The effects of bone density and crestal cortical bone thickness on micromotion and peri-implant bone strain distribution in an immediately loaded implant: a nonlinear finite element analysis

Identifieur interne : 000731 ( Pmc/Corpus ); précédent : 000730; suivant : 000732

The effects of bone density and crestal cortical bone thickness on micromotion and peri-implant bone strain distribution in an immediately loaded implant: a nonlinear finite element analysis

Auteurs : Tsutomu Sugiura ; Kazuhiko Yamamoto ; Satoshi Horita ; Kazuhiro Murakami ; Sadami Tsutsumi ; Tadaaki Kirita

Source :

RBID : PMC:4928204

Abstract

Purpose

This study investigated the effects of bone density and crestal cortical bone thickness at the implant-placement site on micromotion (relative displacement between the implant and bone) and the peri-implant bone strain distribution under immediate-loading conditions.

Methods

A three-dimensional finite element model of the posterior mandible with an implant was constructed. Various bone parameters were simulated, including low or high cancellous bone density, low or high crestal cortical bone density, and crestal cortical bone thicknesses ranging from 0.5 to 2.5 mm. Delayed- and immediate-loading conditions were simulated. A buccolingual oblique load of 200 N was applied to the top of the abutment.

Results

The maximum extent of micromotion was approximately 100 μm in the low-density cancellous bone models, whereas it was under 30 μm in the high-density cancellous bone models. Crestal cortical bone thickness significantly affected the maximum micromotion in the low-density cancellous bone models. The minimum principal strain in the peri-implant cortical bone was affected by the density of the crestal cortical bone and cancellous bone to the same degree for both delayed and immediate loading. In the low-density cancellous bone models under immediate loading, the minimum principal strain in the peri-implant cortical bone decreased with an increase in crestal cortical bone thickness.

Conclusions

Cancellous bone density may be a critical factor for avoiding excessive micromotion in immediately loaded implants. Crestal cortical bone thickness significantly affected the maximum extent of micromotion and peri-implant bone strain in simulations of low-density cancellous bone under immediate loading.


Url:
DOI: 10.5051/jpis.2016.46.3.152
PubMed: 27382504
PubMed Central: 4928204

Links to Exploration step

PMC:4928204

Le document en format XML

<record>
<TEI>
<teiHeader>
<fileDesc>
<titleStmt>
<title xml:lang="en">The effects of bone density and crestal cortical bone thickness on micromotion and peri-implant bone strain distribution in an immediately loaded implant: a nonlinear finite element analysis</title>
<author>
<name sortKey="Sugiura, Tsutomu" sort="Sugiura, Tsutomu" uniqKey="Sugiura T" first="Tsutomu" last="Sugiura">Tsutomu Sugiura</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Surgery, Nara Medical University, Nara,
<country>Japan</country>
.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Yamamoto, Kazuhiko" sort="Yamamoto, Kazuhiko" uniqKey="Yamamoto K" first="Kazuhiko" last="Yamamoto">Kazuhiko Yamamoto</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Surgery, Nara Medical University, Nara,
<country>Japan</country>
.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Horita, Satoshi" sort="Horita, Satoshi" uniqKey="Horita S" first="Satoshi" last="Horita">Satoshi Horita</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Surgery, Nara Medical University, Nara,
<country>Japan</country>
.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Murakami, Kazuhiro" sort="Murakami, Kazuhiro" uniqKey="Murakami K" first="Kazuhiro" last="Murakami">Kazuhiro Murakami</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Surgery, Nara Medical University, Nara,
<country>Japan</country>
.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Tsutsumi, Sadami" sort="Tsutsumi, Sadami" uniqKey="Tsutsumi S" first="Sadami" last="Tsutsumi">Sadami Tsutsumi</name>
<affiliation>
<nlm:aff id="A2">Applied Electronics Laboratory, Kanazawa Institute of Technology, Tokyo,
<country>Japan</country>
.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Kirita, Tadaaki" sort="Kirita, Tadaaki" uniqKey="Kirita T" first="Tadaaki" last="Kirita">Tadaaki Kirita</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Surgery, Nara Medical University, Nara,
<country>Japan</country>
.</nlm:aff>
</affiliation>
</author>
</titleStmt>
<publicationStmt>
<idno type="wicri:source">PMC</idno>
<idno type="pmid">27382504</idno>
<idno type="pmc">4928204</idno>
<idno type="url">http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4928204</idno>
<idno type="RBID">PMC:4928204</idno>
<idno type="doi">10.5051/jpis.2016.46.3.152</idno>
<date when="2016">2016</date>
<idno type="wicri:Area/Pmc/Corpus">000731</idno>
<idno type="wicri:explorRef" wicri:stream="Pmc" wicri:step="Corpus" wicri:corpus="PMC">000731</idno>
</publicationStmt>
<sourceDesc>
<biblStruct>
<analytic>
<title xml:lang="en" level="a" type="main">The effects of bone density and crestal cortical bone thickness on micromotion and peri-implant bone strain distribution in an immediately loaded implant: a nonlinear finite element analysis</title>
<author>
<name sortKey="Sugiura, Tsutomu" sort="Sugiura, Tsutomu" uniqKey="Sugiura T" first="Tsutomu" last="Sugiura">Tsutomu Sugiura</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Surgery, Nara Medical University, Nara,
<country>Japan</country>
.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Yamamoto, Kazuhiko" sort="Yamamoto, Kazuhiko" uniqKey="Yamamoto K" first="Kazuhiko" last="Yamamoto">Kazuhiko Yamamoto</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Surgery, Nara Medical University, Nara,
<country>Japan</country>
.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Horita, Satoshi" sort="Horita, Satoshi" uniqKey="Horita S" first="Satoshi" last="Horita">Satoshi Horita</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Surgery, Nara Medical University, Nara,
<country>Japan</country>
.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Murakami, Kazuhiro" sort="Murakami, Kazuhiro" uniqKey="Murakami K" first="Kazuhiro" last="Murakami">Kazuhiro Murakami</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Surgery, Nara Medical University, Nara,
<country>Japan</country>
.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Tsutsumi, Sadami" sort="Tsutsumi, Sadami" uniqKey="Tsutsumi S" first="Sadami" last="Tsutsumi">Sadami Tsutsumi</name>
<affiliation>
<nlm:aff id="A2">Applied Electronics Laboratory, Kanazawa Institute of Technology, Tokyo,
<country>Japan</country>
.</nlm:aff>
</affiliation>
</author>
<author>
<name sortKey="Kirita, Tadaaki" sort="Kirita, Tadaaki" uniqKey="Kirita T" first="Tadaaki" last="Kirita">Tadaaki Kirita</name>
<affiliation>
<nlm:aff id="A1">Department of Oral and Maxillofacial Surgery, Nara Medical University, Nara,
<country>Japan</country>
.</nlm:aff>
</affiliation>
</author>
</analytic>
<series>
<title level="j">Journal of Periodontal & Implant Science</title>
<idno type="ISSN">2093-2278</idno>
<idno type="eISSN">2093-2286</idno>
<imprint>
<date when="2016">2016</date>
</imprint>
</series>
</biblStruct>
</sourceDesc>
</fileDesc>
<profileDesc>
<textClass></textClass>
</profileDesc>
</teiHeader>
<front>
<div type="abstract" xml:lang="en">
<sec>
<title>Purpose</title>
<p>This study investigated the effects of bone density and crestal cortical bone thickness at the implant-placement site on micromotion (relative displacement between the implant and bone) and the peri-implant bone strain distribution under immediate-loading conditions.</p>
</sec>
<sec>
<title>Methods</title>
<p>A three-dimensional finite element model of the posterior mandible with an implant was constructed. Various bone parameters were simulated, including low or high cancellous bone density, low or high crestal cortical bone density, and crestal cortical bone thicknesses ranging from 0.5 to 2.5 mm. Delayed- and immediate-loading conditions were simulated. A buccolingual oblique load of 200 N was applied to the top of the abutment.</p>
</sec>
<sec>
<title>Results</title>
<p>The maximum extent of micromotion was approximately 100 μm in the low-density cancellous bone models, whereas it was under 30 μm in the high-density cancellous bone models. Crestal cortical bone thickness significantly affected the maximum micromotion in the low-density cancellous bone models. The minimum principal strain in the peri-implant cortical bone was affected by the density of the crestal cortical bone and cancellous bone to the same degree for both delayed and immediate loading. In the low-density cancellous bone models under immediate loading, the minimum principal strain in the peri-implant cortical bone decreased with an increase in crestal cortical bone thickness.</p>
</sec>
<sec>
<title>Conclusions</title>
<p>Cancellous bone density may be a critical factor for avoiding excessive micromotion in immediately loaded implants. Crestal cortical bone thickness significantly affected the maximum extent of micromotion and peri-implant bone strain in simulations of low-density cancellous bone under immediate loading.</p>
</sec>
</div>
</front>
<back>
<div1 type="bibliography">
<listBibl>
<biblStruct>
<analytic>
<author>
<name sortKey="Misch, Ce" uniqKey="Misch C">CE Misch</name>
</author>
<author>
<name sortKey="Wang, Hl" uniqKey="Wang H">HL Wang</name>
</author>
<author>
<name sortKey="Misch, Cm" uniqKey="Misch C">CM Misch</name>
</author>
<author>
<name sortKey="Sharawy, M" uniqKey="Sharawy M">M Sharawy</name>
</author>
<author>
<name sortKey="Lemons, J" uniqKey="Lemons J">J Lemons</name>
</author>
<author>
<name sortKey="Judy, Kw" uniqKey="Judy K">KW Judy</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Laviv, A" uniqKey="Laviv A">A Laviv</name>
</author>
<author>
<name sortKey="Levin, L" uniqKey="Levin L">L Levin</name>
</author>
<author>
<name sortKey="Usiel, Y" uniqKey="Usiel Y">Y Usiel</name>
</author>
<author>
<name sortKey="Schwartz Arad, D" uniqKey="Schwartz Arad D">D Schwartz-Arad</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Pessoa, Rs" uniqKey="Pessoa R">RS Pessoa</name>
</author>
<author>
<name sortKey="Coelho, Pg" uniqKey="Coelho P">PG Coelho</name>
</author>
<author>
<name sortKey="Muraru, L" uniqKey="Muraru L">L Muraru</name>
</author>
<author>
<name sortKey="Marcantonio, E" uniqKey="Marcantonio E">E Marcantonio</name>
</author>
<author>
<name sortKey="Vaz, Lg" uniqKey="Vaz L">LG Vaz</name>
</author>
<author>
<name sortKey="Vander Sloten, J" uniqKey="Vander Sloten J">J Vander Sloten</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Brunski, Jb" uniqKey="Brunski J">JB Brunski</name>
</author>
<author>
<name sortKey="Puleo, Da" uniqKey="Puleo D">DA Puleo</name>
</author>
<author>
<name sortKey="Nanci, A" uniqKey="Nanci A">A Nanci</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Duyck, J" uniqKey="Duyck J">J Duyck</name>
</author>
<author>
<name sortKey="R Nold, Hj" uniqKey="R Nold H">HJ Rønold</name>
</author>
<author>
<name sortKey="Van Oosterwyck, H" uniqKey="Van Oosterwyck H">H Van Oosterwyck</name>
</author>
<author>
<name sortKey="Naert, I" uniqKey="Naert I">I Naert</name>
</author>
<author>
<name sortKey="Vander Sloten, J" uniqKey="Vander Sloten J">J Vander Sloten</name>
</author>
<author>
<name sortKey="Ellingsen, Je" uniqKey="Ellingsen J">JE Ellingsen</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Pilliar, Rm" uniqKey="Pilliar R">RM Pilliar</name>
</author>
<author>
<name sortKey="Lee, Jm" uniqKey="Lee J">JM Lee</name>
</author>
<author>
<name sortKey="Maniatopoulos, C" uniqKey="Maniatopoulos C">C Maniatopoulos</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Tabassum, A" uniqKey="Tabassum A">A Tabassum</name>
</author>
<author>
<name sortKey="Meijer, Gj" uniqKey="Meijer G">GJ Meijer</name>
</author>
<author>
<name sortKey="Wolke, Jg" uniqKey="Wolke J">JG Wolke</name>
</author>
<author>
<name sortKey="Jansen, Ja" uniqKey="Jansen J">JA Jansen</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Bayarchimeg, D" uniqKey="Bayarchimeg D">D Bayarchimeg</name>
</author>
<author>
<name sortKey="Namgoong, H" uniqKey="Namgoong H">H Namgoong</name>
</author>
<author>
<name sortKey="Kim, Bk" uniqKey="Kim B">BK Kim</name>
</author>
<author>
<name sortKey="Kim, Md" uniqKey="Kim M">MD Kim</name>
</author>
<author>
<name sortKey="Kim, S" uniqKey="Kim S">S Kim</name>
</author>
<author>
<name sortKey="Kim, Ti" uniqKey="Kim T">TI Kim</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Ikumi, N" uniqKey="Ikumi N">N Ikumi</name>
</author>
<author>
<name sortKey="Tsutsumi, S" uniqKey="Tsutsumi S">S Tsutsumi</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Roze, J" uniqKey="Roze J">J Rozé</name>
</author>
<author>
<name sortKey="Babu, S" uniqKey="Babu S">S Babu</name>
</author>
<author>
<name sortKey="Saffarzadeh, A" uniqKey="Saffarzadeh A">A Saffarzadeh</name>
</author>
<author>
<name sortKey="Gayet Delacroix, M" uniqKey="Gayet Delacroix M">M Gayet-Delacroix</name>
</author>
<author>
<name sortKey="Hoornaert, A" uniqKey="Hoornaert A">A Hoornaert</name>
</author>
<author>
<name sortKey="Layrolle, P" uniqKey="Layrolle P">P Layrolle</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Hsu, Jt" uniqKey="Hsu J">JT Hsu</name>
</author>
<author>
<name sortKey="Fuh, Lj" uniqKey="Fuh L">LJ Fuh</name>
</author>
<author>
<name sortKey="Tu, Mg" uniqKey="Tu M">MG Tu</name>
</author>
<author>
<name sortKey="Li, Yf" uniqKey="Li Y">YF Li</name>
</author>
<author>
<name sortKey="Chen, Kt" uniqKey="Chen K">KT Chen</name>
</author>
<author>
<name sortKey="Huang, Hl" uniqKey="Huang H">HL Huang</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Miyamoto, I" uniqKey="Miyamoto I">I Miyamoto</name>
</author>
<author>
<name sortKey="Tsuboi, Y" uniqKey="Tsuboi Y">Y Tsuboi</name>
</author>
<author>
<name sortKey="Wada, E" uniqKey="Wada E">E Wada</name>
</author>
<author>
<name sortKey="Suwa, H" uniqKey="Suwa H">H Suwa</name>
</author>
<author>
<name sortKey="Iizuka, T" uniqKey="Iizuka T">T Iizuka</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Bardyn, T" uniqKey="Bardyn T">T Bardyn</name>
</author>
<author>
<name sortKey="Gedet, P" uniqKey="Gedet P">P Gédet</name>
</author>
<author>
<name sortKey="Hallermann, W" uniqKey="Hallermann W">W Hallermann</name>
</author>
<author>
<name sortKey="Buchler, P" uniqKey="Buchler P">P Büchler</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Nkenke, E" uniqKey="Nkenke E">E Nkenke</name>
</author>
<author>
<name sortKey="Hahn, M" uniqKey="Hahn M">M Hahn</name>
</author>
<author>
<name sortKey="Weinzierl, K" uniqKey="Weinzierl K">K Weinzierl</name>
</author>
<author>
<name sortKey="Radespiel Troger, M" uniqKey="Radespiel Troger M">M Radespiel-Tröger</name>
</author>
<author>
<name sortKey="Neukam, Fw" uniqKey="Neukam F">FW Neukam</name>
</author>
<author>
<name sortKey="Engelke, K" uniqKey="Engelke K">K Engelke</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Marquezan, M" uniqKey="Marquezan M">M Marquezan</name>
</author>
<author>
<name sortKey="Lima, I" uniqKey="Lima I">I Lima</name>
</author>
<author>
<name sortKey="Lopes, Rt" uniqKey="Lopes R">RT Lopes</name>
</author>
<author>
<name sortKey="Sant Nna, Ef" uniqKey="Sant Nna E">EF Sant’Anna</name>
</author>
<author>
<name sortKey="De Souza, Mm" uniqKey="De Souza M">MM de Souza</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Palma Carri, C" uniqKey="Palma Carri C">C Palma-Carrió</name>
</author>
<author>
<name sortKey="Maestre Ferrin, L" uniqKey="Maestre Ferrin L">L Maestre-Ferrín</name>
</author>
<author>
<name sortKey="Pe Arrocha Oltra, D" uniqKey="Pe Arrocha Oltra D">D Peñarrocha-Oltra</name>
</author>
<author>
<name sortKey="Pe Arrocha Diago, Ma" uniqKey="Pe Arrocha Diago M">MA Peñarrocha-Diago</name>
</author>
<author>
<name sortKey="Pe Arrocha Diago, M" uniqKey="Pe Arrocha Diago M">M Peñarrocha-Diago</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Petrie, Cs" uniqKey="Petrie C">CS Petrie</name>
</author>
<author>
<name sortKey="Williams, Jl" uniqKey="Williams J">JL Williams</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Guan, H" uniqKey="Guan H">H Guan</name>
</author>
<author>
<name sortKey="Van Staden, R" uniqKey="Van Staden R">R van Staden</name>
</author>
<author>
<name sortKey="Loo, Yc" uniqKey="Loo Y">YC Loo</name>
</author>
<author>
<name sortKey="Johnson, N" uniqKey="Johnson N">N Johnson</name>
</author>
<author>
<name sortKey="Ivanovski, S" uniqKey="Ivanovski S">S Ivanovski</name>
</author>
<author>
<name sortKey="Meredith, N" uniqKey="Meredith N">N Meredith</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Sugiura, T" uniqKey="Sugiura T">T Sugiura</name>
</author>
<author>
<name sortKey="Yamamoto, K" uniqKey="Yamamoto K">K Yamamoto</name>
</author>
<author>
<name sortKey="Kawakami, M" uniqKey="Kawakami M">M Kawakami</name>
</author>
<author>
<name sortKey="Horita, S" uniqKey="Horita S">S Horita</name>
</author>
<author>
<name sortKey="Murakami, K" uniqKey="Murakami K">K Murakami</name>
</author>
<author>
<name sortKey="Kirita, T" uniqKey="Kirita T">T Kirita</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Morton, D" uniqKey="Morton D">D Morton</name>
</author>
<author>
<name sortKey="Jaffin, R" uniqKey="Jaffin R">R Jaffin</name>
</author>
<author>
<name sortKey="Weber, Hp" uniqKey="Weber H">HP Weber</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Lekholm, U" uniqKey="Lekholm U">U Lekholm</name>
</author>
<author>
<name sortKey="Zarb, Ga" uniqKey="Zarb G">GA Zarb</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Kurniawan, D" uniqKey="Kurniawan D">D Kurniawan</name>
</author>
<author>
<name sortKey="Nor, Fm" uniqKey="Nor F">FM Nor</name>
</author>
<author>
<name sortKey="Lee, Hy" uniqKey="Lee H">HY Lee</name>
</author>
<author>
<name sortKey="Lim, Jy" uniqKey="Lim J">JY Lim</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Keyak, Jh" uniqKey="Keyak J">JH Keyak</name>
</author>
<author>
<name sortKey="Rossi, Sa" uniqKey="Rossi S">SA Rossi</name>
</author>
<author>
<name sortKey="Jones, Ka" uniqKey="Jones K">KA Jones</name>
</author>
<author>
<name sortKey="Skinner, Hb" uniqKey="Skinner H">HB Skinner</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Ding, X" uniqKey="Ding X">X Ding</name>
</author>
<author>
<name sortKey="Liao, Sh" uniqKey="Liao S">SH Liao</name>
</author>
<author>
<name sortKey="Zhu, Xh" uniqKey="Zhu X">XH Zhu</name>
</author>
<author>
<name sortKey="Zhang, Xh" uniqKey="Zhang X">XH Zhang</name>
</author>
<author>
<name sortKey="Zhang, L" uniqKey="Zhang L">L Zhang</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Mericske Stern, R" uniqKey="Mericske Stern R">R Mericske-Stern</name>
</author>
<author>
<name sortKey="Assal, P" uniqKey="Assal P">P Assal</name>
</author>
<author>
<name sortKey="Mericske, E" uniqKey="Mericske E">E Mericske</name>
</author>
<author>
<name sortKey="Burgin, W" uniqKey="Burgin W">W Bürgin</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Hudieb, M" uniqKey="Hudieb M">M Hudieb</name>
</author>
<author>
<name sortKey="Wakabayashi, N" uniqKey="Wakabayashi N">N Wakabayashi</name>
</author>
<author>
<name sortKey="Suzuki, T" uniqKey="Suzuki T">T Suzuki</name>
</author>
<author>
<name sortKey="Kasugai, S" uniqKey="Kasugai S">S Kasugai</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Gonda, T" uniqKey="Gonda T">T Gonda</name>
</author>
<author>
<name sortKey="Yasuda, D" uniqKey="Yasuda D">D Yasuda</name>
</author>
<author>
<name sortKey="Ikebe, K" uniqKey="Ikebe K">K Ikebe</name>
</author>
<author>
<name sortKey="Maeda, Y" uniqKey="Maeda Y">Y Maeda</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Vandamme, K" uniqKey="Vandamme K">K Vandamme</name>
</author>
<author>
<name sortKey="Naert, I" uniqKey="Naert I">I Naert</name>
</author>
<author>
<name sortKey="Geris, L" uniqKey="Geris L">L Geris</name>
</author>
<author>
<name sortKey="Vander Sloten, J" uniqKey="Vander Sloten J">J Vander Sloten</name>
</author>
<author>
<name sortKey="Puers, R" uniqKey="Puers R">R Puers</name>
</author>
<author>
<name sortKey="Duyck, J" uniqKey="Duyck J">J Duyck</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Trisi, P" uniqKey="Trisi P">P Trisi</name>
</author>
<author>
<name sortKey="Perfetti, G" uniqKey="Perfetti G">G Perfetti</name>
</author>
<author>
<name sortKey="Baldoni, E" uniqKey="Baldoni E">E Baldoni</name>
</author>
<author>
<name sortKey="Berardi, D" uniqKey="Berardi D">D Berardi</name>
</author>
<author>
<name sortKey="Colagiovanni, M" uniqKey="Colagiovanni M">M Colagiovanni</name>
</author>
<author>
<name sortKey="Scogna, G" uniqKey="Scogna G">G Scogna</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Cha, Jy" uniqKey="Cha J">JY Cha</name>
</author>
<author>
<name sortKey="Kil, Jk" uniqKey="Kil J">JK Kil</name>
</author>
<author>
<name sortKey="Yoon, Tm" uniqKey="Yoon T">TM Yoon</name>
</author>
<author>
<name sortKey="Hwang, Cj" uniqKey="Hwang C">CJ Hwang</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Shen, Wl" uniqKey="Shen W">WL Shen</name>
</author>
<author>
<name sortKey="Chen, Cs" uniqKey="Chen C">CS Chen</name>
</author>
<author>
<name sortKey="Hsu, Ml" uniqKey="Hsu M">ML Hsu</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Frost, Hm" uniqKey="Frost H">HM Frost</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Pattin, Ca" uniqKey="Pattin C">CA Pattin</name>
</author>
<author>
<name sortKey="Caler, We" uniqKey="Caler W">WE Caler</name>
</author>
<author>
<name sortKey="Carter, Dr" uniqKey="Carter D">DR Carter</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Mellal, A" uniqKey="Mellal A">A Mellal</name>
</author>
<author>
<name sortKey="Wiskott, Hw" uniqKey="Wiskott H">HW Wiskott</name>
</author>
<author>
<name sortKey="Botsis, J" uniqKey="Botsis J">J Botsis</name>
</author>
<author>
<name sortKey="Scherrer, Ss" uniqKey="Scherrer S">SS Scherrer</name>
</author>
<author>
<name sortKey="Belser, Uc" uniqKey="Belser U">UC Belser</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Huang, Hl" uniqKey="Huang H">HL Huang</name>
</author>
<author>
<name sortKey="Fuh, Lj" uniqKey="Fuh L">LJ Fuh</name>
</author>
<author>
<name sortKey="Hsu, Jt" uniqKey="Hsu J">JT Hsu</name>
</author>
<author>
<name sortKey="Tu, Mg" uniqKey="Tu M">MG Tu</name>
</author>
<author>
<name sortKey="Shen, Yw" uniqKey="Shen Y">YW Shen</name>
</author>
<author>
<name sortKey="Wu, Cl" uniqKey="Wu C">CL Wu</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Ferreira, Mb" uniqKey="Ferreira M">MB Ferreira</name>
</author>
<author>
<name sortKey="Barao, Va" uniqKey="Barao V">VA Barão</name>
</author>
<author>
<name sortKey="Delben, Ja" uniqKey="Delben J">JA Delben</name>
</author>
<author>
<name sortKey="Faverani, Lp" uniqKey="Faverani L">LP Faverani</name>
</author>
<author>
<name sortKey="Hip Lito, Ac" uniqKey="Hip Lito A">AC Hipólito</name>
</author>
<author>
<name sortKey="Assuncao, Wg" uniqKey="Assuncao W">WG Assunção</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Duaibis, R" uniqKey="Duaibis R">R Duaibis</name>
</author>
<author>
<name sortKey="Kusnoto, B" uniqKey="Kusnoto B">B Kusnoto</name>
</author>
<author>
<name sortKey="Natarajan, R" uniqKey="Natarajan R">R Natarajan</name>
</author>
<author>
<name sortKey="Zhao, L" uniqKey="Zhao L">L Zhao</name>
</author>
<author>
<name sortKey="Evans, C" uniqKey="Evans C">C Evans</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Chiapasco, M" uniqKey="Chiapasco M">M Chiapasco</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Seker, E" uniqKey="Seker E">E Seker</name>
</author>
<author>
<name sortKey="Ulusoy, M" uniqKey="Ulusoy M">M Ulusoy</name>
</author>
<author>
<name sortKey="Ozan, O" uniqKey="Ozan O">O Ozan</name>
</author>
<author>
<name sortKey="Do An, Do" uniqKey="Do An D">DÖ Doğan</name>
</author>
<author>
<name sortKey="Seker, Bk" uniqKey="Seker B">BK Seker</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Chang, Pk" uniqKey="Chang P">PK Chang</name>
</author>
<author>
<name sortKey="Chen, Yc" uniqKey="Chen Y">YC Chen</name>
</author>
<author>
<name sortKey="Huang, Cc" uniqKey="Huang C">CC Huang</name>
</author>
<author>
<name sortKey="Lu, Wh" uniqKey="Lu W">WH Lu</name>
</author>
<author>
<name sortKey="Chen, Yc" uniqKey="Chen Y">YC Chen</name>
</author>
<author>
<name sortKey="Tsai, Hh" uniqKey="Tsai H">HH Tsai</name>
</author>
</analytic>
</biblStruct>
</listBibl>
</div1>
</back>
</TEI>
<pmc article-type="research-article">
<pmc-dir>properties open_access</pmc-dir>
<front>
<journal-meta>
<journal-id journal-id-type="nlm-ta">J Periodontal Implant Sci</journal-id>
<journal-id journal-id-type="iso-abbrev">J Periodontal Implant Sci</journal-id>
<journal-id journal-id-type="publisher-id">JPIS</journal-id>
<journal-title-group>
<journal-title>Journal of Periodontal & Implant Science</journal-title>
</journal-title-group>
<issn pub-type="ppub">2093-2278</issn>
<issn pub-type="epub">2093-2286</issn>
<publisher>
<publisher-name>Korean Academy of Periodontology</publisher-name>
</publisher>
</journal-meta>
<article-meta>
<article-id pub-id-type="pmid">27382504</article-id>
<article-id pub-id-type="pmc">4928204</article-id>
<article-id pub-id-type="doi">10.5051/jpis.2016.46.3.152</article-id>
<article-categories>
<subj-group subj-group-type="heading">
<subject>Research Article</subject>
</subj-group>
</article-categories>
<title-group>
<article-title>The effects of bone density and crestal cortical bone thickness on micromotion and peri-implant bone strain distribution in an immediately loaded implant: a nonlinear finite element analysis</article-title>
</title-group>
<contrib-group>
<contrib contrib-type="author" corresp="yes">
<contrib-id contrib-id-type="orcid" authenticated="true">http://orcid.org/0000-0001-9601-9248</contrib-id>
<name>
<surname>Sugiura</surname>
<given-names>Tsutomu</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
</contrib>
<contrib contrib-type="author">
<contrib-id contrib-id-type="orcid" authenticated="true">http://orcid.org/0000-0002-9931-889X</contrib-id>
<name>
<surname>Yamamoto</surname>
<given-names>Kazuhiko</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
</contrib>
<contrib contrib-type="author">
<contrib-id contrib-id-type="orcid" authenticated="true">http://orcid.org/0000-0002-1061-1986</contrib-id>
<name>
<surname>Horita</surname>
<given-names>Satoshi</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
</contrib>
<contrib contrib-type="author">
<contrib-id contrib-id-type="orcid" authenticated="true">http://orcid.org/0000-0002-2144-5020</contrib-id>
<name>
<surname>Murakami</surname>
<given-names>Kazuhiro</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
</contrib>
<contrib contrib-type="author">
<contrib-id contrib-id-type="orcid" authenticated="true">http://orcid.org/0000-0002-0672-5892</contrib-id>
<name>
<surname>Tsutsumi</surname>
<given-names>Sadami</given-names>
</name>
<xref ref-type="aff" rid="A2">2</xref>
</contrib>
<contrib contrib-type="author">
<contrib-id contrib-id-type="orcid" authenticated="true">http://orcid.org/0000-0002-6900-0397</contrib-id>
<name>
<surname>Kirita</surname>
<given-names>Tadaaki</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
</contrib>
</contrib-group>
<aff id="A1">
<label>1</label>
Department of Oral and Maxillofacial Surgery, Nara Medical University, Nara,
<country>Japan</country>
.</aff>
<aff id="A2">
<label>2</label>
Applied Electronics Laboratory, Kanazawa Institute of Technology, Tokyo,
<country>Japan</country>
.</aff>
<author-notes>
<corresp>Correspondence to Tsutomu Sugiura. Department of Oral and Maxillofacial Surgery, Nara Medical University, 840 Shijo-cho, Kashihara City, Nara 634-8522, Japan.
<email>sugiurat@naramed-u.ac.jp</email>
, Tel: +81-744-29-8876, Fax: +81-744-29-8876</corresp>
</author-notes>
<pub-date pub-type="ppub">
<month>6</month>
<year>2016</year>
</pub-date>
<pub-date pub-type="epub">
<day>28</day>
<month>6</month>
<year>2016</year>
</pub-date>
<volume>46</volume>
<issue>3</issue>
<fpage>152</fpage>
<lpage>165</lpage>
<history>
<date date-type="received">
<day>30</day>
<month>4</month>
<year>2016</year>
</date>
<date date-type="accepted">
<day>09</day>
<month>6</month>
<year>2016</year>
</date>
</history>
<permissions>
<copyright-statement>Copyright © 2016 Korean Academy of Periodontology</copyright-statement>
<copyright-year>2016</copyright-year>
<license license-type="open-access" xlink:href="http://creativecommons.org/licenses/by-nc/3.0/">
<license-p>This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (
<ext-link ext-link-type="uri" xlink:href="http://creativecommons.org/licenses/by-nc/3.0/">http://creativecommons.org/licenses/by-nc/3.0/</ext-link>
).</license-p>
</license>
</permissions>
<abstract>
<sec>
<title>Purpose</title>
<p>This study investigated the effects of bone density and crestal cortical bone thickness at the implant-placement site on micromotion (relative displacement between the implant and bone) and the peri-implant bone strain distribution under immediate-loading conditions.</p>
</sec>
<sec>
<title>Methods</title>
<p>A three-dimensional finite element model of the posterior mandible with an implant was constructed. Various bone parameters were simulated, including low or high cancellous bone density, low or high crestal cortical bone density, and crestal cortical bone thicknesses ranging from 0.5 to 2.5 mm. Delayed- and immediate-loading conditions were simulated. A buccolingual oblique load of 200 N was applied to the top of the abutment.</p>
</sec>
<sec>
<title>Results</title>
<p>The maximum extent of micromotion was approximately 100 μm in the low-density cancellous bone models, whereas it was under 30 μm in the high-density cancellous bone models. Crestal cortical bone thickness significantly affected the maximum micromotion in the low-density cancellous bone models. The minimum principal strain in the peri-implant cortical bone was affected by the density of the crestal cortical bone and cancellous bone to the same degree for both delayed and immediate loading. In the low-density cancellous bone models under immediate loading, the minimum principal strain in the peri-implant cortical bone decreased with an increase in crestal cortical bone thickness.</p>
</sec>
<sec>
<title>Conclusions</title>
<p>Cancellous bone density may be a critical factor for avoiding excessive micromotion in immediately loaded implants. Crestal cortical bone thickness significantly affected the maximum extent of micromotion and peri-implant bone strain in simulations of low-density cancellous bone under immediate loading.</p>
</sec>
</abstract>
<abstract abstract-type="graphical">
<title>Graphical Abstract</title>
<p>
<graphic xlink:href="jpis-46-152-abf001.jpg" position="float" orientation="portrait"></graphic>
</p>
</abstract>
<kwd-group kwd-group-type="author">
<kwd>Bone density</kwd>
<kwd>Dental implants</kwd>
<kwd>Finite element analysis</kwd>
</kwd-group>
<funding-group>
<award-group>
<funding-source country="JP">
<institution-wrap>
<institution>Japan Society for Promotion of Science</institution>
</institution-wrap>
</funding-source>
<award-id>24592963</award-id>
</award-group>
</funding-group>
</article-meta>
</front>
<body>
<sec sec-type="intro">
<title>INTRODUCTION</title>
<p>Osseointegrated dental implants are widely used for functional and aesthetic rehabilitation. Conventional implants are loaded after a long period of healing. Although this approach has proven to be highly predictable and successful, the extended treatment period may be perceived as a considerable inconvenience by patients desiring rapid rehabilitation. Immediate loading has been proposed to reduce the cost and duration of the implant treatment [
<xref rid="B1" ref-type="bibr">1</xref>
], and it has shown a clinical success rate of over 95% [
<xref rid="B2" ref-type="bibr">2</xref>
].</p>
<p>Despite reports of promising results in experimental and clinical studies of immediate-loading protocols, failures can still occur and have been suggested to arise from biomechanical factors [
<xref rid="B3" ref-type="bibr">3</xref>
]. For example, excessive micromotion may lead to encapsulation of the implants, resulting in the failure of osseointegration between the bone and implant. Even for implants that have already become osseointegrated, the accumulation of excessive stress and strain can cause microdamage to accumulate, inducing bone resorption [
<xref rid="B4" ref-type="bibr">4</xref>
<xref rid="B5" ref-type="bibr">5</xref>
].</p>
<p>Primary stability has been characterized as one of the most important variables affecting the success of immediately loaded implants. Excessive micromotion (relative displacement between the implant and bone) may cause osseointegration to fail between the bone and implant [
<xref rid="B4" ref-type="bibr">4</xref>
<xref rid="B6" ref-type="bibr">6</xref>
]. The design of the implant, the quantity and density of bone, and the insertion torque influence primary stability. Studies have reported correlations of bone quantity and bone density with primary implant stability using measurements of insertion torque, removal torque, cutting torque, Periotest values (PTV), and the implant stability quotient (ISQ) derived from resonance frequency analysis [
<xref rid="B7" ref-type="bibr">7</xref>
<xref rid="B8" ref-type="bibr">8</xref>
<xref rid="B9" ref-type="bibr">9</xref>
<xref rid="B10" ref-type="bibr">10</xref>
<xref rid="B11" ref-type="bibr">11</xref>
<xref rid="B12" ref-type="bibr">12</xref>
]. However, every bone parameter, including cancellous bone density, cortical bone density, and cortical bone thickness, has an influence on primary stability. Moreover, most of the above measuring techniques are either low-sensitivity or show somewhat questionable correlations with other techniques [
<xref rid="B13" ref-type="bibr">13</xref>
<xref rid="B14" ref-type="bibr">14</xref>
]. Thus, the relative importance of bone parameters remains controversial [
<xref rid="B10" ref-type="bibr">10</xref>
<xref rid="B13" ref-type="bibr">13</xref>
<xref rid="B15" ref-type="bibr">15</xref>
].</p>
<p>In addition, a key factor for the success of both delayed-loading and immediately loaded implants is the manner in which the stresses and strains are transmitted to the surrounding bone. Bone density is important for the success rate of implants, because it influences the transmission of loads to implants and bone, and a high failure rate for delayed-loading implants has been reported in areas with low bone density [
<xref rid="B16" ref-type="bibr">16</xref>
]. Therefore, the effects of bone quantity and density on the peri-implant distribution of stress and strain have been investigated under delayed-loading situations [
<xref rid="B17" ref-type="bibr">17</xref>
<xref rid="B18" ref-type="bibr">18</xref>
<xref rid="B19" ref-type="bibr">19</xref>
]. Poor bone quantity and density also have been identified as risk factors with implications for the immediate loading of implants [
<xref rid="B20" ref-type="bibr">20</xref>
]. However, the effects of bone quantity and density on biomechanical behavior have not been yet clarified quantitatively for immediately loaded implants.</p>
<p>Clinically, it is impossible to introduce any device into the bone-implant interface to investigate the level of micromotion between the bone and implant under masticatory forces. Finite element analysis (FEA) is an efficient technique of evaluating both the extent of micromotion and the distribution of peri-implant bone strain. The purpose of this study was to investigate the effects of bone density at the implant-placement site on micromotion and the peri-implant bone strain distribution under immediate-loading conditions.</p>
</sec>
<sec sec-type="materials|methods">
<title>MATERIALS AND METHODS</title>
<sec>
<title>Finite element models</title>
<p>The three-dimensional (3D) geometry of the right posterior mandible was obtained from the computed tomography (CT) scan data of a 62-year-old male, using FEA software (Mechanical Finder, Version 6.2, Research Center of Computational Mechanics, Tokyo, Japan). The cortical bone comprised the crestal, lateral, and inferior cortical bones. The cortical bone of the buccal, lingual, and inferior border of the mandible was reconstructed with a thickness of 2.5 mm [
<xref rid="B19" ref-type="bibr">19</xref>
]. Crestal cortical bone thickness at the implant placement site of the mandible ranges from approximately 0.4 mm to 2.8 mm [
<xref rid="B10" ref-type="bibr">10</xref>
<xref rid="B12" ref-type="bibr">12</xref>
<xref rid="B19" ref-type="bibr">19</xref>
]. In previous FEA studies simulating bone types based on the Lekholm and Zarb classification [
<xref rid="B21" ref-type="bibr">21</xref>
], crestal cortical bone thicknesses of 1 mm and 2 mm have been characterized as thin and thick, respectively [
<xref rid="B22" ref-type="bibr">22</xref>
]. Based on previous biomechanical studies [
<xref rid="B8" ref-type="bibr">8</xref>
<xref rid="B13" ref-type="bibr">13</xref>
], crestal cortical bone thicknesses of 0.5 mm, 1.0 mm, 2.0 mm, and 2.5 mm were simulated in the present study.</p>
<p>A Straumann threaded implant (Institut Straumann, Waldenburg, Switzerland) with a diameter of 4.1 mm and a length of 10 mm was simulated in this study. An implant and a 6-mm abutment were modeled as one piece using 3D modeling software and exported into the FEA software to complete the models (
<xref ref-type="fig" rid="F1">Figure 1</xref>
). In the study, two conditions of the implant-bone interface (a bonded interface and a contact interface) were constructed. For delayed-loading implant models, the implant was assumed to achieve complete osseointegration at the implant-bone interface. For immediately loaded implant models, contact interfaces (non-osseointegration) between the implant and bone were simulated. The friction coefficient was set to 0.3 [
<xref rid="B3" ref-type="bibr">3</xref>
]. The material properties were assumed to be homogeneous, isotropic, and linearly elastic. The finite element model was constructed with 4-node tetrahedral elements, and had approximately 102,000 elements and 19,000 nodes.</p>
<fig id="F1" fig-type="figure" orientation="portrait" position="float">
<label>Figure 1</label>
<caption>
<p>Finite element model. (A) The right posterior mandible model consisted of cancellous bone and cortical bone. The cortical bone comprised the crestal cortical bone and the buccal, lingual, and inferior borders of the mandible. The three independent parameters included: (1) the density of cancellous bone, (2) the density of crestal cortical bone, and (3) the thickness of crestal cortical bone. (B) Implant and abutment.</p>
</caption>
<graphic xlink:href="jpis-46-152-g001"></graphic>
</fig>
</sec>
<sec>
<title>Definition of bone density and material properties</title>
<p>A detailed description of the measurement and definition of bone density has been published earlier [
<xref rid="B19" ref-type="bibr">19</xref>
]. The bone densities of 75 potential implant sites in the posterior mandible of 34 patients were measured using a spiral CT machine. The results of Kolmogorov-Smirnov tests showed that the densities of the cancellous bone and crestal cortical bone had normal statistical distributions. We defined the 5th and 95th percentiles of bone density (150 Hounsfield units [HU] and 850 HU for the cancellous bone, and 950 HU and 1750 HU for the crestal cortical bone) as low and high, respectively. Since the cortical bone densities of the buccal, lingual, and inferior border of the mandible showed similar values, the mean value (1765 HU) was used as the bone density of these areas. A linear regression equation was created based on the CT values of the calibration phantom. Using these calibrated CT data, each CT value was converted to apparent bone density expressed in g/cm
<sup>3</sup>
. Young’s modulus for apparent bone density was calculated using the equations proposed by Keyak (
<xref ref-type="table" rid="T1">Table 1</xref>
) [
<xref rid="B23" ref-type="bibr">23</xref>
].</p>
<table-wrap id="T1" orientation="portrait" position="float">
<label>Table 1</label>
<caption>
<title>Bone density and material properties</title>
</caption>
<alternatives>
<graphic xlink:href="jpis-46-152-i001"></graphic>
<table frame="hsides" rules="rows">
<col width="2.41%" span="1"></col>
<col width="2.41%" span="1"></col>
<col width="26.31%" span="1"></col>
<col width="15.26%" span="1"></col>
<col width="20.08%" span="1"></col>
<col width="18.27%" span="1"></col>
<col width="15.26%" span="1"></col>
<thead>
<tr>
<th valign="top" align="left" rowspan="1" colspan="3" style="background-color:rgb(227,240,212)">Material</th>
<th valign="top" align="center" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">CT value (HU)</th>
<th valign="top" align="center" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">Bone density (g/cm
<sup>3</sup>
)</th>
<th valign="top" align="left" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">Young's modulus (GPa)</th>
<th valign="top" align="center" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">Poisson's ratio</th>
</tr>
</thead>
<tbody>
<tr>
<td valign="top" align="left" rowspan="1" colspan="3">Titanium</td>
<td valign="top" align="right" rowspan="1" colspan="1"></td>
<td valign="top" align="right" rowspan="1" colspan="1"></td>
<td valign="top" align="right" rowspan="1" colspan="1">110</td>
<td valign="top" align="right" rowspan="1" colspan="1">0.35</td>
</tr>
<tr>
<td valign="top" align="left" rowspan="1" colspan="3" style="background-color:rgb(227,240,212)">Cortical bone</td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
</tr>
<tr>
<td valign="top" align="left" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
<td valign="top" align="left" rowspan="1" colspan="2" style="background-color:rgb(227,240,212)">Alveolar crest</td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
</tr>
<tr>
<td valign="top" align="left" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
<td valign="top" align="center" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
<td valign="top" align="left" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">Low density</td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">950</td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">0.639</td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">4.14</td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">0.30</td>
</tr>
<tr>
<td valign="top" align="left" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
<td valign="top" align="center" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
<td valign="top" align="left" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">High density</td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">1,750</td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">1.168</td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">13.94</td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">0.30</td>
</tr>
<tr>
<td valign="top" align="left" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
<td valign="top" align="left" rowspan="1" colspan="2" style="background-color:rgb(227,240,212)">Buccal, lingual, and lower borders</td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">1,765</td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">1.178</td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">14.18</td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">0.30</td>
</tr>
<tr>
<td valign="top" align="left" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
<td valign="top" align="left" rowspan="1" colspan="2" style="background-color:rgb(227,240,212)">Standard model</td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)"></td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">13.70</td>
<td valign="top" align="right" rowspan="1" colspan="1" style="background-color:rgb(227,240,212)">0.30</td>
</tr>
<tr>
<td valign="top" align="left" rowspan="1" colspan="3">Cancellous bone</td>
<td valign="top" align="right" rowspan="1" colspan="1"></td>
<td valign="top" align="right" rowspan="1" colspan="1"></td>
<td valign="top" align="right" rowspan="1" colspan="1"></td>
<td valign="top" align="right" rowspan="1" colspan="1"></td>
</tr>
<tr>
<td valign="top" align="right" rowspan="1" colspan="1"></td>
<td valign="top" align="right" rowspan="1" colspan="1"></td>
<td valign="top" align="left" rowspan="1" colspan="1">Low density</td>
<td valign="top" align="right" rowspan="1" colspan="1">150</td>
<td valign="top" align="right" rowspan="1" colspan="1">0.109</td>
<td valign="top" align="right" rowspan="1" colspan="1">0.259</td>
<td valign="top" align="right" rowspan="1" colspan="1">0.30</td>
</tr>
<tr>
<td valign="top" align="right" rowspan="1" colspan="1"></td>
<td valign="top" align="right" rowspan="1" colspan="1"></td>
<td valign="top" align="left" rowspan="1" colspan="1">High density</td>
<td valign="top" align="right" rowspan="1" colspan="1">850</td>
<td valign="top" align="right" rowspan="1" colspan="1">0.572</td>
<td valign="top" align="right" rowspan="1" colspan="1">3.507</td>
<td valign="top" align="right" rowspan="1" colspan="1">0.30</td>
</tr>
<tr>
<td valign="top" align="left" rowspan="1" colspan="1"></td>
<td valign="top" align="left" rowspan="1" colspan="2">Standard model</td>
<td valign="top" align="right" rowspan="1" colspan="1"></td>
<td valign="top" align="right" rowspan="1" colspan="1"></td>
<td valign="top" align="right" rowspan="1" colspan="1">1.37</td>
<td valign="top" align="right" rowspan="1" colspan="1">0.30</td>
</tr>
</tbody>
</table>
</alternatives>
</table-wrap>
<p>Young’s modulus for the cancellous bone and cortical bone of the mandible has been frequently assumed to be 1,370 MPa and 13,700 MPa, respectively [
<xref rid="B3" ref-type="bibr">3</xref>
<xref rid="B24" ref-type="bibr">24</xref>
]. Thus, a model having these Young’s moduli was prepared as the standard model. Poisson’s ratio of bones and the material property of the titanium implant was obtained from previous data (
<xref ref-type="table" rid="T1">Table 1</xref>
) [
<xref rid="B24" ref-type="bibr">24</xref>
]. All experimental procedures were conducted with the ethical approval of Nara Medical University.</p>
</sec>
<sec>
<title>Loads and constraints</title>
<p>With fixed prostheses supported by implants, the average maximum occlusal force was approximately 200 N for the first premolar and molars [
<xref rid="B25" ref-type="bibr">25</xref>
]. A buccolingual oblique load of 200 N was applied to the top of the abutment to simulate the average maximum occlusal load and direction on the first premolar tooth (
<xref ref-type="fig" rid="F1">Figure 1</xref>
). The loading angle was defined as 30° to the axis of the implant [
<xref rid="B26" ref-type="bibr">26</xref>
]. The boundary conditions were established by the nodes of the mesial and distal end of the model in all directions.</p>
</sec>
<sec>
<title>Analysis of micromotion and bone strain</title>
<p>Analyses were performed to calculate the micromotion at the bone-implant interface in the immediate-loading models, as well as the minimum principal strain around the implants. The micromotion was computed as the relative displacement between two nodes (one node on the bone side and one node on the implant side) of elements on the interface. The minimum principal strain (third principal strain) represented the most negative strain, typically the peak compressive strain. In order to evaluate the minimum principal strain, eight 1-mm spheres were placed at the alveolar crest around the implant in a similar manner as described in a previous report [
<xref rid="B27" ref-type="bibr">27</xref>
]. The mean value of the minimum principal strain in multiple solid elements contained in the sphere was used as the typical strain value. The absolute maximum value in these spheres was used as the peak strain value of the model.</p>
</sec>
<sec>
<title>Convergence test</title>
<p>A convergence test of the finite element models was performed to verify the mesh quality, and the convergence criterion was set to be less than 1% in changes of the total strain energy of all elements (
<xref ref-type="fig" rid="F2">Figure 2</xref>
). Based on the results of the convergence test, an average element size of 0.6 mm was set for meshing in all finite element models.</p>
<fig id="F2" fig-type="figure" orientation="portrait" position="float">
<label>Figure 2</label>
<caption>
<p>The results of convergence tests in the standard model with 1.0-mm crestal cortical bone under delayed loading.</p>
</caption>
<graphic xlink:href="jpis-46-152-g002"></graphic>
</fig>
</sec>
</sec>
<sec sec-type="results">
<title>RESULTS</title>
<sec>
<title>Micromotion</title>
<p>In the low-density cancellous bone models, the implant was displaced lingually at the neck and buccally at the apex of the implant. In the high-density cancellous bone models, the implant was less displaced at the apex than at the neck of the implant (
<xref ref-type="fig" rid="F3">Figure 3</xref>
). Consequently, the maximum micromotion was observed at the apex of the implant in the low-density cancellous bone models, whereas it was observed at the implant neck in the high-density cancellous bone models.</p>
<fig id="F3" fig-type="figure" orientation="portrait" position="float">
<label>Figure 3</label>
<caption>
<p>Displacement of the implant and abutment in the immediate-loading models with 2.0-mm crestal cortical bone. The cortical bone and the implant and abutment before deformation are also illustrated. Displacement of the implant and abutment is represented at 15× magnification in all models. B, buccal; L, lingual.</p>
</caption>
<graphic xlink:href="jpis-46-152-g003"></graphic>
</fig>
<p>The maximum micromotion was affected more by the cancellous bone density than by the crestal cortical bone density. The maximum micromotion values in the low-density cancellous bone models were 2.8-5.5-fold higher than those in the high-density cancellous bone models. The maximum micromotion values in the low-density crestal cortical bone models were 1.2-2.0-fold higher than those in the high-density crestal cortical bone models. The greatest extent of micromotion was approximately 100 μm in the low-density cancellous bone models, whereas it was less than 30 μm in the high-density cancellous bone models. Crestal cortical bone thickness greatly affected the maximum micromotion in the low-density cancellous bone models, whereas it did not affect the maximum micromotion in the high-density cancellous bone models. In the standard models, the maximum micromotion decreased by 15.4%-38.5% with increases in the crestal bone cortical thickness, but only to 26 μm (
<xref ref-type="fig" rid="F4">Figure 4</xref>
).</p>
<fig id="F4" fig-type="figure" orientation="portrait" position="float">
<label>Figure 4</label>
<caption>
<p>Maximum micromotion in the immediate-loading models.</p>
</caption>
<graphic xlink:href="jpis-46-152-g004"></graphic>
</fig>
</sec>
<sec>
<title>Strain distribution in the bone around the implant</title>
<p>The minimum principal strain concentration was located at the lingual crestal cortical bone in all models for both delayed and immediate loading (
<xref ref-type="fig" rid="F5">Figure 5</xref>
and
<xref ref-type="fig" rid="F6">6</xref>
). The peak minimum principal strain values in the immediate-loading models were 1.51-2.34-fold higher than those in the delayed-loading models (1.61-1.98-fold in the standard models). The minimum principal strain values were affected by the density of the cancellous bone and the crestal cortical bone to the same degree for both delayed and immediate loading. The peak minimum principal strain values in the low-density bone models were higher than those in the high-density bone models, by 1.35-1.84-fold in the delayed-loading model and 1.46-2.30-fold in the immediate-loading model regarding cancellous bone density, and 1.73-2.28-fold in the delayed-loading model and 1.88-2.63-fold in the immediate-loading model regarding crestal cortical bone density (
<xref ref-type="fig" rid="F7">Figure 7</xref>
).</p>
<fig id="F5" fig-type="figure" orientation="portrait" position="float">
<label>Figure 5</label>
<caption>
<p>Minimum principal strain distribution in the cortical bone in 2.0-mm crestal cortical bone models. (A) Delayed loading. (B) Immediate loading. B, buccal; L, lingual.</p>
</caption>
<graphic xlink:href="jpis-46-152-g005"></graphic>
</fig>
<fig id="F6" fig-type="figure" orientation="portrait" position="float">
<label>Figure 6</label>
<caption>
<p>The effect of crestal cortical bone thickness on the minimum principal strain distribution in the cortical bone in the standard model under immediate loading. (A) 0.5 mm of thickness. (B) 1.0 mm of thickness. (C) 2.0 mm of thickness. (D) 2.5 mm of thickness. Similar trends in the minimum principal strain distributions were observed in the delayed-loading models, although the strain levels were different (data not shown).</p>
</caption>
<graphic xlink:href="jpis-46-152-g006"></graphic>
</fig>
<fig id="F7" fig-type="figure" orientation="portrait" position="float">
<label>Figure 7</label>
<caption>
<p>The peak values of minimum principal strains in the peri-implant cortical bone.</p>
</caption>
<graphic xlink:href="jpis-46-152-g007"></graphic>
</fig>
<p>In the low-density cancellous bone models, the minimum principal strain decreased with an increase in the crestal cortical bone thickness. This tendency was particularly notable under immediate loading in comparison with delayed loading. In contrast, in the high-density cancellous bone models, the peak minimum principal strain values did not significantly depend upon crestal cortical bone thickness (
<xref ref-type="fig" rid="F7">Figure 7</xref>
).</p>
<p>In the standard model, the peak minimum principal strain decreased by 9.4%-26.0% with increases in crestal cortical bone thickness under immediate loading, whereas it was not affected by crestal cortical bone thickness under delayed loading (
<xref ref-type="fig" rid="F6">Figure 6</xref>
and
<xref ref-type="fig" rid="F7">7</xref>
).</p>
</sec>
</sec>
<sec sec-type="discussion">
<title>DISCUSSION</title>
<p>Primary implant stability is essential for the successful formation of bone tissue at the bone-implant interface. The success of dental implants is not related to the timing of loading, but rather to the critical function of micromotion [
<xref rid="B4" ref-type="bibr">4</xref>
]. Micromotion of more than 150 μm can induce the formation of fibrous connective tissue, preventing the osseointegration of an immediately loaded implant [
<xref rid="B4" ref-type="bibr">4</xref>
<xref rid="B6" ref-type="bibr">6</xref>
]. In contrast, Vandamme et al. [
<xref rid="B28" ref-type="bibr">28</xref>
], in a bone chamber experiment, demonstrated that displacement of an implant between 30 μm and 90 μm stimulated bone formation at the implant surface. Therefore, a micromotion level from 100 μm to 150 μm may be the maximum threshold.</p>
<p>We found the maximum extent of micromotion to be significantly influenced by cancellous bone density. The maximum micromotion in the low-density cancellous bone models reached around 100 μm. In contrast, in the high-density cancellous bone, the maximum micromotion values were less than 30 μm. These results indicate that cancellous bone density may be a critical factor for avoiding excessive micromotion in immediately loaded implants.</p>
<p>A correlation between bone density and primary implant stability has been reported in previous studies. Bone density influences the ISQ [
<xref rid="B11" ref-type="bibr">11</xref>
<xref rid="B13" ref-type="bibr">13</xref>
<xref rid="B15" ref-type="bibr">15</xref>
], PTV [
<xref rid="B11" ref-type="bibr">11</xref>
<xref rid="B15" ref-type="bibr">15</xref>
], and insertion torque value (ITV) [
<xref rid="B8" ref-type="bibr">8</xref>
<xref rid="B15" ref-type="bibr">15</xref>
]. Ikumi and Tsutsumi [
<xref rid="B9" ref-type="bibr">9</xref>
] showed a significant correlation between bone density obtained by CT and cutting torque values. An in vitro study found that implant displacement in soft bone was significantly higher than in normal or hard bone [
<xref rid="B29" ref-type="bibr">29</xref>
]. Previous studies have also reported significant correlations between cortical bone thickness and implant stability. Increasing the crestal cortical bone thickness has been found to increase the values of the ISQ [
<xref rid="B10" ref-type="bibr">10</xref>
<xref rid="B11" ref-type="bibr">11</xref>
<xref rid="B12" ref-type="bibr">12</xref>
] and the ITV [
<xref rid="B15" ref-type="bibr">15</xref>
], as well as to reduce the PTV [
<xref rid="B11" ref-type="bibr">11</xref>
]. For example, Miyamoto et al. [
<xref rid="B12" ref-type="bibr">12</xref>
] found a significant correlation between the ISQ and crestal cortical bone thickness. Several in vitro studies using artificial bone models have shown a correlation between crestal cortical bone thickness and primary stability [
<xref rid="B7" ref-type="bibr">7</xref>
<xref rid="B11" ref-type="bibr">11</xref>
<xref rid="B13" ref-type="bibr">13</xref>
].</p>
<p>Studies employing FEA have shown that, when a lateral or oblique force is applied to the implant, most of the force is concentrated on the cortical bone [
<xref rid="B24" ref-type="bibr">24</xref>
]. For this reason, previous studies have focused on the anatomic background related to cortical bone thickness rather than bone density [
<xref rid="B15" ref-type="bibr">15</xref>
<xref rid="B30" ref-type="bibr">30</xref>
]. Several studies have demonstrated no correlations between bone density and implant stability. In one study, 22 implants were inserted into the maxillae and mandibles of human cadavers, and no correlations were found between the ISQ and histomorphometric parameters of cancellous bone analyzed using micro-CT [
<xref rid="B10" ref-type="bibr">10</xref>
]. Similarly, Nkenke et al. [
<xref rid="B14" ref-type="bibr">14</xref>
] demonstrated the importance of cortical bone thickness rather than bone density in implant stability. However, Marquezan et al. [
<xref rid="B15" ref-type="bibr">15</xref>
] showed that cancellous bone played an important role in primary stability in the presence and in the absence of cortical bone based on an analysis of histomorphometric bone parameters using bovine bone blocks.</p>
<p>An in vitro study using biomechanical test materials for implant stability showed the absence of correlations among methods such as resonance frequency analysis, removal torque, and axial testing [
<xref rid="B13" ref-type="bibr">13</xref>
]. Hsu et al. [
<xref rid="B11" ref-type="bibr">11</xref>
] found that initial stability was influenced by both the crestal cortical bone thickness and the Young’s modulus of cancellous bone using artificial bone, but these factors were mostly nonlinearly correlated with the ITV, PTV, and ISQ. Such nonlinear relationships among bone parameters and implant stability, as well as variations in bone density and cortical bone thickness measured in clinical studies or established using biomechanical tests, may be responsible for the different interpretations presented in previous studies.</p>
<p>In the present study, crestal cortical bone thickness affected the maximum micromotion in the standard model and in the low-density cancellous bone models, in agreement with previous studies measuring the ITV, PTV, and ISQ [
<xref rid="B7" ref-type="bibr">7</xref>
<xref rid="B10" ref-type="bibr">10</xref>
<xref rid="B11" ref-type="bibr">11</xref>
<xref rid="B12" ref-type="bibr">12</xref>
]. However, even in the standard model, the maximum micromotion values were much less than 100 μm. Therefore, we believe that crestal cortical bone thickness plays an important role in reducing micromotion, only when the cancellous bone density is low.</p>
<p>Minimum principal strains were calculated for the cortical bone around the implant. Since stress and strain may induce the form of marginal bone loss known as saucerization, the stresses and strains of the crestal cortical bone adjacent to the implant neck must play a major role in the analysis [
<xref rid="B24" ref-type="bibr">24</xref>
<xref rid="B31" ref-type="bibr">31</xref>
]. Potential fatigue damage of the bone occurs with excessive dynamic loading that exceeds 4,000 με in compression and 2,500 με in tension [
<xref rid="B32" ref-type="bibr">32</xref>
<xref rid="B33" ref-type="bibr">33</xref>
]. In particular, bone resorption under compressive strain is attributed to the accumulation of induced microdamage that exceeds the capacity of the bone for repair [
<xref rid="B4" ref-type="bibr">4</xref>
<xref rid="B26" ref-type="bibr">26</xref>
<xref rid="B32" ref-type="bibr">32</xref>
]. Therefore, the minimum principal strain values in the peri-implant cortical bone were evaluated in this study.</p>
<p>The distributions of the minimum principal strain in the delayed and immediate-loading models were consistent with previous reports, which have shown the concentration of stress and strain to be on the compressive side [
<xref rid="B26" ref-type="bibr">26</xref>
<xref rid="B34" ref-type="bibr">34</xref>
]. In the immediately loaded implants, greater stress and strain develop in the cortical bone and cancellous bone because only compressive and frictional forces are transferred via the contact interfaces, compared with the bonded interfaces of the delayed implants [
<xref rid="B35" ref-type="bibr">35</xref>
]. The results of the present study agree with the findings reported by Ferreira et al. [
<xref rid="B36" ref-type="bibr">36</xref>
], who showed that the minimum principal stress in the peri-implant bone in an immediate-loading model was approximately twofold higher than was observed in a delayed-loading model.</p>
<p>As bone density affects peri-implant bone stress and strain in delayed-loading conditions [
<xref rid="B17" ref-type="bibr">17</xref>
<xref rid="B18" ref-type="bibr">18</xref>
], the minimum principal strain values in the peri-implant cortical bone under immediate loading were also affected by the density of the cancellous and crestal cortical bone. The effects of crestal cortical bone thickness on peri-implant bone stress and strain also have been investigated under delayed-loading conditions. Crestal cortical bone thickness has either a minimal effect or no effect on peri-implant bone stress and strain [
<xref rid="B18" ref-type="bibr">18</xref>
<xref rid="B31" ref-type="bibr">31</xref>
]. Our results are consistent with those findings in the delayed-loading models. However, in the present FEA, even in the standard model under immediate loading, the minimum principal strain was affected by crestal cortical bone thickness. In the low-density cancellous bone models, the crestal cortical bone thickness affected the peak minimum principal strain values to a greater extent. This is most likely because crestal cortical bone bears a greater load on the compression side under immediate loading than under delayed loading.</p>
<p>The peak minimum principal strain exceeded 4,000 με for all immediate-loading models. However, this does not necessarily imply bone overloading and implant loss, because in addition to strain amplitude, loading frequency and the number of loading cycles can have a significant effect on the adaptive response of cortical bone [
<xref rid="B3" ref-type="bibr">3</xref>
<xref rid="B5" ref-type="bibr">5</xref>
<xref rid="B32" ref-type="bibr">32</xref>
]. Nonetheless, the results of the present study biomechanically confirm that low-density bone and thin crestal cortical bone at the implant placement site are risk factors for overloads of immediate-loading implants, as well as providing quantitative evidence to support pre-existing recommendations in the implant community. Clinically, in immediate-loading protocols, it is emphasized that implants should be placed in areas of “good-quality” bone (types 1-3 according to the Lekholm and Zarb classification) [
<xref rid="B38" ref-type="bibr">38</xref>
] or at sites where only good primary stability is achieved [
<xref rid="B2" ref-type="bibr">2</xref>
]. Selecting implant placement sites in this way may reduce micromotion and peri-implant strain, resulting in high survival rates for immediately loaded implants.</p>
<p>Biomechanical evaluations of dental implants using FEA have been performed by many researchers. The validity of the simulations depends on morphology, material properties, boundary conditions, and the bone-implant interface. The most important factors in the outcome are the material properties of the materials used, such as Young’s modulus and Poisson’s ratio [
<xref rid="B39" ref-type="bibr">39</xref>
]. Therefore, we defined the degree of bone density and range of crestal cortical bone thickness using the CT data of preoperative patients [
<xref rid="B19" ref-type="bibr">19</xref>
]. This enabled us to evaluate the effects of bone parameters quantitatively.</p>
<p>Chang et al. [
<xref rid="B40" ref-type="bibr">40</xref>
] analyzed micromotion in an immediately loaded implant model similar to that in the present study, with a vertical load of 300 N. They showed a maximum micromotion of 8.5 μm to 15.0 μm in the model with a cortical bone thickness of 2 mm. The maximum micromotion of the standard model with 2.0-mm crestal cortical bone thickness in the present study was 17 μm. This microstrain value is also compatible with the value found by Huang et al. [
<xref rid="B35" ref-type="bibr">35</xref>
], who reported that the maximum micromotion in the high-density cancellous model was 7.9 μm, under a 30-degree oblique load of 129 N. Although the model geometry, material properties, and boundary and loading conditions were not identical, the results of this study were similar to those of other FEA studies; hence, the model of this study should be considered to have been validated accurately by previously published reports.</p>
<p>This finite element model had limitations, such as homogeneities and the isotropic linear elasticity of the material properties. Altering the mandibular properties with the anisotropic assumption may result in different strain distributions [
<xref rid="B35" ref-type="bibr">35</xref>
]. The load was applied in a fixed direction. The boundary conditions were set as fixed, so they did not accurately reproduce the complex forces exerted during chewing. These assumptions do not completely reflect clinical scenarios. Thus, the values of strain and micromotion could not be directly compared with clinical threshold values. Furthermore, the threshold value for micromotion may be influenced by the implant surface treatment. Nonetheless, in agreement with other quantitative studies [
<xref rid="B3" ref-type="bibr">3</xref>
<xref rid="B17" ref-type="bibr">17</xref>
<xref rid="B35" ref-type="bibr">35</xref>
<xref rid="B36" ref-type="bibr">36</xref>
<xref rid="B37" ref-type="bibr">37</xref>
], the present assumptions can be accepted, in a computational sense, to assess the biomechanical behavior of implants. The above limitations should be considered when applying our results to clinical situations.</p>
<p>In conclusion, cancellous bone density may be a critical factor for avoiding excessive micromotion in immediately loaded implants. Crestal cortical bone thickness greatly influenced the maximum micromotion only when the cancellous bone density was low. In cases with low-density cancellous bone, the minimum principal strain in the peri-implant bone was more sensitive to changes in the crestal cortical bone thickness under immediate loading than under delayed loading.</p>
</sec>
</body>
<back>
<ack>
<title>ACNOWLEDGEMENTS</title>
<p>This work was supported by a Grant-in Aid for Scientific Research from the Japan Society for Promotion of Science (No. 24592963).</p>
</ack>
<fn-group>
<fn fn-type="conflict">
<p>
<bold>Conflict of Interest:</bold>
No potential conflict of interest relevant to this article was reported.</p>
</fn>
</fn-group>
<ref-list>
<ref id="B1">
<label>1</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Misch</surname>
<given-names>CE</given-names>
</name>
<name>
<surname>Wang</surname>
<given-names>HL</given-names>
</name>
<name>
<surname>Misch</surname>
<given-names>CM</given-names>
</name>
<name>
<surname>Sharawy</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Lemons</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Judy</surname>
<given-names>KW</given-names>
</name>
</person-group>
<article-title>Rationale for the application of immediate load in implant dentistry: Part I</article-title>
<source>Implant Dent</source>
<year>2004</year>
<volume>13</volume>
<fpage>207</fpage>
<lpage>217</lpage>
<pub-id pub-id-type="pmid">15359155</pub-id>
</element-citation>
</ref>
<ref id="B2">
<label>2</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Laviv</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Levin</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Usiel</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Schwartz-Arad</surname>
<given-names>D</given-names>
</name>
</person-group>
<article-title>Survival of immediately provisionalized dental implants: a case-control study with up to 5 years follow-up</article-title>
<source>Clin Implant Dent Relat Res</source>
<year>2010</year>
<volume>12</volume>
<issue>Suppl 1</issue>
<fpage>e23</fpage>
<lpage>7</lpage>
<pub-id pub-id-type="pmid">19220849</pub-id>
</element-citation>
</ref>
<ref id="B3">
<label>3</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Pessoa</surname>
<given-names>RS</given-names>
</name>
<name>
<surname>Coelho</surname>
<given-names>PG</given-names>
</name>
<name>
<surname>Muraru</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Marcantonio</surname>
<given-names>E</given-names>
<suffix>Jr</suffix>
</name>
<name>
<surname>Vaz</surname>
<given-names>LG</given-names>
</name>
<name>
<surname>Vander Sloten</surname>
<given-names>J</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Influence of implant design on the biomechanical environment of immediately placed implants: computed tomography-based nonlinear three-dimensional finite element analysis</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2011</year>
<volume>26</volume>
<fpage>1279</fpage>
<lpage>1287</lpage>
<pub-id pub-id-type="pmid">22167434</pub-id>
</element-citation>
</ref>
<ref id="B4">
<label>4</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Brunski</surname>
<given-names>JB</given-names>
</name>
<name>
<surname>Puleo</surname>
<given-names>DA</given-names>
</name>
<name>
<surname>Nanci</surname>
<given-names>A</given-names>
</name>
</person-group>
<article-title>Biomaterials and biomechanics of oral and maxillofacial implants: current status and future developments</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2000</year>
<volume>15</volume>
<fpage>15</fpage>
<lpage>46</lpage>
<pub-id pub-id-type="pmid">10697938</pub-id>
</element-citation>
</ref>
<ref id="B5">
<label>5</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Duyck</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Rønold</surname>
<given-names>HJ</given-names>
</name>
<name>
<surname>Van Oosterwyck</surname>
<given-names>H</given-names>
</name>
<name>
<surname>Naert</surname>
<given-names>I</given-names>
</name>
<name>
<surname>Vander Sloten</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Ellingsen</surname>
<given-names>JE</given-names>
</name>
</person-group>
<article-title>The influence of static and dynamic loading on marginal bone reactions around osseointegrated implants: an animal experimental study</article-title>
<source>Clin Oral Implants Res</source>
<year>2001</year>
<volume>12</volume>
<fpage>207</fpage>
<lpage>218</lpage>
<pub-id pub-id-type="pmid">11359477</pub-id>
</element-citation>
</ref>
<ref id="B6">
<label>6</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Pilliar</surname>
<given-names>RM</given-names>
</name>
<name>
<surname>Lee</surname>
<given-names>JM</given-names>
</name>
<name>
<surname>Maniatopoulos</surname>
<given-names>C</given-names>
</name>
</person-group>
<article-title>Observations on the effect of movement on bone ingrowth into porous-surfaced implants</article-title>
<source>Clin Orthop Relat Res</source>
<year>1986</year>
<issue>208</issue>
<fpage>108</fpage>
<lpage>113</lpage>
<pub-id pub-id-type="pmid">3720113</pub-id>
</element-citation>
</ref>
<ref id="B7">
<label>7</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Tabassum</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Meijer</surname>
<given-names>GJ</given-names>
</name>
<name>
<surname>Wolke</surname>
<given-names>JG</given-names>
</name>
<name>
<surname>Jansen</surname>
<given-names>JA</given-names>
</name>
</person-group>
<article-title>Influence of surgical technique and surface roughness on the primary stability of an implant in artificial bone with different cortical thickness: a laboratory study</article-title>
<source>Clin Oral Implants Res</source>
<year>2010</year>
<volume>21</volume>
<fpage>213</fpage>
<lpage>220</lpage>
<pub-id pub-id-type="pmid">20070754</pub-id>
</element-citation>
</ref>
<ref id="B8">
<label>8</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Bayarchimeg</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Namgoong</surname>
<given-names>H</given-names>
</name>
<name>
<surname>Kim</surname>
<given-names>BK</given-names>
</name>
<name>
<surname>Kim</surname>
<given-names>MD</given-names>
</name>
<name>
<surname>Kim</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Kim</surname>
<given-names>TI</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Evaluation of the correlation between insertion torque and primary stability of dental implants using a block bone test</article-title>
<source>J Periodontal Implant Sci</source>
<year>2013</year>
<volume>43</volume>
<fpage>30</fpage>
<lpage>36</lpage>
<pub-id pub-id-type="pmid">23508040</pub-id>
</element-citation>
</ref>
<ref id="B9">
<label>9</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Ikumi</surname>
<given-names>N</given-names>
</name>
<name>
<surname>Tsutsumi</surname>
<given-names>S</given-names>
</name>
</person-group>
<article-title>Assessment of correlation between computerized tomography values of the bone and cutting torque values at implant placement: a clinical study</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2005</year>
<volume>20</volume>
<fpage>253</fpage>
<lpage>260</lpage>
<pub-id pub-id-type="pmid">15839119</pub-id>
</element-citation>
</ref>
<ref id="B10">
<label>10</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Rozé</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Babu</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Saffarzadeh</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Gayet-Delacroix</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Hoornaert</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Layrolle</surname>
<given-names>P</given-names>
</name>
</person-group>
<article-title>Correlating implant stability to bone structure</article-title>
<source>Clin Oral Implants Res</source>
<year>2009</year>
<volume>20</volume>
<fpage>1140</fpage>
<lpage>1145</lpage>
<pub-id pub-id-type="pmid">19519789</pub-id>
</element-citation>
</ref>
<ref id="B11">
<label>11</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Hsu</surname>
<given-names>JT</given-names>
</name>
<name>
<surname>Fuh</surname>
<given-names>LJ</given-names>
</name>
<name>
<surname>Tu</surname>
<given-names>MG</given-names>
</name>
<name>
<surname>Li</surname>
<given-names>YF</given-names>
</name>
<name>
<surname>Chen</surname>
<given-names>KT</given-names>
</name>
<name>
<surname>Huang</surname>
<given-names>HL</given-names>
</name>
</person-group>
<article-title>The effects of cortical bone thickness and trabecular bone strength on noninvasive measures of the implant primary stability using synthetic bone models</article-title>
<source>Clin Implant Dent Relat Res</source>
<year>2013</year>
<volume>15</volume>
<fpage>251</fpage>
<lpage>261</lpage>
<pub-id pub-id-type="pmid">21599830</pub-id>
</element-citation>
</ref>
<ref id="B12">
<label>12</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Miyamoto</surname>
<given-names>I</given-names>
</name>
<name>
<surname>Tsuboi</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Wada</surname>
<given-names>E</given-names>
</name>
<name>
<surname>Suwa</surname>
<given-names>H</given-names>
</name>
<name>
<surname>Iizuka</surname>
<given-names>T</given-names>
</name>
</person-group>
<article-title>Influence of cortical bone thickness and implant length on implant stability at the time of surgery--clinical, prospective, biomechanical, and imaging study</article-title>
<source>Bone</source>
<year>2005</year>
<volume>37</volume>
<fpage>776</fpage>
<lpage>780</lpage>
<pub-id pub-id-type="pmid">16154396</pub-id>
</element-citation>
</ref>
<ref id="B13">
<label>13</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Bardyn</surname>
<given-names>T</given-names>
</name>
<name>
<surname>Gédet</surname>
<given-names>P</given-names>
</name>
<name>
<surname>Hallermann</surname>
<given-names>W</given-names>
</name>
<name>
<surname>Büchler</surname>
<given-names>P</given-names>
</name>
</person-group>
<article-title>Quantifying the influence of bone density and thickness on resonance frequency analysis: an in vitro study of biomechanical test materials</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2009</year>
<volume>24</volume>
<fpage>1006</fpage>
<lpage>1014</lpage>
<pub-id pub-id-type="pmid">20162104</pub-id>
</element-citation>
</ref>
<ref id="B14">
<label>14</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Nkenke</surname>
<given-names>E</given-names>
</name>
<name>
<surname>Hahn</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Weinzierl</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Radespiel-Tröger</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Neukam</surname>
<given-names>FW</given-names>
</name>
<name>
<surname>Engelke</surname>
<given-names>K</given-names>
</name>
</person-group>
<article-title>Implant stability and histomorphometry: a correlation study in human cadavers using stepped cylinder implants</article-title>
<source>Clin Oral Implants Res</source>
<year>2003</year>
<volume>14</volume>
<fpage>601</fpage>
<lpage>609</lpage>
<pub-id pub-id-type="pmid">12969364</pub-id>
</element-citation>
</ref>
<ref id="B15">
<label>15</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Marquezan</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Lima</surname>
<given-names>I</given-names>
</name>
<name>
<surname>Lopes</surname>
<given-names>RT</given-names>
</name>
<name>
<surname>Sant’Anna</surname>
<given-names>EF</given-names>
</name>
<name>
<surname>de Souza</surname>
<given-names>MM</given-names>
</name>
</person-group>
<article-title>Is trabecular bone related to primary stability of miniscrews?</article-title>
<source>Angle Orthod</source>
<year>2014</year>
<volume>84</volume>
<fpage>500</fpage>
<lpage>507</lpage>
<pub-id pub-id-type="pmid">24245816</pub-id>
</element-citation>
</ref>
<ref id="B16">
<label>16</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Palma-Carrió</surname>
<given-names>C</given-names>
</name>
<name>
<surname>Maestre-Ferrín</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Peñarrocha-Oltra</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Peñarrocha-Diago</surname>
<given-names>MA</given-names>
</name>
<name>
<surname>Peñarrocha-Diago</surname>
<given-names>M</given-names>
</name>
</person-group>
<article-title>Risk factors associated with early failure of dental implants. A literature review</article-title>
<source>Med Oral Patol Oral Cir Bucal</source>
<year>2011</year>
<volume>16</volume>
<fpage>e514</fpage>
<lpage>7</lpage>
<pub-id pub-id-type="pmid">21196891</pub-id>
</element-citation>
</ref>
<ref id="B17">
<label>17</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Petrie</surname>
<given-names>CS</given-names>
</name>
<name>
<surname>Williams</surname>
<given-names>JL</given-names>
</name>
</person-group>
<article-title>Probabilistic analysis of peri-implant strain predictions as influenced by uncertainties in bone properties and occlusal forces</article-title>
<source>Clin Oral Implants Res</source>
<year>2007</year>
<volume>18</volume>
<fpage>611</fpage>
<lpage>619</lpage>
<pub-id pub-id-type="pmid">17590159</pub-id>
</element-citation>
</ref>
<ref id="B18">
<label>18</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Guan</surname>
<given-names>H</given-names>
</name>
<name>
<surname>van Staden</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Loo</surname>
<given-names>YC</given-names>
</name>
<name>
<surname>Johnson</surname>
<given-names>N</given-names>
</name>
<name>
<surname>Ivanovski</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Meredith</surname>
<given-names>N</given-names>
</name>
</person-group>
<article-title>Influence of bone and dental implant parameters on stress distribution in the mandible: a finite element study</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2009</year>
<volume>24</volume>
<fpage>866</fpage>
<lpage>876</lpage>
<pub-id pub-id-type="pmid">19865627</pub-id>
</element-citation>
</ref>
<ref id="B19">
<label>19</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Sugiura</surname>
<given-names>T</given-names>
</name>
<name>
<surname>Yamamoto</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Kawakami</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Horita</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Murakami</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Kirita</surname>
<given-names>T</given-names>
</name>
</person-group>
<article-title>Influence of bone parameters on peri-implant bone strain distribution in the posterior mandible</article-title>
<source>Med Oral Patol Oral Cir Bucal</source>
<year>2015</year>
<volume>20</volume>
<fpage>e66</fpage>
<lpage>73</lpage>
<pub-id pub-id-type="pmid">25129251</pub-id>
</element-citation>
</ref>
<ref id="B20">
<label>20</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Morton</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Jaffin</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Weber</surname>
<given-names>HP</given-names>
</name>
</person-group>
<article-title>Immediate restoration and loading of dental implants: clinical considerations and protocols</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2004</year>
<volume>19</volume>
<issue>Suppl</issue>
<fpage>103</fpage>
<lpage>108</lpage>
<pub-id pub-id-type="pmid">15635950</pub-id>
</element-citation>
</ref>
<ref id="B21">
<label>21</label>
<element-citation publication-type="book">
<person-group person-group-type="author">
<name>
<surname>Lekholm</surname>
<given-names>U</given-names>
</name>
<name>
<surname>Zarb</surname>
<given-names>GA</given-names>
</name>
</person-group>
<chapter-title>Patient selection and preparation</chapter-title>
<person-group person-group-type="editor">
<name>
<surname>Brånemark</surname>
<given-names>PI</given-names>
</name>
<name>
<surname>Zarb</surname>
<given-names>GA</given-names>
</name>
<name>
<surname>Albrektsson</surname>
<given-names>T</given-names>
</name>
</person-group>
<source>Tissue-integrated prostheses: osseointegration in clinical dentistry</source>
<publisher-loc>Chicago (IL)</publisher-loc>
<publisher-name>Quintessence</publisher-name>
<year>1985</year>
<fpage>199</fpage>
<lpage>209</lpage>
</element-citation>
</ref>
<ref id="B22">
<label>22</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Kurniawan</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Nor</surname>
<given-names>FM</given-names>
</name>
<name>
<surname>Lee</surname>
<given-names>HY</given-names>
</name>
<name>
<surname>Lim</surname>
<given-names>JY</given-names>
</name>
</person-group>
<article-title>Finite element analysis of bone-implant biomechanics: refinement through featuring various osseointegration conditions</article-title>
<source>Int J Oral Maxillofac Surg</source>
<year>2012</year>
<volume>41</volume>
<fpage>1090</fpage>
<lpage>1096</lpage>
<pub-id pub-id-type="pmid">22260789</pub-id>
</element-citation>
</ref>
<ref id="B23">
<label>23</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Keyak</surname>
<given-names>JH</given-names>
</name>
<name>
<surname>Rossi</surname>
<given-names>SA</given-names>
</name>
<name>
<surname>Jones</surname>
<given-names>KA</given-names>
</name>
<name>
<surname>Skinner</surname>
<given-names>HB</given-names>
</name>
</person-group>
<article-title>Prediction of femoral fracture load using automated finite element modeling</article-title>
<source>J Biomech</source>
<year>1998</year>
<volume>31</volume>
<fpage>125</fpage>
<lpage>133</lpage>
<pub-id pub-id-type="pmid">9593205</pub-id>
</element-citation>
</ref>
<ref id="B24">
<label>24</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Ding</surname>
<given-names>X</given-names>
</name>
<name>
<surname>Liao</surname>
<given-names>SH</given-names>
</name>
<name>
<surname>Zhu</surname>
<given-names>XH</given-names>
</name>
<name>
<surname>Zhang</surname>
<given-names>XH</given-names>
</name>
<name>
<surname>Zhang</surname>
<given-names>L</given-names>
</name>
</person-group>
<article-title>Effect of diameter and length on stress distribution of the alveolar crest around immediate loading implants</article-title>
<source>Clin Implant Dent Relat Res</source>
<year>2009</year>
<volume>11</volume>
<fpage>279</fpage>
<lpage>287</lpage>
<pub-id pub-id-type="pmid">18783411</pub-id>
</element-citation>
</ref>
<ref id="B25">
<label>25</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Mericske-Stern</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Assal</surname>
<given-names>P</given-names>
</name>
<name>
<surname>Mericske</surname>
<given-names>E</given-names>
</name>
<name>
<surname>Bürgin</surname>
<given-names>W</given-names>
</name>
</person-group>
<article-title>Occlusal force and oral tactile sensibility measured in partially edentulous patients with ITI implants</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>1995</year>
<volume>10</volume>
<fpage>345</fpage>
<lpage>353</lpage>
<pub-id pub-id-type="pmid">7615331</pub-id>
</element-citation>
</ref>
<ref id="B26">
<label>26</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Hudieb</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Wakabayashi</surname>
<given-names>N</given-names>
</name>
<name>
<surname>Suzuki</surname>
<given-names>T</given-names>
</name>
<name>
<surname>Kasugai</surname>
<given-names>S</given-names>
</name>
</person-group>
<article-title>Morphologic classification and stress analysis of the mandibular bone in the premolar region for implant placement</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2010</year>
<volume>25</volume>
<fpage>482</fpage>
<lpage>490</lpage>
<pub-id pub-id-type="pmid">20556246</pub-id>
</element-citation>
</ref>
<ref id="B27">
<label>27</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Gonda</surname>
<given-names>T</given-names>
</name>
<name>
<surname>Yasuda</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Ikebe</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Maeda</surname>
<given-names>Y</given-names>
</name>
</person-group>
<article-title>Biomechanical factors associated with mandibular cantilevers: analysis with three-dimensional finite element models</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2014</year>
<volume>29</volume>
<fpage>e275</fpage>
<lpage>82</lpage>
<pub-id pub-id-type="pmid">25397809</pub-id>
</element-citation>
</ref>
<ref id="B28">
<label>28</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Vandamme</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Naert</surname>
<given-names>I</given-names>
</name>
<name>
<surname>Geris</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Vander Sloten</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Puers</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Duyck</surname>
<given-names>J</given-names>
</name>
</person-group>
<article-title>The effect of micro-motion on the tissue response around immediately loaded roughened titanium implants in the rabbit</article-title>
<source>Eur J Oral Sci</source>
<year>2007</year>
<volume>115</volume>
<fpage>21</fpage>
<lpage>29</lpage>
<pub-id pub-id-type="pmid">17305713</pub-id>
</element-citation>
</ref>
<ref id="B29">
<label>29</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Trisi</surname>
<given-names>P</given-names>
</name>
<name>
<surname>Perfetti</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Baldoni</surname>
<given-names>E</given-names>
</name>
<name>
<surname>Berardi</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Colagiovanni</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Scogna</surname>
<given-names>G</given-names>
</name>
</person-group>
<article-title>Implant micromotion is related to peak insertion torque and bone density</article-title>
<source>Clin Oral Implants Res</source>
<year>2009</year>
<volume>20</volume>
<fpage>467</fpage>
<lpage>471</lpage>
<pub-id pub-id-type="pmid">19522976</pub-id>
</element-citation>
</ref>
<ref id="B30">
<label>30</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Cha</surname>
<given-names>JY</given-names>
</name>
<name>
<surname>Kil</surname>
<given-names>JK</given-names>
</name>
<name>
<surname>Yoon</surname>
<given-names>TM</given-names>
</name>
<name>
<surname>Hwang</surname>
<given-names>CJ</given-names>
</name>
</person-group>
<article-title>Miniscrew stability evaluated with computerized tomography scanning</article-title>
<source>Am J Orthod Dentofacial Orthop</source>
<year>2010</year>
<volume>137</volume>
<fpage>73</fpage>
<lpage>79</lpage>
<pub-id pub-id-type="pmid">20122434</pub-id>
</element-citation>
</ref>
<ref id="B31">
<label>31</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Shen</surname>
<given-names>WL</given-names>
</name>
<name>
<surname>Chen</surname>
<given-names>CS</given-names>
</name>
<name>
<surname>Hsu</surname>
<given-names>ML</given-names>
</name>
</person-group>
<article-title>Influence of implant collar design on stress and strain distribution in the crestal compact bone: a three-dimensional finite element analysis</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2010</year>
<volume>25</volume>
<fpage>901</fpage>
<lpage>910</lpage>
<pub-id pub-id-type="pmid">20862403</pub-id>
</element-citation>
</ref>
<ref id="B32">
<label>32</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Frost</surname>
<given-names>HM</given-names>
</name>
</person-group>
<article-title>Skeletal structural adaptations to mechanical usage (SATMU): 2. Redefining Wolff’s law: the remodeling problem</article-title>
<source>Anat Rec</source>
<year>1990</year>
<volume>226</volume>
<fpage>414</fpage>
<lpage>422</lpage>
<pub-id pub-id-type="pmid">2184696</pub-id>
</element-citation>
</ref>
<ref id="B33">
<label>33</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Pattin</surname>
<given-names>CA</given-names>
</name>
<name>
<surname>Caler</surname>
<given-names>WE</given-names>
</name>
<name>
<surname>Carter</surname>
<given-names>DR</given-names>
</name>
</person-group>
<article-title>Cyclic mechanical property degradation during fatigue loading of cortical bone</article-title>
<source>J Biomech</source>
<year>1996</year>
<volume>29</volume>
<fpage>69</fpage>
<lpage>79</lpage>
<pub-id pub-id-type="pmid">8839019</pub-id>
</element-citation>
</ref>
<ref id="B34">
<label>34</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Mellal</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Wiskott</surname>
<given-names>HW</given-names>
</name>
<name>
<surname>Botsis</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Scherrer</surname>
<given-names>SS</given-names>
</name>
<name>
<surname>Belser</surname>
<given-names>UC</given-names>
</name>
</person-group>
<article-title>Stimulating effect of implant loading on surrounding bone. Comparison of three numerical models and validation by in vivo data</article-title>
<source>Clin Oral Implants Res</source>
<year>2004</year>
<volume>15</volume>
<fpage>239</fpage>
<lpage>248</lpage>
<pub-id pub-id-type="pmid">15008937</pub-id>
</element-citation>
</ref>
<ref id="B35">
<label>35</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Huang</surname>
<given-names>HL</given-names>
</name>
<name>
<surname>Fuh</surname>
<given-names>LJ</given-names>
</name>
<name>
<surname>Hsu</surname>
<given-names>JT</given-names>
</name>
<name>
<surname>Tu</surname>
<given-names>MG</given-names>
</name>
<name>
<surname>Shen</surname>
<given-names>YW</given-names>
</name>
<name>
<surname>Wu</surname>
<given-names>CL</given-names>
</name>
</person-group>
<article-title>Effects of implant surface roughness and stiffness of grafted bone on an immediately loaded maxillary implant: a 3D numerical analysis</article-title>
<source>J Oral Rehabil</source>
<year>2008</year>
<volume>35</volume>
<fpage>283</fpage>
<lpage>290</lpage>
<pub-id pub-id-type="pmid">18321264</pub-id>
</element-citation>
</ref>
<ref id="B36">
<label>36</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Ferreira</surname>
<given-names>MB</given-names>
</name>
<name>
<surname>Barão</surname>
<given-names>VA</given-names>
</name>
<name>
<surname>Delben</surname>
<given-names>JA</given-names>
</name>
<name>
<surname>Faverani</surname>
<given-names>LP</given-names>
</name>
<name>
<surname>Hipólito</surname>
<given-names>AC</given-names>
</name>
<name>
<surname>Assunção</surname>
<given-names>WG</given-names>
</name>
</person-group>
<article-title>Non-linear 3D finite element analysis of full-arch implant-supported fixed dentures</article-title>
<source>Mater Sci Eng C</source>
<year>2014</year>
<volume>38</volume>
<fpage>306</fpage>
<lpage>314</lpage>
</element-citation>
</ref>
<ref id="B37">
<label>37</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Duaibis</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Kusnoto</surname>
<given-names>B</given-names>
</name>
<name>
<surname>Natarajan</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Zhao</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Evans</surname>
<given-names>C</given-names>
</name>
</person-group>
<article-title>Factors affecting stresses in cortical bone around miniscrew implants: a three-dimensional finite element study</article-title>
<source>Angle Orthod</source>
<year>2012</year>
<volume>82</volume>
<fpage>875</fpage>
<lpage>880</lpage>
<pub-id pub-id-type="pmid">22390634</pub-id>
</element-citation>
</ref>
<ref id="B38">
<label>38</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Chiapasco</surname>
<given-names>M</given-names>
</name>
</person-group>
<article-title>Early and immediate restoration and loading of implants in completely edentulous patients</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2004</year>
<volume>19</volume>
<issue>Suppl</issue>
<fpage>76</fpage>
<lpage>91</lpage>
<pub-id pub-id-type="pmid">15635948</pub-id>
</element-citation>
</ref>
<ref id="B39">
<label>39</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Seker</surname>
<given-names>E</given-names>
</name>
<name>
<surname>Ulusoy</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Ozan</surname>
<given-names>O</given-names>
</name>
<name>
<surname>Doğan</surname>
<given-names></given-names>
</name>
<name>
<surname>Seker</surname>
<given-names>BK</given-names>
</name>
</person-group>
<article-title>Biomechanical effects of different fixed partial denture designs planned on bicortically anchored short, graft-supported long, or 45-degree-inclined long implants in the posterior maxilla: a three-dimensional finite element analysis</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2014</year>
<volume>29</volume>
<fpage>e1</fpage>
<lpage>9</lpage>
<pub-id pub-id-type="pmid">24451875</pub-id>
</element-citation>
</ref>
<ref id="B40">
<label>40</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Chang</surname>
<given-names>PK</given-names>
</name>
<name>
<surname>Chen</surname>
<given-names>YC</given-names>
</name>
<name>
<surname>Huang</surname>
<given-names>CC</given-names>
</name>
<name>
<surname>Lu</surname>
<given-names>WH</given-names>
</name>
<name>
<surname>Chen</surname>
<given-names>YC</given-names>
</name>
<name>
<surname>Tsai</surname>
<given-names>HH</given-names>
</name>
</person-group>
<article-title>Distribution of micromotion in implants and alveolar bone with different thread profiles in immediate loading: a finite element study</article-title>
<source>Int J Oral Maxillofac Implants</source>
<year>2012</year>
<volume>27</volume>
<fpage>e96</fpage>
<lpage>101</lpage>
<pub-id pub-id-type="pmid">23189316</pub-id>
</element-citation>
</ref>
</ref-list>
</back>
</pmc>
</record>

Pour manipuler ce document sous Unix (Dilib)

EXPLOR_STEP=$WICRI_ROOT/Wicri/Santé/explor/EdenteV2/Data/Pmc/Corpus
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 000731 | SxmlIndent | more

Ou

HfdSelect -h $EXPLOR_AREA/Data/Pmc/Corpus/biblio.hfd -nk 000731 | SxmlIndent | more

Pour mettre un lien sur cette page dans le réseau Wicri

{{Explor lien
   |wiki=    Wicri/Santé
   |area=    EdenteV2
   |flux=    Pmc
   |étape=   Corpus
   |type=    RBID
   |clé=     PMC:4928204
   |texte=   The effects of bone density and crestal cortical bone thickness on micromotion and peri-implant bone strain distribution in an immediately loaded implant: a nonlinear finite element analysis
}}

Pour générer des pages wiki

HfdIndexSelect -h $EXPLOR_AREA/Data/Pmc/Corpus/RBID.i   -Sk "pubmed:27382504" \
       | HfdSelect -Kh $EXPLOR_AREA/Data/Pmc/Corpus/biblio.hfd   \
       | NlmPubMed2Wicri -a EdenteV2 

Wicri

This area was generated with Dilib version V0.6.32.
Data generation: Thu Nov 30 15:26:48 2017. Site generation: Tue Mar 8 16:36:20 2022