Safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (enVista®) – results of a European and Asian-Pacific study

Identifieur interne : 002E73 ( Ncbi/Merge ); précédent : 002E72; suivant : 002E74

Safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (enVista®) – results of a European and Asian-Pacific study

Auteurs : Peter Heiner [Australie] ; Edoardo Ligabue [Italie] ; Alex Fan [Hong Kong] ; Dennis Lam [Hong Kong]

Source :

RBID : PMC:3974695

Abstract

Purpose

To evaluate the safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (IOL) (enVista® MX60; Bausch and Lomb Incorporated, Rochester, NY, USA) following implantation to correct aphakia subsequent to extracapsular cataract extraction in adults.

Subjects and methods

This was an open-label, non-interventional, observational study conducted in 19 university and private-practice settings in Europe and the Asia-Pacific region to investigate clinical outcomes of the MX60 IOL in standard practice. Eligible subjects were at least 18 years of age and had undergone standard phacoemulsification and extracapsular cataract extraction with implantation of the MX60 IOL. The primary safety endpoint was the occurrence of adverse events, and the primary effectiveness endpoints included visual and refractive outcomes and stability, with data collected up to 2 years post-procedure.

Results

In this multicenter study, pooled data of 255 eyes were collected and analyzed. Excellent visual and refractive outcomes and stability were demonstrated. At postoperative visit 4 (61–180 days postoperative), 62.2% of subjects achieved a Snellen best-corrected distance visual acuity (CDVA) of 20/20 (decimal 1.00), and 97.8% of subjects achieved a CDVA of 20/40 (decimal 0.50) or better. One eye (1.0%) underwent neodymium:yttrium aluminum garnet capsulotomy at 12 months post-procedure. No glistenings of any grade were reported for any subject at any visit. Adverse events were infrequent and were consistent with incidences generally reported with cataract surgery.

Conclusion

This study, which enrolled all comers, provided evidence of the excellent safety and effectiveness of the MX60 IOL in standard practice. Favorable clinical outcomes included outstanding visual and refractive outcomes and stability. No glistenings were reported at any postoperative visit.


Url:
DOI: 10.2147/OPTH.S56135
PubMed: 24729678
PubMed Central: 3974695

Links toward previous steps (curation, corpus...)


Links to Exploration step

PMC:3974695

Le document en format XML

<record>
<TEI>
<teiHeader>
<fileDesc>
<titleStmt>
<title xml:lang="en">Safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (enVista®) – results of a European and Asian-Pacific study</title>
<author>
<name sortKey="Heiner, Peter" sort="Heiner, Peter" uniqKey="Heiner P" first="Peter" last="Heiner">Peter Heiner</name>
<affiliation wicri:level="1">
<nlm:aff id="af1-opth-8-629">Vision Eye Institute, Southport, QLD, Australia</nlm:aff>
<country xml:lang="fr">Australie</country>
<wicri:regionArea>Vision Eye Institute, Southport, QLD</wicri:regionArea>
<wicri:noRegion>QLD</wicri:noRegion>
</affiliation>
</author>
<author>
<name sortKey="Ligabue, Edoardo" sort="Ligabue, Edoardo" uniqKey="Ligabue E" first="Edoardo" last="Ligabue">Edoardo Ligabue</name>
<affiliation wicri:level="1">
<nlm:aff id="af2-opth-8-629">Ophthalmic Center, San Siro Clinic, Milan, Italy</nlm:aff>
<country xml:lang="fr">Italie</country>
<wicri:regionArea>Ophthalmic Center, San Siro Clinic, Milan</wicri:regionArea>
<wicri:noRegion>Milan</wicri:noRegion>
</affiliation>
</author>
<author>
<name sortKey="Fan, Alex" sort="Fan, Alex" uniqKey="Fan A" first="Alex" last="Fan">Alex Fan</name>
<affiliation wicri:level="4">
<nlm:aff id="af3-opth-8-629">Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Shatin, New Territories, Hong Kong</nlm:aff>
<country xml:lang="fr">Hong Kong</country>
<wicri:regionArea>Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Shatin, New Territories</wicri:regionArea>
<orgName type="university">Université chinoise de Hong Kong</orgName>
<placeName>
<settlement type="city">Sha Tin</settlement>
</placeName>
</affiliation>
</author>
<author>
<name sortKey="Lam, Dennis" sort="Lam, Dennis" uniqKey="Lam D" first="Dennis" last="Lam">Dennis Lam</name>
<affiliation wicri:level="4">
<nlm:aff id="af3-opth-8-629">Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Shatin, New Territories, Hong Kong</nlm:aff>
<country xml:lang="fr">Hong Kong</country>
<wicri:regionArea>Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Shatin, New Territories</wicri:regionArea>
<orgName type="university">Université chinoise de Hong Kong</orgName>
<placeName>
<settlement type="city">Sha Tin</settlement>
</placeName>
</affiliation>
</author>
</titleStmt>
<publicationStmt>
<idno type="wicri:source">PMC</idno>
<idno type="pmid">24729678</idno>
<idno type="pmc">3974695</idno>
<idno type="url">http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3974695</idno>
<idno type="RBID">PMC:3974695</idno>
<idno type="doi">10.2147/OPTH.S56135</idno>
<date when="2014">2014</date>
<idno type="wicri:Area/Pmc/Corpus">001C41</idno>
<idno type="wicri:Area/Pmc/Curation">001C41</idno>
<idno type="wicri:Area/Pmc/Checkpoint">000972</idno>
<idno type="wicri:Area/Ncbi/Merge">002E73</idno>
</publicationStmt>
<sourceDesc>
<biblStruct>
<analytic>
<title xml:lang="en" level="a" type="main">Safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (enVista®) – results of a European and Asian-Pacific study</title>
<author>
<name sortKey="Heiner, Peter" sort="Heiner, Peter" uniqKey="Heiner P" first="Peter" last="Heiner">Peter Heiner</name>
<affiliation wicri:level="1">
<nlm:aff id="af1-opth-8-629">Vision Eye Institute, Southport, QLD, Australia</nlm:aff>
<country xml:lang="fr">Australie</country>
<wicri:regionArea>Vision Eye Institute, Southport, QLD</wicri:regionArea>
<wicri:noRegion>QLD</wicri:noRegion>
</affiliation>
</author>
<author>
<name sortKey="Ligabue, Edoardo" sort="Ligabue, Edoardo" uniqKey="Ligabue E" first="Edoardo" last="Ligabue">Edoardo Ligabue</name>
<affiliation wicri:level="1">
<nlm:aff id="af2-opth-8-629">Ophthalmic Center, San Siro Clinic, Milan, Italy</nlm:aff>
<country xml:lang="fr">Italie</country>
<wicri:regionArea>Ophthalmic Center, San Siro Clinic, Milan</wicri:regionArea>
<wicri:noRegion>Milan</wicri:noRegion>
</affiliation>
</author>
<author>
<name sortKey="Fan, Alex" sort="Fan, Alex" uniqKey="Fan A" first="Alex" last="Fan">Alex Fan</name>
<affiliation wicri:level="4">
<nlm:aff id="af3-opth-8-629">Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Shatin, New Territories, Hong Kong</nlm:aff>
<country xml:lang="fr">Hong Kong</country>
<wicri:regionArea>Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Shatin, New Territories</wicri:regionArea>
<orgName type="university">Université chinoise de Hong Kong</orgName>
<placeName>
<settlement type="city">Sha Tin</settlement>
</placeName>
</affiliation>
</author>
<author>
<name sortKey="Lam, Dennis" sort="Lam, Dennis" uniqKey="Lam D" first="Dennis" last="Lam">Dennis Lam</name>
<affiliation wicri:level="4">
<nlm:aff id="af3-opth-8-629">Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Shatin, New Territories, Hong Kong</nlm:aff>
<country xml:lang="fr">Hong Kong</country>
<wicri:regionArea>Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Shatin, New Territories</wicri:regionArea>
<orgName type="university">Université chinoise de Hong Kong</orgName>
<placeName>
<settlement type="city">Sha Tin</settlement>
</placeName>
</affiliation>
</author>
</analytic>
<series>
<title level="j">Clinical Ophthalmology (Auckland, N.Z.)</title>
<idno type="ISSN">1177-5467</idno>
<idno type="eISSN">1177-5483</idno>
<imprint>
<date when="2014">2014</date>
</imprint>
</series>
</biblStruct>
</sourceDesc>
</fileDesc>
<profileDesc>
<textClass></textClass>
</profileDesc>
</teiHeader>
<front>
<div type="abstract" xml:lang="en">
<sec>
<title>Purpose</title>
<p>To evaluate the safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (IOL) (enVista® MX60; Bausch and Lomb Incorporated, Rochester, NY, USA) following implantation to correct aphakia subsequent to extracapsular cataract extraction in adults.</p>
</sec>
<sec>
<title>Subjects and methods</title>
<p>This was an open-label, non-interventional, observational study conducted in 19 university and private-practice settings in Europe and the Asia-Pacific region to investigate clinical outcomes of the MX60 IOL in standard practice. Eligible subjects were at least 18 years of age and had undergone standard phacoemulsification and extracapsular cataract extraction with implantation of the MX60 IOL. The primary safety endpoint was the occurrence of adverse events, and the primary effectiveness endpoints included visual and refractive outcomes and stability, with data collected up to 2 years post-procedure.</p>
</sec>
<sec>
<title>Results</title>
<p>In this multicenter study, pooled data of 255 eyes were collected and analyzed. Excellent visual and refractive outcomes and stability were demonstrated. At postoperative visit 4 (61–180 days postoperative), 62.2% of subjects achieved a Snellen best-corrected distance visual acuity (CDVA) of 20/20 (decimal 1.00), and 97.8% of subjects achieved a CDVA of 20/40 (decimal 0.50) or better. One eye (1.0%) underwent neodymium:yttrium aluminum garnet capsulotomy at 12 months post-procedure. No glistenings of any grade were reported for any subject at any visit. Adverse events were infrequent and were consistent with incidences generally reported with cataract surgery.</p>
</sec>
<sec>
<title>Conclusion</title>
<p>This study, which enrolled all comers, provided evidence of the excellent safety and effectiveness of the MX60 IOL in standard practice. Favorable clinical outcomes included outstanding visual and refractive outcomes and stability. No glistenings were reported at any postoperative visit.</p>
</sec>
</div>
</front>
<back>
<div1 type="bibliography">
<listBibl>
<biblStruct>
<analytic>
<author>
<name sortKey="Steinert, Rf" uniqKey="Steinert R">RF Steinert</name>
</author>
<author>
<name sortKey="Puliafito, Ca" uniqKey="Puliafito C">CA Puliafito</name>
</author>
<author>
<name sortKey="Kumar, Sr" uniqKey="Kumar S">SR Kumar</name>
</author>
<author>
<name sortKey="Dudak, Sd" uniqKey="Dudak S">SD Dudak</name>
</author>
<author>
<name sortKey="Patel, S" uniqKey="Patel S">S Patel</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Rickman Barger, L" uniqKey="Rickman Barger L">L Rickman-Barger</name>
</author>
<author>
<name sortKey="Florine, Cw" uniqKey="Florine C">CW Florine</name>
</author>
<author>
<name sortKey="Larson, Rs" uniqKey="Larson R">RS Larson</name>
</author>
<author>
<name sortKey="Lindstrom, Rl" uniqKey="Lindstrom R">RL Lindstrom</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Newland, Tj" uniqKey="Newland T">TJ Newland</name>
</author>
<author>
<name sortKey="Mcdermott, Ml" uniqKey="Mcdermott M">ML McDermott</name>
</author>
<author>
<name sortKey="Eliott, D" uniqKey="Eliott D">D Eliott</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Holweger, Rr" uniqKey="Holweger R">RR Holweger</name>
</author>
<author>
<name sortKey="Marefat, B" uniqKey="Marefat B">B Marefat</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Werner, L" uniqKey="Werner L">L Werner</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Christiansen, G" uniqKey="Christiansen G">G Christiansen</name>
</author>
<author>
<name sortKey="Durcan, Ej" uniqKey="Durcan E">EJ Durcan</name>
</author>
<author>
<name sortKey="Olson, Rj" uniqKey="Olson R">RJ Olson</name>
</author>
<author>
<name sortKey="Christiansen, K" uniqKey="Christiansen K">K Christiansen</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Packer, M" uniqKey="Packer M">M Packer</name>
</author>
<author>
<name sortKey="Fry, L" uniqKey="Fry L">L Fry</name>
</author>
<author>
<name sortKey="Lavery, Kt" uniqKey="Lavery K">KT Lavery</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Dixon, Jr" uniqKey="Dixon J">JR Dixon</name>
</author>
</analytic>
</biblStruct>
<biblStruct></biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Mentak, K" uniqKey="Mentak K">K Mentak</name>
</author>
<author>
<name sortKey="Goldberg, E" uniqKey="Goldberg E">E Goldberg</name>
</author>
<author>
<name sortKey="El Achchabi, A" uniqKey="El Achchabi A">A El-Achchabi</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Eom, Y" uniqKey="Eom Y">Y Eom</name>
</author>
<author>
<name sortKey="Lee, Js" uniqKey="Lee J">JS Lee</name>
</author>
<author>
<name sortKey="Rhim, Jw" uniqKey="Rhim J">JW Rhim</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Caporossi, A" uniqKey="Caporossi A">A Caporossi</name>
</author>
<author>
<name sortKey="Martone, G" uniqKey="Martone G">G Martone</name>
</author>
<author>
<name sortKey="Casprini, F" uniqKey="Casprini F">F Casprini</name>
</author>
<author>
<name sortKey="Rapisarda, L" uniqKey="Rapisarda L">L Rapisarda</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Pepose, Js" uniqKey="Pepose J">JS Pepose</name>
</author>
<author>
<name sortKey="Qazi, Ma" uniqKey="Qazi M">MA Qazi</name>
</author>
<author>
<name sortKey="Edwards, Kh" uniqKey="Edwards K">KH Edwards</name>
</author>
<author>
<name sortKey="Sanderson, Jp" uniqKey="Sanderson J">JP Sanderson</name>
</author>
<author>
<name sortKey="Sarver, Ej" uniqKey="Sarver E">EJ Sarver</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Santhiago, Mr" uniqKey="Santhiago M">MR Santhiago</name>
</author>
<author>
<name sortKey="Netto, Mv" uniqKey="Netto M">MV Netto</name>
</author>
<author>
<name sortKey="Barreto, J" uniqKey="Barreto J">J Barreto</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Nixon, Dr" uniqKey="Nixon D">DR Nixon</name>
</author>
<author>
<name sortKey="Woodcock, Mg" uniqKey="Woodcock M">MG Woodcock</name>
</author>
</analytic>
</biblStruct>
<biblStruct></biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Parulekar, Mv" uniqKey="Parulekar M">MV Parulekar</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Mckellar, Mj" uniqKey="Mckellar M">MJ McKellar</name>
</author>
<author>
<name sortKey="Elder, Mj" uniqKey="Elder M">MJ Elder</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Ram, J" uniqKey="Ram J">J Ram</name>
</author>
<author>
<name sortKey="Kaushik, S" uniqKey="Kaushik S">S Kaushik</name>
</author>
<author>
<name sortKey="Brar, Gs" uniqKey="Brar G">GS Brar</name>
</author>
<author>
<name sortKey="Gupta, A" uniqKey="Gupta A">A Gupta</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Biber, Jm" uniqKey="Biber J">JM Biber</name>
</author>
<author>
<name sortKey="Sandoval, Hp" uniqKey="Sandoval H">HP Sandoval</name>
</author>
<author>
<name sortKey="Trivedi, Rh" uniqKey="Trivedi R">RH Trivedi</name>
</author>
<author>
<name sortKey="Fernandez De Castro, Le" uniqKey="Fernandez De Castro L">LE Fernández de Castro</name>
</author>
<author>
<name sortKey="French, Jw" uniqKey="French J">JW French</name>
</author>
<author>
<name sortKey="Solomon, Kd" uniqKey="Solomon K">KD Solomon</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Apple, Dj" uniqKey="Apple D">DJ Apple</name>
</author>
<author>
<name sortKey="Peng, Q" uniqKey="Peng Q">Q Peng</name>
</author>
<author>
<name sortKey="Visessook, N" uniqKey="Visessook N">N Visessook</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Tetz, Mr" uniqKey="Tetz M">MR Tetz</name>
</author>
<author>
<name sortKey="Werner, L" uniqKey="Werner L">L Werner</name>
</author>
<author>
<name sortKey="Schwahn Bendig, S" uniqKey="Schwahn Bendig S">S Schwahn-Bendig</name>
</author>
<author>
<name sortKey="Batlle, Jf" uniqKey="Batlle J">JF Batlle</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Dhaliwal, Dk" uniqKey="Dhaliwal D">DK Dhaliwal</name>
</author>
<author>
<name sortKey="Mamalis, N" uniqKey="Mamalis N">N Mamalis</name>
</author>
<author>
<name sortKey="Olson, Rj" uniqKey="Olson R">RJ Olson</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Gunenc, U" uniqKey="Gunenc U">U Gunenc</name>
</author>
<author>
<name sortKey="Oner, Fh" uniqKey="Oner F">FH Oner</name>
</author>
<author>
<name sortKey="Tongal, S" uniqKey="Tongal S">S Tongal</name>
</author>
<author>
<name sortKey="Ferliel, M" uniqKey="Ferliel M">M Ferliel</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Tognetto, D" uniqKey="Tognetto D">D Tognetto</name>
</author>
<author>
<name sortKey="Toto, L" uniqKey="Toto L">L Toto</name>
</author>
<author>
<name sortKey="Sanguinetti, G" uniqKey="Sanguinetti G">G Sanguinetti</name>
</author>
<author>
<name sortKey="Ravalico, G" uniqKey="Ravalico G">G Ravalico</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Behndig, A" uniqKey="Behndig A">A Behndig</name>
</author>
<author>
<name sortKey="Monestam, E" uniqKey="Monestam E">E Monestam</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Colin, J" uniqKey="Colin J">J Colin</name>
</author>
<author>
<name sortKey="Praud, D" uniqKey="Praud D">D Praud</name>
</author>
<author>
<name sortKey="Touboul, D" uniqKey="Touboul D">D Touboul</name>
</author>
<author>
<name sortKey="Schweitzer, C" uniqKey="Schweitzer C">C Schweitzer</name>
</author>
</analytic>
</biblStruct>
<biblStruct>
<analytic>
<author>
<name sortKey="Packer, M" uniqKey="Packer M">M Packer</name>
</author>
<author>
<name sortKey="Fry, L" uniqKey="Fry L">L Fry</name>
</author>
<author>
<name sortKey="Lavery, Kt" uniqKey="Lavery K">KT Lavery</name>
</author>
<author>
<name sortKey="Lehmann, R" uniqKey="Lehmann R">R Lehmann</name>
</author>
</analytic>
</biblStruct>
</listBibl>
</div1>
</back>
</TEI>
<pmc article-type="research-article">
<pmc-dir>properties open_access</pmc-dir>
<front>
<journal-meta>
<journal-id journal-id-type="nlm-ta">Clin Ophthalmol</journal-id>
<journal-id journal-id-type="iso-abbrev">Clin Ophthalmol</journal-id>
<journal-id journal-id-type="publisher-id">Clinical Ophthalmology</journal-id>
<journal-title-group>
<journal-title>Clinical Ophthalmology (Auckland, N.Z.)</journal-title>
</journal-title-group>
<issn pub-type="ppub">1177-5467</issn>
<issn pub-type="epub">1177-5483</issn>
<publisher>
<publisher-name>Dove Medical Press</publisher-name>
</publisher>
</journal-meta>
<article-meta>
<article-id pub-id-type="pmid">24729678</article-id>
<article-id pub-id-type="pmc">3974695</article-id>
<article-id pub-id-type="doi">10.2147/OPTH.S56135</article-id>
<article-id pub-id-type="publisher-id">opth-8-629</article-id>
<article-categories>
<subj-group subj-group-type="heading">
<subject>Original Research</subject>
</subj-group>
</article-categories>
<title-group>
<article-title>Safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (enVista®) – results of a European and Asian-Pacific study</article-title>
</title-group>
<contrib-group>
<contrib contrib-type="author">
<name>
<surname>Heiner</surname>
<given-names>Peter</given-names>
</name>
<xref ref-type="aff" rid="af1-opth-8-629">1</xref>
<xref ref-type="corresp" rid="c1-opth-8-629"></xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Ligabue</surname>
<given-names>Edoardo</given-names>
</name>
<xref ref-type="aff" rid="af2-opth-8-629">2</xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Fan</surname>
<given-names>Alex</given-names>
</name>
<xref ref-type="aff" rid="af3-opth-8-629">3</xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Lam</surname>
<given-names>Dennis</given-names>
</name>
<xref ref-type="aff" rid="af3-opth-8-629">3</xref>
</contrib>
</contrib-group>
<aff id="af1-opth-8-629">
<label>1</label>
Vision Eye Institute, Southport, QLD, Australia</aff>
<aff id="af2-opth-8-629">
<label>2</label>
Ophthalmic Center, San Siro Clinic, Milan, Italy</aff>
<aff id="af3-opth-8-629">
<label>3</label>
Department of Ophthalmology and Visual Sciences, The Chinese University of Hong Kong, Shatin, New Territories, Hong Kong</aff>
<author-notes>
<corresp id="c1-opth-8-629">Correspondence: Peter Heiner, Vision Eye Institute, Level 4, 2 Short Street, Southport, QLD 4215, Australia, Tel +61 7 5557 8300, Fax +61 7 5562 0772, Email
<email>pheiner77@hotmail.com</email>
</corresp>
</author-notes>
<pub-date pub-type="collection">
<year>2014</year>
</pub-date>
<pub-date pub-type="epub">
<day>27</day>
<month>3</month>
<year>2014</year>
</pub-date>
<volume>8</volume>
<fpage>629</fpage>
<lpage>635</lpage>
<permissions>
<copyright-statement>© 2014 Heiner et al. This work is published by Dove Medical Press Ltd, and licensed under Creative Commons Attribution – Non Commercial (unported, v3.0) License</copyright-statement>
<copyright-year>2014</copyright-year>
<license>
<license-p>The full terms of the License are available at
<ext-link ext-link-type="uri" xlink:href="http://creativecommons.org/licenses/by-nc/3.0/">http://creativecommons.org/licenses/by-nc/3.0/</ext-link>
. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Ltd, provided the work is properly attributed.</license-p>
</license>
</permissions>
<abstract>
<sec>
<title>Purpose</title>
<p>To evaluate the safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (IOL) (enVista® MX60; Bausch and Lomb Incorporated, Rochester, NY, USA) following implantation to correct aphakia subsequent to extracapsular cataract extraction in adults.</p>
</sec>
<sec>
<title>Subjects and methods</title>
<p>This was an open-label, non-interventional, observational study conducted in 19 university and private-practice settings in Europe and the Asia-Pacific region to investigate clinical outcomes of the MX60 IOL in standard practice. Eligible subjects were at least 18 years of age and had undergone standard phacoemulsification and extracapsular cataract extraction with implantation of the MX60 IOL. The primary safety endpoint was the occurrence of adverse events, and the primary effectiveness endpoints included visual and refractive outcomes and stability, with data collected up to 2 years post-procedure.</p>
</sec>
<sec>
<title>Results</title>
<p>In this multicenter study, pooled data of 255 eyes were collected and analyzed. Excellent visual and refractive outcomes and stability were demonstrated. At postoperative visit 4 (61–180 days postoperative), 62.2% of subjects achieved a Snellen best-corrected distance visual acuity (CDVA) of 20/20 (decimal 1.00), and 97.8% of subjects achieved a CDVA of 20/40 (decimal 0.50) or better. One eye (1.0%) underwent neodymium:yttrium aluminum garnet capsulotomy at 12 months post-procedure. No glistenings of any grade were reported for any subject at any visit. Adverse events were infrequent and were consistent with incidences generally reported with cataract surgery.</p>
</sec>
<sec>
<title>Conclusion</title>
<p>This study, which enrolled all comers, provided evidence of the excellent safety and effectiveness of the MX60 IOL in standard practice. Favorable clinical outcomes included outstanding visual and refractive outcomes and stability. No glistenings were reported at any postoperative visit.</p>
</sec>
</abstract>
<kwd-group>
<title>Keywords</title>
<kwd>cataract surgery</kwd>
<kwd>glistenings</kwd>
<kwd>IOL</kwd>
<kwd>MX60</kwd>
</kwd-group>
</article-meta>
</front>
<body>
<sec>
<title>Introduction</title>
<p>Recent advances in intraocular len (IOL) material and design offer enhanced treatment options to surgeons and patients. As excellent visual and refractive outcomes have become more expected, increased awareness has focused on visual and refractive stability and on reductions in posterior capsule opacification (PCO) and glistenings. Although neodymium:yttrium aluminium garnet (Nd:YAG) capsulotomy is generally performed in-office with a high index of safety, it can result in complications.
<xref rid="b1-opth-8-629" ref-type="bibr">1</xref>
<xref rid="b4-opth-8-629" ref-type="bibr">4</xref>
</p>
<p>The visual significance of glistenings, fluid-filled microvacuoles that form within the IOL optic when in an aqueous environment,
<xref rid="b5-opth-8-629" ref-type="bibr">5</xref>
is a matter that has been actively investigated. A review of published literature on glistenings describes them in association with hydrophobic acrylic IOLs; however, the hydrophobic acrylic IOLs on the market are not manufactured from the same material and by the same manufacturing process.
<xref rid="b5-opth-8-629" ref-type="bibr">5</xref>
The impact of glistenings on visual performance is controversial, but there is the suspicion that severe presentations of glistenings may affect visual function.
<xref rid="b6-opth-8-629" ref-type="bibr">6</xref>
In any case, a pristine IOL is the preferred outcome.</p>
<p>The MX60 (enVista® MX60; Bausch and Lomb Incorporated, Rochester, NY, USA), an IOL in which both the optic and haptics are formed from a proprietary glistening-free hydrophobic acrylic polymer, received Conformité Européene (CE [European Conformity]) mark approval in August 2010. Following a multicenter, prospective, randomized clinical trial,
<xref rid="b7-opth-8-629" ref-type="bibr">7</xref>
the MX60 was approved by the US Food and Drug Administration (FDA) as a single-piece hydrophobic acrylic IOL in May 2012. To obtain further data characterizing the safety and effectiveness of this IOL, a multicenter, observational study was initiated in Europe and the Asia-Pacific region to explore real-world clinical practice outcomes.</p>
</sec>
<sec sec-type="methods">
<title>Subjects and methods</title>
<sec sec-type="methods">
<title>Study design</title>
<p>This was an open-label, non-interventional clinical study conducted at 15 sites in Europe, two sites in Hong Kong, and two sites in Australia, involving 21 surgeon investigators. The study protocol was reviewed and approved by independent ethics committees prior to initiation of the study at each center. The study was conducted in accordance with International Conference on Harmonisation Good Clinical Practice
<xref rid="b8-opth-8-629" ref-type="bibr">8</xref>
and applicable local regulations. Written informed consent was obtained from subjects prior to having any study-related procedure.</p>
<p>The study protocol allowed flexibility in patient selection and surgical technique to reflect each investigator’s current practice of cataract surgery. The primary safety endpoint of this study was the occurrence of adverse events (AEs), described in terms of statistical incidence rates. The primary effectiveness endpoints included visual and refractive outcomes and stability.</p>
</sec>
<sec>
<title>Eligibility criteria</title>
<p>Subjects 18 years of age or older who underwent standard phacoemulsifcation/extracapsular cataract extraction with implantation of the MX60 IOL were enrolled in this study. As the intent of this study was to obtain clinical outcomes found in standard practice, there were no other specific inclusion or exclusion criteria. Both eyes of subjects were eligible for inclusion in this study.</p>
</sec>
<sec sec-type="methods">
<title>IOL material and design</title>
<p>The MX60 IOL optic and haptics are lathed and milled from a single button made from a proprietary soft hydrophobic acrylic polymer, which incorporates an ultraviolet-absorbing chromophore. Surface energy and contact angle measurements have demonstrated that the IOL material is a true hydrophobic polymer with water contact angle values comparable to those of other hydrophobic acrylic IOLs.
<xref rid="b9-opth-8-629" ref-type="bibr">9</xref>
To ensure that the IOL remains glistening-free, this IOL is prehydrated to equilibrium water content.</p>
<p>The surface hardness of the material is 11.0 MPa, which compares favorably with the 0.24 MPa, 0.68 MPa, and 0.43 MPa reported for AcrySof (Alcon Laboratories, Inc., Fort Worth, TX, USA), Acryfold (Hoya Corporation, Santa Clara, CA, USA), and Sensar (Abbott Medical Optics, Abbott Park, IL, USA) IOL materials, respectively.
<xref rid="b10-opth-8-629" ref-type="bibr">10</xref>
The high surface hardness of the material results in a lens which is difficult to either scratch or deform. The reduced compliance of the material also results in a requirement for greater force in folding the lens for insertion into the injection cartridge.
<xref rid="b7-opth-8-629" ref-type="bibr">7</xref>
The material from which the MX60 lens is manufactured has a relatively high glass transition temperature (28°C), which results in slow unfolding in the eye. The lens material is highly temperature sensitive, so that slight warming will result in more rapid unfolding in the eye.
<xref rid="b11-opth-8-629" ref-type="bibr">11</xref>
</p>
<p>The biconvex lens optic has a body diameter of 6.0 mm and an overall length (diameter) of 12.5 mm (
<xref ref-type="fig" rid="f1-opth-8-629">Figure 1</xref>
). The MX60 IOL has aberration-free aspheric optics and modified C-loop haptics.
<xref rid="b12-opth-8-629" ref-type="bibr">12</xref>
<xref rid="b14-opth-8-629" ref-type="bibr">14</xref>
The anterior edge is designed to allow anterior capsulorhexis adherence for improved lens centration. The 360° square posterior edge is designed to minimize the development of PCO (
<xref ref-type="fig" rid="f2-opth-8-629">Figure 2</xref>
).
<xref rid="b15-opth-8-629" ref-type="bibr">15</xref>
The MX60 IOL features step-vaulted haptics that translate the optic posteriorly for direct contact with the capsular bag in addition to a sharp 360° square barrier edge to inhibit lens epithelial cell migration. In addition, the haptics have fenestrations intended to prevent transfer of stress from the haptic to the optic.</p>
</sec>
<sec sec-type="methods">
<title>Study procedures</title>
<p>Eligible subjects who provided written informed consent had a preoperative visit, an operative visit (day 0), and up to six postoperative visits (visit 1, 1–2 days postoperative; visit 2, 3–14 days postoperative; visit 3, 15–60 days postoperative; visit 4, 61–180 days postoperative; visit 5, 181–365 days postoperative; and visit 6, 366–730 days postoperative). The preoperative visit included determinations of uncorrected distance visual acuity (UDVA), best-corrected distance visual acuity (CDVA), manifest refraction, keratometry, axial length, target postoperative refraction, and biometry.</p>
<p>Postoperative assessments included evaluation of UDVA, CDVA, manifest refraction, glistenings assessment through slit-lamp examination, and monitoring for AEs and secondary interventions.</p>
<sec>
<title>Evaluation of glistenings</title>
<p>Glistenings, which are fluid-filled microvacuoles that can form within the IOL optic when in an aqueous environment, are more common in hydrophobic IOLs than in other lens materials.
<xref rid="b5-opth-8-629" ref-type="bibr">5</xref>
Glistenings assessment was performed by slit-lamp examination using high-powered magnification and were graded according to either absent or, if present, according to the following scale (grade 1, less than five glistening observed; grade 2, more than five glistenings observed but not affecting vision; grade 3, glistenings affecting vision).</p>
</sec>
</sec>
<sec sec-type="methods">
<title>Statistical methods</title>
<p>Descriptive statistics were used to summarize the data. Continuous variables were summarized by the sample size, mean, standard deviation (SD), and the minimum and maximum. Categorical variables were summarized by frequency and percentage. All AEs reported at all visits were summarized cumulatively.</p>
</sec>
</sec>
<sec sec-type="results">
<title>Results</title>
<sec>
<title>Demographics and surgery</title>
<p>A total of 255 eyes were included in the pooled data analysis. The mean ± SD age of patients included in the analysis was 73.3±9.6 years. The majority of eyes (59.2%) were from female subjects, and more procedures were completed in the right eye of subjects (54.1%). At baseline, 27 eyes (10.6%) were reported as having macular disease, 15 eyes (5.9%) had glaucoma, six eyes (2.4%) had corneal guttata, five eyes (2.0%) had floppy iris syndrome, two eyes (0.8%) had pseudoexfoliation, and one eye (0.4%) had prior refractive surgery.</p>
<p>For all subjects, the primary incision type was clear cornea, and the mean estimated incision size ± SD was 2.72±0.22 mm. The targeted mean ± SD postoperative manifest refraction spherical equivalent (MRSE) was −0.53±0.67 diopters (D). IOL powers used ranged from +15.00 to +28.50 D. Data acquired at visit 6 (366–730 days postoperative) is reported.</p>
</sec>
<sec>
<title>Visual and refractive outcomes</title>
<p>
<xref ref-type="table" rid="t1-opth-8-629">Table 1</xref>
shows UDVA and CDVA results in logMAR (logarithm of the minimum angle of resolution) and decimal visual acuity at each interval in addition to MRSE outcomes. Visual outcomes indicated good visual and refractive stability, with gradual improvement over time. A graphical presentation of UDVA results at postoperative visits 4 (61–180 days postoperative) and 5 (181–365 days postoperative) can be found in
<xref ref-type="fig" rid="f3-opth-8-629">Figure 3</xref>
. A graph of the number of eyes achieving a Snellen CDVA of 20/40 (decimal 0.50) or better is presented in
<xref ref-type="fig" rid="f4-opth-8-629">Figure 4</xref>
.</p>
<p>The change in mean MRSE was 0.01 D between postoperative visits 4 and 5, and 0.09 D between visits 5 and 6, demonstrating refractive stability. Because the target MRSE was not a plano-refraction in all cases, the analysis was performed using typical inclusion criteria for monofocal studies. Eyes included in the accuracy to target analysis had a postoperative MRSE target of ±0.37 D, preoperative keratometric cylinder of ≤0.75 D, and did not undergo additional refractive procedures such as limbal relaxing incisions (five eyes [2.0%]). The accuracy to target MRSE results are presented in
<xref ref-type="table" rid="t2-opth-8-629">Table 2</xref>
.</p>
</sec>
<sec>
<title>AEs and secondary interventions</title>
<p>The incidence of AEs experienced during the study was comparable to or lower than the incidence reported in the historical control population per FDA draft guidance.
<xref rid="b16-opth-8-629" ref-type="bibr">16</xref>
All AEs and complications were consistent with typical incidences accompanying cataract or implant surgery. There were no unique reports associated with the MX60 IOL.</p>
<p>No glistenings of any grade were reported for any subject at any visit. One eye (1.0%) received an Nd:YAG capsulotomy at 363 days post-procedure. Three eyes (1.2%) had secondary interventions, which included an intravitreal injection of bevacizumab (Avastin®; Genentech, South San Francisco, CA, USA), suture removal, and a paracentesis to lower the intraocular pressure. No IOL explantations were performed.</p>
</sec>
</sec>
<sec sec-type="discussion">
<title>Discussion</title>
<p>The results of this study provide evidence of the safety and effectiveness of the MX60 IOL in a real-world clinical setting. At visit 4, 62.2% of the subjects achieved a Snellen CDVA of 20/20, and 97.8% of the subjects achieved a CDVA of 20/40 or better. In addition, the MX60 displayed visual and refractive stability over time. This IOL is designed to maximize the haptic contact angle under compression, preventing ovalization and/or striae of the capsular bag. This is intended to provide refractive stability under compression. In regards to safety, intraoperative and postoperative ocular AEs were of the type and frequency generally observed in patients who have had cataract surgery.
<xref rid="b17-opth-8-629" ref-type="bibr">17</xref>
,
<xref rid="b18-opth-8-629" ref-type="bibr">18</xref>
As study outcomes were intended to reflect those of real-life practices, the results of this open-label, nonrandomized study may be subject to selection bias.</p>
<p>PCO is the most common visually disabling sequela of modern cataract surgery,
<xref rid="b19-opth-8-629" ref-type="bibr">19</xref>
caused primarily by the migration of lens epithelial cells that persist after cataract surgery.
<xref rid="b20-opth-8-629" ref-type="bibr">20</xref>
In this study, only one eye (1.0%) underwent Nd:YAG capsulotomy at 12 months post-procedure. In contrast, a retrospective study involving AcrySof IOLs reported an Nd:YAG capsulotomy rate of 17.3% (13 of 75 eyes) for the monofocal spherical group (Natural SN60AT; Alcon Laboratories, Inc.) and 4.0% (3 of 75 eyes) for the monofocal aspheric group (IQ SN60WF; Alcon Laboratories, Inc.); the mean time from surgery to Nd:YAG capsulotomy was 13.0±9.3 months in the monofocal spherical group and 9.3±6.4 months in the monofocal aspheric group.
<xref rid="b20-opth-8-629" ref-type="bibr">20</xref>
</p>
<p>Nixon and Woodcock
<xref rid="b15-opth-8-629" ref-type="bibr">15</xref>
reported that IOL design rather than material is the critical factor in minimizing lens epithelial cell migration across the posterior capsule. In their study, characteristics that led to lower levels of PCO with a continuous-edge IOL were the 360° square edge, angled haptics, increased optic–haptic space, and increased resistance to compression. These features help to position the IOL against the posterior capsule and encourage complete circumferential shrink-wrapping of the IOL by the capsule. Apple et al
<xref rid="b21-opth-8-629" ref-type="bibr">21</xref>
surmised that PCO can develop when a potential space is available into which lens epithelial cells can migrate and proliferate. The MX60 features step-vaulted haptics that translate the optic posteriorly for direct contact with the capsular bag. It also has a 360° square barrier edge to inhibit lens epithelial cell migration, thus limiting the development of PCO.</p>
<p>In the study reported here, no glistenings of any grade were reported for any subject at any visit. This observation is corroborated by results from other studies
<xref rid="b7-opth-8-629" ref-type="bibr">7</xref>
,
<xref rid="b22-opth-8-629" ref-type="bibr">22</xref>
investigating this same IOL material. Christiansen et al
<xref rid="b6-opth-8-629" ref-type="bibr">6</xref>
reported that the Snellen acuity in eyes with severe glistenings (grade ≥2+) was half a line lower than in eyes with mild glistenings (
<italic>P</italic>
=0.01), a finding supported by other studies.
<xref rid="b23-opth-8-629" ref-type="bibr">23</xref>
,
<xref rid="b24-opth-8-629" ref-type="bibr">24</xref>
Investigations involving other hydrophobic acrylic IOLs have reported that glistenings appear to increase in severity with time,
<xref rid="b5-opth-8-629" ref-type="bibr">5</xref>
,
<xref rid="b25-opth-8-629" ref-type="bibr">25</xref>
<xref rid="b27-opth-8-629" ref-type="bibr">27</xref>
which was not found with the MX60 IOL. The MX60 IOL is packaged in physiologic saline to eliminate fluid exchange with the aqueous humor. Prehydration of this IOL to equilibrium water content ensures that it remains glistening-free. Following cataract surgery, a pristine IOL is clearly the ideal outcome.</p>
<p>When used to correct aphakia following cataract surgery, the MX60 IOL has demonstrated a well established safety profile with predictable visual outcomes, stable refractive outcomes, a low incidence of Nd:YAG capsulotomies to 12 months post-procedure, and no glistenings.</p>
</sec>
</body>
<back>
<ack>
<title>Acknowledgments</title>
<p>We would like to acknowledge the following participating investigators: Joerg Koch, MD, St Franziskus-Hospital Muenster, Münster, Germany; Michael Kuechle, MD, Augenpraxis-Klinik Erlangen, Erlangen, Germany; Egbert Pool, MD, Ommelander Ziekenhuis Groep, Delfzijl, the Netherlands; Richard Packard, MD, Prince Charles Eye Unit, London, England; Alessandro Franchini, MD, Università Ospedaliera di Firenze Careggi, Firenze, Italy; Javier Mendicute, MD, Hospital de Donostia, San Sebastian, Spain; Manuel Correia, MD, Hospital Narciso Ferreira, Porto, Portugal; Dominique Monnet, MD, PhD, Université Paris Descartes, Paris, France; Pisella Pierre Jean, MD, Hôpital Bretonneau, Tours, France; Andras Berta, MD, University of Debrecen, Debrecen, Hungary; Petr Novak, MD, Hospital Nemocnice Na Homolce, Prague, Czech Republic; Bjorn Johansson, MD, Linköpings Universitetssjukhus, Linköping, Sweden; Gisela Wejde, MD, S:t Eriks ögonsjukhus, Stockholm, Sweden; and Brad Horsburgh, MD, Northside Eye Centre, Nundah, QLD, Australia.</p>
<p>Funding for writing and editorial support was provided by Bausch and Lomb Incorporated, Rochester, NY, USA. Writing and editorial assistance was provided by Stephanie N Baba, OD, Alameda, CA, USA and Gary Mosehauer, MS, Senior Statistician of Bausch and Lomb Incorporated.</p>
</ack>
<fn-group>
<fn>
<p>
<bold>Disclosure</bold>
</p>
<p>This study was sponsored by Bausch and Lomb Incorporated. The authors have no conflicts of interest to disclose.</p>
</fn>
</fn-group>
<ref-list>
<title>References</title>
<ref id="b1-opth-8-629">
<label>1</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Steinert</surname>
<given-names>RF</given-names>
</name>
<name>
<surname>Puliafito</surname>
<given-names>CA</given-names>
</name>
<name>
<surname>Kumar</surname>
<given-names>SR</given-names>
</name>
<name>
<surname>Dudak</surname>
<given-names>SD</given-names>
</name>
<name>
<surname>Patel</surname>
<given-names>S</given-names>
</name>
</person-group>
<article-title>Cystoid macular edema, retinal detachment, and glaucoma after Nd:YAG laser posterior capsulotomy</article-title>
<source>Am J Ophthalmol</source>
<year>1991</year>
<volume>112</volume>
<issue>4</issue>
<fpage>373</fpage>
<lpage>380</lpage>
<pub-id pub-id-type="pmid">1928237</pub-id>
</element-citation>
</ref>
<ref id="b2-opth-8-629">
<label>2</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Rickman-Barger</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Florine</surname>
<given-names>CW</given-names>
</name>
<name>
<surname>Larson</surname>
<given-names>RS</given-names>
</name>
<name>
<surname>Lindstrom</surname>
<given-names>RL</given-names>
</name>
</person-group>
<article-title>Retinal detachment after neodymium:YAG laser posterior capsulotomy</article-title>
<source>Am J Ophthalmol</source>
<year>1989</year>
<issue>5</issue>
<volume>107</volume>
<fpage>531</fpage>
<lpage>536</lpage>
<pub-id pub-id-type="pmid">2712134</pub-id>
</element-citation>
</ref>
<ref id="b3-opth-8-629">
<label>3</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Newland</surname>
<given-names>TJ</given-names>
</name>
<name>
<surname>McDermott</surname>
<given-names>ML</given-names>
</name>
<name>
<surname>Eliott</surname>
<given-names>D</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Experimental neodymium:YAG laser damage to acrylic, poly(methyl methacrylate), and silicone intraocular lens materials</article-title>
<source>J Cataract Refract Surg</source>
<year>1999</year>
<volume>25</volume>
<issue>1</issue>
<fpage>72</fpage>
<lpage>76</lpage>
<pub-id pub-id-type="pmid">9888080</pub-id>
</element-citation>
</ref>
<ref id="b4-opth-8-629">
<label>4</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Holweger</surname>
<given-names>RR</given-names>
</name>
<name>
<surname>Marefat</surname>
<given-names>B</given-names>
</name>
</person-group>
<article-title>Intraocular pressure change after neodymium: YAG capsulotomy</article-title>
<source>J Cataract Refract Surg</source>
<year>1997</year>
<volume>23</volume>
<issue>1</issue>
<fpage>115</fpage>
<lpage>121</lpage>
<pub-id pub-id-type="pmid">9100118</pub-id>
</element-citation>
</ref>
<ref id="b5-opth-8-629">
<label>5</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Werner</surname>
<given-names>L</given-names>
</name>
</person-group>
<article-title>Glistenings and surface light scattering in intraocular lenses</article-title>
<source>J Cataract Refract Surg</source>
<year>2010</year>
<volume>36</volume>
<issue>8</issue>
<fpage>1398</fpage>
<lpage>1420</lpage>
<pub-id pub-id-type="pmid">20656166</pub-id>
</element-citation>
</ref>
<ref id="b6-opth-8-629">
<label>6</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Christiansen</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Durcan</surname>
<given-names>EJ</given-names>
</name>
<name>
<surname>Olson</surname>
<given-names>RJ</given-names>
</name>
<name>
<surname>Christiansen</surname>
<given-names>K</given-names>
</name>
</person-group>
<article-title>Glistenings in the AcrySof intraocular lens: pilot study</article-title>
<source>J Cataract Refract Surg</source>
<year>2001</year>
<volume>27</volume>
<issue>5</issue>
<fpage>728</fpage>
<lpage>733</lpage>
<pub-id pub-id-type="pmid">11377904</pub-id>
</element-citation>
</ref>
<ref id="b7-opth-8-629">
<label>7</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Packer</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Fry</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Lavery</surname>
<given-names>KT</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Safety and effectiveness of a glistening-free single-piece hydrophobic acrylic intraocular lens (enVista™)</article-title>
<source>Clin Ophthalmol</source>
<year>2013</year>
<volume>7</volume>
<fpage>1905</fpage>
<lpage>1912</lpage>
<pub-id pub-id-type="pmid">24109169</pub-id>
</element-citation>
</ref>
<ref id="b8-opth-8-629">
<label>8</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Dixon</surname>
<given-names>JR</given-names>
<suffix>Jr</suffix>
</name>
</person-group>
<article-title>The International Conference on Harmonisation Good Clinical Practice Guideline</article-title>
<source>Qual Assur</source>
<year>1998</year>
<volume>6</volume>
<issue>2</issue>
<fpage>65</fpage>
<lpage>74</lpage>
<pub-id pub-id-type="pmid">10386329</pub-id>
</element-citation>
</ref>
<ref id="b9-opth-8-629">
<label>9</label>
<element-citation publication-type="book">
<source>EnVista HydroStable Acrylic AO Lens [package insert]</source>
<publisher-loc>Rochester, NY</publisher-loc>
<publisher-name>Bausch and Lomb Incorporated</publisher-name>
<year>2012</year>
</element-citation>
</ref>
<ref id="b10-opth-8-629">
<label>10</label>
<element-citation publication-type="confproc">
<person-group person-group-type="author">
<name>
<surname>Mentak</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Goldberg</surname>
<given-names>E</given-names>
</name>
<name>
<surname>El-Achchabi</surname>
<given-names>A</given-names>
</name>
</person-group>
<source>Nanoindentation studies on hydrophobic acrylic iols to evaluate surface mechanical properties</source>
<conf-name>Paper presented at the XXV Congress of the European Society of Cataract and Refractive Surgeons – ESCRS</conf-name>
<conf-date>Sep 2007</conf-date>
<conf-loc>Stockholm, Sweden</conf-loc>
</element-citation>
</ref>
<ref id="b11-opth-8-629">
<label>11</label>
<element-citation publication-type="confproc">
<person-group person-group-type="author">
<name>
<surname>Eom</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Lee</surname>
<given-names>JS</given-names>
</name>
<name>
<surname>Rhim</surname>
<given-names>JW</given-names>
</name>
<etal></etal>
</person-group>
<source>A simple method to shorten the unfolding time of pre-hydrated hydrophobic intraocular lens</source>
<conf-name>Presented in part at the Asia-Pacific Association of Cataract and Refractive Surgeons annual meeting in Singapore (July 11–14, 2013) and the 109th meeting of the Korean Ophthalmological Society in South Korea (April 20–21, 2013)</conf-name>
</element-citation>
</ref>
<ref id="b12-opth-8-629">
<label>12</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Caporossi</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Martone</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Casprini</surname>
<given-names>F</given-names>
</name>
<name>
<surname>Rapisarda</surname>
<given-names>L</given-names>
</name>
</person-group>
<article-title>Prospective randomized study of clinical performance of 3 aspheric and 2 spherical intraocular lenses in 250 eyes</article-title>
<source>J Refract Surg</source>
<year>2007</year>
<volume>23</volume>
<issue>7</issue>
<fpage>639</fpage>
<lpage>648</lpage>
<pub-id pub-id-type="pmid">17912933</pub-id>
</element-citation>
</ref>
<ref id="b13-opth-8-629">
<label>13</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Pepose</surname>
<given-names>JS</given-names>
</name>
<name>
<surname>Qazi</surname>
<given-names>MA</given-names>
</name>
<name>
<surname>Edwards</surname>
<given-names>KH</given-names>
</name>
<name>
<surname>Sanderson</surname>
<given-names>JP</given-names>
</name>
<name>
<surname>Sarver</surname>
<given-names>EJ</given-names>
</name>
</person-group>
<article-title>Comparison of contrast sensitivity, depth of field and ocular wavefront aberrations in eyes with an IOL with zero versus positive spherical aberration</article-title>
<source>Graefes Arch Clin Exp Ophthalmol</source>
<year>2009</year>
<volume>247</volume>
<issue>7</issue>
<fpage>965</fpage>
<lpage>973</lpage>
<pub-id pub-id-type="pmid">19277694</pub-id>
</element-citation>
</ref>
<ref id="b14-opth-8-629">
<label>14</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Santhiago</surname>
<given-names>MR</given-names>
</name>
<name>
<surname>Netto</surname>
<given-names>MV</given-names>
</name>
<name>
<surname>Barreto</surname>
<given-names>J</given-names>
<suffix>Jr</suffix>
</name>
<etal></etal>
</person-group>
<article-title>Wavefront analysis, contrast sensitivity, and depth of focus after cataract surgery with aspherical intraocular lens implantation</article-title>
<source>Am J Ophthalmol</source>
<year>2010</year>
<volume>149</volume>
<issue>3</issue>
<fpage>383</fpage>
<lpage>389</lpage>
<pub-id pub-id-type="pmid">20035922</pub-id>
</element-citation>
</ref>
<ref id="b15-opth-8-629">
<label>15</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Nixon</surname>
<given-names>DR</given-names>
</name>
<name>
<surname>Woodcock</surname>
<given-names>MG</given-names>
</name>
</person-group>
<article-title>Pattern of posterior capsule opacification models 2 years postoperatively with 2 single-piece acrylic intraocular lenses</article-title>
<source>J Cataract Refract Surg</source>
<year>2010</year>
<volume>36</volume>
<issue>6</issue>
<fpage>929</fpage>
<lpage>934</lpage>
<pub-id pub-id-type="pmid">20494763</pub-id>
</element-citation>
</ref>
<ref id="b16-opth-8-629">
<label>16</label>
<element-citation publication-type="other">
<source>Guidance for Industry and for FDA Reviewers: Intraocular Lens Guidance Document</source>
<date-in-citation>Draft Guidance accessed Mar 2013</date-in-citation>
</element-citation>
</ref>
<ref id="b17-opth-8-629">
<label>17</label>
<element-citation publication-type="book">
<person-group person-group-type="author">
<name>
<surname>Parulekar</surname>
<given-names>MV</given-names>
</name>
</person-group>
<article-title>Complications following cataract surgery</article-title>
<person-group person-group-type="editor">
<name>
<surname>Malhotra</surname>
<given-names>R</given-names>
</name>
</person-group>
<source>Cataract</source>
<publisher-loc>Philadelphia</publisher-loc>
<publisher-name>Elsevier Ltd</publisher-name>
<year>2008</year>
<fpage>141</fpage>
<lpage>182</lpage>
</element-citation>
</ref>
<ref id="b18-opth-8-629">
<label>18</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>McKellar</surname>
<given-names>MJ</given-names>
</name>
<name>
<surname>Elder</surname>
<given-names>MJ</given-names>
</name>
</person-group>
<article-title>The early complications of cataract surgery: is routine review of patients 1 week after cataract extraction necessary?</article-title>
<source>Ophthalmology</source>
<year>2001</year>
<volume>108</volume>
<issue>5</issue>
<fpage>930</fpage>
<lpage>935</lpage>
<pub-id pub-id-type="pmid">11320024</pub-id>
</element-citation>
</ref>
<ref id="b19-opth-8-629">
<label>19</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Ram</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Kaushik</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Brar</surname>
<given-names>GS</given-names>
</name>
<name>
<surname>Gupta</surname>
<given-names>A</given-names>
</name>
</person-group>
<article-title>Neodymium:YAG capsulotomy rates following phacoemulsification with implantation of PMMA, silicone, and acrylic intraocular lenses</article-title>
<source>Ophthalmic Surg Lasers</source>
<year>2001</year>
<volume>32</volume>
<issue>5</issue>
<fpage>375</fpage>
<lpage>382</lpage>
<pub-id pub-id-type="pmid">11563781</pub-id>
</element-citation>
</ref>
<ref id="b20-opth-8-629">
<label>20</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Biber</surname>
<given-names>JM</given-names>
</name>
<name>
<surname>Sandoval</surname>
<given-names>HP</given-names>
</name>
<name>
<surname>Trivedi</surname>
<given-names>RH</given-names>
</name>
<name>
<surname>Fernández de Castro</surname>
<given-names>LE</given-names>
</name>
<name>
<surname>French</surname>
<given-names>JW</given-names>
</name>
<name>
<surname>Solomon</surname>
<given-names>KD</given-names>
</name>
</person-group>
<article-title>Comparison of the incidence and visual significance of posterior capsule opacification between multifocal spherical, monofocal spherical, and monofocal aspheric intraocular lenses</article-title>
<source>J Cataract Refract Surg</source>
<year>2009</year>
<volume>35</volume>
<issue>7</issue>
<fpage>1234</fpage>
<lpage>1238</lpage>
<pub-id pub-id-type="pmid">19545814</pub-id>
</element-citation>
</ref>
<ref id="b21-opth-8-629">
<label>21</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Apple</surname>
<given-names>DJ</given-names>
</name>
<name>
<surname>Peng</surname>
<given-names>Q</given-names>
</name>
<name>
<surname>Visessook</surname>
<given-names>N</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Eradication of posterior capsule opacification; documentation of a marked decrease in Nd:YAG laser posterior capsulotomy rates noted in an analysis of 5,416 pseudophakic human eyes obtained postmortem</article-title>
<source>Ophthalmology</source>
<year>2001</year>
<volume>108</volume>
<issue>3</issue>
<fpage>505</fpage>
<lpage>518</lpage>
<pub-id pub-id-type="pmid">11237905</pub-id>
</element-citation>
</ref>
<ref id="b22-opth-8-629">
<label>22</label>
<element-citation publication-type="confproc">
<person-group person-group-type="author">
<name>
<surname>Tetz</surname>
<given-names>MR</given-names>
</name>
<name>
<surname>Werner</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Schwahn-Bendig</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Batlle</surname>
<given-names>JF</given-names>
</name>
</person-group>
<source>Prospective Clinical Study to Quantify Glistenings in New Hydrophobic Acrylic IOL</source>
<conf-name>Paper presented at the ASCRS Symposium on Cataract, IOL and Refractive Surgery</conf-name>
<conf-loc>San Francisco, California, USA</conf-loc>
<conf-date>Apr 2009</conf-date>
</element-citation>
</ref>
<ref id="b23-opth-8-629">
<label>23</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Dhaliwal</surname>
<given-names>DK</given-names>
</name>
<name>
<surname>Mamalis</surname>
<given-names>N</given-names>
</name>
<name>
<surname>Olson</surname>
<given-names>RJ</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Visual significance of glistenings seen in the AcrySof intraocular lens</article-title>
<source>J Cataract Refract Surg</source>
<year>1996</year>
<volume>22</volume>
<issue>4</issue>
<fpage>452</fpage>
<lpage>457</lpage>
<pub-id pub-id-type="pmid">8733849</pub-id>
</element-citation>
</ref>
<ref id="b24-opth-8-629">
<label>24</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Gunenc</surname>
<given-names>U</given-names>
</name>
<name>
<surname>Oner</surname>
<given-names>FH</given-names>
</name>
<name>
<surname>Tongal</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Ferliel</surname>
<given-names>M</given-names>
</name>
</person-group>
<article-title>Effects on visual function of glistenings and folding marks in AcrySof intraocular lenses</article-title>
<source>J Cataract Refract Surg</source>
<year>2001</year>
<volume>27</volume>
<issue>10</issue>
<fpage>1611</fpage>
<lpage>1614</lpage>
<pub-id pub-id-type="pmid">11687360</pub-id>
</element-citation>
</ref>
<ref id="b25-opth-8-629">
<label>25</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Tognetto</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Toto</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Sanguinetti</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Ravalico</surname>
<given-names>G</given-names>
</name>
</person-group>
<article-title>Glistenings in foldable intraocular lenses</article-title>
<source>J Cataract Refract Surg</source>
<year>2002</year>
<volume>28</volume>
<issue>7</issue>
<fpage>1211</fpage>
<lpage>1216</lpage>
<pub-id pub-id-type="pmid">12106730</pub-id>
</element-citation>
</ref>
<ref id="b26-opth-8-629">
<label>26</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Behndig</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Monestam</surname>
<given-names>E</given-names>
</name>
</person-group>
<article-title>Quantification of glistenings in intraocular lenses using Scheimpflug photography</article-title>
<source>J Cataract Refract Surg</source>
<year>2009</year>
<volume>35</volume>
<issue>1</issue>
<fpage>14</fpage>
<lpage>17</lpage>
<pub-id pub-id-type="pmid">19101419</pub-id>
</element-citation>
</ref>
<ref id="b27-opth-8-629">
<label>27</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Colin</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Praud</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Touboul</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Schweitzer</surname>
<given-names>C</given-names>
</name>
</person-group>
<article-title>Incidence of glistening with the latest generation of yellow-tinted hydrophobic acrylic intraocular lenses</article-title>
<source>J Cataract Refract Surg</source>
<year>2012</year>
<volume>38</volume>
<issue>7</issue>
<fpage>1140</fpage>
<lpage>1146</lpage>
<pub-id pub-id-type="pmid">22727284</pub-id>
</element-citation>
</ref>
<ref id="b28-opth-8-629">
<label>28</label>
<element-citation publication-type="journal">
<person-group person-group-type="author">
<name>
<surname>Packer</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Fry</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Lavery</surname>
<given-names>KT</given-names>
</name>
<name>
<surname>Lehmann</surname>
<given-names>R</given-names>
</name>
<etal></etal>
</person-group>
<article-title>Safety and effectiveness of a glistening-free single-piece hydrophobic acrylic intraocular lens (enVista)</article-title>
<source>Clin Ophthalmol</source>
<year>2013</year>
<volume>7</volume>
<fpage>1905</fpage>
<lpage>1912</lpage>
<pub-id pub-id-type="pmid">24109169</pub-id>
</element-citation>
</ref>
</ref-list>
</back>
<floats-group>
<fig id="f1-opth-8-629" position="float">
<label>Figure 1</label>
<caption>
<p>The design of the MX60 (Bausch and Lomb Incorporated, Rochester, NY, USA) intraocular lens.</p>
</caption>
<graphic xlink:href="opth-8-629Fig1"></graphic>
</fig>
<fig id="f2-opth-8-629" position="float">
<label>Figure 2</label>
<caption>
<p>The MX60 (Bausch and Lomb Incorporated, Rochester, NY, USA) intraocular lens. The diagram shows the posterior surface facing toward the right side of the page. The haptics are offset anteriorly with respect to the optic body, which enables consistent posterior movement of lens optic under haptic compression. Image courtesy of David Spalton, FRCS, FRCP, FRCOphth. © 2013 Dove Medical Press Ltd. Reproduced with permission from Packer M, Fry L, Lavery KT, Lehmann R, et al. Safety and effectiveness of a glistening-free single-piece hydrophobic acrylic intraocular lens (enVista).
<italic>Clin Ophthalmol</italic>
. 2013;7:1905–1912.
<xref rid="b28-opth-8-629" ref-type="bibr">28</xref>
</p>
</caption>
<graphic xlink:href="opth-8-629Fig2"></graphic>
</fig>
<fig id="f3-opth-8-629" position="float">
<label>Figure 3</label>
<caption>
<p>Graph displaying uncorrected distance visual acuity results preoperatively and at three postoperative visits.</p>
</caption>
<graphic xlink:href="opth-8-629Fig3"></graphic>
</fig>
<fig id="f4-opth-8-629" position="float">
<label>Figure 4</label>
<caption>
<p>Graph presenting the percentage of eyes having Snellen best-corrected visual acuity of ≥20/40 (decimal 0.50) preoperatively and at four postoperative visits.</p>
</caption>
<graphic xlink:href="opth-8-629Fig4"></graphic>
</fig>
<table-wrap id="t1-opth-8-629" position="float">
<label>Table 1</label>
<caption>
<p>Visual and refractive outcomes</p>
</caption>
<table frame="hsides" rules="groups">
<thead>
<tr>
<th align="left" valign="top" rowspan="1" colspan="1"></th>
<th align="left" valign="top" rowspan="1" colspan="1">Preoperative</th>
<th align="left" valign="top" rowspan="1" colspan="1">3–14 days postoperative</th>
<th align="left" valign="top" rowspan="1" colspan="1">15–60 days postoperative</th>
<th align="left" valign="top" rowspan="1" colspan="1">61–180 days postoperative</th>
<th align="left" valign="top" rowspan="1" colspan="1">181–365 days postoperative</th>
<th align="left" valign="top" rowspan="1" colspan="1">366–730 days postoperative</th>
</tr>
</thead>
<tbody>
<tr>
<td align="left" valign="top" rowspan="1" colspan="1">UDVA, n</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=204</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=127</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=180</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=121</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=105</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=53</td>
</tr>
<tr>
<td align="left" valign="top" rowspan="1" colspan="1">Mean logMAR ± SD</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.66±0.32</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.30±0.25</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.27±0.30</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.28±0.33</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.25±0.30</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.16±0.22</td>
</tr>
<tr>
<td align="left" valign="top" rowspan="1" colspan="1">Mean decimal VA ± SD</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.27±0.17</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.56±0.24</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.64±0.32</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.63±0.30</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.66±0.29</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.77±0.31</td>
</tr>
<tr>
<td align="left" valign="top" rowspan="1" colspan="1">CDVA, n</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=160</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=86</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=141</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=90</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=72</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=38</td>
</tr>
<tr>
<td align="left" valign="top" rowspan="1" colspan="1">Mean logMAR ± SD</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.56±0.26</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.09±0.14</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.05±0.14</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.05±0.14</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.03±0.11</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.00±0.11</td>
</tr>
<tr>
<td align="left" valign="top" rowspan="1" colspan="1">Mean decimal VA ± SD</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.27±0.17</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.84±0.22</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.92±0.27</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.94±0.25</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.96±0.22</td>
<td align="left" valign="top" rowspan="1" colspan="1">1.04±0.25</td>
</tr>
<tr>
<td align="left" valign="top" rowspan="1" colspan="1">MRSE in diopters, n</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=166</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=105</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=147</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=88</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=67</td>
<td align="left" valign="top" rowspan="1" colspan="1">n=39</td>
</tr>
<tr>
<td align="left" valign="top" rowspan="1" colspan="1">Mean ± SD</td>
<td align="left" valign="top" rowspan="1" colspan="1">N/A</td>
<td align="left" valign="top" rowspan="1" colspan="1">−0.29±1.00</td>
<td align="left" valign="top" rowspan="1" colspan="1">−0.30±0.96</td>
<td align="left" valign="top" rowspan="1" colspan="1">−0.10±0.97</td>
<td align="left" valign="top" rowspan="1" colspan="1">−0.09±0.84</td>
<td align="left" valign="top" rowspan="1" colspan="1">0.00±0.69</td>
</tr>
</tbody>
</table>
<table-wrap-foot>
<fn id="tfn1-opth-8-629">
<p>
<bold>Note:</bold>
n = number of eyes.</p>
</fn>
<fn id="tfn2-opth-8-629">
<p>
<bold>Abbreviations:</bold>
CDVA, best-corrected distance visual acuity; logMAR, logarithm of the minimum angle of resolution; MRSE, manifest refraction spherical equivalent; N/A, not applicable; SD, standard deviation; UDVA, uncorrected distance visual acuity; VA, visual acuity.</p>
</fn>
</table-wrap-foot>
</table-wrap>
<table-wrap id="t2-opth-8-629" position="float">
<label>Table 2</label>
<caption>
<p>Accuracy to target manifest refraction spherical equivalent</p>
</caption>
<table frame="hsides" rules="groups">
<thead>
<tr>
<th align="left" valign="top" rowspan="1" colspan="1"></th>
<th align="left" valign="top" rowspan="1" colspan="1">15–60 days postoperative
<break></break>
N=147
<break></break>
n (%)</th>
<th align="left" valign="top" rowspan="1" colspan="1">61–180 days postoperative
<break></break>
N=88
<break></break>
n (%)</th>
<th align="left" valign="top" rowspan="1" colspan="1">181–365 days postoperative
<break></break>
N=67
<break></break>
n (%)</th>
<th align="left" valign="top" rowspan="1" colspan="1">366–730 days postoperative
<break></break>
N=39
<break></break>
n (%)</th>
</tr>
</thead>
<tbody>
<tr>
<td align="left" valign="top" rowspan="1" colspan="1">±0.50 diopter</td>
<td align="left" valign="top" rowspan="1" colspan="1">94 (63.9)</td>
<td align="left" valign="top" rowspan="1" colspan="1">57 (64.8)</td>
<td align="left" valign="top" rowspan="1" colspan="1">51 (76.1)</td>
<td align="left" valign="top" rowspan="1" colspan="1">26 (66.7)</td>
</tr>
<tr>
<td align="left" valign="top" rowspan="1" colspan="1">±1.00 diopter</td>
<td align="left" valign="top" rowspan="1" colspan="1">136 (92.5)</td>
<td align="left" valign="top" rowspan="1" colspan="1">77 (87.5)</td>
<td align="left" valign="top" rowspan="1" colspan="1">62 (92.5)</td>
<td align="left" valign="top" rowspan="1" colspan="1">36 (92.3)</td>
</tr>
<tr>
<td align="left" valign="top" rowspan="1" colspan="1">>1.00 diopter</td>
<td align="left" valign="top" rowspan="1" colspan="1">11 (7.5)</td>
<td align="left" valign="top" rowspan="1" colspan="1">11 (12.5)</td>
<td align="left" valign="top" rowspan="1" colspan="1">5 (7.5)</td>
<td align="left" valign="top" rowspan="1" colspan="1">3 (7.7)</td>
</tr>
</tbody>
</table>
</table-wrap>
</floats-group>
</pmc>
<affiliations>
<list>
<country>
<li>Australie</li>
<li>Hong Kong</li>
<li>Italie</li>
</country>
<settlement>
<li>Sha Tin</li>
</settlement>
<orgName>
<li>Université chinoise de Hong Kong</li>
</orgName>
</list>
<tree>
<country name="Australie">
<noRegion>
<name sortKey="Heiner, Peter" sort="Heiner, Peter" uniqKey="Heiner P" first="Peter" last="Heiner">Peter Heiner</name>
</noRegion>
</country>
<country name="Italie">
<noRegion>
<name sortKey="Ligabue, Edoardo" sort="Ligabue, Edoardo" uniqKey="Ligabue E" first="Edoardo" last="Ligabue">Edoardo Ligabue</name>
</noRegion>
</country>
<country name="Hong Kong">
<noRegion>
<name sortKey="Fan, Alex" sort="Fan, Alex" uniqKey="Fan A" first="Alex" last="Fan">Alex Fan</name>
</noRegion>
<name sortKey="Lam, Dennis" sort="Lam, Dennis" uniqKey="Lam D" first="Dennis" last="Lam">Dennis Lam</name>
</country>
</tree>
</affiliations>
</record>

Pour manipuler ce document sous Unix (Dilib)

EXPLOR_STEP=$WICRI_ROOT/Ticri/CIDE/explor/HapticV1/Data/Ncbi/Merge
HfdSelect -h $EXPLOR_STEP/biblio.hfd -nk 002E73 | SxmlIndent | more

Ou

HfdSelect -h $EXPLOR_AREA/Data/Ncbi/Merge/biblio.hfd -nk 002E73 | SxmlIndent | more

Pour mettre un lien sur cette page dans le réseau Wicri

{{Explor lien
   |wiki=    Ticri/CIDE
   |area=    HapticV1
   |flux=    Ncbi
   |étape=   Merge
   |type=    RBID
   |clé=     PMC:3974695
   |texte=   Safety and effectiveness of a single-piece hydrophobic acrylic intraocular lens (enVista®) – results of a European and Asian-Pacific study
}}

Pour générer des pages wiki

HfdIndexSelect -h $EXPLOR_AREA/Data/Ncbi/Merge/RBID.i   -Sk "pubmed:24729678" \
       | HfdSelect -Kh $EXPLOR_AREA/Data/Ncbi/Merge/biblio.hfd   \
       | NlmPubMed2Wicri -a HapticV1 

Wicri

This area was generated with Dilib version V0.6.23.
Data generation: Mon Jun 13 01:09:46 2016. Site generation: Wed Mar 6 09:54:07 2024