A photoelastic stress analysis of screw- and cement-retained implant prostheses with marginal gaps.

Identifieur interne : 000B61 ( PubMed/Checkpoint ); précédent : 000B60; suivant : 000B62

A photoelastic stress analysis of screw- and cement-retained implant prostheses with marginal gaps.

Auteurs : Jae-In Lee [Corée du Sud] ; Yoon Lee ; Nan-Young Kim ; Yu-Lee Kim ; Hye-Won Cho

Source :

RBID : pubmed:23927066

Descripteurs français

English descriptors

Abstract

The precise fit of an implant prosthesis is considered to be a prerequisite for the success and maintenance of osseointegration. It is unknown how much static stress can be tolerated at the implant-bone interface with ill-fitting prostheses for the two different types of retention (cement vs screw).

DOI: 10.1111/cid.12134
PubMed: 23927066


Affiliations:


Links toward previous steps (curation, corpus...)


Links to Exploration step

pubmed:23927066

Le document en format XML

<record>
<TEI>
<teiHeader>
<fileDesc>
<titleStmt>
<title xml:lang="en">A photoelastic stress analysis of screw- and cement-retained implant prostheses with marginal gaps.</title>
<author>
<name sortKey="Lee, Jae In" sort="Lee, Jae In" uniqKey="Lee J" first="Jae-In" last="Lee">Jae-In Lee</name>
<affiliation wicri:level="1">
<nlm:affiliation>Department of Prosthodontics, College of Dentistry, Wonkwang University, Iksan, Korea.</nlm:affiliation>
<country xml:lang="fr">Corée du Sud</country>
<wicri:regionArea>Department of Prosthodontics, College of Dentistry, Wonkwang University, Iksan</wicri:regionArea>
<wicri:noRegion>Iksan</wicri:noRegion>
</affiliation>
</author>
<author>
<name sortKey="Lee, Yoon" sort="Lee, Yoon" uniqKey="Lee Y" first="Yoon" last="Lee">Yoon Lee</name>
</author>
<author>
<name sortKey="Kim, Nan Young" sort="Kim, Nan Young" uniqKey="Kim N" first="Nan-Young" last="Kim">Nan-Young Kim</name>
</author>
<author>
<name sortKey="Kim, Yu Lee" sort="Kim, Yu Lee" uniqKey="Kim Y" first="Yu-Lee" last="Kim">Yu-Lee Kim</name>
</author>
<author>
<name sortKey="Cho, Hye Won" sort="Cho, Hye Won" uniqKey="Cho H" first="Hye-Won" last="Cho">Hye-Won Cho</name>
</author>
</titleStmt>
<publicationStmt>
<idno type="wicri:source">PubMed</idno>
<date when="2013">2013</date>
<idno type="RBID">pubmed:23927066</idno>
<idno type="pmid">23927066</idno>
<idno type="doi">10.1111/cid.12134</idno>
<idno type="wicri:Area/PubMed/Corpus">000478</idno>
<idno type="wicri:explorRef" wicri:stream="PubMed" wicri:step="Corpus" wicri:corpus="PubMed">000478</idno>
<idno type="wicri:Area/PubMed/Curation">000478</idno>
<idno type="wicri:explorRef" wicri:stream="PubMed" wicri:step="Curation">000478</idno>
<idno type="wicri:Area/PubMed/Checkpoint">000478</idno>
<idno type="wicri:explorRef" wicri:stream="Checkpoint" wicri:step="PubMed">000478</idno>
</publicationStmt>
<sourceDesc>
<biblStruct>
<analytic>
<title xml:lang="en">A photoelastic stress analysis of screw- and cement-retained implant prostheses with marginal gaps.</title>
<author>
<name sortKey="Lee, Jae In" sort="Lee, Jae In" uniqKey="Lee J" first="Jae-In" last="Lee">Jae-In Lee</name>
<affiliation wicri:level="1">
<nlm:affiliation>Department of Prosthodontics, College of Dentistry, Wonkwang University, Iksan, Korea.</nlm:affiliation>
<country xml:lang="fr">Corée du Sud</country>
<wicri:regionArea>Department of Prosthodontics, College of Dentistry, Wonkwang University, Iksan</wicri:regionArea>
<wicri:noRegion>Iksan</wicri:noRegion>
</affiliation>
</author>
<author>
<name sortKey="Lee, Yoon" sort="Lee, Yoon" uniqKey="Lee Y" first="Yoon" last="Lee">Yoon Lee</name>
</author>
<author>
<name sortKey="Kim, Nan Young" sort="Kim, Nan Young" uniqKey="Kim N" first="Nan-Young" last="Kim">Nan-Young Kim</name>
</author>
<author>
<name sortKey="Kim, Yu Lee" sort="Kim, Yu Lee" uniqKey="Kim Y" first="Yu-Lee" last="Kim">Yu-Lee Kim</name>
</author>
<author>
<name sortKey="Cho, Hye Won" sort="Cho, Hye Won" uniqKey="Cho H" first="Hye-Won" last="Cho">Hye-Won Cho</name>
</author>
</analytic>
<series>
<title level="j">Clinical implant dentistry and related research</title>
<idno type="eISSN">1708-8208</idno>
<imprint>
<date when="2013" type="published">2013</date>
</imprint>
</series>
</biblStruct>
</sourceDesc>
</fileDesc>
<profileDesc>
<textClass>
<keywords scheme="KwdEn" xml:lang="en">
<term>Bone Screws</term>
<term>Dental Cements</term>
<term>Dental Implants</term>
<term>Humans</term>
<term>Stress, Mechanical</term>
</keywords>
<keywords scheme="KwdFr" xml:lang="fr">
<term>Ciments dentaires</term>
<term>Contrainte mécanique</term>
<term>Humains</term>
<term>Implants dentaires</term>
<term>Vis orthopédiques</term>
</keywords>
<keywords scheme="MESH" type="chemical" xml:lang="en">
<term>Dental Cements</term>
<term>Dental Implants</term>
</keywords>
<keywords scheme="MESH" xml:lang="en">
<term>Bone Screws</term>
<term>Humans</term>
<term>Stress, Mechanical</term>
</keywords>
<keywords scheme="MESH" xml:lang="fr">
<term>Ciments dentaires</term>
<term>Contrainte mécanique</term>
<term>Humains</term>
<term>Implants dentaires</term>
<term>Vis orthopédiques</term>
</keywords>
</textClass>
</profileDesc>
</teiHeader>
<front>
<div type="abstract" xml:lang="en">The precise fit of an implant prosthesis is considered to be a prerequisite for the success and maintenance of osseointegration. It is unknown how much static stress can be tolerated at the implant-bone interface with ill-fitting prostheses for the two different types of retention (cement vs screw).</div>
</front>
</TEI>
<pubmed>
<MedlineCitation Status="MEDLINE" Owner="NLM">
<PMID Version="1">23927066</PMID>
<DateCompleted>
<Year>2015</Year>
<Month>09</Month>
<Day>10</Day>
</DateCompleted>
<DateRevised>
<Year>2013</Year>
<Month>10</Month>
<Day>11</Day>
</DateRevised>
<Article PubModel="Print-Electronic">
<Journal>
<ISSN IssnType="Electronic">1708-8208</ISSN>
<JournalIssue CitedMedium="Internet">
<Volume>15</Volume>
<Issue>5</Issue>
<PubDate>
<Year>2013</Year>
<Month>Oct</Month>
</PubDate>
</JournalIssue>
<Title>Clinical implant dentistry and related research</Title>
<ISOAbbreviation>Clin Implant Dent Relat Res</ISOAbbreviation>
</Journal>
<ArticleTitle>A photoelastic stress analysis of screw- and cement-retained implant prostheses with marginal gaps.</ArticleTitle>
<Pagination>
<MedlinePgn>735-49</MedlinePgn>
</Pagination>
<ELocationID EIdType="doi" ValidYN="Y">10.1111/cid.12134</ELocationID>
<Abstract>
<AbstractText Label="BACKGROUND" NlmCategory="BACKGROUND">The precise fit of an implant prosthesis is considered to be a prerequisite for the success and maintenance of osseointegration. It is unknown how much static stress can be tolerated at the implant-bone interface with ill-fitting prostheses for the two different types of retention (cement vs screw).</AbstractText>
<AbstractText Label="PURPOSE" NlmCategory="OBJECTIVE">The purpose of this study was to evaluate the stress pattern and magnitude in the supporting tissues around ITI (Straumann AG, Waldenburg, Switzerland) implants with screw- or cement-retained prostheses with marginal gaps by photoelastic analysis.</AbstractText>
<AbstractText Label="MATERIALS AND METHODS" NlmCategory="METHODS">A photoelastic model of a human mandible, partially edentulous distal to the canine, was made of PL-2 resin. Three ITI implants (4.1 × 10 mm, Straumann AG, Waldenburg, Switzerland) were placed in the posterior edentulous region, and screw- or cement-retained three-unit fixed partial dentures (FPDs) were fabricated. Ill-fitting prostheses were made by placing a 100-μm gap between the abutments and the superstructures on the second premolar or the first molar. A static vertical force of 134 N was applied at three loading points on each prosthesis. Photoelastic stress analysis was carried out to measure the fringe order around the implant-supporting structures.</AbstractText>
<AbstractText Label="RESULTS" NlmCategory="RESULTS">Even in the unloaded condition, low-level stresses were generated around the implants after screw tightening or cementing the three-unit FPDs with marginal gaps. Loading on the terminal implants developed high concentrated stresses around the loaded implant, regardless of the types of restorations or the presence of gaps. However, when the middle implant was loaded, moderate stresses were distributed to the anterior and posterior implants.</AbstractText>
<AbstractText Label="CONCLUSIONS" NlmCategory="CONCLUSIONS">Screw-retained FPDs with gaps exhibited a wider range of stresses on the interproximal region of adjacent implants than cement-retained FPDs. However, severe misfit in the prosthesis caused the nonaxial stress transfer to the adjacent implants in the cement-retained FPDs with gaps.</AbstractText>
<CopyrightInformation>© 2013 Wiley Periodicals, Inc.</CopyrightInformation>
</Abstract>
<AuthorList CompleteYN="Y">
<Author ValidYN="Y">
<LastName>Lee</LastName>
<ForeName>Jae-In</ForeName>
<Initials>JI</Initials>
<AffiliationInfo>
<Affiliation>Department of Prosthodontics, College of Dentistry, Wonkwang University, Iksan, Korea.</Affiliation>
</AffiliationInfo>
</Author>
<Author ValidYN="Y">
<LastName>Lee</LastName>
<ForeName>Yoon</ForeName>
<Initials>Y</Initials>
</Author>
<Author ValidYN="Y">
<LastName>Kim</LastName>
<ForeName>Nan-Young</ForeName>
<Initials>NY</Initials>
</Author>
<Author ValidYN="Y">
<LastName>Kim</LastName>
<ForeName>Yu-Lee</ForeName>
<Initials>YL</Initials>
</Author>
<Author ValidYN="Y">
<LastName>Cho</LastName>
<ForeName>Hye-Won</ForeName>
<Initials>HW</Initials>
</Author>
</AuthorList>
<Language>eng</Language>
<PublicationTypeList>
<PublicationType UI="D016428">Journal Article</PublicationType>
<PublicationType UI="D013485">Research Support, Non-U.S. Gov't</PublicationType>
</PublicationTypeList>
<ArticleDate DateType="Electronic">
<Year>2013</Year>
<Month>08</Month>
<Day>09</Day>
</ArticleDate>
</Article>
<MedlineJournalInfo>
<Country>United States</Country>
<MedlineTA>Clin Implant Dent Relat Res</MedlineTA>
<NlmUniqueID>100888977</NlmUniqueID>
<ISSNLinking>1523-0899</ISSNLinking>
</MedlineJournalInfo>
<ChemicalList>
<Chemical>
<RegistryNumber>0</RegistryNumber>
<NameOfSubstance UI="D003738">Dental Cements</NameOfSubstance>
</Chemical>
<Chemical>
<RegistryNumber>0</RegistryNumber>
<NameOfSubstance UI="D015921">Dental Implants</NameOfSubstance>
</Chemical>
</ChemicalList>
<CitationSubset>D</CitationSubset>
<CitationSubset>IM</CitationSubset>
<MeshHeadingList>
<MeshHeading>
<DescriptorName UI="D001863" MajorTopicYN="Y">Bone Screws</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D003738" MajorTopicYN="Y">Dental Cements</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D015921" MajorTopicYN="Y">Dental Implants</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D006801" MajorTopicYN="N">Humans</DescriptorName>
</MeshHeading>
<MeshHeading>
<DescriptorName UI="D013314" MajorTopicYN="Y">Stress, Mechanical</DescriptorName>
</MeshHeading>
</MeshHeadingList>
<KeywordList Owner="NOTNLM">
<Keyword MajorTopicYN="N">cement-retained FPD</Keyword>
<Keyword MajorTopicYN="N">dental implant</Keyword>
<Keyword MajorTopicYN="N">gap prosthesis</Keyword>
<Keyword MajorTopicYN="N">photoelasticity</Keyword>
<Keyword MajorTopicYN="N">screw-retained FPD</Keyword>
</KeywordList>
</MedlineCitation>
<PubmedData>
<History>
<PubMedPubDate PubStatus="entrez">
<Year>2013</Year>
<Month>8</Month>
<Day>10</Day>
<Hour>6</Hour>
<Minute>0</Minute>
</PubMedPubDate>
<PubMedPubDate PubStatus="pubmed">
<Year>2013</Year>
<Month>8</Month>
<Day>10</Day>
<Hour>6</Hour>
<Minute>0</Minute>
</PubMedPubDate>
<PubMedPubDate PubStatus="medline">
<Year>2015</Year>
<Month>9</Month>
<Day>12</Day>
<Hour>6</Hour>
<Minute>0</Minute>
</PubMedPubDate>
</History>
<PublicationStatus>ppublish</PublicationStatus>
<ArticleIdList>
<ArticleId IdType="pubmed">23927066</ArticleId>
<ArticleId IdType="doi">10.1111/cid.12134</ArticleId>
</ArticleIdList>
</PubmedData>
</pubmed>
<affiliations>
<list>
<country>
<li>Corée du Sud</li>
</country>
</list>
<tree>
<noCountry>
<name sortKey="Cho, Hye Won" sort="Cho, Hye Won" uniqKey="Cho H" first="Hye-Won" last="Cho">Hye-Won Cho</name>
<name sortKey="Kim, Nan Young" sort="Kim, Nan Young" uniqKey="Kim N" first="Nan-Young" last="Kim">Nan-Young Kim</name>
<name sortKey="Kim, Yu Lee" sort="Kim, Yu Lee" uniqKey="Kim Y" first="Yu-Lee" last="Kim">Yu-Lee Kim</name>
<name sortKey="Lee, Yoon" sort="Lee, Yoon" uniqKey="Lee Y" first="Yoon" last="Lee">Yoon Lee</name>
</noCountry>
<country name="Corée du Sud">
<noRegion>
<name sortKey="Lee, Jae In" sort="Lee, Jae In" uniqKey="Lee J" first="Jae-In" last="Lee">Jae-In Lee</name>
</noRegion>
</country>
</tree>
</affiliations>
</record>

Pour manipuler ce document sous Unix (Dilib)

EXPLOR_STEP=$WICRI_ROOT/Wicri/Santé/explor/EdenteV1/Data/PubMed/Checkpoint
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 000B61 | SxmlIndent | more

Ou

HfdSelect -h $EXPLOR_AREA/Data/PubMed/Checkpoint/biblio.hfd -nk 000B61 | SxmlIndent | more

Pour mettre un lien sur cette page dans le réseau Wicri

{{Explor lien
   |wiki=    Wicri/Santé
   |area=    EdenteV1
   |flux=    PubMed
   |étape=   Checkpoint
   |type=    RBID
   |clé=     pubmed:23927066
   |texte=   A photoelastic stress analysis of screw- and cement-retained implant prostheses with marginal gaps.
}}

Pour générer des pages wiki

HfdIndexSelect -h $EXPLOR_AREA/Data/PubMed/Checkpoint/RBID.i   -Sk "pubmed:23927066" \
       | HfdSelect -Kh $EXPLOR_AREA/Data/PubMed/Checkpoint/biblio.hfd   \
       | NlmPubMed2Wicri -a EdenteV1 

Wicri

This area was generated with Dilib version V0.6.33.
Data generation: Mon Dec 4 11:02:15 2017. Site generation: Tue Sep 29 19:14:38 2020