqKey="Brunski J">JB Brunski</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Frost, Hm" uniqKey="Frost H">HM Frost</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Chang, Pk" uniqKey="Chang P">PK Chang</name>
</author>
<author>
<name sortKey="Chen, Yc" uniqKey="Chen Y">YC Chen</name>
</author>
<author>
<name sortKey="Huang, Cc" uniqKey="Huang C">CC Huang</name>
</author>
<author>
<name sortKey="Lu, Wh" uniqKey="Lu W">WH Lu</name>
</author>
<author>
<name sortKey="Chen, Yc" uniqKey="Chen Y">YC Chen</name>
</author>
<author>
<name sortKey="Tsai, Hh" uniqKey="Tsai H">HH Tsai</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Eraslan, O" uniqKey="Eraslan O">O Eraslan</name>
</author>
<author>
<name sortKey="Inan, O" uniqKey="Inan O">O Inan</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Steigenga, J" uniqKey="Steigenga J">J Steigenga</name>
</author>
<author>
<name sortKey="Al Shammari, K" uniqKey="Al Shammari K">K Al-Shammari</name>
</author>
<author>
<name sortKey="Misch, C" uniqKey="Misch C">C Misch</name>
</author>
<author>
<name sortKey="Nociti, Fh" uniqKey="Nociti F">FH Nociti</name>
</author>
<author>
<name sortKey="Wang, Hl" uniqKey="Wang H">HL Wang</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Geng, Jp" uniqKey="Geng J">JP Geng</name>
</author>
<author>
<name sortKey="Ma, Qs" uniqKey="Ma Q">QS Ma</name>
</author>
<author>
<name sortKey="Xu, W" uniqKey="Xu W">W Xu</name>
</author>
<author>
<name sortKey="Tan, Kb" uniqKey="Tan K">KB Tan</name>
</author>
<author>
<name sortKey="Liu, Gr" uniqKey="Liu G">GR Liu</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Chun, Hj" uniqKey="Chun H">HJ Chun</name>
</author>
<author>
<name sortKey="Cheong, Sy" uniqKey="Cheong S">SY Cheong</name>
</author>
<author>
<name sortKey="Han, Jh" uniqKey="Han J">JH Han</name>
</author>
<author>
<name sortKey="Heo, Sj" uniqKey="Heo S">SJ Heo</name>
</author>
<author>
<name sortKey="Chung, Jp" uniqKey="Chung J">JP Chung</name>
</author>
<author>
<name sortKey="Rhyu, Ic" uniqKey="Rhyu I">IC Rhyu</name>
</author>
<author>
<name sortKey="Choi, Yc" uniqKey="Choi Y">YC Choi</name>
</author>
<author>
<name sortKey="Baik, Hk" uniqKey="Baik H">HK Baik</name>
</author>
<author>
<name sortKey="Ku, Y" uniqKey="Ku Y">Y Ku</name>
</author>
<author>
<name sortKey="Kim, Mh" uniqKey="Kim M">MH Kim</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Mcallister, Bs" uniqKey="Mcallister B">BS McAllister</name>
</author>
<author>
<name sortKey="Cherry, Je" uniqKey="Cherry J">JE Cherry</name>
</author>
<author>
<name sortKey="Kolinski, Ml" uniqKey="Kolinski M">ML Kolinski</name>
</author>
<author>
<name sortKey="Parrish, Kd" uniqKey="Parrish K">KD Parrish</name>
</author>
<author>
<name sortKey="Pumphrey, Dw" uniqKey="Pumphrey D">DW Pumphrey</name>
</author>
<author>
<name sortKey="Schroering, Rl" uniqKey="Schroering R">RL Schroering</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Arnhart, C" uniqKey="Arnhart C">C Arnhart</name>
</author>
<author>
<name sortKey="Kielbassa, Am" uniqKey="Kielbassa A">AM Kielbassa</name>
</author>
<author>
<name sortKey="Martinez De Fuentes, R" uniqKey="Martinez De Fuentes R">R Martinez-de Fuentes</name>
</author>
<author>
<name sortKey="Goldstein, M" uniqKey="Goldstein M">M Goldstein</name>
</author>
<author>
<name sortKey="Jackowski, J" uniqKey="Jackowski J">J Jackowski</name>
</author>
<author>
<name sortKey="Lorenzoni, M" uniqKey="Lorenzoni M">M Lorenzoni</name>
</author>
<author>
<name sortKey="Maiorana, C" uniqKey="Maiorana C">C Maiorana</name>
</author>
<author>
<name sortKey="Mericske Stern, R" uniqKey="Mericske Stern R">R Mericske-Stern</name>
</author>
<author>
<name sortKey="Pozzi, A" uniqKey="Pozzi A">A Pozzi</name>
</author>
<author>
<name sortKey="Rompen, E" uniqKey="Rompen E">E Rompen</name>
</author>
<author>
<name sortKey="Sanz, M" uniqKey="Sanz M">M Sanz</name>
</author>
<author>
<name sortKey="Strub, Jr" uniqKey="Strub J">JR Strub</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Ivanoff, Cj" uniqKey="Ivanoff C">CJ Ivanoff</name>
</author>
<author>
<name sortKey="Grondahl, K" uniqKey="Grondahl K">K Gröndahl</name>
</author>
<author>
<name sortKey="Sennerby, L" uniqKey="Sennerby L">L Sennerby</name>
</author>
<author>
<name sortKey="Bergstrom, C" uniqKey="Bergstrom C">C Bergström</name>
</author>
<author>
<name sortKey="Lekholm, U" uniqKey="Lekholm U">U Lekholm</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Misch, Ce" uniqKey="Misch C">CE Misch</name>
</author>
<author>
<name sortKey="Steignga, J" uniqKey="Steignga J">J Steignga</name>
</author>
<author>
<name sortKey="Barboza, E" uniqKey="Barboza E">E Barboza</name>
</author>
<author>
<name sortKey="Misch Dietsh, F" uniqKey="Misch Dietsh F">F Misch-Dietsh</name>
</author>
<author>
<name sortKey="Cianciola, Lj" uniqKey="Cianciola L">LJ Cianciola</name>
</author>
<author>
<name sortKey="Kazor, C" uniqKey="Kazor C">C Kazor</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Steigenga, Jt" uniqKey="Steigenga J">JT Steigenga</name>
</author>
<author>
<name sortKey="Al Shammari, Kf" uniqKey="Al Shammari K">KF al-Shammari</name>
</author>
<author>
<name sortKey="Nociti, Fh" uniqKey="Nociti F">FH Nociti</name>
</author>
<author>
<name sortKey="Misch, Ce" uniqKey="Misch C">CE Misch</name>
</author>
<author>
<name sortKey="Wang, Hl" uniqKey="Wang H">HL Wang</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Orsini, E" uniqKey="Orsini E">E Orsini</name>
</author>
<author>
<name sortKey="Giavaresi, G" uniqKey="Giavaresi G">G Giavaresi</name>
</author>
<author>
<name sortKey="Trire, A" uniqKey="Trire A">A Trirè</name>
</author>
<author>
<name sortKey="Ottani, V" uniqKey="Ottani V">V Ottani</name>
</author>
<author>
<name sortKey="Salgarello, S" uniqKey="Salgarello S">S Salgarello</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Chung, Sh" uniqKey="Chung S">SH Chung</name>
</author>
<author>
<name sortKey="Heo, Sj" uniqKey="Heo S">SJ Heo</name>
</author>
<author>
<name sortKey="Koak, Jy" uniqKey="Koak J">JY Koak</name>
</author>
<author>
<name sortKey="Kim, Sk" uniqKey="Kim S">SK Kim</name>
</author>
<author>
<name sortKey="Lee, Jb" uniqKey="Lee J">JB Lee</name>
</author>
<author>
<name sortKey="Han, Js" uniqKey="Han J">JS Han</name>
</author>
<author>
<name sortKey="Han, Ch" uniqKey="Han C">CH Han</name>
</author>
<author>
<name sortKey="Rhyu, Ic" uniqKey="Rhyu I">IC Rhyu</name>
</author>
<author>
<name sortKey="Lee, Sj" uniqKey="Lee S">SJ Lee</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Ma, P" uniqKey="Ma P">P Ma</name>
</author>
<author>
<name sortKey="Liu, Hc" uniqKey="Liu H">HC Liu</name>
</author>
<author>
<name sortKey="Li, Dh" uniqKey="Li D">DH Li</name>
</author>
<author>
<name sortKey="Lin, S" uniqKey="Lin S">S Lin</name>
</author>
<author>
<name sortKey="Shi, Z" uniqKey="Shi Z">Z Shi</name>
</author>
<author>
<name sortKey="Peng, Qj" uniqKey="Peng Q">QJ Peng</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Motoyoshi, M" uniqKey="Motoyoshi M">M Motoyoshi</name>
</author>
<author>
<name sortKey="Yano, S" uniqKey="Yano S">S Yano</name>
</author>
<author>
<name sortKey="Tsuruoka, T" uniqKey="Tsuruoka T">T Tsuruoka</name>
</author>
<author>
<name sortKey="Shimizu, N" uniqKey="Shimizu N">N Shimizu</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Lan, Th" uniqKey="Lan T">TH Lan</name>
</author>
<author>
<name sortKey="Du, Jk" uniqKey="Du J">JK Du</name>
</author>
<author>
<name sortKey="Pan, Cy" uniqKey="Pan C">CY Pan</name>
</author>
<author>
<name sortKey="Lee, He" uniqKey="Lee H">HE Lee</name>
</author>
<author>
<name sortKey="Chung, Wh" uniqKey="Chung W">WH Chung</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Kong, L" uniqKey="Kong L">L Kong</name>
</author>
<author>
<name sortKey="Liu, Bl" uniqKey="Liu B">BL Liu</name>
</author>
<author>
<name sortKey="Hu, Kj" uniqKey="Hu K">KJ Hu</name>
</author>
<author>
<name sortKey="Li, Dh" uniqKey="Li D">DH Li</name>
</author>
<author>
<name sortKey="Song, Yl" uniqKey="Song Y">YL Song</name>
</author>
<author>
<name sortKey="Ma, P" uniqKey="Ma P">P Ma</name>
</author>
<author>
<name sortKey="Yang, J" uniqKey="Yang J">J Yang</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Misch, Ce" uniqKey="Misch C">CE Misch</name>
</author>
<author>
<name sortKey="Strong, T" uniqKey="Strong T">T Strong</name>
</author>
<author>
<name sortKey="Bidez, Mw" uniqKey="Bidez M">MW Bidez</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Ao, J" uniqKey="Ao J">J Ao</name>
</author>
<author>
<name sortKey="Li, T" uniqKey="Li T">T Li</name>
</author>
<author>
<name sortKey="Liu, Y" uniqKey="Liu Y">Y Liu</name>
</author>
<author>
<name sortKey="Ding, Y" uniqKey="Ding Y">Y Ding</name>
</author>
<author>
<name sortKey="Wu, G" uniqKey="Wu G">G Wu</name>
</author>
<author>
<name sortKey="Hu, K" uniqKey="Hu K">K Hu</name>
</author>
<author>
<name sortKey="Kong, L" uniqKey="Kong L">L Kong</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Kong, L" uniqKey="Kong L">L Kong</name>
</author>
<author>
<name sortKey="Hu, K" uniqKey="Hu K">K Hu</name>
</author>
<author>
<name sortKey="Li, D" uniqKey="Li D">D Li</name>
</author>
<author>
<name sortKey="Song, Y" uniqKey="Song Y">Y Song</name>
</author>
<author>
<name sortKey="Yang, J" uniqKey="Yang J">J Yang</name>
</author>
<author>
<name sortKey="Wu, Z" uniqKey="Wu Z">Z Wu</name>
</author>
<author>
<name sortKey="Liu, B" uniqKey="Liu B">B Liu</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Quirynen, M" uniqKey="Quirynen M">M Quirynen</name>
</author>
<author>
<name sortKey="Naert, I" uniqKey="Naert I">I Naert</name>
</author>
<author>
<name sortKey="Van Steenberghe, D" uniqKey="Van Steenberghe D">D van Steenberghe</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Shimada, E" uniqKey="Shimada E">E Shimada</name>
</author>
<author>
<name sortKey="Pilliar, Rm" uniqKey="Pilliar R">RM Pilliar</name>
</author>
<author>
<name sortKey="Deporter, Da" uniqKey="Deporter D">DA Deporter</name>
</author>
<author>
<name sortKey="Schroering, R" uniqKey="Schroering R">R Schroering</name>
</author>
<author>
<name sortKey="Atenafu, E" uniqKey="Atenafu E">E Atenafu</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Hermann, Js" uniqKey="Hermann J">JS Hermann</name>
</author>
<author>
<name sortKey="Schoolfield, Jd" uniqKey="Schoolfield J">JD Schoolfield</name>
</author>
<author>
<name sortKey="Nummikoski, Pv" uniqKey="Nummikoski P">PV Nummikoski</name>
</author>
<author>
<name sortKey="Buser, D" uniqKey="Buser D">D Buser</name>
</author>
<author>
<name sortKey="Schenk, Rk" uniqKey="Schenk R">RK Schenk</name>
</author>
<author>
<name sortKey="Cochran, Dl" uniqKey="Cochran D">DL Cochran</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="H Nggi, Mp" uniqKey="H Nggi M">MP Hänggi</name>
</author>
<author>
<name sortKey="H Nggi, Dc" uniqKey="H Nggi D">DC Hänggi</name>
</author>
<author>
<name sortKey="Schoolfield, Jd" uniqKey="Schoolfield J">JD Schoolfield</name>
</author>
<author>
<name sortKey="Meyer, J" uniqKey="Meyer J">J Meyer</name>
</author>
<author>
<name sortKey="Cochran, Dl" uniqKey="Cochran D">DL Cochran</name>
</author>
<author>
<name sortKey="Hermann, Js" uniqKey="Hermann J">JS Hermann</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Vaillancourt, H" uniqKey="Vaillancourt H">H Vaillancourt</name>
</author>
<author>
<name sortKey="Pilliar, Rm" uniqKey="Pilliar R">RM Pilliar</name>
</author>
<author>
<name sortKey="Mccammond, D" uniqKey="Mccammond D">D McCammond</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Hansson, S" uniqKey="Hansson S">S Hansson</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Chowdhary, R" uniqKey="Chowdhary R">R Chowdhary</name>
</author>
<author>
<name sortKey="Halldin, A" uniqKey="Halldin A">A Halldin</name>
</author>
<author>
<name sortKey="Jimbo, R" uniqKey="Jimbo R">R Jimbo</name>
</author>
<author>
<name sortKey="Wennerberg, A" uniqKey="Wennerberg A">A Wennerberg</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Negri, B" uniqKey="Negri B">B Negri</name>
</author>
<author>
<name sortKey="Calvo Guirado, Jl" uniqKey="Calvo Guirado J">JL Calvo Guirado</name>
</author>
<author>
<name sortKey="Mate Sanchez De Val, Je" uniqKey="Mate Sanchez De Val J">JE Maté Sánchez de Val</name>
</author>
<author>
<name sortKey="Delgado Gt Ruiz, Ra" uniqKey="Delgado Gt Ruiz R">RA Delgado >Ruíz</name>
</author>
<author>
<name sortKey="Ramirez Fernandez, Mp" uniqKey="Ramirez Fernandez M">MP Ramírez Fernández</name>
</author>
<author>
<name sortKey="Barona Dorado, C" uniqKey="Barona Dorado C">C Barona Dorado</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Yun, Hj" uniqKey="Yun H">HJ Yun</name>
</author>
<author>
<name sortKey="Park, Jc" uniqKey="Park J">JC Park</name>
</author>
<author>
<name sortKey="Yun, Jh" uniqKey="Yun J">JH Yun</name>
</author>
<author>
<name sortKey="Jung, Uw" uniqKey="Jung U">UW Jung</name>
</author>
<author>
<name sortKey="Kim, Cs" uniqKey="Kim C">CS Kim</name>
</author>
<author>
<name sortKey="Choi, Sh" uniqKey="Choi S">SH Choi</name>
</author>
<author>
<name sortKey="Cho, Ks" uniqKey="Cho K">KS Cho</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Song, Dw" uniqKey="Song D">DW Song</name>
</author>
<author>
<name sortKey="Lee, Dw" uniqKey="Lee D">DW Lee</name>
</author>
<author>
<name sortKey="Kim, Ck" uniqKey="Kim C">CK Kim</name>
</author>
<author>
<name sortKey="Park, Kh" uniqKey="Park K">KH Park</name>
</author>
<author>
<name sortKey="Moon, Is" uniqKey="Moon I">IS Moon</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Lee, Dw" uniqKey="Lee D">DW Lee</name>
</author>
<author>
<name sortKey="Choi, Ys" uniqKey="Choi Y">YS Choi</name>
</author>
<author>
<name sortKey="Park, Kh" uniqKey="Park K">KH Park</name>
</author>
<author>
<name sortKey="Kim, Cs" uniqKey="Kim C">CS Kim</name>
</author>
<author>
<name sortKey="Moon, Is" uniqKey="Moon I">IS Moon</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Calvo Guirado, Jl" uniqKey="Calvo Guirado J">JL Calvo-Guirado</name>
</author>
<author>
<name sortKey="G Mez Moreno, G" uniqKey="G Mez Moreno G">G Gómez-Moreno</name>
</author>
<author>
<name sortKey="Aguilar Salvatierra, A" uniqKey="Aguilar Salvatierra A">A Aguilar-Salvatierra</name>
</author>
<author>
<name sortKey="Guardia, J" uniqKey="Guardia J">J Guardia</name>
</author>
<author>
<name sortKey="Delgado Ruiz, Ra" uniqKey="Delgado Ruiz R">RA Delgado-Ruiz</name>
</author>
<author>
<name sortKey="Romanos, Ge" uniqKey="Romanos G">GE Romanos</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Amid, R" uniqKey="Amid R">R Amid</name>
</author>
<author>
<name sortKey="Raoofi, S" uniqKey="Raoofi S">S Raoofi</name>
</author>
<author>
<name sortKey="Kadkhodazadeh, M" uniqKey="Kadkhodazadeh M">M Kadkhodazadeh</name>
</author>
<author>
<name sortKey="Movahhedi, Mr" uniqKey="Movahhedi M">MR Movahhedi</name>
</author>
<author>
<name sortKey="Khademi, M" uniqKey="Khademi M">M Khademi</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Schrotenboer, J" uniqKey="Schrotenboer J">J Schrotenboer</name>
</author>
<author>
<name sortKey="Tsao, Yp" uniqKey="Tsao Y">YP Tsao</name>
</author>
<author>
<name sortKey="Kinariwala, V" uniqKey="Kinariwala V">V Kinariwala</name>
</author>
<author>
<name sortKey="Wang, Hl" uniqKey="Wang H">HL Wang</name>
</author>
</analytic>
</biblStruct>
</listBibl>
</div1>
</back>
</TEI>
<pmc article-type="research-article">
<pmc-dir>properties open_access</pmc-dir>
<front>
<journal-meta>
<journal-id journal-id-type="nlm-ta">J Adv Prosthodont</journal-id>
<journal-id journal-id-type="iso-abbrev">J Adv Prosthodont</journal-id>
<journal-id journal-id-type="publisher-id">JAP</journal-id>
<journal-title-group>
<journal-title>The Journal of Advanced Prosthodontics</journal-title>
</journal-title-group>
<issn pub-type="ppub">2005-7806</issn>
<issn pub-type="epub">2005-7814</issn>
<publisher>
<publisher-name>The Korean Academy of Prosthodontics</publisher-name>
</publisher>
</journal-meta>
<article-meta>
<article-id pub-id-type="pmid">25551016</article-id>
<article-id pub-id-type="pmc">4279055</article-id>
<article-id pub-id-type="doi">10.4047/jap.2014.6.6.547</article-id>
<article-categories>
<subj-group subj-group-type="heading">
<subject>Original Article</subject>
</subj-group>
</article-categories>
<title-group>
<article-title>The influence of thread geometry on implant osseointegration under immediate loading: a literature review</article-title>
</title-group>
<contrib-group>
<contrib contrib-type="author">
<name>
<surname>Ryu</surname>
<given-names>Hyo-Sook</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Namgung</surname>
<given-names>Cheol</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Lee</surname>
<given-names>Jong-Ho</given-names>
</name>
<xref ref-type="aff" rid="A2">2</xref>
</contrib>
<contrib contrib-type="author" corresp="yes">
<name>
<surname>Lim</surname>
<given-names>Young-Jun</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
</contrib>
</contrib-group>
<aff id="A1">
<label>1</label>
Department of Prosthodontics and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Republic of Korea.</aff>
<aff id="A2">
<label>2</label>
Department of Oral Maxillofacial Surgery and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Republic of Korea.</aff>
<author-notes>
<corresp>Corresponding author: Young-Jun Lim. Department of Prosthodontics and Dental Research Institute, School of Dentistry, Seoul National University 101 Daehak-ro, Jongno-gu, Seoul 110-749, Republic of Korea. Tel. 82 2 2072 2940:
<email>limdds@snu.ac.kr</email>
</corresp>
</author-notes>
<pub-date pub-type="ppub">
<month>12</month>
<year>2014</year>
</pub-date>
<pub-date pub-type="epub">
<day>17</day>
<month>12</month>
<year>2014</year>
</pub-date>
<volume>6</volume>
<issue>6</issue>
<fpage>547</fpage>
<lpage>554</lpage>
<history>
<date date-type="received">
<day>10</day>
<month>6</month>
<year>2014</year>
</date>
<date date-type="rev-recd">
<day>18</day>
<month>8</month>
<year>2014</year>
</date>
<date date-type="accepted">
<day>06</day>
<month>10</month>
<year>2014</year>
</date>
</history>
<permissions>
<copyright-statement>© 2014 The Korean Academy of Prosthodontics</copyright-statement>
<copyright-year>2014</copyright-year>
<license license-type="open-access" xlink:href="http://creativecommons.org/licenses/by-nc/3.0/">
<license-p>This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (
<ext-link ext-link-type="uri" xlink:href="http://creativecommons.org/licenses/by-nc/3.0/">http://creativecommons.org/licenses/by-nc/3.0/</ext-link>
) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.</license-p>
</license>
</permissions>
<abstract>
<p>Implant success is achieved by the synergistic combination of numerous biomechanical factors. This report examines the mechanical aspect of implants. In particular, it is focused on macrodesign such as thread shape, pitch, width and depth, and crestal module of implants. This study reviews the literature regarding the effect of implant thread geometry on primary stability and osseointegration under immediate loading. The search strategy included both
<italic>in vitro</italic>
and
<italic>in vivo</italic>
studies published in the MEDLINE database from January 2000 to June 2014. Various geometrical parameters are analyzed to evaluate their significance for optimal stress distribution, implant surface area, and bone remodeling responses during the process of osseointegration.</p>
</abstract>
<kwd-group>
<kwd>Dental implants</kwd>
<kwd>Implant macrodesign</kwd>
<kwd>Thread geometry</kwd>
<kwd>Primary stability</kwd>
<kwd>Immediate loading</kwd>
</kwd-group>
<funding-group>
<award-group>
<funding-source country="KR">Ministry of Health and Welfare</funding-source>
<award-id>HI12C0064</award-id>
</award-group>
</funding-group>
</article-meta>
</front>
<body>
<sec sec-type="intro">
<title>INTRODUCTION</title>
<p>Dental implants have gained popularity as a successful treatment modality for restoring edentulism. Many longterm implant studies have revealed their survival rates exceeding 95%.
<xref rid="B1" ref-type="bibr">1</xref>
,
<xref rid="B2" ref-type="bibr">2</xref>
,
<xref rid="B3" ref-type="bibr">3</xref>
However, the potential for clinical failures and limiting factors is still present as a great concern for clinicians. Traditionally, a healing period of 3 to 6 months has been recommended for osseointegration after implant placement. Despite high predictability of this traditional protocol, a long, unloaded period is a major drawback especially for patients with strong esthetic demand. Even if patients have decent provisional prostheses fabricated, they must undergo a series of follow-up appointments for adjustments and maintenance. Therefore, considerable scientific interest has been focused on reducing the healing period and loading of the implant.
<xref rid="B4" ref-type="bibr">4</xref>
</p>
<p>The definitions of immediate loading vary based on its indications and protocols.
<xref rid="B5" ref-type="bibr">5</xref>
,
<xref rid="B6" ref-type="bibr">6</xref>
Immediate loading frequently refers to the placement of a prosthesis in occlusion within 48 hours after implant surgery.
<xref rid="B7" ref-type="bibr">7</xref>
However, prosthesis can often be fabricated immediately after the surgery to restore a single tooth or partially edentulous dentition. This protocol may show a significant advantage to both clinician and patient. Previous studies on immediate loading have reported survival rates comparable to conventional loading.
<xref rid="B8" ref-type="bibr">8</xref>
,
<xref rid="B9" ref-type="bibr">9</xref>
,
<xref rid="B10" ref-type="bibr">10</xref>
However, the extent of loading and primary stability required for successful osseointegration under immediate loading is still open to debate and it is necessary to set standardized protocol guidelines.</p>
<p>The presence of osseointegration is critical to evaluate implant success. The effective bond between an implant and its surrounding bone is created by various mechanical factors. One of the key factors is implant design since it determines primary stability and stress distribution during osseointegration. The geometric features of an implant influence sufficient initial contact to facilitate primary stability of the implant.
<xref rid="B11" ref-type="bibr">11</xref>
It also plays an important role on implant capacity to withstand forces during the process of osseointegration. Therefore, the optimal implant design itself can improve the potential osseointegration process and the primary and secondary stability of the implant.</p>
<p>There are two types of the implant design; macrodesign and microdesign. The former includes thread geometry and body shape while the latter consists of implant material and surface treatment and morphology.
<xref rid="B12" ref-type="bibr">12</xref>
Dental implants on the market today are available in many different thread patterns. In particular, this literature review was focused on thread geometry such as shape, pitch, width and depth, and crestal module as described in
<xref ref-type="fig" rid="F1">Fig. 1</xref>
. The purpose of this report sought to evaluate the effect of different implant thread designs on load distribution and primary stability at the implant-bone interface in an immediate loading situation.</p>
</sec>
<sec sec-type="materials|methods">
<title>MATERIALS AND METHODS</title>
<p>This study reviews the literature dealing with the influence of implant thread geometry on primary stability of osseointegrated implants under immediate loading. A MEDLINE search was conducted using the PubMed search engine. The keywords used included; implant thread, implant macrodesign, immediate loading, thread geometry, thread shape, thread pitch, thread depth, thread width, microthreads, crest module and combinations of these keywords. The articles published from January 2000 to June 2014 were accepted. Both
<italic>in vitro</italic>
and
<italic>in vivo</italic>
studies were included in this review.</p>
</sec>
<sec>
<title>THREAD SHAPE AND STRESS DISTRIBUTION</title>
<p>Most dental implants can be found in various thread shapes developed for effective inserting and force transmission. Threaded implants are inserted into the osteotomy site by creating linear motion through rotation. Thread shapes available for screw-retained implants include square shape, V-shape, buttress and reverse buttress threads, which are defined by the thread thickness and face angle.
<xref rid="B13" ref-type="bibr">13</xref>
</p>
<p>Once an implant is inserted, bone undergoes constant remodeling against external stress, called bone homeostasis. When an implant receives optimal functional load, the surrounding bone experiences remodeling and produces woven bone. However, under extreme adverse stresses, microfractures occur in the alveolar bone inducing "osteoclastogenesis".
<xref rid="B14" ref-type="bibr">14</xref>
Since bone formation is not fast enough to fill in the damage, the defect becomes worse, resulting in severe bone loss and ultimately implant failure.
<xref rid="B15" ref-type="bibr">15</xref>
,
<xref rid="B16" ref-type="bibr">16</xref>
However, the optimal stress distribution is difficult to achieve, and too little or much stress can induce bone resorption.
<xref rid="B14" ref-type="bibr">14</xref>
,
<xref rid="B17" ref-type="bibr">17</xref>
Therefore, implant threads should be fabricated to increase surface contact area and favorable forces while reducing adverse stimuli.</p>
<p>Recently, the finite element analysis (FEA) has been utilized to understand the effect of those geometric parameters on the load distribution at the surrounding regions. Using FEA, Chang and his collegues evaluated the pattern of micromotion within implants and surrounding bone with different thread designs (trapezoidal, buttress, square, and standard V-thread) under immediate loading of 300 N axial load.
<xref rid="B18" ref-type="bibr">18</xref>
The results revealed that all micromotion was located near the interface of cortical and cancellous bone and the square thread profile had the most favorable micromotion value. Eraslan et al.
<xref rid="B19" ref-type="bibr">19</xref>
performed the similar study with four different thread forms under a static axial load of 100 N. The study reported that maximum stress was concentrated at the cervical cortical regions around the first thread and the stress value was lowest in the square thread type. Likewise, other previous FEA and animal studies showed the most effective stress distribution and bone-to-implant contact area (BIC) in the square thread shape.
<xref rid="B20" ref-type="bibr">20</xref>
,
<xref rid="B21" ref-type="bibr">21</xref>
,
<xref rid="B22" ref-type="bibr">22</xref>
Very few randomized clinical trials have evaluated the effect of thread design on osseointegration under immediate loading. However, McAllister et al.
<xref rid="B23" ref-type="bibr">23</xref>
and Arnhart et al.
<xref rid="B24" ref-type="bibr">24</xref>
recently performed multi-center clinical trials using variable-thread tapered implants (NobelActive, Nobel Biocare, Gothenburg, Sweden) respectively. The data from both studies demonstrated that the variable-thread tapered implants exhibited successful clinical results under demanding conditions of immediate loading. The summary of these studies is shown in
<xref ref-type="table" rid="T1">Table 1</xref>
.</p>
<p>The FEA and
<italic>in vivo</italic>
studies discussed above indicate that stress distribution varies with the thread shape. However, regardless of different thread designs, the maximum stress was concentrated near the cervical bone region around the first thread of the implant. In particular, the data from the FEA studies demonstrated that the square thread type had more favorable stress distribution. Using the variable-thread tapered implants, two multi-center clinical trials also showed clinical results comparable to the standard design. Although clinical evidence is limited, it is possible to conclude that the thread shape plays an important role in stress transfer between the implant and the surrounding tissue and determines the primary stability of the implant.</p>
</sec>
<sec>
<title>THREAD PITCH</title>
<p>Thread design should maximize implant surface area and create a better spreading of stress and primary stability.
<xref rid="B25" ref-type="bibr">25</xref>
Like thread shape, pitch is another important geometric factor that determines the bone-to-implant contact and the biomechanical load distribution. Thread pitch is defined as the distance between two neighboring threads, measured on the same side of the axis (
<xref ref-type="fig" rid="F1">Fig. 1</xref>
).
<xref rid="B26" ref-type="bibr">26</xref>
It also refers to the number of threads per unit length.
<xref rid="B26" ref-type="bibr">26</xref>
Therefore, when implants have the same length, smaller pitch indicates more threads, leading to greater surface area. Another geometric parameter related to thread pitch is lead. Lead is the distance within the same thread between before and after one complete rotation in the axial direction. That is, for single-, double-, and triple-threaded implants, lead increases by one, two, and three times the pitch. Since lead indicates the distance that an implant would move after one turn, it plays an important role on determining the speed of implant insertion. Hence, thread pitch is clinically significant due to its effect on surface area and insertion speed.
<xref rid="B27" ref-type="bibr">27</xref>
</p>
<p>Multiple studies demonstrated that implants with smaller pitch showed the greater surface area and better stress distribution particularly in low-density bone as summarized in
<xref ref-type="table" rid="T2">Table 2</xref>
. Orsini et al.
<xref rid="B28" ref-type="bibr">28</xref>
conducted an animal study placing "narrow-pitch" (0.5 mm) and "wide-pitch" (1.7 mm) implants in a sheep iliac crest model. The findings suggested that the greater BIC gained by reducing thread pitch could improve initial anchorage and primary stability in cancellous bone. Another animal study performed by Chung et al.
<xref rid="B29" ref-type="bibr">29</xref>
also revealed that implants with a 0.6-mm pitch created more crestal bone resorption than those with a 0.5-mm pitch. In addition, the same results were found in the studies using FEA models. Authors independently reported that smaller pitch presented better load resistance and less effective stress in their three-dimensional FEA models.
<xref rid="B22" ref-type="bibr">22</xref>
,
<xref rid="B30" ref-type="bibr">30</xref>
,
<xref rid="B31" ref-type="bibr">31</xref>
However, the optimal pitch values vary in different thread shapes. In the FEA study using triangular and trapezoidal thread forms with various pitches, Lan and his collegues recommended 1.2 mm and 1.6 mm as the optimal thread pitches of triangular- and trapezoidal-threaded implants respectively.
<xref rid="B32" ref-type="bibr">32</xref>
For V-shape threads, the implants with a 0.8-mm pitch produced the most favorable stress distribution.
<xref rid="B33" ref-type="bibr">33</xref>
</p>
<p>The thread pitch determines the surface area available for load transfer to the peri-implant tissue. The common findings indicate that smaller pitch exhibits more favorable stress distribution and reinforces the primary stability of the implant. However, the optimal level of thread pitch varies in different thread design; for example, a pitch of 0.8 mm had the most effective stress distribution for V-shape threads. Lastly, thread pitch plays a more critical role in enhancing the primary stability in low-density bone than in high-density bone, so clinicians should pay close attention to this geometric factor, especially when the bone quality is poor.</p>
</sec>
<sec>
<title>THREAD DEPTH AND WIDTH</title>
<p>In addition to thread shape and pitch, its depth and width are important design parameters that affect the stress distribution around endosteal implants. According to Misch, thread depth is the distance from the outermost tip to the innermost body of the thread (
<xref ref-type="fig" rid="F1">Fig. 1</xref>
).
<xref rid="B34" ref-type="bibr">34</xref>
The same author defines thread width as the distance between the superiormost and inferior-most tip of a single thread measured axially (
<xref ref-type="fig" rid="F1">Fig. 1</xref>
).
<xref rid="B34" ref-type="bibr">34</xref>
In other words, thread depth can also be calculated by difference between the major and minor diameter of the thread.
<xref rid="B34" ref-type="bibr">34</xref>
</p>
<p>Like the previously discussed geometric variables, thread depth and width clinically influence implant insertion and surface area. The shallower the thread depth, the easier the implantation procedure, especially in the highdensity bone.
<xref rid="B34" ref-type="bibr">34</xref>
It may be able to eliminate the need for tapping during the surgery. On the contrary, deep threads increase the functional surface area at the bone-implant interface, which can improve primary stability in the low-density bone or the region with high occlusal load. Therefore, various implant systems are available in the market using progressive threads; for example, Ankylos (Dentsply Friadent, Mannheim, Germany).
<xref rid="B12" ref-type="bibr">12</xref>
In this thread form, thread depth gradually decreases from the apical end to the coronal neck of the implant. It is allegedly claimed that it may transfer the stress away from the crestal cortical region, preventing possible bone resorption.
<xref rid="B12" ref-type="bibr">12</xref>
</p>
<p>Despite the advantageous effect of thread depth and width, there have been few studies designed to examine these geometric parameters. Ao's study is one of the few which evaluated the maximum von Mises stresses in the implants with various thread depths and widths under immediate loading.
<xref rid="B35" ref-type="bibr">35</xref>
A three-dimensional FEA model was created for immediate loading applied at an axial load of 100 N and a 45-degree buccolingual load of 30 N. The thread depths and widths used in this study respectively ranged from 0.2-0.6 mm and 0.1-0.4 mm. The results revealed that thread depth affected the stress distribution more significantly than thread width. Threads with depth of more than 0.44 mm and width of 0.19-0.23 mm showed the most favorable biomechanical behavior for the immediately loaded cylinder implants designed in this study. A similar study was performed by Kong et al.
<xref rid="B36" ref-type="bibr">36</xref>
Instead of a 45-degree buccolingual load of 30 N, 50 N was used and the rest of the FEA modeling conditions were the same as Ao's. The results were consistent with Ao's data. The optimal values of thread depth and width were 0.34-0.5 mm and 0.18-0.3 mm respectively and the thread depth was also a more sensitive factor to reduce the peak stress concentration within the bone. A more detailed description of Ao's and Kong's studies is available in
<xref ref-type="table" rid="T3">Table 3</xref>
.</p>
<p>As a result of a lack of clinical evidence, it is difficult to establish consensus guidelines for the optimal level of thread depth and width. However, similar values were obtained by both Ao and Kong's studies using the immediately loaded cylinder implants. Another common finding was that thread depth had a more significant impact on effective stress distribution than thread width. Additional
<italic>in vivo</italic>
studies and clinical trials are necessary to confirm the observations made in these studies.</p>
</sec>
<sec>
<title>CRESTAL MODULE</title>
<p>Crest module refers to the neck portion of the implant. Implant neck configurations can be critical for minimizing the marginal bone loss. Previous studies have reported a strong association between crestal module and marginal bone resorption.
<xref rid="B37" ref-type="bibr">37</xref>
,
<xref rid="B38" ref-type="bibr">38</xref>
This area is important for implant primary stability because the transition from an endosteal environment to an oral cavity occurs here. In addition, this is the region where the thick cortical bone is present and the occlusal stress is concentrated.
<xref rid="B27" ref-type="bibr">27</xref>
</p>
<p>Previously, smooth crest module was common to prevent plaque accumulation because it was exposed above the crestal plate. However, Hermann et al.
<xref rid="B39" ref-type="bibr">39</xref>
and Hanggi et al.
<xref rid="B40" ref-type="bibr">40</xref>
reported that shear forces increased at the crestal region when a smooth neck was inserted below the bone crest, leading to marginal bone resorption and pocket formation. Furthermore, some authors raised the concept of "disuse atrophy".
<xref rid="B41" ref-type="bibr">41</xref>
Marginal bone resorption is caused by a smooth neck system due to its lack of exertion of the mechanical stress into the marginal bone. Therefore, Hansson introduced microthreads as the retentive elements at the neck portion to prevent the bone loss (
<xref ref-type="fig" rid="F1">Fig. 1</xref>
).
<xref rid="B42" ref-type="bibr">42</xref>
</p>
<p>Studies reporting association between microthreads and marginal bone loss are listed in
<xref ref-type="table" rid="T4">Table 4</xref>
. Chowdhary et al.
<xref rid="B43" ref-type="bibr">43</xref>
conducted an FEA and
<italic>in vivo</italic>
study to analyze the influence of microthreads on osseointegration and primary stability of implants. They placed oxidized implants with and without microthreads in the femur and tibia of rabbits. The results showed that the test group had more favorable stress concentration, bone volume and BIC than the control group. A recent animal study also compared polished and microthreaded collars and the microthreaded group exhibited less crestal bone resorption.
<xref rid="B44" ref-type="bibr">44</xref>
</p>
<p>In addition to the aforementioned animal study, previous clinical studies have already demonstrated reduced marginal bone loss for the microthreaded implants under conventional loading.
<xref rid="B45" ref-type="bibr">45</xref>
,
<xref rid="B46" ref-type="bibr">46</xref>
,
<xref rid="B47" ref-type="bibr">47</xref>
Although few clinical data are available on the immediate loading in the microthreaded implants, Calvo-Guirado et al.
<xref rid="B48" ref-type="bibr">48</xref>
recently performed a prospective study for 3 years after the implant placement under immediate loading. The implant system used in this study had microthreads up to the prosthetic platform at the neck region. The results showed 100% survival rate and minimal marginal bone loss.</p>
<p>Several FEA models were designed to evaluate the effect of microthread design on load transfer in the surrounding bone. The results of the two independent analyses by Amid et al.
<xref rid="B49" ref-type="bibr">49</xref>
and Schrotenboer et al.
<xref rid="B50" ref-type="bibr">50</xref>
indicated that addition of a microthread design increased maximum stress value at the crestal bone. This supports Wolff's law as a possible theoretical foundation for the crestal bone maintenance but still remains open to debate.</p>
<p>All of the studies reviewed indicate that microthreads promote bone formation and contribute to the effective stress distribution and less crestal bone resorption. Like thread pitch, microthreads plays a more critical role in preventing marginal bone loss in the cancellous bone than in the cortical bone. Therefore, the implant crestal module is an important macrodesign parameter that should be considered especially when dealing with low-density bone.</p>
</sec>
<sec sec-type="conclusions">
<title>CONCLUSION</title>
<p>In conclusion, the biomechanical behavior of implant macrodesign factors significantly influences stress distribution at the cervical crestal bone region and ultimately primary stability under immediate loading in cancellous bone. Based on the results presented in this literature review, it was possible to draw several useful clinical conclusions about various implant macrodesign features. First, for thread shape, the square thread profile may provide the best primary stability in an immediate loading situation. When primary stability is a concern, implants with smaller pitch are beneficial by increasing BIC. Studies have shown that thread depth is more critical for dissipating peak stresses within the bone than thread width and the optimal values of thread depth and width may vary depending on thread shape and other geometric factors. Lastly, the microthread configuration at the implant neck may improve bone formation and stress distribution for the implants inserted in the cancellous bone under immediate loading.</p>
<p>However, information about the effect of different macrodesign characteristics should be carefully interpreted. Optimal primary stability and stress concentration are the combined results of various mechanical and biological factors. For a specific implant, the benefits from a single design feature could be enhanced or weakened by the other variables of the implant. Hence, it is important to evaluate patient's biological condition and consider various mechanical features as a whole for a particular clinical situation. Clinicians should understand that a certain design factor alone will not guarantee implant success and survival.</p>
<p>Few
<italic>in vivo</italic>
clinical studies have been conducted to investigate the influence of implant macrodesign features on primary stability of immediately loaded implants. The results of FEA and animal models may not show the same outcome in a clinical setting because they cannot completely duplicate the human clinical situation. Each individual patient has a unique biological condition such as bone quality, force distribution, masticatory activity, etc. However, these studies are relatively a simple, fast, and inexpensive approach to predict the biomechanical behavior of implants in an oral cavity compared to human studies, so they can still be an initial guide for further clinical research.</p>
</sec>
</body>
<back>
<fn-group>
<fn fn-type="supported-by">
<p>This research was supported by a grant of the Korea Health Technology R&D Project through the Korea Health Industry Development Institute (KHIDI), funded by the Ministry of Health & Welfare, Republic of Korea (grant number : HI12C0064).</p>
</fn>
</fn-group>
<ref-list>
<ref id="B1">
<label>1</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Haas</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Mensdorff-Pouilly</surname>
<given-names>N</given-names>
</name>
<name>
<surname>Mailath</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Watzek</surname>
<given-names>G</given-names>
</name>
</person-group>
<article-title>Brånemark single tooth implants: a preliminary report of 76 implants</article-title>
<source>J Prosthet Dent</source>
<year>1995</year>
<volume>73</volume>
<fpage>274</fpage>
<lpage>279</lpage>
<pub-id pub-id-type="pmid">7760277</pub-id>
</element-citation>
</ref>
<ref id="B2">
<label>2</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Goodacre</surname>
<given-names>CJ</given-names>
</name>
<name>
<surname>Bernal</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Rungcharassaeng</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Kan</surname>
<given-names>JY</given-names>
</name>
</person-group>
<article-title>Clinical complications with implants and implant prostheses</article-title>
<source>J Prosthet Dent</source>
<year>2003</year>
<volume>90</volume>
<fpage>121</fpage>
<lpage>132</lpage>
<pub-id pub-id-type="pmid">12886205</pub-id>
</element-citation>
</ref>
<ref id="B3">
<label>3</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Fugazzotto</surname>
<given-names>PA</given-names>
</name>
</person-group>
<article-title>Success and failure rates of osseointegrated implants in function in regenerated bone for 72 to 133 months</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2005</year>
<volume>20</volume>
<fpage>77</fpage>
<lpage>83</lpage>
<pub-id pub-id-type="pmid">15747677</pub-id>
</element-citation>
</ref>
<ref id="B4">
<label>4</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Atieh</surname>
<given-names>MA</given-names>
</name>
<name>
<surname>Atieh</surname>
<given-names>AH</given-names>
</name>
<name>
<surname>Payne</surname>
<given-names>AG</given-names>
</name>
<name>
<surname>Duncan</surname>
<given-names>WJ</given-names>
</name>
</person-group>
<article-title>Immediate loading with single implant crowns: a systematic review and meta-analysis</article-title>
<source>Int J Prosthodont</source>
<year>2009</year>
<volume>22</volume>
<fpage>378</fpage>
<lpage>387</lpage>
<pub-id pub-id-type="pmid">19639076</pub-id>
</element-citation>
</ref>
<ref id="B5">
<label>5</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Cochran</surname>
<given-names>DL</given-names>
</name>
<name>
<surname>Morton</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Weber</surname>
<given-names>HP</given-names>
</name>
</person-group>
<article-title>Consensus statements and recommended clinical procedures regarding loading protocols for endosseous dental implants</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2004</year>
<volume>19</volume>
<fpage>109</fpage>
<lpage>113</lpage>
<pub-id pub-id-type="pmid">15635951</pub-id>
</element-citation>
</ref>
<ref id="B6">
<label>6</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Glauser</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Zembic</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Hämmerle</surname>
<given-names>CH</given-names>
</name>
</person-group>
<article-title>A systematic review of marginal soft tissue at implants subjected to immediate loading or immediate restoration</article-title>
<source>Clin Oral Implants Res</source>
<year>2006</year>
<volume>17</volume>
<fpage>82</fpage>
<lpage>92</lpage>
<pub-id pub-id-type="pmid">16968384</pub-id>
</element-citation>
</ref>
<ref id="B7">
<label>7</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Romanos</surname>
<given-names>GE</given-names>
</name>
<name>
<surname>Toh</surname>
<given-names>CG</given-names>
</name>
<name>
<surname>Siar</surname>
<given-names>CH</given-names>
</name>
<name>
<surname>Swaminathan</surname>
<given-names>D</given-names>
</name>
</person-group>
<article-title>Histologic and histomorphometric evaluation of peri-implant bone subjected to immediate loading: an experimental study with Macaca fascicularis</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2002</year>
<volume>17</volume>
<fpage>44</fpage>
<lpage>51</lpage>
<pub-id pub-id-type="pmid">11858574</pub-id>
</element-citation>
</ref>
<ref id="B8">
<label>8</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Ibañez</surname>
<given-names>JC</given-names>
</name>
<name>
<surname>Jalbout</surname>
<given-names>ZN</given-names>
</name>
</person-group>
<article-title>Immediate loading of osseotite implants: two-year results</article-title>
<source>Implant Dent</source>
<year>2002</year>
<volume>11</volume>
<fpage>128</fpage>
<lpage>136</lpage>
<pub-id pub-id-type="pmid">12078594</pub-id>
</element-citation>
</ref>
<ref id="B9">
<label>9</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Rocci</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Rocci</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Rocci</surname>
<given-names>C</given-names>
</name>
<name>
<surname>Scoccia</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Gargari</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Martignoni</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Gottlow</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Sennerby</surname>
<given-names>L</given-names>
</name>
</person-group>
<article-title>Immediate loading of Brånemark system TiUnite and machined-surface implants in the posterior mandible, part II: a randomized open-ended 9-year follow-up clinical trial</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2013</year>
<volume>28</volume>
<fpage>891</fpage>
<lpage>895</lpage>
<pub-id pub-id-type="pmid">23748324</pub-id>
</element-citation>
</ref>
<ref id="B10">
<label>10</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Ostman</surname>
<given-names>PO</given-names>
</name>
<name>
<surname>Hellman</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Sennerby</surname>
<given-names>L</given-names>
</name>
</person-group>
<article-title>Direct implant loading in the edentulous maxilla using a bone density-adapted surgical protocol and primary implant stability criteria for inclusion</article-title>
<source>Clin Implant Dent Relat Res</source>
<year>2005</year>
<volume>7</volume>
<fpage>S60</fpage>
<lpage>S69</lpage>
<pub-id pub-id-type="pmid">16137089</pub-id>
</element-citation>
</ref>
<ref id="B11">
<label>11</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Sennerby</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Meredith</surname>
<given-names>N</given-names>
</name>
</person-group>
<article-title>Implant stability measurements using resonance frequency analysis: biological and biomechanical aspects and clinical implications</article-title>
<source>Periodontol 2000</source>
<year>2008</year>
<volume>47</volume>
<fpage>51</fpage>
<lpage>66</lpage>
<pub-id pub-id-type="pmid">18412573</pub-id>
</element-citation>
</ref>
<ref id="B12">
<label>12</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Abuhussein</surname>
<given-names>H</given-names>
</name>
<name>
<surname>Pagni</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Rebaudi</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Wang</surname>
<given-names>HL</given-names>
</name>
</person-group>
<article-title>The effect of thread pattern upon implant osseointegration</article-title>
<source>Clin Oral Implants Res</source>
<year>2010</year>
<volume>21</volume>
<fpage>129</fpage>
<lpage>136</lpage>
<pub-id pub-id-type="pmid">19709058</pub-id>
</element-citation>
</ref>
<ref id="B13">
<label>13</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Boggan</surname>
<given-names>RS</given-names>
</name>
<name>
<surname>Strong</surname>
<given-names>JT</given-names>
</name>
<name>
<surname>Misch</surname>
<given-names>CE</given-names>
</name>
<name>
<surname>Bidez</surname>
<given-names>MW</given-names>
</name>
</person-group>
<article-title>Influence of hex geometry and prosthetic table width on static and fatigue strength of dental implants</article-title>
<source>J Prosthet Dent</source>
<year>1999</year>
<volume>82</volume>
<fpage>436</fpage>
<lpage>440</lpage>
<pub-id pub-id-type="pmid">10512962</pub-id>
</element-citation>
</ref>
<ref id="B14">
<label>14</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Hansson</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Werke</surname>
<given-names>M</given-names>
</name>
</person-group>
<article-title>The implant thread as a retention element in cortical bone: the effect of thread size and thread profile: a finite element study</article-title>
<source>J Biomech</source>
<year>2003</year>
<volume>36</volume>
<fpage>1247</fpage>
<lpage>1258</lpage>
<pub-id pub-id-type="pmid">12893033</pub-id>
</element-citation>
</ref>
<ref id="B15">
<label>15</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Prendergast</surname>
<given-names>PJ</given-names>
</name>
<name>
<surname>Huiskes</surname>
<given-names>R</given-names>
</name>
</person-group>
<article-title>Microdamage and osteocyte-lacuna strain in bone: a microstructural finite element analysis</article-title>
<source>J Biomech Eng</source>
<year>1996</year>
<volume>118</volume>
<fpage>240</fpage>
<lpage>246</lpage>
<pub-id pub-id-type="pmid">8738790</pub-id>
</element-citation>
</ref>
<ref id="B16">
<label>16</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Brunski</surname>
<given-names>JB</given-names>
</name>
</person-group>
<article-title>In vivo bone response to biomechanical loading at the bone/dental-implant interface</article-title>
<source>Adv Dent Res</source>
<year>1999</year>
<volume>13</volume>
<fpage>99</fpage>
<lpage>119</lpage>
<pub-id pub-id-type="pmid">11276755</pub-id>
</element-citation>
</ref>
<ref id="B17">
<label>17</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Frost</surname>
<given-names>HM</given-names>
</name>
</person-group>
<article-title>Skeletal structural adaptations to mechanical usage (SATMU): 1. Redefining Wolff's law: the bone modeling problem</article-title>
<source>Anat Rec</source>
<year>1990</year>
<volume>226</volume>
<fpage>403</fpage>
<lpage>413</lpage>
<pub-id pub-id-type="pmid">2184695</pub-id>
</element-citation>
</ref>
<ref id="B18">
<label>18</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Chang</surname>
<given-names>PK</given-names>
</name>
<name>
<surname>Chen</surname>
<given-names>YC</given-names>
</name>
<name>
<surname>Huang</surname>
<given-names>CC</given-names>
</name>
<name>
<surname>Lu</surname>
<given-names>WH</given-names>
</name>
<name>
<surname>Chen</surname>
<given-names>YC</given-names>
</name>
<name>
<surname>Tsai</surname>
<given-names>HH</given-names>
</name>
</person-group>
<article-title>Distribution of micromotion in implants and alveolar bone with different thread profiles in immediate loading: a finite element study</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2012</year>
<volume>27</volume>
<fpage>e96</fpage>
<lpage>e101</lpage>
<pub-id pub-id-type="pmid">23189316</pub-id>
</element-citation>
</ref>
<ref id="B19">
<label>19</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Eraslan</surname>
<given-names>O</given-names>
</name>
<name>
<surname>Inan</surname>
<given-names>O</given-names>
</name>
</person-group>
<article-title>The effect of thread design on stress distribution in a solid screw implant: a 3D finite element analysis</article-title>
<source>Clin Oral Investig</source>
<year>2010</year>
<volume>14</volume>
<fpage>411</fpage>
<lpage>416</lpage>
</element-citation>
</ref>
<ref id="B20">
<label>20</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Steigenga</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Al-Shammari</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Misch</surname>
<given-names>C</given-names>
</name>
<name>
<surname>Nociti</surname>
<given-names>FH</given-names>
<suffix>Jr</suffix>
</name>
<name>
<surname>Wang</surname>
<given-names>HL</given-names>
</name>
</person-group>
<article-title>Effects of implant thread geometry on percentage of osseointegration and resistance to reverse torque in the tibia of rabbits</article-title>
<source>J Periodontol</source>
<year>2004</year>
<volume>75</volume>
<fpage>1233</fpage>
<lpage>1241</lpage>
<pub-id pub-id-type="pmid">15515339</pub-id>
</element-citation>
</ref>
<ref id="B21">
<label>21</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Geng</surname>
<given-names>JP</given-names>
</name>
<name>
<surname>Ma</surname>
<given-names>QS</given-names>
</name>
<name>
<surname>Xu</surname>
<given-names>W</given-names>
</name>
<name>
<surname>Tan</surname>
<given-names>KB</given-names>
</name>
<name>
<surname>Liu</surname>
<given-names>GR</given-names>
</name>
</person-group>
<article-title>Finite element analysis of four thread-form configurations in a stepped screw implant</article-title>
<source>J Oral Rehabil</source>
<year>2004</year>
<volume>31</volume>
<fpage>233</fpage>
<lpage>239</lpage>
<pub-id pub-id-type="pmid">15025655</pub-id>
</element-citation>
</ref>
<ref id="B22">
<label>22</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Chun</surname>
<given-names>HJ</given-names>
</name>
<name>
<surname>Cheong</surname>
<given-names>SY</given-names>
</name>
<name>
<surname>Han</surname>
<given-names>JH</given-names>
</name>
<name>
<surname>Heo</surname>
<given-names>SJ</given-names>
</name>
<name>
<surname>Chung</surname>
<given-names>JP</given-names>
</name>
<name>
<surname>Rhyu</surname>
<given-names>IC</given-names>
</name>
<name>
<surname>Choi</surname>
<given-names>YC</given-names>
</name>
<name>
<surname>Baik</surname>
<given-names>HK</given-names>
</name>
<name>
<surname>Ku</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Kim</surname>
<given-names>MH</given-names>
</name>
</person-group>
<article-title>Evaluation of design parameters of osseointegrated dental implants using finite element analysis</article-title>
<source>J Oral Rehabil</source>
<year>2002</year>
<volume>29</volume>
<fpage>565</fpage>
<lpage>574</lpage>
<pub-id pub-id-type="pmid">12071926</pub-id>
</element-citation>
</ref>
<ref id="B23">
<label>23</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>McAllister</surname>
<given-names>BS</given-names>
</name>
<name>
<surname>Cherry</surname>
<given-names>JE</given-names>
</name>
<name>
<surname>Kolinski</surname>
<given-names>ML</given-names>
</name>
<name>
<surname>Parrish</surname>
<given-names>KD</given-names>
</name>
<name>
<surname>Pumphrey</surname>
<given-names>DW</given-names>
</name>
<name>
<surname>Schroering</surname>
<given-names>RL</given-names>
</name>
</person-group>
<article-title>Two-year evaluation of a variable-thread tapered implant in extraction sites with immediate temporization: a multicenter clinical trial</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2012</year>
<volume>27</volume>
<fpage>611</fpage>
<lpage>618</lpage>
<pub-id pub-id-type="pmid">22616055</pub-id>
</element-citation>
</ref>
<ref id="B24">
<label>24</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Arnhart</surname>
<given-names>C</given-names>
</name>
<name>
<surname>Kielbassa</surname>
<given-names>AM</given-names>
</name>
<name>
<surname>Martinez-de Fuentes</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Goldstein</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Jackowski</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Lorenzoni</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Maiorana</surname>
<given-names>C</given-names>
</name>
<name>
<surname>Mericske-Stern</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Pozzi</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Rompen</surname>
<given-names>E</given-names>
</name>
<name>
<surname>Sanz</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Strub</surname>
<given-names>JR</given-names>
</name>
</person-group>
<article-title>Comparison of variable-thread tapered implant designs to a standard tapered implant design after immediate loading. A 3-year multicentre randomised controlled trial</article-title>
<source>Eur J Oral Implantol</source>
<year>2012</year>
<volume>5</volume>
<fpage>123</fpage>
<lpage>136</lpage>
<pub-id pub-id-type="pmid">22866289</pub-id>
</element-citation>
</ref>
<ref id="B25">
<label>25</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Ivanoff</surname>
<given-names>CJ</given-names>
</name>
<name>
<surname>Gröndahl</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Sennerby</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Bergström</surname>
<given-names>C</given-names>
</name>
<name>
<surname>Lekholm</surname>
<given-names>U</given-names>
</name>
</person-group>
<article-title>Influence of variations in implant diameters: a 3- to 5-year retrospective clinical report</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>1999</year>
<volume>14</volume>
<fpage>173</fpage>
<lpage>180</lpage>
<pub-id pub-id-type="pmid">10212533</pub-id>
</element-citation>
</ref>
<ref id="B26">
<label>26</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Misch</surname>
<given-names>CE</given-names>
</name>
<name>
<surname>Steignga</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Barboza</surname>
<given-names>E</given-names>
</name>
<name>
<surname>Misch-Dietsh</surname>
<given-names>F</given-names>
</name>
<name>
<surname>Cianciola</surname>
<given-names>LJ</given-names>
</name>
<name>
<surname>Kazor</surname>
<given-names>C</given-names>
</name>
</person-group>
<article-title>Short dental implants in posterior partial edentulism: a multicenter retrospective 6-year case series study</article-title>
<source>J Periodontol</source>
<year>2006</year>
<volume>77</volume>
<fpage>1340</fpage>
<lpage>1347</lpage>
<pub-id pub-id-type="pmid">16937587</pub-id>
</element-citation>
</ref>
<ref id="B27">
<label>27</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Steigenga</surname>
<given-names>JT</given-names>
</name>
<name>
<surname>al-Shammari</surname>
<given-names>KF</given-names>
</name>
<name>
<surname>Nociti</surname>
<given-names>FH</given-names>
</name>
<name>
<surname>Misch</surname>
<given-names>CE</given-names>
</name>
<name>
<surname>Wang</surname>
<given-names>HL</given-names>
</name>
</person-group>
<article-title>Dental implant design and its relationship to long-term implant success</article-title>
<source>Implant Dent</source>
<year>2003</year>
<volume>12</volume>
<fpage>306</fpage>
<lpage>317</lpage>
<pub-id pub-id-type="pmid">14752967</pub-id>
</element-citation>
</ref>
<ref id="B28">
<label>28</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Orsini</surname>
<given-names>E</given-names>
</name>
<name>
<surname>Giavaresi</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Trirè</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Ottani</surname>
<given-names>V</given-names>
</name>
<name>
<surname>Salgarello</surname>
<given-names>S</given-names>
</name>
</person-group>
<article-title>Dental implant thread pitch and its influence on the osseointegration process: an in vivo comparison study</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2012</year>
<volume>27</volume>
<fpage>383</fpage>
<lpage>392</lpage>
<pub-id pub-id-type="pmid">22442779</pub-id>
</element-citation>
</ref>
<ref id="B29">
<label>29</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Chung</surname>
<given-names>SH</given-names>
</name>
<name>
<surname>Heo</surname>
<given-names>SJ</given-names>
</name>
<name>
<surname>Koak</surname>
<given-names>JY</given-names>
</name>
<name>
<surname>Kim</surname>
<given-names>SK</given-names>
</name>
<name>
<surname>Lee</surname>
<given-names>JB</given-names>
</name>
<name>
<surname>Han</surname>
<given-names>JS</given-names>
</name>
<name>
<surname>Han</surname>
<given-names>CH</given-names>
</name>
<name>
<surname>Rhyu</surname>
<given-names>IC</given-names>
</name>
<name>
<surname>Lee</surname>
<given-names>SJ</given-names>
</name>
</person-group>
<article-title>Effects of implant geometry and surface treatment on osseointegration after functional loading: a dog study</article-title>
<source>J Oral Rehabil</source>
<year>2008</year>
<volume>35</volume>
<fpage>229</fpage>
<lpage>236</lpage>
<pub-id pub-id-type="pmid">18254802</pub-id>
</element-citation>
</ref>
<ref id="B30">
<label>30</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Ma</surname>
<given-names>P</given-names>
</name>
<name>
<surname>Liu</surname>
<given-names>HC</given-names>
</name>
<name>
<surname>Li</surname>
<given-names>DH</given-names>
</name>
<name>
<surname>Lin</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Shi</surname>
<given-names>Z</given-names>
</name>
<name>
<surname>Peng</surname>
<given-names>QJ</given-names>
</name>
</person-group>
<article-title>Influence of helix angle and density on primary stability of immediately loaded dental implants: three-dimensional finite element analysis</article-title>
<source>Zhonghua Kou Qiang Yi Xue Za Zhi</source>
<year>2007</year>
<volume>42</volume>
<fpage>618</fpage>
<lpage>621</lpage>
<pub-id pub-id-type="pmid">18215373</pub-id>
</element-citation>
</ref>
<ref id="B31">
<label>31</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Motoyoshi</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Yano</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Tsuruoka</surname>
<given-names>T</given-names>
</name>
<name>
<surname>Shimizu</surname>
<given-names>N</given-names>
</name>
</person-group>
<article-title>Biomechanical effect of abutment on stability of orthodontic mini-implant. A finite element analysis</article-title>
<source>Clin Oral Implants Res</source>
<year>2005</year>
<volume>16</volume>
<fpage>480</fpage>
<lpage>485</lpage>
<pub-id pub-id-type="pmid">16117774</pub-id>
</element-citation>
</ref>
<ref id="B32">
<label>32</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Lan</surname>
<given-names>TH</given-names>
</name>
<name>
<surname>Du</surname>
<given-names>JK</given-names>
</name>
<name>
<surname>Pan</surname>
<given-names>CY</given-names>
</name>
<name>
<surname>Lee</surname>
<given-names>HE</given-names>
</name>
<name>
<surname>Chung</surname>
<given-names>WH</given-names>
</name>
</person-group>
<article-title>Biomechanical analysis of alveolar bone stress around implants with different thread designs and pitches in the mandibular molar area</article-title>
<source>Clin Oral Investig</source>
<year>2012</year>
<volume>16</volume>
<fpage>363</fpage>
<lpage>369</lpage>
</element-citation>
</ref>
<ref id="B33">
<label>33</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Kong</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Liu</surname>
<given-names>BL</given-names>
</name>
<name>
<surname>Hu</surname>
<given-names>KJ</given-names>
</name>
<name>
<surname>Li</surname>
<given-names>DH</given-names>
</name>
<name>
<surname>Song</surname>
<given-names>YL</given-names>
</name>
<name>
<surname>Ma</surname>
<given-names>P</given-names>
</name>
<name>
<surname>Yang</surname>
<given-names>J</given-names>
</name>
</person-group>
<article-title>Optimized thread pitch design and stress analysis of the cylinder screwed dental implant</article-title>
<source>Hua Xi Kou Qiang Yi Xue Za Zhi</source>
<year>2006</year>
<volume>24</volume>
<fpage>509</fpage>
<lpage>512</lpage>
<fpage>515</fpage>
<pub-id pub-id-type="pmid">17334071</pub-id>
</element-citation>
</ref>
<ref id="B34">
<label>34</label>
<element-citation publication-type="book">
<person-group person-group-type="author">
<name>
<surname>Misch</surname>
<given-names>CE</given-names>
</name>
<name>
<surname>Strong</surname>
<given-names>T</given-names>
</name>
<name>
<surname>Bidez</surname>
<given-names>MW</given-names>
</name>
</person-group>
<chapter-title>Scientific rationale for dental implant design</chapter-title>
<person-group person-group-type="editor">
<name>
<surname>Misch</surname>
<given-names>CE</given-names>
</name>
</person-group>
<source>Contemporary Implant Dentistry</source>
<edition>3rd ed</edition>
<publisher-loc>St. Louis</publisher-loc>
<publisher-name>Mosby</publisher-name>
<year>2008</year>
<fpage>200</fpage>
<lpage>229</lpage>
</element-citation>
</ref>
<ref id="B35">
<label>35</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Ao</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Li</surname>
<given-names>T</given-names>
</name>
<name>
<surname>Liu</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Ding</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Wu</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Hu</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Kong</surname>
<given-names>L</given-names>
</name>
</person-group>
<article-title>Optimal design of thread height and width on an immediately loaded cylinder implant: a finite element analysis</article-title>
<source>Comput Biol Med</source>
<year>2010</year>
<volume>40</volume>
<fpage>681</fpage>
<lpage>686</lpage>
<pub-id pub-id-type="pmid">20599193</pub-id>
</element-citation>
</ref>
<ref id="B36">
<label>36</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Kong</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Hu</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Li</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Song</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Yang</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Wu</surname>
<given-names>Z</given-names>
</name>
<name>
<surname>Liu</surname>
<given-names>B</given-names>
</name>
</person-group>
<article-title>Evaluation of the cylinder implant thread height and width: a 3-dimensional finite element analysis</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2008</year>
<volume>23</volume>
<fpage>65</fpage>
<lpage>74</lpage>
<pub-id pub-id-type="pmid">18416414</pub-id>
</element-citation>
</ref>
<ref id="B37">
<label>37</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Quirynen</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Naert</surname>
<given-names>I</given-names>
</name>
<name>
<surname>van Steenberghe</surname>
<given-names>D</given-names>
</name>
</person-group>
<article-title>Fixture design and overload influence marginal bone loss and fixture success in the Brånemark system</article-title>
<source>Clin Oral Implants Res</source>
<year>1992</year>
<volume>3</volume>
<fpage>104</fpage>
<lpage>111</lpage>
<pub-id pub-id-type="pmid">1290790</pub-id>
</element-citation>
</ref>
<ref id="B38">
<label>38</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Shimada</surname>
<given-names>E</given-names>
</name>
<name>
<surname>Pilliar</surname>
<given-names>RM</given-names>
</name>
<name>
<surname>Deporter</surname>
<given-names>DA</given-names>
</name>
<name>
<surname>Schroering</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Atenafu</surname>
<given-names>E</given-names>
</name>
</person-group>
<article-title>A pilot study to assess the performance of a partially threaded sintered porous-surfaced dental implant in the dog mandible</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2007</year>
<volume>22</volume>
<fpage>948</fpage>
<lpage>954</lpage>
<pub-id pub-id-type="pmid">18271376</pub-id>
</element-citation>
</ref>
<ref id="B39">
<label>39</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Hermann</surname>
<given-names>JS</given-names>
</name>
<name>
<surname>Schoolfield</surname>
<given-names>JD</given-names>
</name>
<name>
<surname>Nummikoski</surname>
<given-names>PV</given-names>
</name>
<name>
<surname>Buser</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Schenk</surname>
<given-names>RK</given-names>
</name>
<name>
<surname>Cochran</surname>
<given-names>DL</given-names>
</name>
</person-group>
<article-title>Crestal bone changes around titanium implants: a methodologic study comparing linear radiographic with histometric measurements</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2001</year>
<volume>16</volume>
<fpage>475</fpage>
<lpage>485</lpage>
<pub-id pub-id-type="pmid">11515994</pub-id>
</element-citation>
</ref>
<ref id="B40">
<label>40</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Hänggi</surname>
<given-names>MP</given-names>
</name>
<name>
<surname>Hänggi</surname>
<given-names>DC</given-names>
</name>
<name>
<surname>Schoolfield</surname>
<given-names>JD</given-names>
</name>
<name>
<surname>Meyer</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Cochran</surname>
<given-names>DL</given-names>
</name>
<name>
<surname>Hermann</surname>
<given-names>JS</given-names>
</name>
</person-group>
<article-title>Crestal bone changes around titanium implants. Part I: A retrospective radiographic evaluation in humans comparing two non-submerged implant designs with different machined collar lengths</article-title>
<source>J Periodontol</source>
<year>2005</year>
<volume>76</volume>
<fpage>791</fpage>
<lpage>802</lpage>
<pub-id pub-id-type="pmid">15898941</pub-id>
</element-citation>
</ref>
<ref id="B41">
<label>41</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Vaillancourt</surname>
<given-names>H</given-names>
</name>
<name>
<surname>Pilliar</surname>
<given-names>RM</given-names>
</name>
<name>
<surname>McCammond</surname>
<given-names>D</given-names>
</name>
</person-group>
<article-title>Finite element analysis of crestal bone loss around porous-coated dental implants</article-title>
<source>J Appl Biomater</source>
<year>1995</year>
<volume>6</volume>
<fpage>267</fpage>
<lpage>282</lpage>
<pub-id pub-id-type="pmid">8589512</pub-id>
</element-citation>
</ref>
<ref id="B42">
<label>42</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Hansson</surname>
<given-names>S</given-names>
</name>
</person-group>
<article-title>The implant neck: smooth or provided with retention elements. A biomechanical approach</article-title>
<source>Clin Oral Implants Res</source>
<year>1999</year>
<volume>10</volume>
<fpage>394</fpage>
<lpage>405</lpage>
<pub-id pub-id-type="pmid">10551064</pub-id>
</element-citation>
</ref>
<ref id="B43">
<label>43</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Chowdhary</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Halldin</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Jimbo</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Wennerberg</surname>
<given-names>A</given-names>
</name>
</person-group>
<article-title>Influence of micro threads alteration on osseointegration and primary stability of implants: an FEA and in vivo analysis in rabbits</article-title>
<source>Clin Implant Dent Relat Res</source>
<year>2013</year>
<month>8</month>
<day>27</day>
</element-citation>
</ref>
<ref id="B44">
<label>44</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Negri</surname>
<given-names>B</given-names>
</name>
<name>
<surname>Calvo Guirado</surname>
<given-names>JL</given-names>
</name>
<name>
<surname>Maté Sánchez de Val</surname>
<given-names>JE</given-names>
</name>
<name>
<surname>Delgado >Ruíz</surname>
<given-names>RA</given-names>
</name>
<name>
<surname>Ramírez Fernández</surname>
<given-names>MP</given-names>
</name>
<name>
<surname>Barona Dorado</surname>
<given-names>C</given-names>
</name>
</person-group>
<article-title>Peri-implant tissue reactions to immediate nonocclusal loaded implants with different collar design: an experimental study in dogs</article-title>
<source>Clin Oral Implants Res</source>
<year>2014</year>
<volume>25</volume>
<fpage>e54</fpage>
<lpage>e63</lpage>
<pub-id pub-id-type="pmid">23057529</pub-id>
</element-citation>
</ref>
<ref id="B45">
<label>45</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Yun</surname>
<given-names>HJ</given-names>
</name>
<name>
<surname>Park</surname>
<given-names>JC</given-names>
</name>
<name>
<surname>Yun</surname>
<given-names>JH</given-names>
</name>
<name>
<surname>Jung</surname>
<given-names>UW</given-names>
</name>
<name>
<surname>Kim</surname>
<given-names>CS</given-names>
</name>
<name>
<surname>Choi</surname>
<given-names>SH</given-names>
</name>
<name>
<surname>Cho</surname>
<given-names>KS</given-names>
</name>
</person-group>
<article-title>A short-term clinical study of marginal bone level change around microthreaded and platform-switched implants</article-title>
<source>J Periodontal Implant Sci</source>
<year>2011</year>
<volume>41</volume>
<fpage>211</fpage>
<lpage>217</lpage>
<pub-id pub-id-type="pmid">22087411</pub-id>
</element-citation>
</ref>
<ref id="B46">
<label>46</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Song</surname>
<given-names>DW</given-names>
</name>
<name>
<surname>Lee</surname>
<given-names>DW</given-names>
</name>
<name>
<surname>Kim</surname>
<given-names>CK</given-names>
</name>
<name>
<surname>Park</surname>
<given-names>KH</given-names>
</name>
<name>
<surname>Moon</surname>
<given-names>IS</given-names>
</name>
</person-group>
<article-title>Comparative analysis of peri-implant marginal bone loss based on microthread location: a 1-year prospective study after loading</article-title>
<source>J Periodontol</source>
<year>2009</year>
<volume>80</volume>
<fpage>1937</fpage>
<lpage>1944</lpage>
<pub-id pub-id-type="pmid">19961377</pub-id>
</element-citation>
</ref>
<ref id="B47">
<label>47</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Lee</surname>
<given-names>DW</given-names>
</name>
<name>
<surname>Choi</surname>
<given-names>YS</given-names>
</name>
<name>
<surname>Park</surname>
<given-names>KH</given-names>
</name>
<name>
<surname>Kim</surname>
<given-names>CS</given-names>
</name>
<name>
<surname>Moon</surname>
<given-names>IS</given-names>
</name>
</person-group>
<article-title>Effect of microthread on the maintenance of marginal bone level: a 3-year prospective study</article-title>
<source>Clin Oral Implants Res</source>
<year>2007</year>
<volume>18</volume>
<fpage>465</fpage>
<lpage>470</lpage>
<pub-id pub-id-type="pmid">17441983</pub-id>
</element-citation>
</ref>
<ref id="B48">
<label>48</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Calvo-Guirado</surname>
<given-names>JL</given-names>
</name>
<name>
<surname>Gómez-Moreno</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Aguilar-Salvatierra</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Guardia</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Delgado-Ruiz</surname>
<given-names>RA</given-names>
</name>
<name>
<surname>Romanos</surname>
<given-names>GE</given-names>
</name>
</person-group>
<article-title>Marginal bone loss evaluation around immediate non-occlusal microthreaded implants placed in fresh extraction sockets in the maxilla: a 3-year study</article-title>
<source>Clin Oral Implants Res</source>
<year>2014</year>
<month>1</month>
<day>15</day>
</element-citation>
</ref>
<ref id="B49">
<label>49</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Amid</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Raoofi</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Kadkhodazadeh</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Movahhedi</surname>
<given-names>MR</given-names>
</name>
<name>
<surname>Khademi</surname>
<given-names>M</given-names>
</name>
</person-group>
<article-title>Effect of microthread design of dental implants on stress and strain patterns: a three-dimensional finite element analysis</article-title>
<source>Biomed Tech (Berl)</source>
<year>2013</year>
<volume>58</volume>
<fpage>457</fpage>
<lpage>467</lpage>
<pub-id pub-id-type="pmid">24021588</pub-id>
</element-citation>
</ref>
<ref id="B50">
<label>50</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Schrotenboer</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Tsao</surname>
<given-names>YP</given-names>
</name>
<name>
<surname>Kinariwala</surname>
<given-names>V</given-names>
</name>
<name>
<surname>Wang</surname>
<given-names>HL</given-names>
</name>
</person-group>
<article-title>Effect of microthreads and platform switching on crestal bone stress levels: a finite element analysis</article-title>
<source>J Periodontol</source>
<year>2008</year>
<volume>79</volume>
<fpage>2166</fpage>
<lpage>2172</lpage>
<pub-id pub-id-type="pmid">18980526</pub-id>
</element-citation>
</ref>
</ref-list>
</back>
<floats-group>
<fig id="F1" orientation="portrait" position="float">
<label>Fig. 1</label>
<caption>
<p>Implant macrodesign parameters used in this report (Implant image from Neobiotech CMI IS-II active®).</p>
</caption>
<graphic xlink:href="jap-6-547-g001"></graphic>
</fig>
<table-wrap id="T1" orientation="portrait" position="float">
<label>Table 1</label>
<caption>
<p>Studies evaluating the effect of thread shape on stress distribution</p>
</caption>
<graphic xlink:href="jap-6-547-i001"></graphic>
<table-wrap-foot>
<fn>
<p>FEA = Finite element analysis; BIC = Bone-to-implant contact.</p>
</fn>
</table-wrap-foot>
</table-wrap>
<table-wrap id="T2" orientation="portrait" position="float">
<label>Table 2</label>
<caption>
<p>Studies evaluating the effect of thread pitch on load transfer and bone-to-implant contact</p>
</caption>
<graphic xlink:href="jap-6-547-i002"></graphic>
<table-wrap-foot>
<fn>
<p>FEA = Finite element analysis; BIC = Bone-to-implant contact.</p>
</fn>
</table-wrap-foot>
</table-wrap>
<table-wrap id="T3" orientation="portrait" position="float">
<label>Table 3</label>
<caption>
<p>Studies evaluating the effect of thread depth and width on stress distribution and bone-to-implant contact</p>
</caption>
<graphic xlink:href="jap-6-547-i003"></graphic>
<table-wrap-foot>
<fn>
<p>FEA = Finite element analysis.</p>
</fn>
</table-wrap-foot>
</table-wrap>
<table-wrap id="T4" orientation="portrait" position="float">
<label>Table 4</label>
<caption>
<p>Studies evaluating the effect of crestal module on marginal bone resorption</p>
</caption>
<graphic xlink:href="jap-6-547-i004"></graphic>
<table-wrap-foot>
<fn>
<p>FEA = Finite element analysis.</p>
</fn>
</table-wrap-foot>
</table-wrap>
</floats-group>
</pmc>
</record>

Pour manipuler ce document sous Unix (Dilib)

EXPLOR_STEP=$WICRI_ROOT/Wicri/Santé/explor/EdenteV2/Data/Pmc/Corpus
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 002F07 | SxmlIndent | more

Ou

HfdSelect -h $EXPLOR_AREA/Data/Pmc/Corpus/biblio.hfd -nk 002F07 | SxmlIndent | more

Pour mettre un lien sur cette page dans le réseau Wicri

{{Explor lien
   |wiki=    Wicri/Santé
   |area=    EdenteV2
   |flux=    Pmc
   |étape=   Corpus
   |type=    RBID
   |clé=     PMC:4279055
   |texte=   The influence of thread geometry on implant osseointegration under immediate loading: a literature review
}}

Pour générer des pages wiki

HfdIndexSelect -h $EXPLOR_AREA/Data/Pmc/Corpus/RBID.i   -Sk "pubmed:25551016" \
       | HfdSelect -Kh $EXPLOR_AREA/Data/Pmc/Corpus/biblio.hfd   \
       | NlmPubMed2Wicri -a EdenteV2 

Wicri

This area was generated with Dilib version V0.6.32.
Data generation: Thu Nov 30 15:26:48 2017. Site generation: Tue Mar 8 16:36:20 2022