In-vitro study on the accuracy of a simple-design CT-guided stent for dental implants

Identifieur interne : 002C89 ( Pmc/Corpus ); précédent : 002C88; suivant : 002C90

In-vitro study on the accuracy of a simple-design CT-guided stent for dental implants

Auteurs : Young-June Huh ; Bo-Ram Choi ; Kyung-Hoe Huh ; Won-Jin Yi ; Min-Suk Heo ; Sam-Sun Lee ; Soon-Chul Choi

Source :

RBID : PMC:3465755

Abstract

Purpose

An individual surgical stent fabricated from computed tomography (CT) data, called a CT-guided stent, would be useful for accurate installation of implants. The purpose of the present study was to introduce a newly developed CT-guided stent with a simple design and evaluate the accuracy of the stent placement.

Materials and Methods

A resin template was fabricated from a hog mandible and a specially designed plastic plate, with 4 metal balls inserted in it for radiographic recognition, was attached to the occlusal surface of the template. With the surgical stent applied, CT images were taken, and virtual implants were placed using software. The spatial positions of the virtually positioned implants were acquired and implant guiding holes were drilled into the surgical stent using a specially designed 5-axis drilling machine. The surgical stent was placed on the mandible and CT images were taken again. The discrepancy between the central axis of the drilled holes on the second CT images and the virtually installed implants on the first CT images was evaluated.

Results

The deviation of the entry point and angulation of the central axis in the reference plane were 0.47±0.27 mm, 0.57±0.23 mm, and 0.64±0.16°, 0.57±0.15°, respectively. However, for the two different angulations in each group, the 20° angulation showed a greater error in the deviation of the entry point than did the 10° angulation.

Conclusion

The CT-guided template proposed in this study was highly accurate. It could replace existing implant guide systems to reduce costs and effort.


Url:
DOI: 10.5624/isd.2012.42.3.139
PubMed: 23071963
PubMed Central: 3465755

Links to Exploration step

PMC:3465755

Le document en format XML

<record>
<TEI>
<teiHeader>
<fileDesc>
<titleStmt>
<title xml:lang="en">In-vitro study on the accuracy of a simple-design CT-guided stent for dental implants</title>
<author>
<name sortKey="Huh, Young June" sort="Huh, Young June" uniqKey="Huh Y" first="Young-June" last="Huh">Young-June Huh</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Radiology and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Choi, Bo Ram" sort="Choi, Bo Ram" uniqKey="Choi B" first="Bo-Ram" last="Choi">Bo-Ram Choi</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Radiology and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Huh, Kyung Hoe" sort="Huh, Kyung Hoe" uniqKey="Huh K" first="Kyung-Hoe" last="Huh">Kyung-Hoe Huh</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Radiology and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Yi, Won Jin" sort="Yi, Won Jin" uniqKey="Yi W" first="Won-Jin" last="Yi">Won-Jin Yi</name>
<affiliation>
<nlm:aff id="A2">Department of Oral and Maxillofacial Radiology, Dental Research Institute and BK21 Craniomaxillofacial Life Science, School of Dentistry, Seoul National University, Seoul, Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Heo, Min Suk" sort="Heo, Min Suk" uniqKey="Heo M" first="Min-Suk" last="Heo">Min-Suk Heo</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Radiology and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Lee, Sam Sun" sort="Lee, Sam Sun" uniqKey="Lee S" first="Sam-Sun" last="Lee">Sam-Sun Lee</name>
<affiliation>
<nlm:aff id="A2">Department of Oral and Maxillofacial Radiology, Dental Research Institute and BK21 Craniomaxillofacial Life Science, School of Dentistry, Seoul National University, Seoul, Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Choi, Soon Chul" sort="Choi, Soon Chul" uniqKey="Choi S" first="Soon-Chul" last="Choi">Soon-Chul Choi</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Radiology and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Korea.</nlm:aff>
</affiliation>
</author>
</titleStmt>
<publicationStmt>
<idno type="wicri:source">PMC</idno>
<idno type="pmid">23071963</idno>
<idno type="pmc">3465755</idno>
<idno type="url">http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3465755</idno>
<idno type="RBID">PMC:3465755</idno>
<idno type="doi">10.5624/isd.2012.42.3.139</idno>
<date when="2012">2012</date>
<idno type="wicri:Area/Pmc/Corpus">002C89</idno>
<idno type="wicri:explorRef" wicri:stream="Pmc" wicri:step="Corpus" wicri:corpus="PMC">002C89</idno>
</publicationStmt>
<sourceDesc>
<biblStruct>
<analytic>
<title xml:lang="en" level="a" type="main">In-vitro study on the accuracy of a simple-design CT-guided stent for dental implants</title>
<author>
<name sortKey="Huh, Young June" sort="Huh, Young June" uniqKey="Huh Y" first="Young-June" last="Huh">Young-June Huh</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Radiology and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Choi, Bo Ram" sort="Choi, Bo Ram" uniqKey="Choi B" first="Bo-Ram" last="Choi">Bo-Ram Choi</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Radiology and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Huh, Kyung Hoe" sort="Huh, Kyung Hoe" uniqKey="Huh K" first="Kyung-Hoe" last="Huh">Kyung-Hoe Huh</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Radiology and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Yi, Won Jin" sort="Yi, Won Jin" uniqKey="Yi W" first="Won-Jin" last="Yi">Won-Jin Yi</name>
<affiliation>
<nlm:aff id="A2">Department of Oral and Maxillofacial Radiology, Dental Research Institute and BK21 Craniomaxillofacial Life Science, School of Dentistry, Seoul National University, Seoul, Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Heo, Min Suk" sort="Heo, Min Suk" uniqKey="Heo M" first="Min-Suk" last="Heo">Min-Suk Heo</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Radiology and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Lee, Sam Sun" sort="Lee, Sam Sun" uniqKey="Lee S" first="Sam-Sun" last="Lee">Sam-Sun Lee</name>
<affiliation>
<nlm:aff id="A2">Department of Oral and Maxillofacial Radiology, Dental Research Institute and BK21 Craniomaxillofacial Life Science, School of Dentistry, Seoul National University, Seoul, Korea.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Choi, Soon Chul" sort="Choi, Soon Chul" uniqKey="Choi S" first="Soon-Chul" last="Choi">Soon-Chul Choi</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Radiology and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Korea.</nlm:aff>
</affiliation>
</author>
</analytic>
<series>
<title level="j">Imaging Science in Dentistry</title>
<idno type="ISSN">2233-7822</idno>
<idno type="eISSN">2233-7830</idno>
<imprint>
<date when="2012">2012</date>
</imprint>
</series>
</biblStruct>
</sourceDesc>
</fileDesc>
<profileDesc>
<textClass></textClass>
</profileDesc>
</teiHeader>
<front>
<div type="abstract" xml:lang="en">
<sec>
<title>Purpose</title>
<p>An individual surgical stent fabricated from computed tomography (CT) data, called a CT-guided stent, would be useful for accurate installation of implants. The purpose of the present study was to introduce a newly developed CT-guided stent with a simple design and evaluate the accuracy of the stent placement.</p>
</sec>
<sec>
<title>Materials and Methods</title>
<p>A resin template was fabricated from a hog mandible and a specially designed plastic plate, with 4 metal balls inserted in it for radiographic recognition, was attached to the occlusal surface of the template. With the surgical stent applied, CT images were taken, and virtual implants were placed using software. The spatial positions of the virtually positioned implants were acquired and implant guiding holes were drilled into the surgical stent using a specially designed 5-axis drilling machine. The surgical stent was placed on the mandible and CT images were taken again. The discrepancy between the central axis of the drilled holes on the second CT images and the virtually installed implants on the first CT images was evaluated.</p>
</sec>
<sec>
<title>Results</title>
<p>The deviation of the entry point and angulation of the central axis in the reference plane were 0.47±0.27 mm, 0.57±0.23 mm, and 0.64±0.16°, 0.57±0.15°, respectively. However, for the two different angulations in each group, the 20° angulation showed a greater error in the deviation of the entry point than did the 10° angulation.</p>
</sec>
<sec>
<title>Conclusion</title>
<p>The CT-guided template proposed in this study was highly accurate. It could replace existing implant guide systems to reduce costs and effort.</p>
</sec>
</div>
</front>
<back>
<div1 type="bibliography">
<listBibl>
<biblStruct>
<analytic>
<author>
<name sortKey="Branemark, Pi" uniqKey="Branemark P">PI Branemark</name>
</author>
<author>
<name sortKey="Adell, R" uniqKey="Adell R">R Adell</name>
</author>
<author>
<name sortKey="Breine, U" uniqKey="Breine U">U Breine</name>
</author>
<author>
<name sortKey="Hansson, Bo" uniqKey="Hansson B">BO Hansson</name>
</author>
<author>
<name sortKey="Lindstrom, J" uniqKey="Lindstrom J">J Lindstrom</name>
</author>
<author>
<name sortKey="Ohlsson, A" uniqKey="Ohlsson A">A Ohlsson</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Israelson, H" uniqKey="Israelson H">H Israelson</name>
</author>
<author>
<name sortKey="Plemons, Jm" uniqKey="Plemons J">JM Plemons</name>
</author>
<author>
<name sortKey="Watkins, P" uniqKey="Watkins P">P Watkins</name>
</author>
<author>
<name sortKey="Sory, C" uniqKey="Sory C">C Sory</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Espinosa Marino, J" uniqKey="Espinosa Marino J">J Espinosa Marino</name>
</author>
<author>
<name sortKey="Alvarez Arenal, A" uniqKey="Alvarez Arenal A">A Alvarez Arenal</name>
</author>
<author>
<name sortKey="Pardo Ceballos, A" uniqKey="Pardo Ceballos A">A Pardo Ceballos</name>
</author>
<author>
<name sortKey="Fernandez Vazquez, Jp" uniqKey="Fernandez Vazquez J">JP Fernandez Vazquez</name>
</author>
<author>
<name sortKey="Ibaseta Diaz, G" uniqKey="Ibaseta Diaz G">G Ibaseta Diaz</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Basten, Ch" uniqKey="Basten C">CH Basten</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Borrow, Jw" uniqKey="Borrow J">JW Borrow</name>
</author>
<author>
<name sortKey="Smith, Jp" uniqKey="Smith J">JP Smith</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Basten, Ch" uniqKey="Basten C">CH Basten</name>
</author>
<author>
<name sortKey="Kois, Jc" uniqKey="Kois J">JC Kois</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Weinberg, La" uniqKey="Weinberg L">LA Weinberg</name>
</author>
<author>
<name sortKey="Kruger, B" uniqKey="Kruger B">B Kruger</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Weinberg, La" uniqKey="Weinberg L">LA Weinberg</name>
</author>
<author>
<name sortKey="Kruger, B" uniqKey="Kruger B">B Kruger</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Becker, Cm" uniqKey="Becker C">CM Becker</name>
</author>
<author>
<name sortKey="Kaiser, Da" uniqKey="Kaiser D">DA Kaiser</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Cucchiara, R" uniqKey="Cucchiara R">R Cucchiara</name>
</author>
<author>
<name sortKey="Franchini, F" uniqKey="Franchini F">F Franchini</name>
</author>
<author>
<name sortKey="Lamma, A" uniqKey="Lamma A">A Lamma</name>
</author>
<author>
<name sortKey="Lamma, E" uniqKey="Lamma E">E Lamma</name>
</author>
<author>
<name sortKey="Sansoni, T" uniqKey="Sansoni T">T Sansoni</name>
</author>
<author>
<name sortKey="Sarti, E" uniqKey="Sarti E">E Sarti</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Cucchiara, R" uniqKey="Cucchiara R">R Cucchiara</name>
</author>
<author>
<name sortKey="Lamma, E" uniqKey="Lamma E">E Lamma</name>
</author>
<author>
<name sortKey="Sansoni, T" uniqKey="Sansoni T">T Sansoni</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Owings, Jr" uniqKey="Owings J">JR Owings</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Marchack, Cb" uniqKey="Marchack C">CB Marchack</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Mupparapu, M" uniqKey="Mupparapu M">M Mupparapu</name>
</author>
<author>
<name sortKey="Singer, Sr" uniqKey="Singer S">SR Singer</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Tardieu, Pb" uniqKey="Tardieu P">PB Tardieu</name>
</author>
<author>
<name sortKey="Vrielinck, L" uniqKey="Vrielinck L">L Vrielinck</name>
</author>
<author>
<name sortKey="Escolano, E" uniqKey="Escolano E">E Escolano</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Widmann, G" uniqKey="Widmann G">G Widmann</name>
</author>
<author>
<name sortKey="Zangerl, A" uniqKey="Zangerl A">A Zangerl</name>
</author>
<author>
<name sortKey="Keiler, M" uniqKey="Keiler M">M Keiler</name>
</author>
<author>
<name sortKey="Stoffner, R" uniqKey="Stoffner R">R Stoffner</name>
</author>
<author>
<name sortKey="Bale, R" uniqKey="Bale R">R Bale</name>
</author>
<author>
<name sortKey="Puelacher, W" uniqKey="Puelacher W">W Puelacher</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Widmann, G" uniqKey="Widmann G">G Widmann</name>
</author>
<author>
<name sortKey="Keiler, M" uniqKey="Keiler M">M Keiler</name>
</author>
<author>
<name sortKey="Zangerl, A" uniqKey="Zangerl A">A Zangerl</name>
</author>
<author>
<name sortKey="Stoffner, R" uniqKey="Stoffner R">R Stoffner</name>
</author>
<author>
<name sortKey="Longato, S" uniqKey="Longato S">S Longato</name>
</author>
<author>
<name sortKey="Bale, R" uniqKey="Bale R">R Bale</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Pongracz, F" uniqKey="Pongracz F">F Pongrácz</name>
</author>
<author>
<name sortKey="Bardosi, Z" uniqKey="Bardosi Z">Z Bárdosi</name>
</author>
<author>
<name sortKey="Szab, L" uniqKey="Szab L">L Szabó</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Koulechov, K" uniqKey="Koulechov K">K Koulechov</name>
</author>
<author>
<name sortKey="Lueth, T" uniqKey="Lueth T">T Lueth</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Parel, Sm" uniqKey="Parel S">SM Parel</name>
</author>
<author>
<name sortKey="Triplett, Rg" uniqKey="Triplett R">RG Triplett</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Besimo, Ce" uniqKey="Besimo C">CE Besimo</name>
</author>
<author>
<name sortKey="Lambrecht, Jt" uniqKey="Lambrecht J">JT Lambrecht</name>
</author>
<author>
<name sortKey="Guindy, Js" uniqKey="Guindy J">JS Guindy</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Fortin, T" uniqKey="Fortin T">T Fortin</name>
</author>
<author>
<name sortKey="Champleboux, G" uniqKey="Champleboux G">G Champleboux</name>
</author>
<author>
<name sortKey="Lormee, J" uniqKey="Lormee J">J Lormee</name>
</author>
<author>
<name sortKey="Coudert, Jl" uniqKey="Coudert J">JL Coudert</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Widmann, G" uniqKey="Widmann G">G Widmann</name>
</author>
<author>
<name sortKey="Widmann, R" uniqKey="Widmann R">R Widmann</name>
</author>
<author>
<name sortKey="Widmann, E" uniqKey="Widmann E">E Widmann</name>
</author>
<author>
<name sortKey="Jaschke, W" uniqKey="Jaschke W">W Jaschke</name>
</author>
<author>
<name sortKey="Bale, Rj" uniqKey="Bale R">RJ Bale</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Brief, J" uniqKey="Brief J">J Brief</name>
</author>
<author>
<name sortKey="Edinger, D" uniqKey="Edinger D">D Edinger</name>
</author>
<author>
<name sortKey="Hassfeld, S" uniqKey="Hassfeld S">S Hassfeld</name>
</author>
<author>
<name sortKey="Eggers, G" uniqKey="Eggers G">G Eggers</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Wanschitz, F" uniqKey="Wanschitz F">F Wanschitz</name>
</author>
<author>
<name sortKey="Birkfellner, W" uniqKey="Birkfellner W">W Birkfellner</name>
</author>
<author>
<name sortKey="Watzinger, F" uniqKey="Watzinger F">F Watzinger</name>
</author>
<author>
<name sortKey="Schopper, C" uniqKey="Schopper C">C Schopper</name>
</author>
<author>
<name sortKey="Patruta, S" uniqKey="Patruta S">S Patruta</name>
</author>
<author>
<name sortKey="Kainberger, F" uniqKey="Kainberger F">F Kainberger</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Van Steenberghe, D" uniqKey="Van Steenberghe D">D Van Steenberghe</name>
</author>
<author>
<name sortKey="Malevez, C" uniqKey="Malevez C">C Malevez</name>
</author>
<author>
<name sortKey="Van Cleynenbreugel, J" uniqKey="Van Cleynenbreugel J">J Van Cleynenbreugel</name>
</author>
<author>
<name sortKey="Bou Serhal, C" uniqKey="Bou Serhal C">C Bou Serhal</name>
</author>
<author>
<name sortKey="Dhoore, E" uniqKey="Dhoore E">E Dhoore</name>
</author>
<author>
<name sortKey="Schutyser, F" uniqKey="Schutyser F">F Schutyser</name>
</author>
</analytic>
</biblStruct>
</listBibl>
</div1>
</back>
</TEI>
<pmc article-type="research-article">
<pmc-dir>properties open_access</pmc-dir>
<front>
<journal-meta>
<journal-id journal-id-type="nlm-ta">Imaging Sci Dent</journal-id>
<journal-id journal-id-type="iso-abbrev">Imaging Sci Dent</journal-id>
<journal-id journal-id-type="publisher-id">ISD</journal-id>
<journal-title-group>
<journal-title>Imaging Science in Dentistry</journal-title>
</journal-title-group>
<issn pub-type="ppub">2233-7822</issn>
<issn pub-type="epub">2233-7830</issn>
<publisher>
<publisher-name>Korean Academy of Oral and Maxillofacial Radiology</publisher-name>
</publisher>
</journal-meta>
<article-meta>
<article-id pub-id-type="pmid">23071963</article-id>
<article-id pub-id-type="pmc">3465755</article-id>
<article-id pub-id-type="doi">10.5624/isd.2012.42.3.139</article-id>
<article-categories>
<subj-group subj-group-type="heading">
<subject>Original Article</subject>
</subj-group>
</article-categories>
<title-group>
<article-title>In-vitro study on the accuracy of a simple-design CT-guided stent for dental implants</article-title>
</title-group>
<contrib-group>
<contrib contrib-type="author">
<name>
<surname>Huh</surname>
<given-names>Young-June</given-names>
</name>
<xref ref-type="aff" rid="A1"></xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Choi</surname>
<given-names>Bo-Ram</given-names>
</name>
<xref ref-type="aff" rid="A1"></xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Huh</surname>
<given-names>Kyung-Hoe</given-names>
</name>
<xref ref-type="aff" rid="A1"></xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Yi</surname>
<given-names>Won-Jin</given-names>
</name>
<xref ref-type="aff" rid="A2">*</xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Heo</surname>
<given-names>Min-Suk</given-names>
</name>
<xref ref-type="aff" rid="A1"></xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Lee</surname>
<given-names>Sam-Sun</given-names>
</name>
<xref ref-type="aff" rid="A2">*</xref>
</contrib>
<contrib contrib-type="author" corresp="yes">
<name>
<surname>Choi</surname>
<given-names>Soon-Chul</given-names>
</name>
<xref ref-type="aff" rid="A1"></xref>
</contrib>
</contrib-group>
<aff id="A1">Department of Oral and Maxillofacial Radiology and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, Korea.</aff>
<aff id="A2">
<label>*</label>
Department of Oral and Maxillofacial Radiology, Dental Research Institute and BK21 Craniomaxillofacial Life Science, School of Dentistry, Seoul National University, Seoul, Korea.</aff>
<author-notes>
<corresp>Correspondence to: Prof. Soon-Chul Choi. Department of Oral and Maxillofacial Radiology, School of Dentistry, Seoul National University, 28 Yeongeon-dong, Jongno-gu, Seoul 110-768, Korea. Tel) 82-2-2072-2622, Fax) 82-2-744-3919,
<email>raychoi@snu.ac.kr</email>
</corresp>
</author-notes>
<pub-date pub-type="ppub">
<month>9</month>
<year>2012</year>
</pub-date>
<pub-date pub-type="epub">
<day>21</day>
<month>9</month>
<year>2012</year>
</pub-date>
<volume>42</volume>
<issue>3</issue>
<fpage>139</fpage>
<lpage>146</lpage>
<history>
<date date-type="received">
<day>20</day>
<month>3</month>
<year>2012</year>
</date>
<date date-type="rev-recd">
<day>01</day>
<month>6</month>
<year>2012</year>
</date>
<date date-type="accepted">
<day>16</day>
<month>6</month>
<year>2012</year>
</date>
</history>
<permissions>
<copyright-statement>Copyright © 2012 by Korean Academy of Oral and Maxillofacial Radiology</copyright-statement>
<copyright-year>2012</copyright-year>
<license license-type="open-access" xlink:href="http://creativecommons.org/licenses/by-nc/3.0">
<license-p>This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (
<ext-link ext-link-type="uri" xlink:href="http://creativecommons.org/licenses/by-nc/3.0">http://creativecommons.org/licenses/by-nc/3.0</ext-link>
) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.</license-p>
</license>
</permissions>
<abstract>
<sec>
<title>Purpose</title>
<p>An individual surgical stent fabricated from computed tomography (CT) data, called a CT-guided stent, would be useful for accurate installation of implants. The purpose of the present study was to introduce a newly developed CT-guided stent with a simple design and evaluate the accuracy of the stent placement.</p>
</sec>
<sec>
<title>Materials and Methods</title>
<p>A resin template was fabricated from a hog mandible and a specially designed plastic plate, with 4 metal balls inserted in it for radiographic recognition, was attached to the occlusal surface of the template. With the surgical stent applied, CT images were taken, and virtual implants were placed using software. The spatial positions of the virtually positioned implants were acquired and implant guiding holes were drilled into the surgical stent using a specially designed 5-axis drilling machine. The surgical stent was placed on the mandible and CT images were taken again. The discrepancy between the central axis of the drilled holes on the second CT images and the virtually installed implants on the first CT images was evaluated.</p>
</sec>
<sec>
<title>Results</title>
<p>The deviation of the entry point and angulation of the central axis in the reference plane were 0.47±0.27 mm, 0.57±0.23 mm, and 0.64±0.16°, 0.57±0.15°, respectively. However, for the two different angulations in each group, the 20° angulation showed a greater error in the deviation of the entry point than did the 10° angulation.</p>
</sec>
<sec>
<title>Conclusion</title>
<p>The CT-guided template proposed in this study was highly accurate. It could replace existing implant guide systems to reduce costs and effort.</p>
</sec>
</abstract>
<kwd-group>
<kwd>Computer-Assisted Diagnosis</kwd>
<kwd>Dental Implant</kwd>
<kwd>Tomography, X-Ray Computed</kwd>
<kwd>In Vitro</kwd>
</kwd-group>
</article-meta>
</front>
<body>
<sec>
<title>Introduction</title>
<p>Ever since the first dental implants have been introduced by Branemark et al
<xref ref-type="bibr" rid="B1">1</xref>
in 1969, more than four decades ago, dental practitioners and researchers have been searching for various methods for improving the accuracy of the surgical placement of implants. Advances in dental imaging technology utilizing computed tomography (CT) has been proven especially useful when determining the installation sites of dental implants.
<xref ref-type="bibr" rid="B2">2</xref>
-
<xref ref-type="bibr" rid="B8">8</xref>
However, even though the CT images may provide accurate anatomical information of the jaws, the final decision is performed by the surgeon placing the implants.
<xref ref-type="bibr" rid="B2">2</xref>
-
<xref ref-type="bibr" rid="B8">8</xref>
Therefore, the operator's experience and skills become a critical factor in successful implant therapy, and for those surgeons lacking sufficient expertise for specific clinical situations, positioning and installing implant fixtures often become highly arduous tasks.
<xref ref-type="bibr" rid="B2">2</xref>
-
<xref ref-type="bibr" rid="B8">8</xref>
To cope with the problem, more attention is now being paid to image-guided implantology, which is designed to aid surgeons when performing implant treatments.
<xref ref-type="bibr" rid="B9">9</xref>
-
<xref ref-type="bibr" rid="B12">12</xref>
</p>
<p>Currently, one of the most popular systems is based on the CAD/CAM concept, which produces a 3-dimensional (3D) rapid processing (RP) jaw model with 3D CT imaging, and fabricates surgical stents for this RP model.
<xref ref-type="bibr" rid="B13">13</xref>
Though this stent, leaving a hole in the position for fixture installation, may be advantageous in placing implants precisely, it has its setbacks due to the complex procedures required, errors during RP model fabrication, and high costs.
<xref ref-type="bibr" rid="B14">14</xref>
,
<xref ref-type="bibr" rid="B15">15</xref>
A modified system aimed to reduce costs and effort was proposed, which attached a framework to the patient's face prior to taking CT images and setting the stent holes on the basis of this framework.
<xref ref-type="bibr" rid="B16">16</xref>
,
<xref ref-type="bibr" rid="B17">17</xref>
However, it was inconvenient to use and there were possible errors during the fixation of the framework on the patient's face. Pongracz and his coworkers developed a software program, allowing intuitive treatment plans using 3D CT images.
<xref ref-type="bibr" rid="B18">18</xref>
With the software, both virtual planning and virtual surgery were made possible. However, transferring virtual information to actual operations had many shortcomings. Koulechov and Lueth proposed tracing the position of an operative drill for automated implant surgery.
<xref ref-type="bibr" rid="B19">19</xref>
This method showed high accuracy in tracing drill position with very limited errors, but its use in the oral cavity was restricted due to the large tracking devices attached to the drills. Another method used radiopaque markers inserted into the dentures of the patients, and CT images were taken to place guide holes based on the marker locations of the dentures. However, this method was only applicable to patients with complete dentures, and produced potential errors from ill-fitting dentures.
<xref ref-type="bibr" rid="B20">20</xref>
</p>
<p>So far, CT-guided template production techniques have been either inclined toward precision, or focused on feasibility regarding costs and easy application. It is necessary to develop a method with both high accuracy and feasibility for clinical settings.</p>
<p>The purpose of the present study was to evaluate the accuracy of a newly developed CT-guided surgical stent with a simple design.</p>
</sec>
<sec sec-type="materials|methods">
<title>Materials and Methods</title>
<p>A hog dry mandible, with a molar extracted on each side, was used throughout the experiment for surgical stent fabrication and CT image taking. An impression of the mandible including the alveolar bone and teeth was taken, using silicone rubber impression material (Fegura-Sil hydro spezial II, Feguramed GmbH, Buchen, Germany) and an impression tray made with baseplate wax. From this impression, a stone cast was prepared. On the stone cast, an occlusal stent that registered the occlusal surface of the hog's teeth was fabricated using a resin plate according to common clinical protocol. With the stent applied, the edentulous area of the mandible was filled with resin up to the level of the occlusal surface of the adjacent teeth. On the edentulous area of the occlusal stent, a specially designed plastic plate (20×35×2 mm) was attached to serve as a reference plane on CT images. For radiographic recognition, a metal ball (2 mm in diameter) was inserted in each of the 4 corners of the plate 2 mm apart from the edge of the plate (
<xref ref-type="fig" rid="F1">Fig. 1</xref>
).</p>
<p>CT images of the hog mandible, with the surgical stents applied on both sides, were taken using a multi-detector spiral CT machine (Somatom Sensation 10, Siemens, Erlangen, Germany). The conditions of the CT images were as follows: Slice thickness of 0.75 mm, 120 kVp, 100 mAs, gantry 0°, 0.254 pixel spacing, and reconstruction interval of 0.6 mm. The CT images were saved in DICOM format, and the data were transferred to the V-Implant (Cybermed, Seoul, Korea) software.</p>
<p>Reformatted and recombinant images including panoramic, cross-sectional, and 3D images as well as axial images were acquired from the CT data. The positions and directions of the implants were planned using the additional functions of the V-Implant software. The software expressed numerically the spatial position of the virtual implant axis with the coordinates of the insertion point and the angle relative to the reference plane. The reference plane on the CT images was determined by connecting the center of the radiopaque metal balls in the plastic plate. The procedure determining the center of the metal balls and the resultant reference plane was automatically performed by the software. The data of the spatial positions for the virtually installed implants were recorded in the software (
<xref ref-type="fig" rid="F2">Fig. 2</xref>
).</p>
<p>The detailed procedure calculating the spatial position of the central axis of the virtually installed implant was as follows. The line connecting the 2 metal balls aligned to the long axis on the plastic plate was set as the X-axis, and the other line aligned to the short axis was set as the Y-axis. The point of intersection was set as the datum point O, thereby constituting the XOY-plane (
<xref ref-type="fig" rid="F3">Fig. 3</xref>
). The X and Y coordinates of an insertion point P(x, y) was determined by the crossing point between the axis of the virtual implant and the XOY-plane, and the angle between the axis of the virtual implant and the XOY-plane was determined as Xθ and Yθ according to the X-axis and Y-axis, respectively (
<xref ref-type="fig" rid="F4">Figs. 4</xref>
and
<xref ref-type="fig" rid="F5">5</xref>
).</p>
<p>The virtual installation of implants was performed with five different inclinations from the occlusal plane, i.e., perpendicular, buccal, lingual, mesial, and distal. Each inclination but the perpendicular one had two different angulations, 10° and 20°. Thus, the planned implants were divided into 9 groups according to their axis inclination and angulation. For each group, 10 fixtures were placed, 5 on each side (5 on the right side of the mandible, 5 on the left side of the mandible). Therefore, a total of 90 fixtures were virtually inserted.</p>
<p>The data of the virtually installed fixtures as mentioned above were transferred to a specially designed 5-axis drilling machine (
<xref ref-type="fig" rid="F6">Fig. 6A</xref>
) and the implant guiding holes were drilled into each surgical stent. The drilling procedure was as follows: The plastic plate, used as a reference plane, of the surgical stent was fixed parallel to the table of the drilling machine. By moving and rotating the table of the machine, which can control the position and angulation of the table by rotating each of its 5 axes, the spatial position of each virtual implant axis was reproduced on the table of the machine. Finally, guide holes were drilled into each of the surgical stents (
<xref ref-type="fig" rid="F6">Fig. 6B</xref>
). A total of 90 surgical stents, each with a guide hole (9 groups according to the inclination and angulation of the virtual implant axis, and 10 stents for each group), were drilled.</p>
<p>After applying the stents to the hog mandible, CT images were taken under the same conditions as taken previously. The spatial position of the drilled holes was acquired on the software. The procedure for determining the spatial position of the drilled holes was as follows: The drilled guide hole was shown as a black cylindrical area on the CT images, and placing a virtual implant into the guide hole was performed on the software (
<xref ref-type="fig" rid="F7">Fig. 7A</xref>
). After confirming the virtually positioned implant to be placed exactly into the image of a guide hole, the position and the angulation of the central axis of the guide hole were calculated automatically by the software (
<xref ref-type="fig" rid="F7">Fig. 7B</xref>
).</p>
<p>The spatial positions of the drilled holes were then compared with those of the virtually installed implant in the planning phase. The discrepancy between them was calculated and recorded as the amount of error for this system.</p>
</sec>
<sec sec-type="results">
<title>Results</title>
<p>
<xref ref-type="table" rid="T1">Table 1</xref>
shows the discrepancy of the X, Y coordinates and the angulation for each group. The insertion point of the implant axis on the plate was described as the X and Y coordinates.</p>
<p>The discrepancy of the insertion point P (x, y) ranged from 0.12 mm to 0.93 mm on the X-axis and from 0.21 mm to 0.91 mm on the Y-axis. The discrepancy of insertion angulations to the XOY-plane (Xθ, Yθ) ranged from 0.35 to 0.89 on the X-axis and from 0.45 to 0.77 on the Y-axis. The mean values of ΔX, ΔY, ΔXθ, and ΔYθ were 0.46mm, 0.57 mm, 0.64°, and 0.57°, respectively.</p>
<p>For the two different angulations in each group, 20 degree angulation showed a greater error in the deviation of the insertion point P (x, y) than 10 degree angulation, but not in the discrepancy of the angle.</p>
</sec>
<sec sec-type="discussion">
<title>Discussion</title>
<p>The present system used a plastic plate with metal balls to determine the reference plane, and the procedure is relatively simple and performed commonly in dental clinics. Even though the procedure was simple and cost-effective, the resultant stent still provided an accurate template.</p>
<p>Previous studies have shown that, in general, the accuracy of implant placement using surgical stents will be higher than that of implant placement without stents. However, its accuracy will depend mostly on the operator's technique.
<xref ref-type="bibr" rid="B21">21</xref>
,
<xref ref-type="bibr" rid="B22">22</xref>
Recently, several studies using objective numerical values have been conducted to evaluate the accuracy of implant guiding systems, specifically, the quantified accuracy of automatic position-trace systems. Widmann et al evaluated the accuracy of automatic position tracing system based on deviation from the reference point, and the average error was 0.42±0.26 mm (1 mm maximum).
<xref ref-type="bibr" rid="B23">23</xref>
Brief et al used a coordinate measuring device called the 'Faro arm', and they reported an error of 0.60 mm and 2.12°.
<xref ref-type="bibr" rid="B24">24</xref>
Using mutual information registration, which evaluated the deviation of the fixture position by superimposing CT images before and after the surgery, Wanschitz et al reported that the accuracy was 0.96±0.72 mm (range 0.0-3.5 mm).
<xref ref-type="bibr" rid="B25">25</xref>
Van Steenberghe et al inserted zygoma implants using a 3D CT system and reported that angulation errors of most implants were not greater than 3° and that the largest error was 2.7 mm.
<xref ref-type="bibr" rid="B26">26</xref>
The deviation of the entry point and angulation for the system proposed in the present study were 0.47±0.27 mm, 0.57±0.23 mm, 0.64±0.16°, and 0.57±0.15°, respectively, and this shows that this system is highly accurate. The range of errors acquired from the present study were similar to, or more accurate than, those obtained from previous studies.</p>
<p>We attempted to exclude the errors that may have been caused by surgeons and only focused on evaluating the in vitro accuracy of the guide holes of the stents. However, any possible accidental errors that may have occurred in the course of the clinical surgery could have affected the results profoundly. Therefore, for clinical applications, additional studies regarding surgeon factors should be performed.</p>
<p>There are some limitations in the present system. First, an error could originate from the partial volume averaging effect of the CT image. This kind of error occurs when the shape of the spherical metal ball, inserted in the plastic plate, is ovoid rather than round on the CT images. Because the virtual reference plane is set by connecting the centers of the metal balls on the CT images, discrepancy between a virtual reference plane and the real one might be caused, when the center of the metal ball on the CT images was different from the center of the real ones. This phenomenon might happen everywhere in the guide stent system, to all the structures, since resolution in the longitudinal direction is lower than that in the axial direction, which means that the voxel of the CT is not isotropic. This is inevitable, due to mechanical limitations of the CT itself. However, in the present study the minimum CT slice thickness of 0.75mm was used to minimize this error. If dental cone beam CT with isotropic voxel and higher resolution can be used, this error will be reduced further.</p>
<p>Second, the error may also originate from the drilling procedure. Before the guide hole is made in the stent using the specially designed 5-axis drilling machine, the reference plane including the 4 metal balls and the datum point O, which was the cross point of the X-axis and the Y-axis, is determined. For drilling, the drill tip should be placed on the datum point of the plastic plate (the datum point settlement), and afterwards the table of the machine should be moved according to the calculated values from the software. In the present study, however, it was not easy to precisely place the drill tip at the datum point because the drill tip was not sharp enough. Especially when the plastic plate was inclined to reflect the angulation of the central axis of the virtual implant fixture, it was not easy to accurately set the datum point with the drill tip spatially. We think that is why 20 degree angulation showed a greater errors in the deviation of the entry point than did 10 degree angulation.</p>
<p>During the drilling procedure in the present system, this kind of error seems to have originated from datum point settlement. Improvements in settlement of the datum point and the design of the plastic plate are needed.</p>
<p>Third, after the datum point settlement, the plastic plate fixed to the machine table should be moved using a scale attached to each axis to avoid any possible errors. The minimum distance of the table movement was 0.02 mm, and the minimum angulation of the table inclination is 1/30; therefore, the values below the limitation cannot be reflected precisely in this drilling machine. However, the amount of error in the present system was comparable to or more accurate than those in previous studies.</p>
<p>In conclusion, the present study proposed a newly developed CT-guided surgical stent system, in which guide holes were drilled into preoperatively planned positions with remarkably high accuracy. This system could be anticipated to replace current implant guide systems in order to reduce costs and effort, and contribute to the popularization of implant treatments.</p>
</sec>
</body>
<back>
<fn-group>
<fn fn-type="supported-by">
<p>This study was supported by the research grant No. 03-2005-001 from the Seoul National University Dental Hospital Research Fund.</p>
</fn>
</fn-group>
<ref-list>
<ref id="B1">
<label>1</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Branemark</surname>
<given-names>PI</given-names>
</name>
<name>
<surname>Adell</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Breine</surname>
<given-names>U</given-names>
</name>
<name>
<surname>Hansson</surname>
<given-names>BO</given-names>
</name>
<name>
<surname>Lindstrom</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Ohlsson</surname>
<given-names>A</given-names>
</name>
</person-group>
<article-title>Intra-osseous anchorage of dental prostheses. I. Experimental studies</article-title>
<source>Scand J Plast Reconstr Surg</source>
<year>1969</year>
<volume>3</volume>
<fpage>81</fpage>
<lpage>100</lpage>
<pub-id pub-id-type="pmid">4924041</pub-id>
</element-citation>
</ref>
<ref id="B2">
<label>2</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Israelson</surname>
<given-names>H</given-names>
</name>
<name>
<surname>Plemons</surname>
<given-names>JM</given-names>
</name>
<name>
<surname>Watkins</surname>
<given-names>P</given-names>
</name>
<name>
<surname>Sory</surname>
<given-names>C</given-names>
</name>
</person-group>
<article-title>Barium-coated surgical stents and computer-assisted tomography in the preoperative assessment of dental implant patients</article-title>
<source>Int J Periodontics Restorative Dent</source>
<year>1992</year>
<volume>12</volume>
<fpage>52</fpage>
<lpage>61</lpage>
<pub-id pub-id-type="pmid">1526711</pub-id>
</element-citation>
</ref>
<ref id="B3">
<label>3</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Espinosa Marino</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Alvarez Arenal</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Pardo Ceballos</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Fernandez Vazquez</surname>
<given-names>JP</given-names>
</name>
<name>
<surname>Ibaseta Diaz</surname>
<given-names>G</given-names>
</name>
</person-group>
<article-title>Fabrication of an implant radiologic-surgical stent for the partially edentulous patient</article-title>
<source>Quintessence Int</source>
<year>1995</year>
<volume>26</volume>
<fpage>111</fpage>
<lpage>114</lpage>
<pub-id pub-id-type="pmid">7568720</pub-id>
</element-citation>
</ref>
<ref id="B4">
<label>4</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Basten</surname>
<given-names>CH</given-names>
</name>
</person-group>
<article-title>The use of radiopaque templates for predictable implant placement</article-title>
<source>Quintessence Int</source>
<year>1995</year>
<volume>26</volume>
<fpage>609</fpage>
<lpage>612</lpage>
<pub-id pub-id-type="pmid">8602440</pub-id>
</element-citation>
</ref>
<ref id="B5">
<label>5</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Borrow</surname>
<given-names>JW</given-names>
</name>
<name>
<surname>Smith</surname>
<given-names>JP</given-names>
</name>
</person-group>
<article-title>Stent marker materials for computerized tomograph-assisted implant planning</article-title>
<source>Int J Periodontics Restorative Dent</source>
<year>1996</year>
<volume>16</volume>
<fpage>60</fpage>
<lpage>67</lpage>
<pub-id pub-id-type="pmid">8631612</pub-id>
</element-citation>
</ref>
<ref id="B6">
<label>6</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Basten</surname>
<given-names>CH</given-names>
</name>
<name>
<surname>Kois</surname>
<given-names>JC</given-names>
</name>
</person-group>
<article-title>The use of barium sulfate for implant templates</article-title>
<source>J Prosthet Dent</source>
<year>1996</year>
<volume>76</volume>
<fpage>451</fpage>
<lpage>454</lpage>
<pub-id pub-id-type="pmid">8897306</pub-id>
</element-citation>
</ref>
<ref id="B7">
<label>7</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Weinberg</surname>
<given-names>LA</given-names>
</name>
<name>
<surname>Kruger</surname>
<given-names>B</given-names>
</name>
</person-group>
<article-title>Three-dimensional guidance system for implant insertion: Part I</article-title>
<source>Implant Dent</source>
<year>1998</year>
<volume>7</volume>
<fpage>81</fpage>
<lpage>93</lpage>
<pub-id pub-id-type="pmid">9743637</pub-id>
</element-citation>
</ref>
<ref id="B8">
<label>8</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Weinberg</surname>
<given-names>LA</given-names>
</name>
<name>
<surname>Kruger</surname>
<given-names>B</given-names>
</name>
</person-group>
<article-title>Three-dimensional guidance system for implant insertion: Part II. Dual axes table-problem solving</article-title>
<source>Implant Dent</source>
<year>1999</year>
<volume>8</volume>
<fpage>255</fpage>
<lpage>264</lpage>
<pub-id pub-id-type="pmid">10709472</pub-id>
</element-citation>
</ref>
<ref id="B9">
<label>9</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Becker</surname>
<given-names>CM</given-names>
</name>
<name>
<surname>Kaiser</surname>
<given-names>DA</given-names>
</name>
</person-group>
<article-title>Surgical guide for dental implant placement</article-title>
<source>J Prosthet Dent</source>
<year>2000</year>
<volume>83</volume>
<fpage>248</fpage>
<lpage>251</lpage>
<pub-id pub-id-type="pmid">10668038</pub-id>
</element-citation>
</ref>
<ref id="B10">
<label>10</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Cucchiara</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Franchini</surname>
<given-names>F</given-names>
</name>
<name>
<surname>Lamma</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Lamma</surname>
<given-names>E</given-names>
</name>
<name>
<surname>Sansoni</surname>
<given-names>T</given-names>
</name>
<name>
<surname>Sarti</surname>
<given-names>E</given-names>
</name>
</person-group>
<article-title>Enhancing implant surgery planning via computerized image processing</article-title>
<source>Int J Comput Dent</source>
<year>2001</year>
<volume>4</volume>
<fpage>9</fpage>
<lpage>24</lpage>
<pub-id pub-id-type="pmid">11441568</pub-id>
</element-citation>
</ref>
<ref id="B11">
<label>11</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Cucchiara</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Lamma</surname>
<given-names>E</given-names>
</name>
<name>
<surname>Sansoni</surname>
<given-names>T</given-names>
</name>
</person-group>
<article-title>An image analysis approach for automatically re-orienteering CT images for dental implants</article-title>
<source>Comput Med Imaging Graph</source>
<year>2004</year>
<volume>28</volume>
<fpage>185</fpage>
<lpage>201</lpage>
<pub-id pub-id-type="pmid">15121208</pub-id>
</element-citation>
</ref>
<ref id="B12">
<label>12</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Owings</surname>
<given-names>JR</given-names>
<suffix>Jr</suffix>
</name>
</person-group>
<article-title>Virtual imaging guiding implant surgery</article-title>
<source>Compend Contin Educ Dent</source>
<year>2003</year>
<volume>24</volume>
<fpage>333</fpage>
<lpage>344</lpage>
<pub-id pub-id-type="pmid">12793218</pub-id>
</element-citation>
</ref>
<ref id="B13">
<label>13</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Marchack</surname>
<given-names>CB</given-names>
</name>
</person-group>
<article-title>An immediately loaded CAD/CAM-guided definitive prosthesis: a clinical report</article-title>
<source>J Prosthet Dent</source>
<year>2005</year>
<volume>93</volume>
<fpage>8</fpage>
<lpage>12</lpage>
<pub-id pub-id-type="pmid">15623990</pub-id>
</element-citation>
</ref>
<ref id="B14">
<label>14</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Mupparapu</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Singer</surname>
<given-names>SR</given-names>
</name>
</person-group>
<article-title>Implant imaging for the dentist</article-title>
<source>J Can Dent Assoc</source>
<year>2004</year>
<volume>70</volume>
<fpage>32</fpage>
<pub-id pub-id-type="pmid">14709253</pub-id>
</element-citation>
</ref>
<ref id="B15">
<label>15</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Tardieu</surname>
<given-names>PB</given-names>
</name>
<name>
<surname>Vrielinck</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Escolano</surname>
<given-names>E</given-names>
</name>
</person-group>
<article-title>Computer-assisted implant placement. A case report: treatment of the mandible</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2003</year>
<volume>18</volume>
<fpage>599</fpage>
<lpage>604</lpage>
<pub-id pub-id-type="pmid">12939016</pub-id>
</element-citation>
</ref>
<ref id="B16">
<label>16</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Widmann</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Zangerl</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Keiler</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Stoffner</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Bale</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Puelacher</surname>
<given-names>W</given-names>
</name>
</person-group>
<article-title>Flapless implant surgery in the edentulous jaw based on three fixed intoral reference points and image-guided surgical templates: accuracy in human cadavers</article-title>
<source>Clin Oral Implants Res</source>
<year>2010</year>
<volume>21</volume>
<fpage>835</fpage>
<lpage>841</lpage>
<pub-id pub-id-type="pmid">20345381</pub-id>
</element-citation>
</ref>
<ref id="B17">
<label>17</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Widmann</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Keiler</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Zangerl</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Stoffner</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Longato</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Bale</surname>
<given-names>R</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Computer-assisted surgery in the edentulous jaw based on 3 fixed intraoral reference points</article-title>
<source>J Oral Maxillofac Surg</source>
<year>2010</year>
<volume>68</volume>
<fpage>1140</fpage>
<lpage>1147</lpage>
<pub-id pub-id-type="pmid">20156663</pub-id>
</element-citation>
</ref>
<ref id="B18">
<label>18</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Pongrácz</surname>
<given-names>F</given-names>
</name>
<name>
<surname>Bárdosi</surname>
<given-names>Z</given-names>
</name>
<name>
<surname>Szabó</surname>
<given-names>L</given-names>
</name>
</person-group>
<article-title>Dentition planning for image-guided implantology</article-title>
<source>Int Congr Ser</source>
<year>2004</year>
<volume>1268</volume>
<fpage>1168</fpage>
<lpage>1173</lpage>
</element-citation>
</ref>
<ref id="B19">
<label>19</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Koulechov</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Lueth</surname>
<given-names>T</given-names>
</name>
</person-group>
<article-title>A new metric for drill location for navigated control in navigated dental implantology</article-title>
<source>Int Congr Ser</source>
<year>2004</year>
<volume>1268</volume>
<fpage>1220</fpage>
<lpage>1225</lpage>
</element-citation>
</ref>
<ref id="B20">
<label>20</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Parel</surname>
<given-names>SM</given-names>
</name>
<name>
<surname>Triplett</surname>
<given-names>RG</given-names>
</name>
</person-group>
<article-title>Interactive imaging for implant planning, placement, and prosthesis construction</article-title>
<source>J Oral Maxillofac Surg</source>
<year>2004</year>
<volume>62</volume>
<issue>9 suppl 2</issue>
<fpage>41</fpage>
<lpage>47</lpage>
<pub-id pub-id-type="pmid">15332180</pub-id>
</element-citation>
</ref>
<ref id="B21">
<label>21</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Besimo</surname>
<given-names>CE</given-names>
</name>
<name>
<surname>Lambrecht</surname>
<given-names>JT</given-names>
</name>
<name>
<surname>Guindy</surname>
<given-names>JS</given-names>
</name>
</person-group>
<article-title>Accuracy of implant treatment planning utilizing template-guided reformatted computed tomography</article-title>
<source>Dentomaxillofac Radiol</source>
<year>2000</year>
<volume>29</volume>
<fpage>46</fpage>
<lpage>51</lpage>
<pub-id pub-id-type="pmid">10654036</pub-id>
</element-citation>
</ref>
<ref id="B22">
<label>22</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Fortin</surname>
<given-names>T</given-names>
</name>
<name>
<surname>Champleboux</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Lormee</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Coudert</surname>
<given-names>JL</given-names>
</name>
</person-group>
<article-title>Precise dental implant placement in bone using surgical guides in conjunction with medical imaging techniques</article-title>
<source>J Oral Implantol</source>
<year>2000</year>
<volume>26</volume>
<fpage>300</fpage>
<lpage>303</lpage>
<pub-id pub-id-type="pmid">11831236</pub-id>
</element-citation>
</ref>
<ref id="B23">
<label>23</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Widmann</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Widmann</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Widmann</surname>
<given-names>E</given-names>
</name>
<name>
<surname>Jaschke</surname>
<given-names>W</given-names>
</name>
<name>
<surname>Bale</surname>
<given-names>RJ</given-names>
</name>
</person-group>
<article-title>In vitro accuracy of a novel registration and targeting technique for image-guided template production</article-title>
<source>Clin Oral Implants Res</source>
<year>2005</year>
<volume>16</volume>
<fpage>502</fpage>
<lpage>508</lpage>
<pub-id pub-id-type="pmid">16117777</pub-id>
</element-citation>
</ref>
<ref id="B24">
<label>24</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Brief</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Edinger</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Hassfeld</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Eggers</surname>
<given-names>G</given-names>
</name>
</person-group>
<article-title>Accuracy of image-guided implantology</article-title>
<source>Clin Oral Implants Res</source>
<year>2005</year>
<volume>16</volume>
<fpage>495</fpage>
<lpage>501</lpage>
<pub-id pub-id-type="pmid">16117776</pub-id>
</element-citation>
</ref>
<ref id="B25">
<label>25</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Wanschitz</surname>
<given-names>F</given-names>
</name>
<name>
<surname>Birkfellner</surname>
<given-names>W</given-names>
</name>
<name>
<surname>Watzinger</surname>
<given-names>F</given-names>
</name>
<name>
<surname>Schopper</surname>
<given-names>C</given-names>
</name>
<name>
<surname>Patruta</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Kainberger</surname>
<given-names>F</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Evaluation of accuracy of computer-aided intraoperative positioning of endosseous oral implants in the edentulous mandible</article-title>
<source>Clin Oral Implants Res</source>
<year>2002</year>
<volume>13</volume>
<fpage>59</fpage>
<lpage>64</lpage>
<pub-id pub-id-type="pmid">12005146</pub-id>
</element-citation>
</ref>
<ref id="B26">
<label>26</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Van Steenberghe</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Malevez</surname>
<given-names>C</given-names>
</name>
<name>
<surname>Van Cleynenbreugel</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Bou Serhal</surname>
<given-names>C</given-names>
</name>
<name>
<surname>Dhoore</surname>
<given-names>E</given-names>
</name>
<name>
<surname>Schutyser</surname>
<given-names>F</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Accuracy of drilling guides for transfer from three-dimensional CT-based planning to placement of zygoma implants in human cadavers</article-title>
<source>Clin Oral Implants Res</source>
<year>2003</year>
<volume>14</volume>
<fpage>131</fpage>
<lpage>136</lpage>
<pub-id pub-id-type="pmid">12562376</pub-id>
</element-citation>
</ref>
</ref-list>
</back>
<floats-group>
<fig id="F1" position="float">
<label>Fig. 1</label>
<caption>
<p>Plastic plate for determination of a reference plane. For radiographic recognition, 4 metal balls (2 mm in diameter) are inserted in each of the 4 corners of the plate 2 mm away from the edge of the plate.</p>
</caption>
<graphic xlink:href="isd-42-139-g001"></graphic>
</fig>
<fig id="F2" position="float">
<label>Fig. 2</label>
<caption>
<p>Implant planning procedure using V-Implant software.</p>
</caption>
<graphic xlink:href="isd-42-139-g002"></graphic>
</fig>
<fig id="F3" position="float">
<label>Fig. 3</label>
<caption>
<p>The reference XOY-plane determined on the plastic plate to calculate the relative position and angulation of the virtual implant.</p>
</caption>
<graphic xlink:href="isd-42-139-g003"></graphic>
</fig>
<fig id="F4" position="float">
<label>Fig. 4</label>
<caption>
<p>These illustrations show the procedure for the determination of the position and angle of the virtual implant. The insertion point P (x, y) is determined by the crossing point between the axis of the virtual implant and the XOY-plane, and ∠X (Xθ) and ∠Y (Yθ) are defined as the degree from the X- and Y-axis, respectively.</p>
</caption>
<graphic xlink:href="isd-42-139-g004"></graphic>
</fig>
<fig id="F5" position="float">
<label>Fig. 5</label>
<caption>
<p>At the insertion points P (x, y), ∠X(Xθ) and ∠Y(Yθ) are calculated automatically by the software.</p>
</caption>
<graphic xlink:href="isd-42-139-g005"></graphic>
</fig>
<fig id="F6" position="float">
<label>Fig. 6</label>
<caption>
<p>A. The specially designed 5-axis drill machine. B. By moving and rotating the table of the machine, a guide hole is made reflecting the determined implant axis.</p>
</caption>
<graphic xlink:href="isd-42-139-g006"></graphic>
</fig>
<fig id="F7" position="float">
<label>Fig. 7</label>
<caption>
<p>A. The guide hole is shown as the black cylindrical area on the CT image. The virtual implant is placed into the guide hole on the software. B. After confirming the virtually positioned implants to be placed exactly into the guide hole, the position and the angulation of the central axis of the implant are calculated.</p>
</caption>
<graphic xlink:href="isd-42-139-g007"></graphic>
</fig>
<table-wrap id="T1" position="float">
<label>Table 1</label>
<caption>
<p>Discrepancy of the insertion point in the X and Y coordinates and the angle for each group</p>
</caption>
<graphic xlink:href="isd-42-139-i001"></graphic>
<table-wrap-foot>
<fn>
<p>ΔX and ΔY represent the discrepancy of the entry point in the X and Y coordinates, respectively. ΔXθ and ΔYθ represent the discrepancy of the angle according to the X-axis and Y-axis, respectively. P: perpendicular, B: buccal, L: lingual, M: mesial, D: distal</p>
</fn>
</table-wrap-foot>
</table-wrap>
</floats-group>
</pmc>
</record>

Pour manipuler ce document sous Unix (Dilib)

EXPLOR_STEP=$WICRI_ROOT/Wicri/Santé/explor/EdenteV2/Data/Pmc/Corpus
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 002C89 | SxmlIndent | more

Ou

HfdSelect -h $EXPLOR_AREA/Data/Pmc/Corpus/biblio.hfd -nk 002C89 | SxmlIndent | more

Pour mettre un lien sur cette page dans le réseau Wicri

{{Explor lien
   |wiki=    Wicri/Santé
   |area=    EdenteV2
   |flux=    Pmc
   |étape=   Corpus
   |type=    RBID
   |clé=     PMC:3465755
   |texte=   In-vitro study on the accuracy of a simple-design CT-guided stent for dental implants
}}

Pour générer des pages wiki

HfdIndexSelect -h $EXPLOR_AREA/Data/Pmc/Corpus/RBID.i   -Sk "pubmed:23071963" \
       | HfdSelect -Kh $EXPLOR_AREA/Data/Pmc/Corpus/biblio.hfd   \
       | NlmPubMed2Wicri -a EdenteV2 

Wicri

This area was generated with Dilib version V0.6.32.
Data generation: Thu Nov 30 15:26:48 2017. Site generation: Tue Mar 8 16:36:20 2022