The Influence of the Tolerance between Mechanical Components on the Accuracy of Implants Inserted with a Stereolithographic Surgical Guide: A Retrospective Clinical Study.

Identifieur interne : 000628 ( PubMed/Corpus ); précédent : 000627; suivant : 000629

The Influence of the Tolerance between Mechanical Components on the Accuracy of Implants Inserted with a Stereolithographic Surgical Guide: A Retrospective Clinical Study.

Auteurs : Michele Cassetta ; Alfonso Di Mambro ; Gianni Di Giorgio ; Luigi V. Stefanelli ; Ersilia Barbato

Source :

RBID : pubmed:23879723

English descriptors

Abstract

The stereolithographic-guided surgery system involves a sequence of diagnostic and therapeutic events, and errors can arise at different stages. In these systems, one of the potentially clinically relevant errors may be the mechanical errors caused by the bur-guide gap due to the presence of a rotational allowance of the drills in the tubes.

DOI: 10.1111/cid.12120
PubMed: 23879723

Links to Exploration step

pubmed:23879723

Le document en format XML

<record>
<TEI>
<teiHeader>
<fileDesc>
<titleStmt>
<title xml:lang="en">The Influence of the Tolerance between Mechanical Components on the Accuracy of Implants Inserted with a Stereolithographic Surgical Guide: A Retrospective Clinical Study.</title>
<author>
<name sortKey="Cassetta, Michele" sort="Cassetta, Michele" uniqKey="Cassetta M" first="Michele" last="Cassetta">Michele Cassetta</name>
<affiliation>
<nlm:affiliation>Department of Oral and Maxillofacial Sciences, School of Dentistry, "Sapienza" University of Rome, Rome, Italy.</nlm:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Di Mambro, Alfonso" sort="Di Mambro, Alfonso" uniqKey="Di Mambro A" first="Alfonso" last="Di Mambro">Alfonso Di Mambro</name>
<affiliation>
<nlm:affiliation>Department of Oral and Maxillofacial Sciences, School of Dentistry, "Sapienza" University of Rome, Rome, Italy.</nlm:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Di Giorgio, Gianni" sort="Di Giorgio, Gianni" uniqKey="Di Giorgio G" first="Gianni" last="Di Giorgio">Gianni Di Giorgio</name>
<affiliation>
<nlm:affiliation>Department of Oral and Maxillofacial Sciences, School of Dentistry, "Sapienza" University of Rome, Rome, Italy.</nlm:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Stefanelli, Luigi V" sort="Stefanelli, Luigi V" uniqKey="Stefanelli L" first="Luigi V" last="Stefanelli">Luigi V. Stefanelli</name>
<affiliation>
<nlm:affiliation>Department of Oral and Maxillofacial Sciences, School of Dentistry, "Sapienza" University of Rome, Rome, Italy.</nlm:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Barbato, Ersilia" sort="Barbato, Ersilia" uniqKey="Barbato E" first="Ersilia" last="Barbato">Ersilia Barbato</name>
<affiliation>
<nlm:affiliation>Department of Oral and Maxillofacial Sciences, School of Dentistry, "Sapienza" University of Rome, Rome, Italy.</nlm:affiliation>
</affiliation>
</author>
</titleStmt>
<publicationStmt>
<idno type="wicri:source">PubMed</idno>
<date when="2015">2015</date>
<idno type="RBID">pubmed:23879723</idno>
<idno type="pmid">23879723</idno>
<idno type="doi">10.1111/cid.12120</idno>
<idno type="wicri:Area/PubMed/Corpus">000628</idno>
<idno type="wicri:explorRef" wicri:stream="PubMed" wicri:step="Corpus" wicri:corpus="PubMed">000628</idno>
</publicationStmt>
<sourceDesc>
<biblStruct>
<analytic>
<title xml:lang="en">The Influence of the Tolerance between Mechanical Components on the Accuracy of Implants Inserted with a Stereolithographic Surgical Guide: A Retrospective Clinical Study.</title>
<author>
<name sortKey="Cassetta, Michele" sort="Cassetta, Michele" uniqKey="Cassetta M" first="Michele" last="Cassetta">Michele Cassetta</name>
<affiliation>
<nlm:affiliation>Department of Oral and Maxillofacial Sciences, School of Dentistry, "Sapienza" University of Rome, Rome, Italy.</nlm:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Di Mambro, Alfonso" sort="Di Mambro, Alfonso" uniqKey="Di Mambro A" first="Alfonso" last="Di Mambro">Alfonso Di Mambro</name>
<affiliation>
<nlm:affiliation>Department of Oral and Maxillofacial Sciences, School of Dentistry, "Sapienza" University of Rome, Rome, Italy.</nlm:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Di Giorgio, Gianni" sort="Di Giorgio, Gianni" uniqKey="Di Giorgio G" first="Gianni" last="Di Giorgio">Gianni Di Giorgio</name>
<affiliation>
<nlm:affiliation>Department of Oral and Maxillofacial Sciences, School of Dentistry, "Sapienza" University of Rome, Rome, Italy.</nlm:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Stefanelli, Luigi V" sort="Stefanelli, Luigi V" uniqKey="Stefanelli L" first="Luigi V" last="Stefanelli">Luigi V. Stefanelli</name>
<affiliation>
<nlm:affiliation>Department of Oral and Maxillofacial Sciences, School of Dentistry, "Sapienza" University of Rome, Rome, Italy.</nlm:affiliation>
</affiliation>
</author>
<author>
<name sortKey="Barbato, Ersilia" sort="Barbato, Ersilia" uniqKey="Barbato E" first="Ersilia" last="Barbato">Ersilia Barbato</name>
<affiliation>
<nlm:affiliation>Department of Oral and Maxillofacial Sciences, School of Dentistry, "Sapienza" University of Rome, Rome, Italy.</nlm:affiliation>
</affiliation>
</author>
</analytic>
<series>
<title level="j">Clinical implant dentistry and related research</title>
<idno type="eISSN">1708-8208</idno>
<imprint>
<date when="2015" type="published">2015</date>
</imprint>
</series>
</biblStruct>
</sourceDesc>
</fileDesc>
<profileDesc>
<textClass>
<keywords scheme="KwdEn" xml:lang="en">
<term>Dental Implantation, Endosseous (instrumentation)</term>
<term>Dental Implantation, Endosseous (methods)</term>
<term>Dental Implants</term>
<term>Dental Instruments</term>
<term>Dental Prosthesis Design</term>
<term>Humans</term>
<term>Image Processing, Computer-Assisted</term>
<term>Imaging, Three-Dimensional</term>
<term>Mandible (diagnostic imaging)</term>
<term>Mandible (surgery)</term>
<term>Maxilla (diagnostic imaging)</term>
<term>Maxilla (surgery)</term>
<term>Mouth, Edentulous (rehabilitation)</term>
<term>Retrospective Studies</term>
<term>Surgery, Computer-Assisted (instrumentation)</term>
<term>Surgery, Computer-Assisted (methods)</term>
<term>Tomography, X-Ray Computed</term>
<term>Treatment Outcome</term>
</keywords>
<keywords scheme="MESH" type="chemical" xml:lang="en">
<term>Dental Implants</term>
</keywords>
<keywords scheme="MESH" qualifier="diagnostic imaging" xml:lang="en">
<term>Mandible</term>
<term>Maxilla</term>
</keywords>
<keywords scheme="MESH" qualifier="instrumentation" xml:lang="en">
<term>Dental Implantation, Endosseous</term>
<term>Surgery, Computer-Assisted</term>
</keywords>
<keywords scheme="MESH" qualifier="methods" xml:lang="en">
<term>Dental Implantation, Endosseous</term>
<term>Surgery, Computer-Assisted</term>
</keywords>
<keywords scheme="MESH" qualifier="rehabilitation" xml:lang="en">
<term>Mouth, Edentulous</term>
</keywords>
<keywords scheme="MESH" qualifier="surgery" xml:lang="en">
<term>Mandible</term>
<term>Maxilla</term>
</keywords>
<keywords scheme="MESH" xml:lang="en">
<term>Dental Instruments</term>
<term>Dental Prosthesis Design</term>
<term>Humans</term>
<term>Image Processing, Computer-Assisted</term>
<term>Imaging, Three-Dimensional</term>
<term>Retrospective Studies</term>
<term>Tomography, X-Ray Computed</term>
<term>Treatment Outcome</term>
</keywords>
</textClass>
</profileDesc>
</teiHeader>
<front>
<div type="abstract" xml:lang="en">The stereolithographic-guided surgery system involves a sequence of diagnostic and therapeutic events, and errors can arise at different stages. In these systems, one of the potentially clinically relevant errors may be the mechanical errors caused by the bur-guide gap due to the presence of a rotational allowance of the drills in the tubes.</div>
</front>
</TEI>
<pubmed>
<MedlineCitation Status="MEDLINE" Owner="NLM">
<PMID Version="1">23879723</PMID>
<DateCompleted>
<Year>2017</Year>
<Month>02</Month>
<Day>22</Day>
</DateCompleted>
<DateRevised>
<Year>2017</Year>
<Month>02</Month>
<Day>22</Day>
</DateRevised>
<Article PubModel="Print-Electronic">
<Journal>
<ISSN IssnType="Electronic">1708-8208</ISSN>
<JournalIssue CitedMedium="Internet">
<Volume>17</Volume>
<Issue>3</Issue>
<PubDate>
<Year>2015</Year>
<Month>Jun</Month>
</PubDate>
</JournalIssue>
<Title>Clinical implant dentistry and related research</Title>
<ISOAbbreviation>Clin Implant Dent Relat Res</ISOAbbreviation>
</Journal>
<ArticleTitle>The Influence of the Tolerance between Mechanical Components on the Accuracy of Implants Inserted with a Stereolithographic Surgical Guide: A Retrospective Clinical Study.</ArticleTitle>
<Pagination>
<MedlinePgn>580-8</MedlinePgn>
</Pagination>
<ELocationID EIdType="doi" ValidYN="Y">10.1111/cid.12120</ELocationID>
<Abstract>
<AbstractText Label="BACKGROUND" NlmCategory="BACKGROUND">The stereolithographic-guided surgery system involves a sequence of diagnostic and therapeutic events, and errors can arise at different stages. In these systems, one of the potentially clinically relevant errors may be the mechanical errors caused by the bur-guide gap due to the presence of a rotational allowance of the drills in the tubes.</AbstractText>
<AbstractText Label="PURPOSE" NlmCategory="OBJECTIVE">The purpose of this retrospective clinical study is to determine if it is possible to reduce the total error by limiting the tolerance among the mechanical components and to evaluate its clinical incidence.</AbstractText>
<AbstractText Label="MATERIALS AND METHODS" NlmCategory="METHODS">Sixty-six implants were inserted using the External Hex Safe® (Materialise Dental, Leuven, Belgium) system (Group A), and 71 implants were inserted using the same system with mechanical components modified to minimize the tolerance (Group B). Regarding only the angular deviation values, the t-test was used to determine the influence of reduced tolerance among the mechanical components on the accuracy values.</AbstractText>
<AbstractText Label="RESULTS" NlmCategory="RESULTS">t-Test showed that there is a statistically significant better accuracy with the modified system (Group B).</AbstractText>
<AbstractText Label="CONCLUSIONS" NlmCategory="CONCLUSIONS">Limiting the error that originates from mechanical components, total error could be statistically significantly reduced. Mechanical error is one of the most important source of error using External Hex Safe stereolithographic surgical guide.</AbstractText>
<CopyrightInformation>© 2013 Wiley Periodicals, Inc.</CopyrightInformation>
</Abstract>
<AuthorList CompleteYN="Y">
<Author ValidYN="Y">
<LastName>Cassetta</LastName>
<ForeName>Michele</ForeName>
<Initials>M</Initials>
<AffiliationInfo>
<Affiliation>Department of Oral and Maxillofacial Sciences, School of Dentistry, "Sapienza" University of Rome, Rome, Italy.</Affiliation>
</AffiliationInfo>
</Author>
<Author ValidYN="Y">
<LastName>Di Mambro</LastName>
<ForeName>Alfonso</ForeName>
<Initials>A</Initials>
<AffiliationInfo>
<Affiliation>Department of Oral and Maxillofacial Sciences, School of Dentistry, "Sapienza" University of Rome, Rome, Italy.</Affiliation>
</AffiliationInfo>
</Author>
<Author ValidYN="Y">
<LastName>Di Giorgio</LastName>
<ForeName>Gianni</ForeName>
<Initials>G</Initials>
<AffiliationInfo>
<Affiliation>Department of Oral and Maxillofacial Sciences, School of Dentistry, "Sapienza" University of Rome, Rome, Italy.</Affiliation>
</AffiliationInfo>
</Author>
<Author ValidYN="Y">
<LastName>Stefanelli</LastName>
<ForeName>Luigi V</ForeName>
<Initials>LV</Initials>
<AffiliationInfo>
<Affiliation>Department of Oral and Maxillofacial Sciences, School of Dentistry, "Sapienza" University of Rome, Rome, Italy.</Affiliation>
</AffiliationInfo>
</Author>
<Author ValidYN="Y">
<LastName>Barbato</LastName>
<ForeName>Ersilia</ForeName>
<Initials>E</Initials>
<AffiliationInfo>
<Affiliation>Department of Oral and Maxillofacial Sciences, School of Dentistry, "Sapienza" University of Rome, Rome, Italy.</Affiliation>
</AffiliationInfo>
</Author>
</AuthorList>
<Language>eng</Language>
<PublicationTypeList>
<PublicationType UI="D000068397">Clinical Study</PublicationType>
<PublicationType UI="D016428">Journal Article</PublicationType>
</PublicationTypeList>
<ArticleDate DateType="Electronic">
<Year>2013</Year>
<Month>07</Month>
<Day>24</Day>
</ArticleDate>
</Article>
<MedlineJournalInfo>
<Country>United States</Country>
<MedlineTA>Clin Implant Dent Relat Res</MedlineTA>
<NlmUniqueID>100888977</NlmUniqueID>
<ISSNLinking>1523-0899</ISSNLinking>
</MedlineJournalInfo>
<ChemicalList>
<Chemical>
<RegistryNumber>0</RegistryNumber>
<NameOfSubstance UI="D015921">Dental Implants</NameOfSubstance>
</Chemical>
</ChemicalList>
<CitationSubset>D</CitationSubset>
<CitationSubset>IM</CitationSubset>
<MeshHeadingList>
<MeshHeading>
<DescriptorName UI="D003758" MajorTopicYN="N">Dental Implantation, Endosseous</DescriptorName>
<QualifierName UI="Q000295" MajorTopicYN="N">instrumentation</QualifierName>
<QualifierName UI="Q000379" MajorTopicYN="Y">methods</QualifierName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D015921" MajorTopicYN="Y">Dental Implants</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D003762" MajorTopicYN="N">Dental Instruments</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D017267" MajorTopicYN="N">Dental Prosthesis Design</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D006801" MajorTopicYN="N">Humans</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D007091" MajorTopicYN="N">Image Processing, Computer-Assisted</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D021621" MajorTopicYN="N">Imaging, Three-Dimensional</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D008334" MajorTopicYN="N">Mandible</DescriptorName>
<QualifierName UI="Q000000981" MajorTopicYN="N">diagnostic imaging</QualifierName>
<QualifierName UI="Q000601" MajorTopicYN="N">surgery</QualifierName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D008437" MajorTopicYN="N">Maxilla</DescriptorName>
<QualifierName UI="Q000000981" MajorTopicYN="N">diagnostic imaging</QualifierName>
<QualifierName UI="Q000601" MajorTopicYN="N">surgery</QualifierName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D009066" MajorTopicYN="N">Mouth, Edentulous</DescriptorName>
<QualifierName UI="Q000534" MajorTopicYN="N">rehabilitation</QualifierName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D012189" MajorTopicYN="N">Retrospective Studies</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D025321" MajorTopicYN="N">Surgery, Computer-Assisted</DescriptorName>
<QualifierName UI="Q000295" MajorTopicYN="N">instrumentation</QualifierName>
<QualifierName UI="Q000379" MajorTopicYN="Y">methods</QualifierName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D014057" MajorTopicYN="N">Tomography, X-Ray Computed</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D016896" MajorTopicYN="N">Treatment Outcome</DescriptorName>
</MeshHeading>
</MeshHeadingList>
<KeywordList Owner="NOTNLM">
<Keyword MajorTopicYN="N">CAD/CAM technology</Keyword>
<Keyword MajorTopicYN="N">accuracy</Keyword>
<Keyword MajorTopicYN="N">clinical study</Keyword>
<Keyword MajorTopicYN="N">computer-assisted</Keyword>
<Keyword MajorTopicYN="N">flapless implant surgery</Keyword>
<Keyword MajorTopicYN="N">implant</Keyword>
<Keyword MajorTopicYN="N">implantology</Keyword>
<Keyword MajorTopicYN="N">retrospective</Keyword>
<Keyword MajorTopicYN="N">stereo lithography</Keyword>
</KeywordList>
</MedlineCitation>
<PubmedData>
<History>
<PubMedPubDate PubStatus="entrez">
<Year>2013</Year>
<Month>7</Month>
<Day>25</Day>
<Hour>6</Hour>
<Minute>0</Minute>
</PubMedPubDate>
<PubMedPubDate PubStatus="pubmed">
<Year>2013</Year>
<Month>7</Month>
<Day>25</Day>
<Hour>6</Hour>
<Minute>0</Minute>
</PubMedPubDate>
<PubMedPubDate PubStatus="medline">
<Year>2017</Year>
<Month>2</Month>
<Day>23</Day>
<Hour>6</Hour>
<Minute>0</Minute>
</PubMedPubDate>
</History>
<PublicationStatus>ppublish</PublicationStatus>
<ArticleIdList>
<ArticleId IdType="pubmed">23879723</ArticleId>
<ArticleId IdType="doi">10.1111/cid.12120</ArticleId>
</ArticleIdList>
</PubmedData>
</pubmed>
</record>

Pour manipuler ce document sous Unix (Dilib)

EXPLOR_STEP=$WICRI_ROOT/Wicri/Santé/explor/EdenteV1/Data/PubMed/Corpus
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 000628 | SxmlIndent | more

Ou

HfdSelect -h $EXPLOR_AREA/Data/PubMed/Corpus/biblio.hfd -nk 000628 | SxmlIndent | more

Pour mettre un lien sur cette page dans le réseau Wicri

{{Explor lien
   |wiki=    Wicri/Santé
   |area=    EdenteV1
   |flux=    PubMed
   |étape=   Corpus
   |type=    RBID
   |clé=     pubmed:23879723
   |texte=   The Influence of the Tolerance between Mechanical Components on the Accuracy of Implants Inserted with a Stereolithographic Surgical Guide: A Retrospective Clinical Study.
}}

Pour générer des pages wiki

HfdIndexSelect -h $EXPLOR_AREA/Data/PubMed/Corpus/RBID.i   -Sk "pubmed:23879723" \
       | HfdSelect -Kh $EXPLOR_AREA/Data/PubMed/Corpus/biblio.hfd   \
       | NlmPubMed2Wicri -a EdenteV1 

Wicri

This area was generated with Dilib version V0.6.33.
Data generation: Mon Dec 4 11:02:15 2017. Site generation: Tue Sep 29 19:14:38 2020