Bioactive and Thermally Compatible Glass Coating on Zirconia Dental Implants
Identifieur interne : 001379 ( Main/Exploration ); précédent : 001378; suivant : 001380Bioactive and Thermally Compatible Glass Coating on Zirconia Dental Implants
Auteurs : A. Kirsten [Allemagne] ; A. Hausmann [Allemagne] ; M. Weber [Allemagne] ; J. Fischer [Suisse] ; H. Fischer [Allemagne]Source :
- Journal of Dental Research [ 0022-0345 ] ; 2015.
Abstract
The healing time of zirconia implants may be reduced by the use of bioactive glass coatings. Unfortunately, existing glasses are either bioactive like Bioglass 45S5 but thermally incompatible with the zirconia substrate, or they are thermally compatible but exhibit only a very low level of bioactivity. In this study, we hypothesized that a tailored substitution of alkaline earth metals and alkaline metals in 45S5 can lead to a glass composition that is both bioactive and thermally compatible with zirconia implants. A novel glass composition was analyzed using x-ray fluorescence spectroscopy, dilatometry, differential scanning calorimetry, and heating microscopy to investigate its chemical, physical, and thermal properties. Bioactivity was tested in vitro using simulated body fluid (SBF). Smooth and microstructured glass coatings were applied using a tailored spray technique with subsequent thermal treatment. Coating adhesion was tested on implants that were inserted in bovine ribs. The cytocompatibility of the coating was analyzed using L929 mouse fibroblasts. The coefficient of thermal expansion of the novel glass was shown to be slightly lower (11.58·10–6 K–1) than that of the zirconia (11.67·10–6 K–1). After storage in SBF, the glass showed reaction layers almost identical to the bioactive glass gold standard, 45S5. A process window between 800 °C and 910 °C was found to result in densely sintered and amorphous coatings. Microstructured glass coatings on zirconia implants survived a minimum insertion torque of 60 Ncm in the in vitro experiment on bovine ribs. Proliferation and cytotoxicity of the glass coatings was comparable with the controls. The novel glass composition showed a strong adhesion to the zirconia substrate and a significant bioactive behavior in the SBF in vitro experiments. Therefore, it holds great potential to significantly reduce the healing time of zirconia dental implants.
Url:
DOI: 10.1177/0022034514559250
PubMed: 25421839
PubMed Central: 4438729
Affiliations:
Links toward previous steps (curation, corpus...)
- to stream Pmc, to step Corpus: 000239
- to stream Pmc, to step Curation: 000239
- to stream Pmc, to step Checkpoint: 000F87
- to stream Ncbi, to step Merge: 005647
- to stream Ncbi, to step Curation: 005647
- to stream Ncbi, to step Checkpoint: 005647
- to stream Main, to step Merge: 001382
- to stream Main, to step Curation: 001379
Le document en format XML
<record><TEI><teiHeader><fileDesc><titleStmt><title xml:lang="en">Bioactive and Thermally Compatible Glass Coating on Zirconia Dental Implants</title>
<author><name sortKey="Kirsten, A" sort="Kirsten, A" uniqKey="Kirsten A" first="A." last="Kirsten">A. Kirsten</name>
<affiliation wicri:level="3"><nlm:aff id="aff1-0022034514559250">Dental Materials and Biomaterials Research, RWTH Aachen University Hospital, Aachen, Germany</nlm:aff>
<country xml:lang="fr">Allemagne</country>
<wicri:regionArea>Dental Materials and Biomaterials Research, RWTH Aachen University Hospital, Aachen</wicri:regionArea>
<placeName><region type="land" nuts="1">Rhénanie-du-Nord-Westphalie</region>
<region type="district" nuts="2">District de Cologne</region>
<settlement type="city">Aix-la-Chapelle</settlement>
</placeName>
</affiliation>
</author>
<author><name sortKey="Hausmann, A" sort="Hausmann, A" uniqKey="Hausmann A" first="A." last="Hausmann">A. Hausmann</name>
<affiliation wicri:level="3"><nlm:aff id="aff1-0022034514559250">Dental Materials and Biomaterials Research, RWTH Aachen University Hospital, Aachen, Germany</nlm:aff>
<country xml:lang="fr">Allemagne</country>
<wicri:regionArea>Dental Materials and Biomaterials Research, RWTH Aachen University Hospital, Aachen</wicri:regionArea>
<placeName><region type="land" nuts="1">Rhénanie-du-Nord-Westphalie</region>
<region type="district" nuts="2">District de Cologne</region>
<settlement type="city">Aix-la-Chapelle</settlement>
</placeName>
</affiliation>
</author>
<author><name sortKey="Weber, M" sort="Weber, M" uniqKey="Weber M" first="M." last="Weber">M. Weber</name>
<affiliation wicri:level="3"><nlm:aff id="aff1-0022034514559250">Dental Materials and Biomaterials Research, RWTH Aachen University Hospital, Aachen, Germany</nlm:aff>
<country xml:lang="fr">Allemagne</country>
<wicri:regionArea>Dental Materials and Biomaterials Research, RWTH Aachen University Hospital, Aachen</wicri:regionArea>
<placeName><region type="land" nuts="1">Rhénanie-du-Nord-Westphalie</region>
<region type="district" nuts="2">District de Cologne</region>
<settlement type="city">Aix-la-Chapelle</settlement>
</placeName>
</affiliation>
</author>
<author><name sortKey="Fischer, J" sort="Fischer, J" uniqKey="Fischer J" first="J." last="Fischer">J. Fischer</name>
<affiliation wicri:level="1"><nlm:aff id="aff2-0022034514559250">Institute of Dental Materials and Engineering, University Hospital of Dental Medicine, University of Basel, Basel, Switzerland</nlm:aff>
<country xml:lang="fr">Suisse</country>
<wicri:regionArea>Institute of Dental Materials and Engineering, University Hospital of Dental Medicine, University of Basel, Basel</wicri:regionArea>
<wicri:noRegion>Basel</wicri:noRegion>
</affiliation>
</author>
<author><name sortKey="Fischer, H" sort="Fischer, H" uniqKey="Fischer H" first="H." last="Fischer">H. Fischer</name>
<affiliation wicri:level="3"><nlm:aff id="aff1-0022034514559250">Dental Materials and Biomaterials Research, RWTH Aachen University Hospital, Aachen, Germany</nlm:aff>
<country xml:lang="fr">Allemagne</country>
<wicri:regionArea>Dental Materials and Biomaterials Research, RWTH Aachen University Hospital, Aachen</wicri:regionArea>
<placeName><region type="land" nuts="1">Rhénanie-du-Nord-Westphalie</region>
<region type="district" nuts="2">District de Cologne</region>
<settlement type="city">Aix-la-Chapelle</settlement>
</placeName>
</affiliation>
</author>
</titleStmt>
<publicationStmt><idno type="wicri:source">PMC</idno>
<idno type="pmid">25421839</idno>
<idno type="pmc">4438729</idno>
<idno type="url">http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4438729</idno>
<idno type="RBID">PMC:4438729</idno>
<idno type="doi">10.1177/0022034514559250</idno>
<date when="2015">2015</date>
<idno type="wicri:Area/Pmc/Corpus">000239</idno>
<idno type="wicri:explorRef" wicri:stream="Pmc" wicri:step="Corpus" wicri:corpus="PMC">000239</idno>
<idno type="wicri:Area/Pmc/Curation">000239</idno>
<idno type="wicri:explorRef" wicri:stream="Pmc" wicri:step="Curation">000239</idno>
<idno type="wicri:Area/Pmc/Checkpoint">000F87</idno>
<idno type="wicri:explorRef" wicri:stream="Pmc" wicri:step="Checkpoint">000F87</idno>
<idno type="wicri:Area/Ncbi/Merge">005647</idno>
<idno type="wicri:Area/Ncbi/Curation">005647</idno>
<idno type="wicri:Area/Ncbi/Checkpoint">005647</idno>
<idno type="wicri:doubleKey">0022-0345:2015:Kirsten A:bioactive:and:thermally</idno>
<idno type="wicri:Area/Main/Merge">001382</idno>
<idno type="wicri:Area/Main/Curation">001379</idno>
<idno type="wicri:Area/Main/Exploration">001379</idno>
</publicationStmt>
<sourceDesc><biblStruct><analytic><title xml:lang="en" level="a" type="main">Bioactive and Thermally Compatible Glass Coating on Zirconia Dental Implants</title>
<author><name sortKey="Kirsten, A" sort="Kirsten, A" uniqKey="Kirsten A" first="A." last="Kirsten">A. Kirsten</name>
<affiliation wicri:level="3"><nlm:aff id="aff1-0022034514559250">Dental Materials and Biomaterials Research, RWTH Aachen University Hospital, Aachen, Germany</nlm:aff>
<country xml:lang="fr">Allemagne</country>
<wicri:regionArea>Dental Materials and Biomaterials Research, RWTH Aachen University Hospital, Aachen</wicri:regionArea>
<placeName><region type="land" nuts="1">Rhénanie-du-Nord-Westphalie</region>
<region type="district" nuts="2">District de Cologne</region>
<settlement type="city">Aix-la-Chapelle</settlement>
</placeName>
</affiliation>
</author>
<author><name sortKey="Hausmann, A" sort="Hausmann, A" uniqKey="Hausmann A" first="A." last="Hausmann">A. Hausmann</name>
<affiliation wicri:level="3"><nlm:aff id="aff1-0022034514559250">Dental Materials and Biomaterials Research, RWTH Aachen University Hospital, Aachen, Germany</nlm:aff>
<country xml:lang="fr">Allemagne</country>
<wicri:regionArea>Dental Materials and Biomaterials Research, RWTH Aachen University Hospital, Aachen</wicri:regionArea>
<placeName><region type="land" nuts="1">Rhénanie-du-Nord-Westphalie</region>
<region type="district" nuts="2">District de Cologne</region>
<settlement type="city">Aix-la-Chapelle</settlement>
</placeName>
</affiliation>
</author>
<author><name sortKey="Weber, M" sort="Weber, M" uniqKey="Weber M" first="M." last="Weber">M. Weber</name>
<affiliation wicri:level="3"><nlm:aff id="aff1-0022034514559250">Dental Materials and Biomaterials Research, RWTH Aachen University Hospital, Aachen, Germany</nlm:aff>
<country xml:lang="fr">Allemagne</country>
<wicri:regionArea>Dental Materials and Biomaterials Research, RWTH Aachen University Hospital, Aachen</wicri:regionArea>
<placeName><region type="land" nuts="1">Rhénanie-du-Nord-Westphalie</region>
<region type="district" nuts="2">District de Cologne</region>
<settlement type="city">Aix-la-Chapelle</settlement>
</placeName>
</affiliation>
</author>
<author><name sortKey="Fischer, J" sort="Fischer, J" uniqKey="Fischer J" first="J." last="Fischer">J. Fischer</name>
<affiliation wicri:level="1"><nlm:aff id="aff2-0022034514559250">Institute of Dental Materials and Engineering, University Hospital of Dental Medicine, University of Basel, Basel, Switzerland</nlm:aff>
<country xml:lang="fr">Suisse</country>
<wicri:regionArea>Institute of Dental Materials and Engineering, University Hospital of Dental Medicine, University of Basel, Basel</wicri:regionArea>
<wicri:noRegion>Basel</wicri:noRegion>
</affiliation>
</author>
<author><name sortKey="Fischer, H" sort="Fischer, H" uniqKey="Fischer H" first="H." last="Fischer">H. Fischer</name>
<affiliation wicri:level="3"><nlm:aff id="aff1-0022034514559250">Dental Materials and Biomaterials Research, RWTH Aachen University Hospital, Aachen, Germany</nlm:aff>
<country xml:lang="fr">Allemagne</country>
<wicri:regionArea>Dental Materials and Biomaterials Research, RWTH Aachen University Hospital, Aachen</wicri:regionArea>
<placeName><region type="land" nuts="1">Rhénanie-du-Nord-Westphalie</region>
<region type="district" nuts="2">District de Cologne</region>
<settlement type="city">Aix-la-Chapelle</settlement>
</placeName>
</affiliation>
</author>
</analytic>
<series><title level="j">Journal of Dental Research</title>
<idno type="ISSN">0022-0345</idno>
<idno type="eISSN">1544-0591</idno>
<imprint><date when="2015">2015</date>
</imprint>
</series>
</biblStruct>
</sourceDesc>
</fileDesc>
<profileDesc><textClass></textClass>
</profileDesc>
</teiHeader>
<front><div type="abstract" xml:lang="en"><p>The healing time of zirconia implants may be reduced by the use of bioactive glass coatings. Unfortunately, existing glasses are either bioactive like Bioglass 45S5 but thermally incompatible with the zirconia substrate, or they are thermally compatible but exhibit only a very low level of bioactivity. In this study, we hypothesized that a tailored substitution of alkaline earth metals and alkaline metals in 45S5 can lead to a glass composition that is both bioactive and thermally compatible with zirconia implants. A novel glass composition was analyzed using x-ray fluorescence spectroscopy, dilatometry, differential scanning calorimetry, and heating microscopy to investigate its chemical, physical, and thermal properties. Bioactivity was tested in vitro using simulated body fluid (SBF). Smooth and microstructured glass coatings were applied using a tailored spray technique with subsequent thermal treatment. Coating adhesion was tested on implants that were inserted in bovine ribs. The cytocompatibility of the coating was analyzed using L929 mouse fibroblasts. The coefficient of thermal expansion of the novel glass was shown to be slightly lower (11.58·10<sup>–6</sup>
K<sup>–1</sup>
) than that of the zirconia (11.67·10<sup>–6</sup>
K<sup>–1</sup>
). After storage in SBF, the glass showed reaction layers almost identical to the bioactive glass gold standard, 45S5. A process window between 800 °C and 910 °C was found to result in densely sintered and amorphous coatings. Microstructured glass coatings on zirconia implants survived a minimum insertion torque of 60 Ncm in the in vitro experiment on bovine ribs. Proliferation and cytotoxicity of the glass coatings was comparable with the controls. The novel glass composition showed a strong adhesion to the zirconia substrate and a significant bioactive behavior in the SBF in vitro experiments. Therefore, it holds great potential to significantly reduce the healing time of zirconia dental implants.</p>
</div>
</front>
</TEI>
<affiliations><list><country><li>Allemagne</li>
<li>Suisse</li>
</country>
<region><li>District de Cologne</li>
<li>Rhénanie-du-Nord-Westphalie</li>
</region>
<settlement><li>Aix-la-Chapelle</li>
</settlement>
</list>
<tree><country name="Allemagne"><region name="Rhénanie-du-Nord-Westphalie"><name sortKey="Kirsten, A" sort="Kirsten, A" uniqKey="Kirsten A" first="A." last="Kirsten">A. Kirsten</name>
</region>
<name sortKey="Fischer, H" sort="Fischer, H" uniqKey="Fischer H" first="H." last="Fischer">H. Fischer</name>
<name sortKey="Hausmann, A" sort="Hausmann, A" uniqKey="Hausmann A" first="A." last="Hausmann">A. Hausmann</name>
<name sortKey="Weber, M" sort="Weber, M" uniqKey="Weber M" first="M." last="Weber">M. Weber</name>
</country>
<country name="Suisse"><noRegion><name sortKey="Fischer, J" sort="Fischer, J" uniqKey="Fischer J" first="J." last="Fischer">J. Fischer</name>
</noRegion>
</country>
</tree>
</affiliations>
</record>
Pour manipuler ce document sous Unix (Dilib)
EXPLOR_STEP=$WICRI_ROOT/Wicri/Santé/explor/EdenteV2/Data/Main/Exploration
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 001379 | SxmlIndent | more
Ou
HfdSelect -h $EXPLOR_AREA/Data/Main/Exploration/biblio.hfd -nk 001379 | SxmlIndent | more
Pour mettre un lien sur cette page dans le réseau Wicri
{{Explor lien |wiki= Wicri/Santé |area= EdenteV2 |flux= Main |étape= Exploration |type= RBID |clé= PMC:4438729 |texte= Bioactive and Thermally Compatible Glass Coating on Zirconia Dental Implants }}
Pour générer des pages wiki
HfdIndexSelect -h $EXPLOR_AREA/Data/Main/Exploration/RBID.i -Sk "pubmed:25421839" \ | HfdSelect -Kh $EXPLOR_AREA/Data/Main/Exploration/biblio.hfd \ | NlmPubMed2Wicri -a EdenteV2
This area was generated with Dilib version V0.6.32. |